Simplexa VZV Swab Direct, Simplexa VZV Positive Control Pack

K192376 · Diasorin Molecular, LLC · PGI · Nov 26, 2019 · Microbiology

Device Facts

Record IDK192376
Device NameSimplexa VZV Swab Direct, Simplexa VZV Positive Control Pack
ApplicantDiasorin Molecular, LLC
Product CodePGI · Microbiology
Decision DateNov 26, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3309
Device ClassClass 2
AttributesPediatric

Indications for Use

The DiaSorin Molecular Simplexa™ VZV Swab Direct assay is intended for use on the LIAJSON® MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA present in cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of VZV infection. This test is intended as an aid in the diagnosis of VZV infection. Negative results do not preclude VZV infection and should not be used as the sole basis for treatment or other patient management decisions. The Simplexa™ VZV Positive Control Pack is intended to be used as a control with the Simplexa™ VZV Direct kit and the Simplexa™ VZV Swab Direct kit on the LIAISON® MDX Instrument. It is not intended for use with other assays or systems.

Device Story

Simplexa VZV Swab Direct is a real-time PCR assay for qualitative detection of VZV DNA from unprocessed cutaneous and mucocutaneous lesion swabs. The system uses the LIAISON MDX instrument and Direct Amplification Disc (DAD). It targets a conserved region of the VZV DNA polymerase gene; fluorescent probes and primers amplify VZV and an internal control. The assay eliminates the need for nucleic acid extraction. It is used in clinical laboratory settings by trained personnel. Results aid in diagnosing VZV infection; negative results do not rule out infection. The device provides rapid, automated detection, assisting clinicians in patient management decisions.

Clinical Evidence

Clinical study included 452 prospective, 60 retrospective, and 60 contrived samples. Prospective study sensitivity 97.8% (92.2-99.4% CI), specificity 99.2% (97.6-99.7% CI) against a 3-part composite reference method (DSFA/culture + 2 PCR/sequencing). Retrospective study PPA 98.4%, NPA 99.2%. Contrived study PPA 100%, NPA 100%. Reproducibility evaluated across 3 sites with 90 replicates per panel member; 100% agreement observed.

Technological Characteristics

Real-time PCR assay; targets VZV DNA polymerase gene; uses fluorescent probes and primers; requires LIAISON MDX instrument and Direct Amplification Disc (DAD); no nucleic acid extraction required; internal control included for PCR inhibition monitoring; qualitative output.

Indications for Use

Indicated for qualitative detection of VZV DNA in cutaneous and mucocutaneous lesion swabs from symptomatic patients of all ages (one month to >60 years) suspected of VZV infection. Rx only.

