NATtrol BD MAX Vaginal Panel External Controls

K201403 · Zeptometrix · PMN · Feb 2, 2022 · Microbiology

Device Facts

Record IDK201403
Device NameNATtrol BD MAX Vaginal Panel External Controls
ApplicantZeptometrix
Product CodePMN · Microbiology
Decision DateFeb 2, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3920
Device ClassClass 2

Indications for Use

NATtrol™ BV Negative Control (NATBVNEG-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol BV Negative Control is a qualitative control containing intact and inactivated Lactobacillus crispatus and intended to be used solely with the BD Vaginal Panel for BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. NATtrol™ BV Positive Control (NATBVPOS-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol™ BV Positive Control is a qualitative control containing intact and inactivated Lactobacillus jensenii, Gardnerella vaginalis, Atopobium vaginae, and Saccharomyces cerevisiae containing BVAB2 (Bacterial Vaginosis-Associated Bacterium 2) sequence and intended to be used solely with the BD Vaginal Panel for BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. NATtrol™ Candida/TV Positive Control (NATCTVPOS-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol Candida/TV Positive Control is a qualitative control containing intact and inactivated Candida albicans, Candida krusei, Candida glabrata, and Trichomonas vaginalis and intended to be used solely with the BD Vaginal Panel for the BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert.

Device Story

NATtrol™ BD MAX Vaginal Panel External Controls consist of purified, intact, chemically inactivated microorganisms; formulated in a matrix containing Fetal Bovine Serum, Human Serum Albumin, and sodium azide. Used in clinical laboratories to monitor the performance of the BD MAX™ System (extraction, amplification, and detection steps) for vaginal panel assays. Controls are ready-to-use liquids; processed by laboratory technicians alongside patient samples. Output is a qualitative result (positive/negative) for specific vaginal pathogens; used by clinicians to verify assay validity and ensure reliable patient testing. Benefits include standardized quality assurance for molecular diagnostic workflows.

Clinical Evidence

Single-site performance study conducted with 6 operators using multiple lots of controls. Evaluated precision and performance against expected results on the BD MAX System. Results showed high agreement: BV Positive Control (97.7-100% agreement), Candida/TV Positive Control (98.1-99.0% agreement), and BV Negative Control (99.0-100% agreement). No clinical patient data; bench testing only.

Technological Characteristics

Formulated with purified, intact, chemically inactivated microorganisms. Matrix includes Fetal Bovine Serum, Human Serum Albumin, and sodium azide. Ready-to-use liquid format. Refrigerator stable. Designed for use with the BD MAX System (automated nucleic acid extraction, amplification, and detection).

Indications for Use

Indicated for use as an assayed quality control material to monitor the performance of in vitro diagnostic nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for the BD MAX™ System. Not intended for patient diagnosis or to replace manufacturer-provided controls.

Regulatory Classification

Identification

An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.

