Browse hierarchy Microbiology (MI) Subpart D — Serological Reagents 21 CFR 866.3309 Product Code PGI K180559 — HSV 1 & 2 ELITe MGB Kit; ELITe InGenius
HSV 1 & 2 ELITe MGB Kit; ELITe InGenius
K180559 · Elitechgroup · PGI · Oct 29, 2018 · Microbiology
Device Facts
Record ID K180559
Device Name HSV 1 & 2 ELITe MGB Kit; ELITe InGenius
Applicant Elitechgroup
Product Code PGI · Microbiology
Decision Date Oct 29, 2018
Decision SESE
Submission Type Traditional
Regulation 21 CFR 866.3309
Device Class Class 2
Indications for Use
The HSV 1&2 ELITe MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-1 and HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-1 or HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections. The HSV 1&2 ELITe MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products.
Device Story
The HSV 1&2 ELITe MGB Assay is a qualitative in vitro diagnostic test performed on the ELITe InGenius system. It processes swab specimens from cutaneous or mucocutaneous lesions. The system automates nucleic acid extraction using the ELITe InGenius SP 200 Extraction Cartridge, followed by real-time PCR amplification of HSV-1 (glycoprotein D gene) and HSV-2 (glycoprotein G gene) DNA. An internal control monitors for inhibition. The device provides qualitative results to clinicians to aid in the differential diagnosis of HSV-1 and HSV-2 infections. It is intended for use in clinical laboratory settings by trained personnel. The output assists healthcare providers in confirming infection status, which informs clinical management and treatment decisions for patients presenting with herpetic lesions.
Clinical Evidence
Clinical performance was evaluated using 1,174 prospectively collected archived swab samples compared to a composite reference method (FDA-cleared PCR and sequencing). For HSV-1, PPA was 98.7-99.2% and NPA was 97.6-98.5%. For HSV-2, PPA was 96.2-97.6% and NPA was 98.2-98.6%. Reproducibility was assessed across 3 sites with >630 samples, showing high agreement. Analytical studies included LoD, inclusivity, cross-reactivity, and interference testing.
Technological Characteristics
Qualitative real-time PCR assay. Targets HSV-1 glycoprotein D and HSV-2 glycoprotein G genes. Uses fluorescent-labeled oligonucleotide probes. Automated extraction and amplification on the ELITe InGenius system (1-12 parallel tracks). Benchtop form factor. Software-controlled processing. Internal control (randomized DNA sequence) included. Compatible with UTM, M4, M4RT, M5, and M6 transport media.
Indications for Use
Indicated for symptomatic male and female patients with cutaneous or mucocutaneous lesions to detect and differentiate HSV-1 and HSV-2 DNA. Not for CSF, prenatal screening, or blood/blood product screening.
Regulatory Classification
Identification A herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel is a qualitative in vitro diagnostic device intended for the simultaneous detection and differentiation of different herpes viruses in cutaneous and mucocutaneous lesion samples from symptomatic patients suspected of Herpetic infections. Negative results do not preclude infection and should not be used as the sole basis for treatment or other patient management decisions. The assay is not intended for use in cerebrospinal fluid samples.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer design and selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation of a clinical study using lesion samples in which Herpes Simplex Virus 1, Herpes Simplex Virus 2, or Varicella Zoster Virus DNA detection was requested. The study must compare the device performance to an appropriate well established reference method.
(4) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(5) The device labeling must include a limitation statement that reads: “The device is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS).”
(6) Premarket notification submissions must include quality assurance protocols and a detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(7) The risk management activities performed as part of the manufacturer's 21 CFR 820.10(c) design and development activities must document an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.
Predicate Devices
Related Devices
K150617 — cobas HSV 1 and 2 Test · Roche Molecular Systems, Inc. · Jun 1, 2015
K173798 — Simplexa HSV 1 & 2 Direct, Simplexa HSV 1 & 2 Positive Control Pack · Diasorin Molecular, LLC · Mar 14, 2018
K142738 — artus HSV-1/2 QS-RGQ MDx Kit · Qiagen · Dec 19, 2014
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K180559
B. Purpose for Submission:
Clearance of New Device.
C. Measurand:
Target DNA Sequences from conserved regions of Herpes Simplex Virus Type 1 (HSV-1) and Herpes Simplex Virus Type 2 (HSV-2)
D. Type of Test:
Qualitative Real-Time PCR Assay
E. Applicant:
ELITechGroup Inc. Molecular Diagnostics
F. Proprietary and Established Names:
HSV 1&2 ELITe MBG Assay
G. Regulatory Information:
1. Regulation section:
21CFR 866.3309
2. Classification:
Class II
3. Product code:
PGI
4. Panel:
83 - Microbiology
{1}
H. Intended Use:
1. Intended use(s):
The HSV 1&2 ELITe MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-1 and HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-1 or HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections.
The HSV 1&2 ELITe MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products.
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
The HSV 1&2 ELITe MGB Assay is indicated for use on the ELITe InGenius ® System.
I. Device Description:
The HSV 1&2 ELITe MGB Assay is a multiplexed qualitative in vitro diagnostic Real-Time PCR Assay that uses unique primer sets and single uniquely labeled probes to amplify and detect:
- The Herpes Simplex Virus (HSV) genotype 1; HSV-1 glycoprotein D encoding gene,
- The HSV-2 glycoprotein G encoding gene, and
- An Internal Control.
Intended for the direct detection of HSV DNA in symptomatic male and female patients using DNA purified from swab specimens collected from cutaneous or mucocutaneous lesions from individuals with herpetic lesions.
The HSV 1&2 ELITe MGB Assay is used on the ELITe InGenius system. The ELITe InGenius system is a bench top instrument integrating all required hardware, reagents and software components to perform nucleic acid sample preparation and real-time PCR operations.