Regulatory Classification

Identification

A herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel is a qualitative in vitro diagnostic device intended for the simultaneous detection and differentiation of different herpes viruses in cutaneous and mucocutaneous lesion samples from symptomatic patients suspected of Herpetic infections. Negative results do not preclude infection and should not be used as the sole basis for treatment or other patient management decisions. The assay is not intended for use in cerebrospinal fluid samples.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer design and selection. (2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination. (3) Premarket notification submissions must include detailed documentation of a clinical study using lesion samples in which Herpes Simplex Virus 1, Herpes Simplex Virus 2, or Varicella Zoster Virus DNA detection was requested. The study must compare the device performance to an appropriate well established reference method. (4) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling. (5) The device labeling must include a limitation statement that reads: “The device is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS).” (6) Premarket notification submissions must include quality assurance protocols and a detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software. (7) The risk management activities performed as part of the manufacturer's 21 CFR 820.10(c) design and development activities must document an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K192376 B Applicant DiaSorin Molecular LLC C Proprietary and Established Names Simplexa VZV Swab Direct, Simplexa VZV Positive Control Pack D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | PGI | Class II | 21 CFR 866.3309 - Herpes Virus Nucleic Acid-Based Cutaneous And Mucocutaneous Lesion Panel | MI - Microbiology | | PMN | Class II | 21 CFR 866.3920 - Assayed quality control material for clinical microbiology assays | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain a substantial equivalence determination and FDA clearance for a new device. B Measurand: Varicella-zoster virus DNA K192376 - Page 1 of 16 {1} C Type of Test: Realtime Polymerase Chain Reaction ## III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: Simplexa™ VZV Swab Direct The DiaSorin Molecular Simplexa™ VZV Swab Direct assay is intended for use on the LIAISON® MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA present in cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of VZV infection. This test is intended as an aid in the diagnosis of VZV infection. Negative results do not preclude VZV infection and should not be used as the sole basis for treatment or other patient management decisions. Simplexa™ VZV Positive Control Pack The Simplexa™ VZV Positive Control Pack is intended to be used as a control with the Simplexa™ VZV Direct kit and the Simplexa™ VZV Swab Direct kit on the LIAISON® MDX instrument. It is not intended for use with other assays or systems. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: LIAISON MDX instrument ## IV Device/System Characteristics: A Device Description: The Simplexa VZV Swab Direct assay is a real-time PCR system that enables the direct amplification and detection of VZV DNA from unprocessed cutaneous and mucocutaneous lesion swab specimens without nucleic acid extraction. The system consists of the Simplexa VZV Swab Direct assay, the LIAISON MDX (with LIAISON MDX Studio Software), the Direct Amplification Disc (DAD) and associated accessories. In the Simplexa VZV Swab Direct assay, fluorescent probes are used together with corresponding forward and reverse primers to amplify VZV and internal control targets. A well- K192376 - Page 2 of 16 {2} conserved region of the VZV DNA polymerase gene is targeted to identify VZV DNA in the specimen. An internal control is used to detect PCR failure and/or inhibition. ## B Principle of Operation: Nucleic acid amplification testing (NAAT) ## C Instrument Description Information: | Modes of Operation | Yes | No | | --- | --- | --- | | Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☐ | ☑ | | Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑ | | Software | | | | FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ | 1. Instrument Name: LIAISON MDX 2. Specimen Identification: N/A 3. Specimen Sampling and Handling: N/A 4. Calibration: N/A 5. Quality Control: N/A ## V Substantial Equivalence Information: ### A Predicate Device Name(s): Solana HSV 1+2/VZV Assay ### B Predicate 510(k) Number(s): K162451 ## C Comparison with Predicate(s): | Device & Predicate Device(s): | K192376 | K162451 | | --- | --- | --- | | Device Trade Name | Simplexa VZV Swab Direct | Solana HSV 1+2/VZV Assay | | General Device Characteristic Similarities | | | | Intended Use/Indications for Use | The DiaSorin Molecular Simplexa VZV Swab Direct | The Solana HSV 1+2/VZV Assay is an in vitro diagnostic | K192376 - Page 3 of 16 {3} | | assay is intended for use on the LIAISON MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA present in cutaneous and mucocutaneous lesion swabs from patients with signs and symptoms of VZV infection. This test is intended as an aid in the diagnosis of VZV infection. Negative results do not preclude VZV infection and should not be used as the sole basis for treatment or other patient management decisions. | test, using isothermal amplification technology (helicase-dependent amplification, HDA), for the qualitative detection and differentiation of herpes simplex virus type 1, herpes simplex virus type 2, and varicella-zoster virus DNA isolated and purified from cutaneous or mucocutaneous lesion samples obtained from symptomatic patients suspected of active herpes simplex virus 1, herpes simplex virus 2 and/or Varicella-zoster infection. The Solana HSV 1+2/VZV Assay is intended to aid in the diagnosis of herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus active cutaneous or mucocutaneous infections. Negative results do not preclude herpes simplex virus 1, herpes simplex virus 2 and varicella-zoster virus infections and should not be used as the sole basis for diagnosis, treatment or other management decisions. The Solana HSV 1+2/VZV Assay is intended for use only with the Solana instrument. | | --- | --- | --- | | Automated System | Yes | Yes | | Technology | Nucleic acid amplification | Same | | General Device Characteristic Differences | | | | Instrument | LIAISON MDX | Solana instrument | VI Standards/Guidance Documents Referenced: N/A VII Performance Characteristics (if/when applicable): A Analytical Performance: K192376 - Page 4 of 16 {4} # 1. Precision/Reproducibility: Reproducibility for the Simplexa VZV Swab Direct assay was evaluated at three (3) investigative sites to assess the device's inter-site, inter/intra-day and inter/intra-assay reproducibility. Each of the laboratories tested a sample panel consisting of Simplexa VZV Swab Direct Positive Control, No Template Control, and four (4) contrived samples in negative matrix. Two (2) strains of VZV were used in the study, 9939 and Ellen. The four (4) contrived samples consisted of a low positive (LP) at $2 \times$ LoD and a medium positive (MP) at $4 \times$ LoD for each VZV strain. Each sample panel member was tested in triplicate per run, for two (2) runs per day by two (2) different operators at each site. Therefore, a total of ninety (90) replicates [three (3) replicates x two (2) runs x five (5) days x three (3) sites] were tested for each sample panel member. A total of six (6) LIAISON MDX instruments [two (2) per site] were used. The combined results for all sites are presented in Table 1. Table 1. Simplexa VZV Swab Direct Reproducibility | Summary of VZV Qualitative Results and VZV Ct Values ± SD (%CV) | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | Site 1 | | Site 2 | | Site 4 | | All Sites | | | | % Agreement With Expected Results | Detected Mean Ct ± SD (%CV) | % Agreement With Expected Results | Detected Mean Ct ± SD (%CV) | % Agreement With Expected Results | Detected Mean Ct ± SD (%CV) | % Agreement With Expected Results | Detected Mean Ct ± SD (%CV) | | 9939 LP | 100.0% (30/30) | 36.6 ± 1.12 (3.1%) | 100.0% (30/30) | 36.8 ± 0.68 (1.9%) | 100.0% (30/30) | 36.4 ± 0.83 (2.3%) | 100.0% (90/90) | 36.6 ± 0.9 (2.5%) | | 9939 MP | 100.0% (30/30) | 35.8 ± 0.86 (2.4%) | 100.0% (30/30) | 35.7 ± 0.54 (1.5%) | 100.0% (30/30) | 35.3 ± 0.78 (2.2%) | 100.0% (90/90) | 35.6 ± 0.76 (2.1%) | | Ellen LP | 100.0% (30/30) | 35.4 ± 1.22 (3.4%) | 100.0% (30/30) | 34.5 ± 1.77 (5.1%) | 100.0% (30/30) | 35.0 ± 0.56 (1.6%) | 100.0% (90/90) | 35.0 ± 1.32 (3.8%) | | Ellen MP | 100.0% (30/30) | 34.5 ± 0.65 (1.9%) | 100.0% (30/30) | 34.5 ± 0.47 (1.4%) | 100.0% (30/30) | 33.5 ± 1.3 (3.9%) | 100.0% (90/90) | 34.1 ± 0.99 (2.9%) | | UTM (NTC) | 0.0% (0/30) | 0.0 ± 0.00 (N/A) | 0.0% (0/30) | 0.0 ± 0.00 (N/A) | 0.0% (0/30) | 0.0 ± 0.00 (N/A) | 0.0% (0/90) | 0.0 ± 0.00 (N/A) | | PC | 100.0% (30/30) | 30.2 ± 0.74 (2.5%) | 100.0% (30/30) | 30.4 ± 0.59 (1.9%) | 100.0% (30/30) | 29.7 ± 0.86 (2.9%) | 100.0% (90/90) | 30.1 ± 0.79 (2.6%) | # 2. Linearity: N/A # 3. Analytical Specificity/Interference: Cross-Reactivity (Analytical Specificity): The Simplexa VZV Swab Direct assay's analytical specificity was evaluated by testing the ability of the assay to exclusively identify VZV with no cross-reactivity to organisms that are closely related, cause similar clinical symptoms or that K192376 - Page 5 of 16 {5} could be found in cutaneous and mucocutaneous lesion swab specimens. Analytical specificity/cross-reactivity was tested with ninety-nine (99) different bacteria, viruses, parasites and fungi and assayed in triplicate. No cross-reactivity was observed with the ninety-nine (99) organisms. The organisms and the concentration at which these were tested are presented in Table 2. Table 2. Simplexa VZV