Special Controls

An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; (ii) Expected values; (iii) Analyte source; (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays; (C) Stating if the material is a surrogate control; and (D) The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.” (iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ February 2, 2022 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bold blue font, and the word "ADMINISTRATION" in a smaller, lighter blue font below. Zeptometrix % Michael Smith Manager, Regulatory and Clinical Affairs MDC Associates Inc. 180 Cabot Street Beverly, Massachusetts 01915 Re: K201403 Trade/Device Name: NATtrol Vaginal Panel External Run Controls Regulation Number: 21 CFR 866.3920 Regulation Name: Assayed Quality Control Material For Clinical Microbiology Assays Regulatory Class: Class II Product Code: PMN Dated: May 26, 2020 Received: May 28, 2020 Dear Michael Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Maria Garcia, Ph.D. Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201403 #### Device Name NATtrol™ BD MAX Vaginal Panel External Controls #### Indications for Use (Describe) NATtrol™ BV Negative Control (NATBVNEG-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol BV Negative Control is a qualitative control containing intact and inactivated Lactobacillus crispatus and intended to be used solely with the BD Vaginal Panel for BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. NATtroI™ BV Positive Control (NATBVPOS-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol™ BV Positive Control is a qualitative control containing intact and inactivated Lactobacillus jensenii, Gardnerella vaginalis, Atopobium vaginae, and Saccharomyces cerevisiae containing BVAB2 (Bacterial Vaginosis-Associated Bacterium 2) sequence and intended to be used solely with the BD Vaginal Panel for BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. NATtrol™ Candida/TV Positive Control (NATCTVPOS-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol CandidaTV Positive Control is a qualitative control containing intact and inactivated Candida krusei, Candida glabrata, and Trichomonas vaginalis and intended to be used solely with the BD Vaginal Panel for the BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------|--| | <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> | | | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | Date of Summary: | November 12, 2021 | |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name: | NATtrol™ BD MAX Vaginal Panel External Controls | | Sponsor: | ZeptoMetrix, LLC<br>878 Main Street<br>Buffalo, NY. 14202 | | Correspondent: | MDC Associates, Inc.<br>Michael Smith, Manager, Regulatory and Clinical Affairs<br>180 Cabot Street<br>Beverly, MA 01915<br>Phone: (978) 927 3808<br>Fax: (866) 540 3448<br>Email: regulatory@mdcassoc.com | | Common Name: | Assayed quality control material for clinical microbiology assays. | | Regulation Number: | 866.3920 | | Classification: | Class II | | Predicate Device: | Microbiologics, Inc. BD MAX CT/GC/TV 20-Day QC Panel (K181683) | # Substantial Equivalency | Description | ZeptoMetrix Corporation<br>NATtrol™ BD MAX Vaginal Panel External<br>Controls (New Device) | Microbiologics, Inc.<br>BD MAX CT/GC/TV 20-Day QC Panel<br>K181683<br>(Predicate Device) | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Similarities | | | | Intended Use | NATtrol™ BV Negative Control<br>(NATBVNEG-BD) is intended for use as an<br>assayed quality control material to<br>monitor the performance of in vitro<br>diagnostic laboratory nucleic acid testing<br>procedures for the qualitative detection of<br>targets on the BD Vaginal Panel for BD<br>MAX™ System. NATtrol BV Negative<br>Control is a qualitative control containing<br>intact and inactivated <i>Lactobacillus crispatus</i> and intended to be used solely<br>with the BD Vaginal Panel for BD MAX™<br>System. This product is not intended to<br>replace manufacturer controls provided in<br>the package insert.<br><br>NATtrol™ BV Positive Control (NATBVPOS-<br>BD) is intended for use as an assayed | The BD MAX CT/GC/TV 20-Day QC Panel is<br>intended for use as an external assayed positive<br>quality control material to monitor the<br>performance of in vitro laboratory nucleic acid<br>testing procedures for the qualitative detection<br>of <i>Chlamydia trachomatis, Neisseria gonorrhoeae</i> and <i>Trichomonas vaginalis</i> with the<br>BD MAX CT/GC/TV Assay on the BD MAX<br>System. The controls comprise cultured and<br>inactivated <i>C. trachomatis, N. gonorrhoeae</i> and<br><i>T. vaginalis</i> . The BD MAX CT/GC/TV 20-Day QC<br>Panel is not intended to replace manufacturer<br>controls provided with the device. | | Description | ZeptoMetrix Corporation<br>NATtrolTM BD MAXTM Vaginal Panel External<br>Controls (New Device) | Microbiologics, Inc.<br>BD MAX CT/GC/TV 20-Day QC Panel<br>K181683<br>(Predicate Device) | | | quality control material to monitor the<br>performance of in vitro diagnostic<br>laboratory nucleic acid testing procedures<br>for the qualitative detection of targets on<br>the BD Vaginal Panel for BD MAXTM System.<br>NATtrolTM BV Positive Control is a<br>qualitative control containing intact and<br>inactivated Lactobacillus jensenii,<br>Gardnerella vaginalis, Atopobium vaginae,<br>and Saccharomyces cerevisiae containing<br>BVAB2 (Bacterial Vaginosis-Associated<br>Bacterium 2) sequence and intended to be<br>used solely with the BD Vaginal Panel for<br>BD MAXTM System. This product is not<br>intended to replace manufacturer controls<br>provided in the package insert.