The ELITe InGenius system can process 1 to 12 samples in 12 parallel tracks, and samples may be loaded in primary tubes or in secondary tubes provided. The system utilizes a universal, cassette-based process for sample extraction and allows for multiple and independent PCRs to be performed from a single nucleic acid eluate. The unused eluates can be saved for future retesting or archiving.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
ARIES® HSV 1&2 Assay
2. Predicate 510(k) number(s):
K151906
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications For Use | The HSV 1&2 ELITE MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-1 and HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-1 or HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections. The HSV 1&2 ELITE MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products. | The ARIES® HSV 1&2 Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV 1 and HSV 2) DNA in cutaneous or mucocutaneous lesion specimens from symptomatic patients. The test is indicated for use as an aid in diagnosis of HSV infection in symptomatic patients. The ARIES® HSV 1&2 Assay is indicated for use on the ARIES® System. WARNING: The ARIES® HSV 1&2 Assay is not FDA cleared for use with cerebrospinal fluid (CSF). The assay is not intended to be used for prenatal screening. |
| Extraction Technology | Same as predicate | Automated |
| Amplification Technology | Same as predicate | Qualitative Real-time PCR |
| Samples Type | Same as predicate | Male and female cutaneous and mucocutaneous lesion swab samples |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Targets | DNA sequences from HSV-1 glycoprotein D gene and HSV-2 glycoprotein G gene. | DNA sequences from Herpes Simplex Virus type 1 (HSV-1) and Herpes Simplex Virus type 2 (HSV-2)- different target |
| Detection Technology | Multiplex assay with paired reporter and quencher fluorescence labeled probes and different reporter dyes for each target. Measures increase in assay fluorescence with each PCR cycle. | Pairs fluorescent-labeled primers with quencher labeled nucleotides. Measures decrease in assay fluorescence with each PCR cycle. |
K. Standard/Guidance Document Referenced (if applicable):
1. Evaluating Substantial Equivalence in Premarket Notifications 510(k)
2. Statistical Guidance on Reporting Results from studies evaluating diagnostic tests
3. Guidance Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (DRAFT 10-2-14)
4. Guidance for Clinical Investigators, Sponsors, and IRBs - Adverse Event Reporting to IRBs - Improving Human Subject Protection
5. Guidance for Evaluation and Reporting of Age-, Race-, and Ethnicity-Specific Data in Medical Device Clinical Studies (1500626) (09-12-17)
6. Guidance for Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable
7. Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
8. Guidance for Refuse to Accept Policy for 510(k)s
9. EP05-A3 Vol. 34 No. 13 - Evaluation of Precision Performance of Quantitative
10. Measurement Methods; Approved Guideline-Third Edition
11. EP12-A2 Vol. 28 No. 3 - User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition
12. EP17-A2 Vol. 32 No. 8 - Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition
13. EP24-A2 Vol. 31 No. 23 - Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Approved Guideline - Second Edition
14. EP25-A Vol. 29 No. 20 - Evaluation of Stability of in Vitro Diagnostic Reagents; Approved Guideline
15. M29-A4 Vol. 34 No. 8 - Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline - 4th Edition
16. MM03 - Molecular Diagnostic Methods for Infectious Diseases; Approved Guideline, 3rd Edition
{4}
17. MM06-A2 Vol. 30 No. 22 - Quantitative Molecular Methods for Infectious Diseases; Approved Guideline Second Edition
18. MM09-A2 Vol. 34 No.4 - Nucleic Acid Sequencing Methods in Diagnostic Laboratory Medicine; Approved Guideline
19. MM13-A Vol. 25 No. 31 - Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods; Approved Guideline
20. MM17 Ed 2 Vol. 38 No. 9 - Verification and Validation of Multiplex Nucleic Acid Assays (NATS); Approved Guideline
# L. Test Principle:
The HSV 1&2 ELITe MGB Assay system is comprised of three major processes:
(1) automated preparation of unprocessed sample to extract nucleic acids from primary swab specimens using the ELITe InGenius SP 200 Extraction Cartridge,
(2) PCR amplification of target DNA sequences using HSV-1 and HSV-2 specific primers,
(3) real-time detection of fluorescent-labeled HSV-1 and HSV-2 specific oligonucleotide detection probes.
An Internal Control (IC), containing unrelated randomized DNA sequence, is added to all samples prior to extraction and monitors the integrity of the reagents, equipment function, and the presence of inhibitors in the samples. A positive signal in the Internal Control channel in the absence of HSV DNA indicates that the PCR has not been inhibited.
The amplification reagents, Positive Control and Internal Control are packaged as part of the HSV 1&2 ELITe MGB Assay.
The following table shows the result interpretation:
Summary of the Run Validity Criteria and Result Interpretation Algorithm:
| HSV-1 CT value AP593, Channel 4 | HSV-2 CT value FAM, Channel 1 | IC CT value AP525, Channel 2 | Status | HSV-1 Result | HSV-2 Result |
| --- | --- | --- | --- | --- | --- |
| Undetermined CT≥45.0 | Undetermined CT>45.0 | CT≤32.0 | Valid | Negative | Negative |
| | | Undetermined or CT>32.0 | Invalid | Invalid | Invalid |
| Determined CT≤45.0 | Undetermined CT>45.0 | NA | Valid | Positive | Negative |
| | Determined CT≤45.0 | NA | | Positive | Positive |
| Undetermined CT>45.0 | Determined CT≤45.0 | NA | Valid | Negative | Positive |
The Sample run is valid when the conditions reported in the table below are met.
{5}
6
| 1) HSV 1&2 ELITE MGB Positive Control | Status |
| --- | --- |
| HSV 1&2 ELITE MGB Positive Control | APPROVED |
| 2) HSV 1&2 ELITE MGB Negative Control | Status |
| HSV 1&2 ELITE MGB Negative Control | APPROVED |
ELITE InGenius software reports as target DNA being “Detected” or “Not Detected”. For a valid run, the specimen results are interpreted as follows.
ELITE InGenius Software Output and Interpretation
| Results of Sample Run | | Interpretation |
| --- | --- | --- |
| HSV1 Result | HSV2 Result | |
| HSV-1 DNA Not Detected | HSV-2 DNA Not Detected | HSV-1 and HSV-2 negative |
| HSV-1 DNA Not Detected | HSV-2 DNA Detected | HSV-1 negative and HSV-2 positive |
| HSV-1 DNA Detected | HSV-2 DNA Not Detected | HSV-1 positive and HSV-2 negative |
| HSV-1 DNA Detected | HSV-2 DNA Detected | HSV-1 positive and HSV-2 positive |
| Invalid - Retest Sample | | Not valid assay result due to Internal Control failure (Incorrect extraction or presence of inhibitor). Specimen Should be retested |
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The reproducibility of the HSV 1&2 ELITE MGB Assay was evaluated in a multi-site investigation using contrived clinical samples. HSV test panels were prepared by spiking HSV-1 (MacIntyre strain) or HSV-2 (MS strain) virus into Universal transport media (UTM) at the concentrations of $<1 \times \mathrm{LoD}$, $1 \times \mathrm{LoD}$ and $3 \times \mathrm{LoD}$. HSV-1 and HSV-2 negative panel members were included as panel member controls. The reproducibility panel composition is shown in the table below:
| Name | Description of Contents | Viral Load | Expected Positivity Rate |
| --- | --- | --- | --- |
| M1 | HSV-1 C50 (High Negative) in UTM | <1× LOD | 20-80% positive |
| M2 | HSV-1 C95 (Low Positive) in UTM | 1× LOD | ≥95% positive |
| M3 | HSV-1 C100 (Moderate Positive) in UTM | 2-3 × LOD | 100% positive |
{6}
Panels were tested at 3 sites by 2 operators per site with 1 run per operator per day, for 10 non-consecutive days using a single lot of HSV 1&2 ELITe MGB Assay. Testing was performed on a minimum of 90 (30 per site) replicates per panel member. Lot-to-Lot variability was assessed only at EGI MDx using three lots of HSV 1&2 ELITe MGB Assay. Controls were run daily and were included in the first run of the day.