Swab Direct Cross-Reactivity | Organism | Tested Concentration | Organism | Tested Concentration | | --- | --- | --- | --- | | Acholeplasma laidlawi (genomic DNA) | 1 x 10^6 copies/mL | Human genomic DNA | 1 x 10^6 copies/mL | | Acinetobacter calcoaceticus | 1 x 10^6 CFU/mL | Human metapneumovirus A1 | 1 x 10^5 TCID50/mL | | Acinetobacter lwoffii | 1 x 10^6 CFU/mL | Human Papilloma Virus 18 | 1 x 10^5 copies/mL | | Adenovirus 7A | 1 x 10^5 TCID50/mL | Influenza A/California/7/2009 | 1 x 10^5 TCID50/mL | | Bacteroides fragilis | 1 x 10^6 CFU/mL | Influenza B/Florida/02/2006 | 1 x 10^5 TCID50/mL | | Bordetella bronchiseptica | 1 x 10^6 CFU/mL | Klebsiella pneumoniae | 1 x 10^6 CFU/mL | | Bordetella pertussis | 1 x 10^6 CFU/mL | Lactobacillus acidophilus | 1 x 10^6 CFU/mL | | Borrelia burgdorferi (genomic DNA) | 1 x 10^6 copies/mL | Legionella pneumophila | 1 x 10^6 CFU/mL | | Candida albicans | 1 x 10^6 CFU/mL | Measles virus | 1 x 10^5 TCID50/mL | | Candida glabrata | 1 x 10^6 CFU/mL | Mobiluncus curtisii | 1 x 10^6 CFU/mL | | Candida guilliermondii | 1 x 10^6 CFU/mL | Mobiluncus mulieris | 1 x 10^6 CFU/mL | | Candida krusei | 1 x 10^6 CFU/mL | Moraxella cartarrhalis | 1 x 10^6 CFU/mL | | Candida lusitaniae | 1 x 10^6 CFU/mL | Mumps virus | 1 x 10^5 TCID50/mL | | Candida parapsilosis | 1 x 10^6 CFU/mL | Mycoplasma genitalium | 1 x 10^6 CCU/mL | | Candida tropicalis | 1 x 10^6 CFU/mL | Mycoplasma hominis | 1 x 10^6 CCU/mL | | Chlamydia trachomatis | 1 x 10^6 IFU/mL | Mycoplasma hyorhinis | 1 x 10^6 CCU/mL | | Chlamydophila pneumoniae | 1 x 10^6 IFU/mL | Mycoplasma orale | 1 x 10^6 CCU/mL | | Clostridium difficile | 1 x 10^6 CFU/mL | Mycoplasma pneumoniae | 1 x 10^6 CCU/mL | | Clostridium perfringens | 1 x 10^6 CFU/mL | Mycoplasma salivarium | 1 x 10^6 CCU/mL | | Clostridium sordellii | 1 x 10^6 CFU/mL | Neisseria gonorrhoeae | 1 x 10^6 CFU/mL | | Coronavirus OC43 | 1 x 10^5 TCID50/mL | Neisseria meningitidis | 1 x 10^6 CFU/mL | | Corynebacterium diphtheriae | 1 x 10^6 CFU/mL | Parainfluenza Type 1 | 1 x 10^5 TCID50/mL | | Corynebacterium genitalium | 1 x 10^6 CFU/mL | Parainfluenza Type 2 | 1 x 10^5 TCID50/mL | | Coxsackievirus B1 | 1 x 10^5 TCID50/mL | Parainfluenza Type 3 | 1 x 10^5 TCID50/mL | | Coxsackievirus B4 | 1 x 10^5 TCID50/mL | Parainfluenza Type 4 | 1 x 10^5 TCID50/mL | | Cytomegalovirus (AD169 strain) | 1 x 10^5 TCID50/mL | Prevotella melaninogenica | 1 x 10^6 CFU/mL | K192376 - Page 6 of 16 {6} | Organism | Tested Concentration | Organism | Tested Concentration | | --- | --- | --- | --- | | Cytomegalovirus (Towne strain) | 1 x 10^{5} TCID_{50}/mL | Proteus mirabilis | 1 x 10^{6} CFU/mL | | Echovirus 11 | 1 x 10^{5} TCID_{50}/mL | Proteus vulgaris | 1 x 10^{6} CFU/mL | | Enterobacter cloacae | 1 x 10^{6} CFU/mL | Pseudomonas aeruginosa | 1 x 10^{6} CFU/mL | | Enterococcus faecalis vanB | 1 x 10^{6} CFU/mL | RSV A Long | 1 x 10^{5} TCID_{50}/mL | | Enterococcus faecium | 1 x 10^{6} CFU/mL | RSV B Washington | 1 x 10^{5} TCID_{50}/mL | | Enterovirus 70 | 1 x 10^{5} TCID_{50}/mL | Rubella Virus | 1 x 10^{5} TCID_{50}/mL | | Enterovirus 71 | 1 x 10^{5} TCID_{50}/mL | Salmonella enteritidis (genomic DNA) | 1 x 10^{6} copies/mL | | Epstein Barr Virus (B95-8 strain) | 1 x 10^{5} copies/mL | Salmonella typhimurium | 1 x 10^{6} CFU/mL | | Escherichia coli O15:H7 | 1 x 10^{6} CFU/mL | Serratia marcescens | 1 x 10^{6} CFU/mL | | Fusobacterium nucleatum | 1 x 10^{6} CFU/mL | Simian Virus type 40 | 1 x 10^{5} TCID_{50}/mL | | Gardnerella vaginalis | 1 x 10^{6} CFU/mL | Staphylococcus aureus (MRSA), ATCC 700699 | 1 x 10^{6} CFU/mL | | Haemophilus ducreyi | 1 x 10^{6} CFU/mL | Staphylococcus aureus (MRSA), COL | 1 x 10^{6} CFU/mL | | Haemophilus influenza type A | 1 x 10^{6} CFU/mL | Staphylococcus epidermidis (MRSE), ATCC 29887 | 1 x 10^{6} CFU/mL | | Hepatitis A virus | 1 x 10^{6} TCID_{50}/mL | Staphylococcus saprophyticus | 1 x 10^{6} CFU/mL | | Hepatitis B virus | 1 x 10^{5} IU/mL | Streptococcus agalactiae | 1 x 10^{6} CFU/mL | | Hepatitis C virus | 1 x 10^{5} IU/mL | Streptococcus mitis | 1 x 10^{6} CFU/mL | | HHV-6 (Z29 strain) | 1 x 10^{5} copies/mL | Streptococcus mutans | 1 x 10^{6} CFU/mL | | HHV-6A | 1 x 10^{5} copies/mL | Streptococcus pneumoniae | 1 x 10^{6} CFU/mL | | HHV-7 SB | 1 x 10^{5} TCID_{50}/mL | Streptococcus pyogenes, M1 | 1 x 10^{6} CFU/mL | | HHV-8 | 1 x 10^{5} copies/mL | Streptococcus salivarius | 1 x 10^{6} CFU/mL | | HIV-1 IIIB | 1 x 10^{5} TCID_{50}/mL | Toxoplasma gondii | 1 x 10^{6} tachyzoites/mL | | HIV-2 NIHZ | 1 x 10^{5} TCID_{50}/mL | Trichomonas vaginalis | 1 x 10^{6} trophozoites/mL | | HSV-1 (McIntyre strain) | 1 x 10^{5} TCID_{50}/mL | Ureaplasma urealyticum | 1 x 10^{6} CCU/mL | | HSV-2 (G strain) | 1 x 10^{5} TCID_{50}/mL | | | Note: Bacteroides ureolyticus, Hepatitis D virus, Treponema pallidum and Tropheryma whipplei were tested using in silico NCBI BLAST analysis due to unavailability of the organism. No cross-reactivity was found. K192376 - Page 7 of 16 {7} Inhibition by other Microorganisms: The Simplexa VZV Swab Direct assay was evaluated by testing the ability to identify VZV virus (Ellen and 9939 strains) when other potentially inhibitory organisms are present. A panel of ninety-nine (99) potential inhibitory organisms was individually spiked into pooled cutaneous and mucocutaneous swab matrix containing a low concentration of VZV at approximately 2X LoD and tested in triplicate. Table below references the microorganisms and their respective tested concentration. No inhibition was observed for the detection of either VZV Ellen or 9939 strain. Table 3. Simplexa VZV Swab Direct Microbial Inhibition | Organism | Tested Concentration | Organism | Tested Concentration | | --- | --- | --- | --- | | Acholeplasma laidlawi (genomic DNA) | 1 x 10^{6} copies/mL | Human genomic DNA | 1 x 10^{6} copies/mL | | Acinetobacter calcoaceticus | 1 x 10^{6} CFU/mL | Human metapneumovirus A1 | 1 x 10^{5} TCID_{50}/mL | | Acinetobacter lwoffii | 1 x 10^{6} CFU/mL | Human papilloma virus 18 | 1 x 10^{5} copies/mL | | Adenovirus 7A | 1 x 10^{5} TCID_{50}/mL | Influenza A/California/7/2009 | 1 x 10^{5} TCID_{50}/mL | | Bacteroides fragilis | 1 x 10^{6} CFU/mL | Influenza B/Florida/02/2006 | 1 x 10^{5} TCID_{50}/mL | | Bordetella bronchiseptica | 1 x 10^{6} CFU/mL | Klebsiella pneumoniae | 1 x 10^{6} CFU/mL | | Bordetella pertussis | 1 x 10^{6} CFU/mL | Lactobacillus acidophilus | 1 x 10^{6} CFU/mL | | Borrelia burgdorferi (genomic DNA) | 1 x 10^{6} copies/mL | Legionella pneumophila | 1 x 10^{6} CFU/mL | | Candida albicans | 1 x 10^{6} CFU/mL | Measles virus | 1 x 10^{5} TCID_{50}/mL | | Candida glabrata | 1 x 10^{6} CFU/mL | Mobiluncus curtisii | 1 x 10^{6} CFU/mL | | Candida guilliermondii | 1 x 10^{6} CFU/mL | Mobiluncus mulieris | 1 x 10^{6} CFU/mL | | Candida krusei | 1 x 10^{6} CFU/mL | Moraxella cartarrhalis | 1 x 10^{6} CFU/mL | | Candida lusitaniae | 1 x 10^{6} CFU/mL | Mumps virus | 1 x 10^{5} TCID_{50}/mL | | Candida parapsilosis | 1 x 10^{6} CFU/mL | Mycoplasma genitalium | 1 x 10^{6} CCU/mL | | Candida tropicalis | 1 x 10^{6} CFU/mL | Mycoplasma hominis | 1 x 10^{6} CCU/mL | | Chlamydia trachomatis | 1 x 10^{6} IFU/mL | Mycoplasma hyorhinis | 1 x 10^{6} CCU/mL | | Chlamydophila pneumoniae | 1 x 10^{6} IFU/mL | Mycoplasma orale | 1 x 10^{6} CCU/mL | | Clostridium difficile | 1 x 10^{6} CFU/mL | Mycoplasma pneumoniae | 1 x 10^{6} CCU/mL | | Clostridium perfringens | 1 x 10^{6} CFU/mL | Mycoplasma salivarium | 1 x 10^{6} CCU/mL | | Clostridium sordellii | 1 x 10^{6} CFU/mL | Neisseria gonorrhoeae | 1 x 10^{6} CFU/mL | | Coronavirus OC43 | 1 x 10^{5} TCID_{50}/mL | Neisseria meningitidis | 1 x 10^{6} CFU/mL | | Corynebacterium diphtheriae | 1 x 10^{6} CFU/mL | Parainfluenza Type 1 | 1 x 10^{5} TCID_{50}/mL | | Corynebacterium genitalium | 1 x 10^{6} CFU/mL | Parainfluenza Type 2 | 1 x 10^{5} TCID_{50}/mL | | Coxsackievirus B1 | 1 x 10^{5} TCID_{50}/mL | Parainfluenza Type 3 | 1 x 10^{5} TCID_{50}/mL | | Coxsackievirus B4 | 1 x 10^{5} TCID_{50}/mL | Parainfluenza Type 4 | 1 x 10^{5} TCID_{50}/mL | | Cytomegalovirus (AD169 strain) | 1 x 10^{5} TCID_{50}/mL | Prevotella melaninogenica | 1 x 10^{6} CFU/mL | K192376 - Page 8 of 16 {8} | Organism | Tested Concentration | Organism | Tested Concentration | | --- | --- | --- | --- | | Cytomegalovirus (Towne strain) | 1 x 10^{5} TCID_{50}/mL | Proteus mirabilis | 1 x 10^{6} CFU/mL | | Echovirus 11 | 1 x 10^{5} TCID_{50}/mL | Proteus vulgaris | 1 x 10^{6} CFU/mL | | Enterobacter cloacae | 1 x 10^{6} CFU/mL | Pseudomonas aeruginosa | 1 x 10^{6} CFU/mL | | Enterococcus faecalis vanB | 1 x 10^{6} CFU/mL | RSV A Long | 1 x 10^{5} TCID_{50}/mL | | Enterococcus faecium | 1 x 10^{6} CFU/mL | RSV B Washington | 1 x 10^{5} TCID_{50}/mL | | Enterovirus 70 | 1 x 10^{5} TCID_{50}/mL | Rubella Virus | 1 x 10^{5} TCID_{50}/mL | | Enterovirus 71 | 1 x 10^{5} TCID_{50}/mL | Salmonella enteritidis (genomic DNA) | 1 x 10^{6} copies/mL | | Epstein Barr Virus (B95-8 strain) | 1 x 10^{5} copies/mL | Salmonella typhimurium | 1 x 10^{6} CFU/mL | | Escherichia coli O15:H7 | 1 x 10^{6} CFU/mL | Serratia marcescens | 1 x 10^{6} CFU/mL | | Fusobacterium nucleatum | 1 x 10^{6} CFU/mL | Simian Virus type 40 | 1 x 10^{5} TCID_{50}/mL | | Gardnerella vaginalis | 1 x 10^{6} CFU/mL | Staphylococcus aureus (MRSA), ATCC 700699 | 1 x 10^{6} CFU/mL | | Haemophilus ducreyi | 1 x 10^{6} CFU/mL | Staphylococcus aureus (MRSA), COL | 1 x 10^{6} CFU/mL | | Haemophilus influenza type A | 1 x 10^{6} CFU/mL | Staphylococcus epidermidis (MRSE), ATCC 29887 | 1 x 10^{6} CFU/mL | | Hepatitis A virus | 1 x 10^{6} TCID_{50}/mL | Staphylococcus saprophyticus | 1 x 10^{6} CFU/mL | | Hepatitis B virus | 1 x 10^{5} IU/mL | Streptococcus agalactiae | 1 x 10^{6} CFU/mL | | Hepatitis C virus | 1 x 10^{5} IU/mL | Streptococcus mitis | 1 x 10^{6} CFU/mL | | HHV-6 (Z29 strain) | 1 x 10^{5} copies/mL | Streptococcus mutans | 1 x 10^{6} CFU/mL | | HHV-6A | 1 x 10^{5} copies/mL | Streptococcus pneumoniae | 1 x 10^{6} CFU/mL | | HHV-7 SB | 1 x 10^{5} TCID_{50}/mL | Streptococcus pyogenes, M1 | 1 x 10^{6} CFU/mL | | HHV-8 | 1 x 10^{5} copies/mL | Streptococcus salivarius | 1 x 10^{6} CFU/mL | | HIV-1 IIIB | 1 x 10^{5} TCID_{50}/mL | Toxoplasma gondii | 1 x 10^{6} tachyzoites/mL | | HIV-2 NIHZ | 1 x 10^{5} TCID_{50}/mL | Trichomonas vaginalis | 1 x 10^{6} trophozoites/mL | | HSV-1 (McIntyre strain) | 1 x 10^{5} TCID_{50}/mL | Ureaplasma urealyticum | 1 x 10^{6} CCU/mL | | HSV-2 (G strain) | 1 x 10^{5} TCID_{50}/mL | | | Note: Bacteroides ureolyticus, Hepatitis D virus, Treponema pallidum and Tropheryma whipplei were tested using in silico NCBI BLAST analysis due to unavailability of the organism. No interference was found. Interference: The performance of the Simplexa VZV Swab Direct assay was evaluated with potentially interfering substances. The tested concentrations of the potential interfering endogenous and exogenous substances are indicated in the table below. A total of forty-five K192376 - Page 9 of 16 {9} (45) potential interfering substances were individually spiked into a pooled cutaneous and mucocutaneous swab matrix containing a low concentration of VZV at approximately 2X LoD and tested in triplicate. No interference was observed as shown below. Table 4. Simplexa VZV Swab Direct Interference | Potentially Interfering Substance. | VZV Strain | Active Ingredient | Tested Concentration | # Detected/# Tested | | --- | --- | --- | --- | --- | | Abreva | 9939 | Docosanol 10% | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Acetaminophen | 9939 | N/A | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Acyclovir | 9939 | N/A | 10 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Albumin | 9939 | N/A | 10 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Balneol lotion | 9939 | Buffers, emulsifiers, PEG, water, mineral oil, lanolin oil, preservatives | 7% (v/v) | 3/3 | | | Ellen | | | 3/3 | | Carmex | 9939 | Camphor, 1.7%; Menthol, 0.7% | 10% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Casein | 9939 | N/A | 10 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Chlor-Trimeton | 9939 | Chlorpheniramine maleate | 5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Cidofovir | 9939 | N/A | 2.5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Clotrimazole Vaginal Cream | 9939 | Clotrimazole | 7% (w/v) | 3/3 | | | Ellen | | 3.5% (w/v) | 3/3 | | Cold-Eeze | Ellen | Zincum Gluconicum 2X | 5% (w/v) | 3/3 | | | 9939 | | 2.5% (w/v) | 3/3 | | Cornstarch | 9939 | N/A | 1.25 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Denavir | 9939 | N/A | 2.5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Desitin | 9939 | Zinc Oxide, 40% | 7% (w/v) | 2/2 | | | | | 3.5% (w/v) | 3/3 | | | Ellen | | 3.5% (w/v) | 3/3 | | Dextromethorphan hydrobromide (Robitussin-DM) | 9939 | Dextromethorphan hydrobromide | 10 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Douche | 9939 | N/A | 7% (v/v) | 3/3 | | | Ellen | | | 3/3 | | Famciclovir | 9939 | N/A | 2.5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Feces | 9939 | N/A | 2.5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Foscarnet | 9939 | N/A | 1.25 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Glucose | 9939 | N/A | 11 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Gynol II contraceptive jelly | 9939 | Nonoxynol-9 (3%) | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Human genomic DNA | 9939 | N/A | 20 μg/mL | 3/3 | K192376 - Page 10 of 16 {10} K192376 - Page 11 of 16 | Potentially Interfering Substance. | VZV Strain | Active Ingredient | Tested Concentration | # Detected/# Tested | | --- | --- | --- | --- | --- | | | Ellen | | | 3/3 | | Immunoglobulin | 9939 | N/A | 10 mg/mL | 3/3 | | | Ellen | | | 3/3 | | KY Jelly | 9939 | N/A | 10 mg/mL | 3/3 | | | Ellen | | 5% (w/v) | 3/3 | | Lactate | 9939 | N/A | 2.2 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Lanacane | 9939 | Benzethonium chloride, 0.2%; Benzocaine, 20% | 7% (v/v) | 3/3 | | | Ellen | | | 3/3 | | Lip-Clear Lysine | 9939 | Zinc Oxide, 1.2% | 7% (w/v) | 3/3 | | | Ellen | | 3.5% (w/v) | 3/3 | | Miconazole 1 | 9939 | Miconazole nitrate, 26% | 10% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Miconazole 3 | 9939 | Miconazole nitrate, 2% | 10% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Monistat 1 insert | 9939 | Miconazole nitrate, 1200 mg | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Monistat 3 cream | 9939 | Miconazole nitrate 2% | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Mouthwash (Listerine) | 9939 | Eucalyptol, 0.092%; Menthol, 0.042%; Methyl salicylate, 0.060%; Thymol, 0.064% | 7% (v/v) | 3/3 | | | Ellen | | | 3/3 | | Mucin | 9939 | N/A | 5% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Preparation H | 9939 | N/A | 10% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Releev | 9939 | N/A | 10% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Seminal