<br><br>NATtrolTM Candida/TV Positive Control<br>(NATCTVPOS-BD) is intended for use as an<br>assayed quality control material to<br>monitor the performance of in vitro<br>diagnostic laboratory nucleic acid testing<br>procedures for the qualitative detection of<br>targets on the BD Vaginal Panel for BD<br>MAXTM System. NATtrolTM Candida/TV<br>Positive Control is a qualitative control<br>containing intact and inactivated Candida<br>albicans, Candida krusei, Candida glabrata,<br>and Trichomonas vaginalis and intended to<br>be used solely with the BD Vaginal Panel<br>for the BD MAXTM System. This product is<br>not intended to replace manufacturer<br>controls provided in the package insert. | | | Composition | Inactivated microorganisms | Same | | Test system | BD MAX System | Same | | Assay Steps Monitored | Extraction, amplification and detection | Same | | Analytes | Inactivated microorganisms | Same | | | Differences | | | Physical Format | Ready-to-use liquid | Lyophilized pellet | | Number of Targets<br>Monitored | 8 | 3 | {4}------------------------------------------------ {5}------------------------------------------------ # Intended Use NATtrol™ BV Negative Control (NATBVNEG-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol BV Negative Control is a qualitative control containing intact and inactivated Lactobacillus crispatus and intended to be used solely with the BD Vaginal Panel for BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. NATtrol™ BV Positive Control (NATBVPOS-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol™ BV Positive Control is a qualitative control containing intact and inactivated Lactobacillus jensenii, Gardnerella vaginalis, Atopobium vaqinae, and Saccharomyces cerevisiae containing BVAB2 (Bacterial Vaginosis-Associated Bacterium 2) sequence and intended to be used solely with the BD Vaginal Panel for BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. NATtrol™ Candida/TV Positive Control (NATCTVPOS-BD) is intended for use as an assayed quality control material to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of targets on the BD Vaginal Panel for BD MAX™ System. NATtrol Candida/TV Positive Control is a qualitative control containing intact and inactivated Candida albicans, Candida krusei, Candida qlabrata, and Trichomonas vaginalis and intended to be used solely with the BD Vaginal Panel for the BD MAX™ System. This product is not intended to replace manufacturer controls provided in the package insert. ### Test Principle Not applicable; this is control material to monitor performance of a test. ### Device Description NATtrol™ BD MAX Vaginal Panel External Controls are formulated with purified, intact microorganisms that have been chemically modified to render them non-infectious and refrigerator stable. The controls are formulated in a proprietary matrix that contains Fetal Bovine Serum, Human Serum Albumin, and sodium azide. NATtrol™ BV Positive Control contains intact and inactivated Lactobacillus jensenii, Gardnerella vaginalis, Atopobium vaginae, and Saccharomyces cerevisiae containing BVAB2 (Bacterial Vaginosis-Associated Bacterium 2) sequence. NATtrol™ Candida/TV Positive Control contains intact and inactivated Candida albicans, Candida krusei, Candida glabrata, and Trichomonas vaginalis. NATtrol™ BV Negative Control contains intact and inactivated Lactobacillus crispatus. {6}------------------------------------------------ # Performance Data A single site performance study was conducted to evaluate precision and performance of the NATtrol™ BD MAX Vaginal Panel External Controls. The Performance of the controls was evaluated by observing the data collected during multiple studies performed by six (6) operators using two (2) lots of BV Positive, two (2) lots of BV Negative and one (1) lot of Candida/TV Positive controls. | | BV Positive Control | | |-------------|-----------------------------------|-----------------------------------------------| | Target | BD MAX Result<br>(Observed/Total) | % Agreement with expected<br>Results (95% CI) | | BV | Positive (84/86) | 97.7%<br>(91.9%-99.4%) | | C group | Negative (86/86) | 100%<br>(95.7%-100%) | | C. krusei | Negative (86/86) | 100%<br>(95.7%-100%) | | C. glabrata | Negative (86/86) | 100%<br>(95.7%-100%) | | TV | Negative (86/86) | 100%<br>(95.7%-100%) | | | Candida/TV Positive Control | | |-------------|-----------------------------------|-----------------------------------------------| | Target | BD MAX Result<br>(Observed/Total) | % Agreement with expected<br>Results (95% CI) | | BV | Negative (102/104) | 98.1%<br>(93.3%-99.5%) | | C group | Positive (103/104) | 99.0%<br>(94.8%-99.8%) | | C. krusei | Positive (103/104) | 99.0%<br>(94.8%-99.8%) | | C. glabrata | Positive (103/104) | 99.0%<br>(94.8%-99.8%) | | TV | Positive (103/104) | 99.0%<br>(94.8%-99.8%) | | | BV Negative Control | | |-------------|-----------------------------------|-----------------------------------------------| | Target | BD MAX Result<br>(Observed/Total) | % Agreement with expected<br>Results (95% CI) | | BV | Negative (191/191) | 100%<br>(98.0%-100%) | | C group | Negative (189/191) | 99.0%<br>(96.3%-99.7%) | | C. krusei | Negative (191/191) | 100%<br>(98.0%-100%) | | C. glabrata | Negative (191/191) | 100%<br>(98.0%-100%) | | TV | Negative (191/191) | 100%<br>(98.0%-100%) | # Conclusion The submitted information in this premarket notification is complete and supports asubstantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...