The results are presented in the following table.
| Target | Sample | Site - 1 | | | Site - 2 | | | Site - 3 | | | % Agree-ment with Expected Results | 95% CI |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | % Agree-ment with Expected Results | Avg. Ct | Total %CV | %Agree-ment with Expected Results | Avg. Ct | Total %CV | %Agree-ment with Expected Results | Avg. Ct | Total %CV | | |
| HSV-1 Result | HSV-1 Low Pos | 100.0% (30/30) | 38.9 | 1.70% | 100.0% (30/30) | 38.3 | 2.10% | 100.0% (30/30) | 38 | 2.00% | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV-1 Mod Pos | 100.0% (30/30) | 36.4 | 1.30% | 100.0% (30/30) | 35.5 | 5.20% | 100.0% (30/30) | 35.6 | 1.50% | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV-2 Low Pos | 100.0% (30/30)* | NA | NA | 100.0% (29/29)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (89/89) | 95.6 to 100.0% |
| | HSV-2 Mod Pos | 100.0% (30/30)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV Neg | 100.0% (60/60) | NA | NA | 100.0% (38/38) | 41.4 | 2.50% | 100.0% (40/40) | NA | NA | 100.0% (138/138) | 97.3 to 100.0% |
| | Pos Control | 100.0% (30/30) | 27.5 | 1.30% | 100.0% (5/5) | 27.5 | 1.20% | 100.0% (5/5) | 27 | 0.80% | 100.0% (40/40) | 91.2 to 100.0% |
| | Total Agreement | 100.0% (210/210) | | | 100.0% (162/162) | | | 100.0% (165/165) | | | 100.0% (537/537) | 98.2 to 100.0% |
{7}
8
| a) Expected Results of HSV-2 Low Positive, HSV-2 Moderate Positive and HSV Negative samples are “Negative” for HSV-1. | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Target | Sample | Site - 1 | | | Site - 2 | | | Site - 3 | | | % Agreement with Expected Results | 95% CI |
| | | % Agreement with Expected Results | Avg. Ct | Total %CV | % Agreement with Expected Results | Avg. Ct | Total %CV | % Agreement with Expected Results | Avg. Ct | Total %CV | | |
| HSV-2 Result | HSV-1 Low Pos | 100.0% (30/30)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV-1 Mod Pos | 100.0% (30/30)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (30/30)* | NA | NA | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV-2 Low Pos | 100.0% (30/30) | 36.8 | 3.10% | 100.0% (29/29) | 37.8 | 2.30% | 100.0% (30/30) | 36.6 | 1.90% | 100.0% (89/89) | 95.9 to 100.0% |
| | HSV-2 Mod Pos | 100.0% (30/30) | 35.2 | 1.30% | 100.0% (30/30) | 35.9 | 1.60% | 100.0% (30/30) | 34.6 | 2.30% | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV Neg | 100.0% (60/60)* | NA | NA | 100.0% (38/38)* | NA | NA | 100.0% (40/40)* | NA | NA | 100.0% (138/138) | 95.9 to 100.0% |
| | Pos Control | 100.0% (30/30) | 27 | 1.30% | 100.0% (5/5) | 27.4 | 1.50% | 100.0% (5/5) | 26.8 | 1.40% | 100.0% (40/40) | 95.9 to 100.0% |
| | Total Agreement | 100.0% (210/210) | | | 100.0% (162/162) | | | 100.0% (165/165) | | | 100.0% (537/537) | 95.9 to 100.0% |
| b) Expected Results of HSV-1 Low Positive, HSV-1 Moderate Positive and HSV Negative samples are “Negative” for HSV-2. | | | | | | | | | | | | |
| Target | Sample | Site - 1 | | | Site - 2 | | | Site - 3 | | | % Agreement with Expected Results | 95% CI |
| | | % Agreement with Expected Results | Avg. Ct | Total %CV | % Agreement with Expected Results | Avg. Ct | Total %CV | % Agreement with Expected Results | Avg. Ct | Total %CV | | |
| IC Result | HSV-1 Low Pos | 100.0% (30/30) | 30.4 | 3.80% | 100.0% (30/30) | 30.3 | 2.00% | 100.0% (30/30) | 30.2 | 1.70% | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV-1 Mod Pos | 100.0% (30/30) | 30.2 | 2.30% | 100.0% (30/30) | 30.4 | 2.80% | 100.0% (30/30) | 30.1 | 0.90% | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV-2 Low Pos | 100.0% (30/30) | 29.9 | 0.50% | 100.0% (29/29) | 30.4 | 2.20% | 100.0% (30/30) | 30.2 | 0.60% | 100.0% (89/89) | 95.9 to 100.0% |
| | HSV-2 Mod Pos | 100.0% (30/30) | 29.7 | 0.80% | 100.0% (30/30) | 30.4 | 1.20% | 100.0% (30/30) | 30.1 | 0.60% | 100.0% (90/90) | 95.9 to 100.0% |
| | HSV Neg | 100.0% (60/60) | 30.2 | 1.10% | 100.0% (40/40) | 30.2 | 1.90% | 100.0% (38/38) | 30.1 | 0.90% | 100.0% (138/138) | 97.3 to 100.0% |
| | Pos Control | 100.0% (30/30) | 29.3 | 1.40% | 100.0% (5/5) | 30.2 | 2.10% | 100.0% (5/5) | 29.4 | 0.90% | 100.0% (40/40) | 91.2 to 100.0% |
| | Total Agreement | 100.0% (210/210) | | | 100.0% (164/164) | | | 100.0% (163/163) | | | 100.0% (537/537) | 98.2 to 100.0% |
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The stability of the HSV 1&2 ELITe MGB Assay was evaluated in a real-time study
{8}
and the data support a shelf-life claim of 10 months for the HSV 1&2 ELITe MGB Assay. A reagent freeze thaw stability study was conducted with 3 reagent lots and demonstrated that the HSV 1&2 ELITe MGB Assay is stable when subjected to 8 freeze-thaw cycles.