Fluid | 9939 | N/A | 10% (v/v) | 3/3 | | | Ellen | | | 3/3 | | Tioconazole 1 | 9939 | N/A | 10% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Toothpaste (Colgate) | 9939 | Sodium fluoride, 0.243% | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Urine | 9939 | N/A | 10% (v/v) | 3/3 | | | Ellen | | | 3/3 | | Vagisil creme | 9939 | Benzocaine (5%), Resorcinol (2%) | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | | Valacyclovir | 9939 | N/A | 2.5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | Valgancyclovir | 9939 | N/A | 2.5 mg/mL | 3/3 | | | Ellen | | | 3/3 | | White blood cells | 9939 | N/A | 5.5x10^{7} cells/mL | 3/3 | | | Ellen | | | 3/3 | | Whole Blood in EDTA | 9939 | N/A | 10% (v/v) | 3/3 | | | Ellen | | | 3/3 | | YeastGard suppositories | 9939 | Candida albicans 27X HPU (Candida albicans), Candida parapsilosis 27X HPUS (Candida parapsilosis), Pulsatilla 27X HPUS (Meadow Anemone) | 7% (w/v) | 3/3 | | | Ellen | | | 3/3 | {11} 4. Assay Reportable Range: N/A 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): N/A 6. Detection Limit: The limit of detection (LoD) was determined for the Simplex VZV Swab Direct assay using quantified stocks of two (2) VZV strains (Ellen and 9939) in a pool of cutaneous and mucocutaneous lesion swab sample types UTM matrix. The LoD was determined to be the lowest concentration that could be detected positive $\geq 95\%$ of the time. The LoD for each strain is presented in Table 5. Table 5. Simplexa VZV Swab Direct Assay Limit of Detection | VZV Strain | LoD (TCID_{50}/mL) | LoD (Copies /mL) | | --- | --- | --- | | 993 | 0.77 | 800 | | Elle | 0.054 | 3500 | 7. Assay Cut-Off: N/A 8. Accuracy (Instrument): N/A 9. Carry-Over: The amplification carry-over for the Simplexa assays including the Simplexa VZV Swab Direct was assessed from the Simplexa Flu A/B & RSV Direct viral assay study. The study can be applied to the Simplexa VZV Swab Direct assay as the study is not analyte specific. In the Simplexa Flu A/B & RSV Direct, the amplification carry-over study was performed to determine the potential for contamination when negative samples are tested adjacent to strong positive samples. The study was designed by alternately placing high positive and negative samples on each disc. No evidence of carry-over contamination was observed. B Comparison Studies: 1. Method Comparison with Predicate Device: Clinical Studies K192376 - Page 12 of 16 {12} The performance of the Simplexa VZV Swab Direct assay was established in a clinical study that included three cohorts based on sample status. Specifically, prospective and retrospective cutaneous and mucocutaneous swab samples from human patients with signs and symptoms of VZV infection, as well as contrived samples, were tested in the clinical agreement study. Prospective Study: A total of four hundred fifty-two (452) cutaneous and mucocutaneous prospective specimens were collected from ten collection sites across the USA during the clinical study (November 2018 – May 2019). The specimens were taken from anorectal, genital, nasal, ocular, oral, skin and urethral locations of the body. The age of the patients ranged from one month to greater than 60 years of age. Of these specimens, $62.4\%$ of the specimens were from female patients and $37.6\%$ of the specimens were from male patients. Ten testing sites performed the Simplexa VZV Swab Direct assay on enrolled specimens and shipped the specimens to two comparator testing sites to test against a three part composite reference method (CRM). The three part CRM consisted of VZV direct stain fluorescent antibody (DSFA) and/or culture isolation with direct fluorescent antibody (DFA) and two validated VZV polymerase chain reaction (PCR) assays followed by bi-directional sequencing. The comparator testing was performed by two sites. One testing site performed the VZV DSFA and/or culture isolation with DFA and another testing site conducted the two validated VZV PCR assays testing followed by bi-directional sequencing. The results of the study are presented in Table 6. Table 6. Simplexa VZV Swab Direct Prospective Agreement Results | Prospective Study | Composite Reference Method (CRM) | | | | Total | Sensitivity 95% CIc | Specificity 95% CI | | --- | --- | --- | --- | --- | --- | --- | --- | | | + | | - | | | | | | | Simplexa VZV Swab Direct | | Simplexa VZV Swab Direct | | | | | | | + | - | + | - | | | | | Mucocutaneous | 7 | 1a | 0 | 171 | 179 | 87.5% (7/8)52.9% - 97.8% | 100.0% (171/171)97.8% - 100.0% | | Cutaneous | 79 | 1b | 3 | 162 | 245 | 98.8% (79/80)93.3% - 99.8% | 98.2% (162/165)94.8% - 99.4% | | Unknown | 1 | 0 | 0 | 27 | 28 | 100.0% (1/1)20.7% - 100.0% | 100.0% (27/27)87.5% - 100.0% | | All | 87 | 2 | 3 | 360 | 452 | 97.8% (87/89)92.2% -99.4% | 99.2% (360/363)97.6% - 99.7% | a The discordant negative mucocutaneous result is from an oral lesion sample. The sample was negative by the Simplexa VZV Swab Direct, DSFA/DFA and by the sites routine culture testing. The sample was positive by the two (2) PCR/Bi-directional sequencing assays. b The discordant negative cutaneous result is from a skin lesion sample. The sample was negative by the Simplexa VZV Swab Direct, DSFA/DFA testing. The sample was positive by the 2 PCR/Bi-directional sequencing assays. ${}^{\mathrm{c}}\mathrm{{CI}} =$ Confidence Interval. The 95% confidence intervals (CI) were calculated following Wilson Score method. Retrospective Study: A total of sixty (60) cutaneous and mucocutaneous retrospective positive swab specimens in UTM were blinded and randomized with one hundred twenty (120) negative masked specimens prior to being tested by Simplexa VZV Swab Direct assay. The Composite Reference Method 2 (CRM 2) utilized a two out of three outcomes from one FDA Cleared NAAT PCR assay for VZV and two validated VZV PCR assays followed by bi-directional sequencing. The FDA Cleared NAAT was performed by one external site. DiaSorin Molecular performed the Simplexa VZV Swab Direct testing and different DiaSorin Molecular operators K192376 - Page 13 of 16 {13} performed the two validated VZV PCR assays followed by bi-directional sequencing. The positive and negative percent agreement (PPA and NPA) results of the study are presented in Table 7. Table 7. Simplexa VZV Swab Direct Retrospective Agreement Results | Retrospective Study | Composite Reference Method 2 (CRM 2) | | | | Total | PPA 95% CIb | NPA 95% CI | | --- | --- | --- | --- | --- | --- | --- | --- | | | + | | - | | | | | | | Simplexa VZV Swab Direct | | Simplexa VZV Swab Direct | | | | | | | + | - | + | - | | | | | Mucocutaneous | 9 | 1a | 0 | 63 | 73 | 90.0% (9/10) 59.6% - 98.2% | 100.0% (63/63) 94.3% - 100.0% | | Cutaneous | 52 | 0 | 1 | 54 | 107 | 100.0% (52/52) 93.1% - 100.0% | 98.2% (54/55) 90.4% - 99.7% | | All | 61 | 1 | 1 | 117 | 180 | 98.4% (61/62) 91.4% - 99.7% | 99.2% (117/118) 95.4% - 99.9% | a The discordant negative mucocutaneous result is from an oral lesion sample. The sample was negative by the Simplexa VZV Swab Direct and NAAT testing. The sample was positive by the two (2) PCR/Bi-directional sequencing assays. $^{\mathrm{b}}$ CI = Confidence Interval. The 95% confidence intervals (CI) were calculated following Wilson Score method. Contrived Sample Study: A total of sixty (60) contrived positive specimens in individual negative UTM mucocutaneous swab matrix were blinded and randomized with sixty (60) masked negative UTM mucocutaneous specimens prior to being tested by Simplexa VZV Swab Direct assay. The results were compared with a two out of three outcomes from one FDA Cleared NAAT assay and two validated VZV PCR assays followed by bi-directional sequencing (Composite Reference Method 2 or CRM 2). Of the sixty (60) contrived specimens, thirty (30) were spiked with VZV Ellen strain and thirty (30) were spiked with VZV 9939 strain at different known concentrations across the detection range. The results are presented in Table 8. Table 8. Simplexa VZV Swab Direct Contrived Agreement Results | Contrived Mucocutaneous Samples | Composite Reference Method 2 (CRM 2) | | | | Total | PPA 95% CIa | NPA 95% CI | | --- | --- | --- | --- | --- | --- | --- | --- | | | + | | - | | | | | | | Simplexa™ VZV Swab Direct | | Simplexa™ VZV Swab Direct | | | | | | | + | - | + | - | | | | | Ellen | 30 | 0 | 0 | 0 | 30 | 100.0% (30/30) 88.6% - 100.0% | N/A | | 9939 | 30 | 0 | 0 | 0 | 30 | 100.0% (30/30) 88.6% - 100.0% | N/A | | Negative | 0 | 0 | 0 | 60 | 60 | N/A | 60/60 (100%) 94.0% -100.0% | | All | 60 | 0 | 0 | 60 | 120 | 100.0% (60/60) 94.0% - 100.0% | 100.0% (60/60) 94.0% - 100.0% | $^{\mathrm{a}}$ CI = Confidence Interval. The 95% confidence intervals (CI) were calculated following Wilson Score method. K192376 - Page 14 of 16 {14} 2. Matrix Comparison: N/A # C Clinical Studies: 1. Clinical Sensitivity: N/A 2. Clinical Specificity: N/A 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): N/A # D Clinical Cut-Off: N/A # E Expected Values/Reference Range: In the prospective clinical study 452 cutaneous and mucocutaneous samples were tested by Simplexa VZV Swab Direct. The values for VZV detection per site varied between $0\%$ and $75\%$ . The overall expected value for all sites is $19.9\%$ . The expected values per site are presented in Table 9 below. Table 9. VZV Incidence as Determined by Simplexa VZV Swab Direct | Site ID | Total | # Detected | VZV Incidence % Detected | | --- | --- | --- | --- | | 3 | 4 | 3 | 75.0% | | 5 | 61 | 24 | 39.9% | | 6 | 48 | 15 | 31.3% | | 7 | 38 | 13 | 34.2% | | 8 | 12 | 3 | 25.0% | | 9 | 108 | 1 | 0.9% | | 10 | 139 | 16 | 11.5% | | 11 | 28 | 15 | 53.6% | | 13 | 14 | 0 | 0.0% | | Total | 452 | 90 | 19.9% | # F Other Supportive Instrument Performance Characteristics Data: N/A K192376 - Page 15 of 16 {15} VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K192376 - Page 16 of 16
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...