# d. Detection limit:
The limit of detection (LoD) of the HSV 1&2 ELITe MGB Assay was determined using 4 HSV strains (two for each target). Quantitated viral strains were obtained and serially diluted into HSV negative pooled human oral cheek matrix in UTM. All dilutions were tested, and the LoD was determined using Probit (Logit) Data Analysis software (Analyse-it for Microsoft Excel v4.80.2, Logistic Function model). LoD for each strain represents the lowest viral titer in $\mathrm{TCID}_{50} / \mathrm{mL}$ at which a positive result can be detected at $\geq 95\%$ confidence. The LoD for each strain was then verified by testing 20 replicates. The results are presented in the following table.
Limit of Detection Results
| Organism | Isolate/Strain | Cell Line | Qualitative results #detected/Total | Mean CT±SD from detected replicates | 1×LoD TCID50/mL |
| --- | --- | --- | --- | --- | --- |
| HSV-1 | MacIntyre strain | Vero | 20/20 | 37.91 ± 0.69 | 59.0 TCID50/mL |
| HSV-1 | Isolate #15 (Zeptometrix) | Vero | 20/20 | 39.94 ± 0.95 | 1.5 TCID50/mL |
| HSV-2 | MS strain | Vero | 20/20 | 37.90 ± 0.92 | 5.4 TCID50/mL |
| HSV-2 | Isolate #2 (Zeptometrix) | Vero | 20/20 | 38.67 ± 1.03 | 0.3 TCID50/mL |
A swab elution efficiency study was also performed by using the same HSV negative pooled human cheek swab matrix and the HSV-1 MacIntyre strain used to verify the LoB. This study showed $100\%$ elution efficiency from the Copan regular flocked swab compared with the same volume of material directly spiked into UTM.
# e. Analytical Reactivity:
The analytical reactivity (Inclusivity) of the HSV 1&2 ELITe MGB Assay was tested on 44 well characterized commercially available HSV-1 and HSV-2 isolates. All strains were tested with the HSV 1&2 ELITE MGB Assay as spiked samples in UTM prepared at close to LoD (3xLOD) level. For HSV-1 and HSV-2 viral isolates not detected (negative) at $3 \times$ LoD concentrations, $2 \times$ incremental concentrations were tested, the lowest detectable level was determined, and the final test concentration was reported. All tested HSV-1 and HSV-2 isolates were detected by the HSV 1&2 ELITe MGB Assay at concentrations of $16.2 - 354 \mathrm{TCID} 50 / \mathrm{mL}$ .
{9}
| # | Isolate | Estimated 1×LoD (TCID50/mL) | ×LoD Tested | Final Test Conc. (TCID50/mL) | Positivity |
| --- | --- | --- | --- | --- | --- |
| 1 | HSV-1 MacIntyre Strain | 59 | 3× | 177 | 3/3 |
| 2 | HSV-1 Isolate #1 | 59 | 3× | 177 | 3/3 |
| 3 | HSV-1 Isolate #2 | 59 | 3× | 177 | 3/3 |
| 4 | HSV-1 Isolate #3 | 59 | 3× | 177 | 3/3 |
| 5 | HSV-1 Isolate #4 | 59 | 3× | 177 | 3/3 |
| 6 | HSV-1 Isolate #5 | 59 | 3× | 177 | 3/3 |
| 7 | HSV-1 Isolate #6 | 59 | 3× | 177 | 0/3 |
| | | 59 | 6× | 354 | 3/3 |
| 8 | HSV-1 Isolate #7 | 59 | 3× | 177 | 3/3 |
| 9 | HSV-1 Isolate #8 | 59 | 3× | 177 | 3/3 |
| 10 | HSV-1 Isolate #9 | 59 | 3× | 177 | 3/3 |
| 11 | HSV-1 Isolate #10 | 59 | 3× | 177 | 3/3 |
| 12 | HSV-1 Isolate #11 | 59 | 3× | 177 | 3/3 |
| 13 | HSV-1 Isolate #12 | 59 | 3× | 177 | 3/3 |
| 14 | HSV-1 Isolate #13 | 59 | 3× | 177 | 3/3 |
| 15 | HSV-1 Isolate #14 | 59 | 3× | 177 | 3/3 |
| 16 | HSV-1 Isolate #15 | 59 | 3× | 177 | 3/3 |
| 17 | HSV-1 Isolate #16 | 59 | 3× | 177 | 3/3 |
| 18 | HSV-1 Isolate #17 | 59 | 3× | 177 | 3/3 |
| 19 | HSV-1 Isolate #18 | 59 | 3× | 177 | 3/3 |
| 20 | HSV-1 Isolate #19 | 59 | 3× | 177 | 3/3 |
| 21 | HSV-1 Isolate #20 | 59 | 3× | 177 | 0/3 |
| | | 59 | 6× | 354 | 3/3 |
| 22 | HSV-1 Isolate #21 | 59 | 3× | 177 | 3/3 |
| 23 | HSV-2 MS Strain | 5.4 | 3× | 16.2 | 3/3 |
| 24 | HSV-2 Isolate #1 | 5.4 | 3× | 16.2 | 3/3 |
| 25 | HSV-2 Isolate #2 | 5.4 | 3× | 16.2 | 3/3 |
| 26 | HSV-2 Isolate #3 | 5.4 | 3× | 16.2 | 3/3 |
| 27 | HSV-2 Isolate #4 | 5.4 | 3× | 16.2 | 3/3 |
| 28 | HSV-2 Isolate #5 | 5.4 | 3× | 16.2 | 3/3 |
| 29 | HSV-2 Isolate #6 | 5.4 | 3× | 16.2 | 3/3 |
| 30 | HSV-2 Isolate #7 | 5.4 | 3× | 16.2 | 3/3 |
| 31 | HSV-2 Isolate #8 | 5.4 | 3× | 16.2 | 2/3 |
| | | 5.4 | 3× | 16.2 | 3/3 |
| 32 | HSV-2 Isolate #9 | 5.4 | 3× | 16.2 | 3/3 |
| 33 | HSV-2 Isolate #10 | 5.4 | 3× | 16.2 | 3/3 |
| 34 | HSV-2 Isolate #11 | 5.4 | 3× | 16.2 | 2/3 |
| | | 5.4 | 6× | 32.4 | 3/3 |
| 35 | HSV-2 Isolate #12 | 5.4 | 3× | 16.2 | 1/3 |
| | | 5.4 | 6× | 32.4 | 3/3 |
| 36 | HSV-2 Isolate #13 | 5.4 | 3× | 16.2 | 0/3 |
{10}
| # | Isolate | Estimated 1×LoD (TCID50/mL) | ×LoD Tested | Final Test Conc. (TCID50/mL) | Positivity |
| --- | --- | --- | --- | --- | --- |
| | | 5.4 | 6× | 32.4 | 2/3 |
| | | 5.4 | 12× | 64.8 | 2/3 |
| | | 5.4 | 24× | 129.6 | 3/3 |
| 37 | HSV-2 Isolate #14 | 5.4 | 3× | 16.2 | 1/3 |
| | | 5.4 | 6× | 32.4 | 3/3 |
| 38 | HSV-2 Isolate #15 | 5.4 | 3× | 16.2 | 0/3 |
| | | 5.4 | 6× | 32.4 | 3/3 |
| 39 | HSV-2 Isolate #16 | 5.4 | 3× | 16.2 | 1/3 |
| | | 5.4 | 6× | 32.4 | 3/3 |
| 40 | HSV-2 Isolate #17 | 5.4 | 3× | 16.2 | 1/3 |
| | | 5.4 | 6× | 32.4 | 1/3 |
| | | 5.4 | 12× | 64.8 | 3/3 |
| 41 | HSV-2 Isolate #18 | 5.4 | 3× | 16.2 | 3/3 |
| 42 | HSV-2 Isolate #19 | 5.4 | 3× | 16.2 | 2/3 |
| | | 5.4 | 6× | 32.4 | 1/3 |
| | | 5.4 | 12× | 64.8 | 3/3 |
| 43 | HSV-2 Isolate #20 | 5.4 | 3× | 16.2 | 0/3 |
| | | 5.4 | 6× | 32.4 | 1/3 |
| | | 5.4 | 12× | 64.8 | 3/3 |
| 44 | HSV-2 Isolate #21 | 5.4 | 3× | 16.2 | 3/3 |
# f. Analytical specificity:
# Cross-Reactivity
Potential cross-reactivity of the HSV 1&2 ELITe MGB Assay was evaluated by testing 49 different microbial species that are either genetically related to HSV1 and HSV2 or cause similar clinical symptoms or may be present in the cutaneous and mucocutaneous sites tested by this device. For each of the 49 potential cross reactants, the sample to be tested was prepared from quantified stock diluted to the required concentration using UTM. The potential cross reactants were tested, the concentrations were evaluated and the results are presented in the following table:
| No. | Potential Cross-Reactants | Tested Concentration | Qualitative Result (#Detected/#Total) | |
| --- | --- | --- | --- | --- |
| | | | HSV-1 | HSV-2 |
| 1 | Acinetobacter calcoaceticus | 1×106CFU/mL | 0/3 | 0/3 |
| 2 | Acinetobacter lwoffii | 1×106CFU/mL | 0/3 | 0/3 |
| 3 | Adenovirus type 2 | 1×105TCID50/mL | 0/3 | 0/3 |
| 4 | Bacteroides fragilis | 1×106CFU/mL | 0/3 | 0/3 |
| 5 | Candida albicans | 1×106CFU/mL | 0/3 | 0/3 |
| 6 | Candida glabrata | 1×106CFU/mL | 0/3 | 0/3 |
| 7 | Candida parapsilosis | 1×106CFU/mL | 0/3 | 0/3 |
| 8 | Candida parapsilosis | 1×106CFU/mL | 0/3 | 0/3 |
| 9 | C. c. c. c. c. | 1×106CFU/mL | 0/3 | 0/3 |
| 10 | C. c. c. c. c. | 1×106CFU/mL | 0/3 | 0/3 |
{11}
| No. | Potential Cross-Reactants | Tested Concentration | Qualitative Result (#Detected/#Total) | |
| --- | --- | --- | --- | --- |
| | | | HSV-1 | HSV-2 |
| 6 | Candida glabrata | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 7 | Candida guilliermondii | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 8 | Candida krusei | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 9 | Candida lusitaniae | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 10 | Candida parapsilosis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 11 | Candida tropicalis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 12 | Chlamydia trachomatis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 13 | Cytomegalovirus | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 14 | Enterobacter cloacae | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 15 | Enterovirus | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 16 | Epstein-Barr Virus | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 17 | Escherichia coli | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 18 | Fusobacterium nucleatum | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 19 | Gardnerella vaginalis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 20 | Haemophilus ducreyi | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 21 | Human Genomic DNA | 500 ng/swab | 0/3 | 0/3 |
| 22 | Human Herpes Virus 6 | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 23 | Human Herpes Virus 7 | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 24 | Human papilloma virus 16 | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 25 | Human papilloma virus 18 | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 26 | Herpes Simplex Virus 1 (HSV-1), isolate 20, ZMC | 1×10^5 TCID50/mL | 3/3 | 0/3 |
| 27 | Herpes Simplex Virus 2 (HSV-2), isolate 20. ZMC | 1×10^5 TCID50/mL | 0/3 | 3/3 |
| 28 | Klebsiella pneumoniae | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 29 | Lactobacillus acidophilus | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 30 | Mobiluncus curtsii | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 31 | Mobiluncus mulieris | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 32 | Moraxella catarrhalis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 33 | Mycoplasma hominis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 34 | Neisseria gonorrhea | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 35 | Neisseria meningitides | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 36 | Prevotella melaninogenica | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 37 | Rubella Virus | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 38 | Staphylococcus aureus (MSSA) | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 39 | Staphylococcus epidermidis (MRSE) | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 44 | Staphylococcus saprophyticus | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 41 | Streptococcus mitis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 42 | Streptococcus mutans | 1×10^6 CFU/mL | 0/3 | 0/3 |
{12}
| No. | Potential Cross-Reactants | Tested Concentration | Qualitative Result (#Detected/#Total) | |
| --- | --- | --- | --- | --- |
| | | | HSV-1 | HSV-2 |
| 43 | Streptococcus pneumoniae | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 44 | Streptococcus pyogenes | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 45 | Streptococcus salivarius | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 46 | Toxoplasma gondii | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 47 | Trichomonas vaginalis | 1×10^6 CFU/mL | 0/3 | 0/3 |
| 48 | Varicella-Zoster Virus (VZV) | 1×10^5 TCID50/mL | 0/3 | 0/3 |
| 49 | Chlamydophila pneumoniae | 1×10^6 CFU/mL | 0/3 | 0/3 |
## Microbial Interference
The microbial interference was evaluated in the presence of either HSV-1 or HSV-2 spiked at $3 \times \mathrm{LoD}$ in UTM and the 49 organisms indicated in the cross reactivity section above. Each microorganism was tested either at $1 \times 10^{6} \mathrm{CFU} / \mathrm{mL}$ or higher for bacterial isolates, or at $1 \times 10^{5} \mathrm{TCID} 50 / \mathrm{mL}$ or higher for viruses. None of the non-target organisms that are reasonably expected to be found in cutaneous and mucocutaneous swab samples interfered with the detection of HSV-1 or HSV-2 species. The only exception is that amplification of HSV-1 is completely inhibited in the presence of HSV-2 titers of $1 \times 10^{3}$ or higher. This observed HSV-2 interference is reported as a limitation in the device package insert.
## Competitive Interference
Competitive interference was studied to evaluate the effects of possible clinically relevant co-infection with both HSV-1 and HSV-2 using HSV 1&2 ELITe MGB Assay.
The study assessed whether a high concentration of one virus in the sample could potentially affect the HSV 1&2 ELITe MGB Assay performance for another target present at low levels. A low positive sample was contrived at approximately $3 \times \mathrm{LoD}$ for each target (HSV-1 McIntryre strain and HSV-2 MS strain), and a baseline Ct was determined for each sample. Each potential concomitant infecting virus was spiked into the low level sample and assayed in triplicate.
Competitive interference of HSV-1 with HSV-2 was observed at HSV-2 titers of $1 \times 10^{3}$, $1 \times 10^{4}$, $1 \times 10^{5} \mathrm{TCID}_{50} / \mathrm{mL}$. This observed interference information is included as a limitation in the device package insert.
No competitive interference of HSV-2 with HSV-1 at the levels tested was observed. The results of the testing are shown in the table below.
{13}
Competitive Interference of HSV-1 and HSV-2 targets at unequal concentrations
| Baseline (Low Level) | | Competitive Interferent (High Concentration) | | Qualitative Results (#Detected/#Total) | |
| --- | --- | --- | --- | --- | --- |
| Strain | Concentration (TCID_{50}/mL) | Strain | Concentration (TCID_{50}/mL) | HSV-1 | HSV-2 |
| HSV-1 McIntyre | 177 | HSV-2 MS | 100000 | 0/3 | 3/3 |
| HSV-1 McIntyre | 177 | HSV-2 MS | 10000 | 1/3 | 3/3 |
| HSV-1 McIntyre | 177 | HSV-2 MS | 1000 | 1/3 | 3/3 |
| HSV-1 McIntyre | 177 | HSV-2 MS | 100 | 3/3 | 3/3 |
| HSV-1 McIntyre | 177 | HSV-2 MS | 0 | 3/3 | 0/3 |
| HSV-2 MS | 16.2 | HSV-1 McIntyre | 100000 | 3/3 | 3/3 |
| HSV-2 MS | 16.2 | HSV-1 McIntyre | 10000 | 3/3 | 3/3 |
| HSV-2 MS | 16.2 | HSV-1 McIntyre | 1000 | 3/3 | 3/3 |
| HSV-2 MS | 16.2 | HSV-1 McIntyre | 100 | 3/3 | 3/3 |
| HSV-2 MS | 16.2 | HSV-1 McIntyre | 0 | 0/3 | 3/3 |
Additionally, in a separate study both strains were tested at similar or equal concentrations of $3 \times \mathrm{LoD}$, $1 \times 10^{3}$ and $1 \times 10^{5}$, and no competitive interference was observed.
Competitive Interference of HSV-1 & HSV-2 targets at equal concentrations
| HSV-1 Concentration | | HSV-2 Concentration | | Qualitative Results (#Detected/#Total) | | Quantitative Results (%CV) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Strain | TCID_{50}/mL | Strain | TCID_{50}/mL | HSV-1 | HSV-2 | HSV-1 | HSV-2 |
| HSV-1 McIntyre | 1×10^{5} | HSV-2 MS | 1×10^{5} | 5/5 | 5/5 | 3.02 % | 1.64 % |
| HSV-1 McIntyre | 1×10^{3} | HSV-2 MS | 1×10^{3} | 5/5 | 5/5 | 1.09 % | 2.95 % |
| HSV-1 McIntyre | 177 (3×LoD) | HSV-2 MS | 16.2 (3×LoD) | 5/5 | 5/5 | 1.74 % | 1.88 % |
## Interfering Substances
The performance of the HSV 1&2 ELITe MGB Assay was evaluated with potentially interfering substances that could be encountered in lesion swab specimens obtained from cutaneous and mucocutaneous locations. A total of 33 substances were individually spiked into negative UTM containing one of each HSV-1 and HSV-2 isolates at 3X LoD level and tested in triplicate with the HSV 1&2 ELITe MGB Assay. There were no invalid results. No interference was observed (see table below).
{14}
| Potential Interferent | Interferent Concentration | #Detected/#Total | | |
| --- | --- | --- | --- | --- |
| | | HSV-1 | HSV-2 | IC |
| Whole blood with EDTA | 5% v/v | 0/3 | 0/3 | 3/3 |
| Buffy coat | 5% v/v | 0/3 | 0/3 | 3/3 |
| Acyclovir | 2.5 mg/mL | 0/3 | 0/3 | 3/3 |
| Albumin | 5 mg/mL | 0/3 | 0/3 | 3/3 |
| Casein | 7 mg/mL | 0/3 | 0/3 | 3/3 |
| Female urine | 10% v/v | 0/3 | 0/3 | 3/3 |
| Male urine | 10% v/v | 0/3 | 0/3 | 3/3 |
| K-Y Brand jelly | 5% w/v | 0/3 | 0/3 | 3/3 |
| Douche | 10% v/v | 0/3 | 0/3 | 3/3 |
| Spermicide | 5% w/v | 0/3 | 0/3 | 3/3 |
| Yeast-Gard | 1% w/v | 0/3 | 0/3 | 3/3 |
| Monistat 1 | 5% w/v | 0/3 | 0/3 | 3/3 |
| Monistat 3 | 5% w/v | 0/3 | 0/3 | 3/3 |
| Vagisil Cream | 1% w/v | 0/3 | 0/3 | 3/3 |
| Tioconazole 1 | 5% w/v | 0/3 | 0/3 | 3/3 |
| Rite Aid Feminine Wash, Sensitive Skin | 10% v/v | 0/3 | 0/3 | 3/3 |
| Clotrimazole-7 vaginal cream | 1% w/v | 0/3 | 0/3 | 3/3 |
| Oral Analgesic Gel | 5% w/v | 0/3 | 0/3 | 3/3 |
| Listerine antiseptic mouthwash | 10% v/v | 0/3 | 0/3 | 3/3 |
| Abreva | 10% v/v | 0/3 | 0/3 | 3/3 |
| Carmex lip balm | 1% w/v | 0/3 | 0/3 | 3/3 |
| Releev cold sore treatment | 1% v/v | 0/3 | 0/3 | 3/3 |
| Lip Clear lysine | 1% w/v | 0/3 | 0/3 | 3/3 |
| Toothpaste | 5% w/v | 0/3 | 0/3 | 3/3 |
| Acetaminophen | 5 mg/mL | 0/3 | 0/3 | 3/3 |
| Wal-Finate | 5 mg/mL | 0/3 | 0/3 | 3/3 |
| Cold-Eeze | 7% w/v | 0/3 | 0/3 | 3/3 |
| Non-GMO Corn Starch | 1.25 mg/mL | 0/3 | 0/3 | 3/3 |
| Zinc Oxide Ointment | 7% w/v | 0/3 | 0/3 | 3/3 |
| Cough DM | 10mg/mL | 0/3 | 0/3 | 3/3 |
| Lanacane Max Strength anti-itch cream | 7% w/v | 0/3 | 0/3 | 3/3 |
| Seminal fluid | 7% v/v | 0/3 | 0/3 | 3/3 |
| Foscarnet sodium | 5% v/v | 0/3 | 0/3 | 3/3 |
g. Assay cut-off:
The assay cut-off analysis was performed on a separate set of 141 clinical samples collected from 3 clinical sites. Each clinical sample was evaluated using HSV 1&2 ELITe MGB Assay in conjunction with the ELITe InGenius instrument and a
{15}
composite reference method (FDA-cleared real-time PCR assay combined with PCR amplification and bi-directional sequencing). Both targets in clinical samples were detected up to cycle 45. Therefore $C_T$ of 45 was established as a diagnostic assay cutoff for both HSV-1 and HSV-2 targets.
h. Sample stability studies:
This study assessed both sample stability and sample freeze-thaw stability. The samples for the stability evaluation were prepared by spiking both the HSV-1 and HSV-2 vendor quantitated viral stocks (HSV-1 MacIntyre strain and HSV-2 MS strain) in UTM, M4, M4RT, M5 and M6 media.
Each stability sample set consisted of:
- 5 replicates spiked at $3 \times \mathrm{LoD}$,
- 5 replicates spiked at $1 \times 10^{3}$ TCID50/mL, and
- 5 replicates spiked at $1 \times 10^{5}$ TCID50/mL (15 replicates total for each sample set).
The stability of each sample set was assessed by sample incubation at $+4^{\circ}\mathrm{C}$ for 1 week. All HSV-1 and HSV-2 samples were confirmed to be stable in UTM, M4, M4RT, M5 and M6 media for 1 week at $+4^{\circ}\mathrm{C}$.
The storage conditions were also validated by re-testing previously analyzed clinical samples that were stored in a $-80^{\circ}\mathrm{C}$ freezer ($\leq -70^{\circ}\mathrm{C}$) for minimum of 4 months. Sample concentrations covered HSV clinical range. Ten HSV-1 or HSV-2 positive samples were tested for each media (except M6 for which only 7 HSV-positive samples were available). Positivity of all samples was confirmed after 4 month storage in a $-80^{\circ}\mathrm{C}$ freezer ($\leq -70^{\circ}\mathrm{C}$).
A freeze thaw study was performed using 5 sample sets prepared as in the above using UTM, M4, M4RT, M5 and M6 media. All samples were subjected to 3 freeze-thaw cycles. All the samples were tested with the HSV 1&2 ELITe MGB Assay on the ELITe InGenius. The data obtained show that HSV-1 and HSV-2 viruses are stable after 3 freeze-thaw cycles in UTM, M4, M4RT, M5 and M6 media.
i. Carryover contamination:
The sample-to-sample carry-over from positive samples into the negative samples for the HSV 1&2 ELITe MGB Assay was studied by performing 5 integrated checkerboard runs; HSV-1 high positive at $2.5 \times 10^{7} \mathrm{TCID}_{50} / \mathrm{mL}$ concentration and negative samples (UTM) interspersed, which were compared to the overall background noise from 2 negative sample runs. All 5 checkerboard runs and 2 complete negative runs were performed by 1 operator. The high positive sample concentration in this study exceeded $95\%$ of the concentration levels obtained from samples of infected patients in the study. No carry-over and cross contamination was
16
{16}
observed (the results are shown below).
Carry-Over and Cross-Contamination Results
| Run description | Positive Samples | | Negative Samples | |
| --- | --- | --- | --- | --- |
| | # Neg | % Neg. | # Pos. | % Pos. |
| Run #1, BLANK | 0 / 0 | NA | 0 / 10 | 0 % |
| Run #2, Checkerboard | 0 / 5 | 0 % | 0 / 6 | 0 % |
| Run #3, Checkerboard | 0 / 6 | 0 % | 0 / 6 | 0 % |
| Run #4, BLANK | 0 / 0 | NA | 0 / 10 | 0 % |
| Run #5, Checkerboard | 0 / 6 | 0 % | 0 / 6 | 0 % |
| Run #6, Checkerboard | 0 / 6 | 0 % | 0 / 6 | 0 % |
| Run #7, Checkerboard | 0 / 6 | 0 % | 0 / 6 | 0 % |
| All runs | 0 / 29 | 0 % | 0 / 50 | 0 % |
# 2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Because all analytical studies were conducted in the UTM and clinical studies were conducted using UTM, M4, M4RT, M5 and M6 media, a the matrix comparison study was performed. The matrix comparison study was conducted using contrived sample panel made by spiking HSV-1 or HSV-2 quantitated viral organisms into each of the recommended media: UTM, M4, M4RT, M5 and M6. Each sample set consisted of 3 replicates spiked at $3 \times \mathrm{LoD}$ , 3 replicates spiked at $1 \times 10^{3}$ TCID50/mL, and 3 replicates spiked at $1 \times 10^{5}$ TCID50/mL (9 replicates total for each sample set). Each sample was processed on the InGenius using the HSV 1&2 ELITe MGB Assay. All replicates in all media were detected and shown comparable results. All tested media showed comparable performance as shown in the table below.
| Target/ Channel | Sample Titer TCID50/ mL | Sample Matrix | | | | | All Media Avg C_T | All Media StDev | All Media %CV |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | UTM, Avg C_T | M4, Avg C_T | M4RT, Avg C_T | M5, Avg C_T | M6, Avg C_T | | | |
| HSV-2 CH1, FAM | 1.00E+05 | 27.15 | 26.76 | 26.42 | 26.82 | 27.23 | 26.88 | 0.33 | 1.21% |
| | 1.00E+03 | 33.86 | 33.59 | 33.76 | 33.51 | 34.15 | 33.77 | 0.25 | 0.74% |
| | 3×LoD | 36.32 | 35.56 | 35.96 | 35.54 | 36.14 | 35.91 | 0.35 | 0.97% |
| HSV-1 CH4, AP593 | 1.00E+05 | 22.02 | 21.13 | 20.82 | 20.77 | 20.63 | 21.08 | 0.56 | 2.66% |
| | 1.00E+03 | 28.01 | 28.47 | 27.97 | 28.58 | 26.72 | 27.95 | 0.74 | 2.64% |
| | 3×LoD | 35.84 | 36.07 | 37.02 | 35.43 | 34.69 | 35.81 | 0.86 | 2.39% |
{17}
18
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
To evaluate the clinical performance of the HSV 1&2 ELITe MGB Assay, device performance was compared to a composite reference method. It consisted of an FDA cleared assay and a validated HSV 1&2 PCR followed by bi-directional sequencing of gel electrophoresis-positive samples). Validated HSV 1&2 PCR targeted genomic regions distinct from the HSV 1&2 ELITe MGB Assay. A positive result by the composite reference method is defined as a positive by the FDA cleared PCR or the validated sequencing. Two negative results are needed to confirm a negative)
A total of 1,174 left-over prospectively collected archived swab samples from cutaneous (546) and mucocutaneous (628) lesions from symptomatic patients were collected and evaluated in the study.
The samples were tested with HSV 1&2 ELITe MGB Assay and the Composite Reference Method. Out of the 1,174 tested samples, 2 samples were found invalid by the ELITe MGB Assay and were excluded from the performance analysis tables. Out of the 1172 remaining samples 1 additional invalid sample result for HSV1 and 2 additional invalid sample results for HSV2 by the composite reference method were removed from the performance analysis tables. Therefore for HSV1, 1171 samples analyzed and for HSV2 1170 samples were analyzed.
**HSV-1 Positive/Negative Percent Agreements (PPA/NPA) - Summary of the Results:**
The PPA/NPA performance of HSV 1&2 ELITe MGB Assay when compared to the Composite Reference Method in detection of HSV-1 DNA in cutaneous and mucocutaneous lesions is summarized in the tables below:
{18}
19
| Summary of HSV-1 Results for Valid Cutaneous Lesion Samples (N=545) | | | |
| --- | --- | --- | --- |
| HSV 1&2 ELITe MGB Kit | Composite reference method | | |
| | Positive | Negative | Total |
| Positive | 78 | 7 | 85 |
| Negative | 1 | 459 | 460 |
| Total | 79 | 466 | 545 |
| | | 95% CI | |
| PPA | 98.7% | 93.2-99.8% | |
| NPA | 98.5% | 96.9-99.3% | |
| Summary of HSV-1 Results for Valid Mucocutaneous Lesion Samples (N=626) | | | |
| --- | --- | --- | --- |
| HSV 1&2 ELITe MGB Kit | Composite reference method | | |
| | Positive | Negative | Total |
| Positive | 126 | 12 | 138 |
| Negative | 1 | 487 | 488 |
| Total | 127 | 499 | 626 |
| | | 95% CI | |
| PPA | 99.2% | 95.7-99.9% | |
| NPA | 97.6% | 95.8-98.6% | |
**HSV-2 Positive/Negative Percent Agreements (PPA/NPA) - Summary of the Results:**
The PPA/NPA performance of HSV 1&2 ELITe MGB Assay when compared to the Composite Reference Method in detection of HSV-2 DNA in cutaneous and mucocutaneous lesions is summarized in the table below:
| Summary of HSV-2 Results for Valid Cutaneous Lesion Samples (N=545) | | | |
| --- | --- | --- | --- |
| HSV 1&2 ELITe MGB Kit | Composite reference method | | |
| | Positive | Negative | Total |
| Positive | 125 | 6 | 131 |
| Negative | 5 | 409 | 414 |
| Total | 130 | 415 | 545 |
| | | 95% CI | |
| PPA | 96.2% | 91.3-98.3% | |
| NPA | 98.6% | 96.9-99.3% | |
| Summary of HSV-2 Results for Valid Mucocutaneous Lesion Samples (N=625) | | | |
| --- | --- | --- | --- |
| HSV 1&2 ELITe MGB Kit | Composite reference method | | |
| | Positive | Negative | Total |
| Positive | 164 | 8 | 172 |
| Negative | 4 | 449 | 453 |
{19}
| Total | 168 | 457 | 625 |
| --- | --- | --- | --- |
| | | 95% CI | |
| PPA | 97.6% | 94.0-99.1% | |
| NPA | 98.2% | 96.6-99.1% | |
## Contrived Sample Study:
Due to the difficulty in obtaining sufficient HSV-2 positive oral samples, testing for HSV-2 was supplemented by using a contrived panel. The panel consisted of 75 individual negative cheek swab samples collected in Universal Transport Media (UTM) and spiked with HSV-2 at concentrations of 3×LoD, 8×LoD, 40×LoD, 200×LoD and 1000×LoD (10 of each), 10 HSV-1 Positive samples spiked at 10×LoD and 15 HSV-1/HSV-2 Negative Oral Samples
All panel members were randomized, blinded to the tester and tested with HSV 1&2 ELITe MGB Assay on the ELITe InGenius instrument according to the clinical study protocol.
The HSV-2 Oral Contrived Panel Study revealed that 49 out of 50 oral HSV-2 contrived samples were positive using HSV 1&2 ELITe MGB Assay (98% detection). All 10 HSV-1 Positive samples confirmed 100% positivity.
## 4. Clinical cut-off:
Not Applicable
## 5. Expected values/Reference range:
The observed expected values for HSV-1 and HSV-2 in the study population using the HSV 1&2 ELITe MGB Assay were calculated and are presented for the combined sample set stratified by age group, gender and by lesion source in the tables below. A total of 6 dual positives for HSV1 and HSV2 were detected by the ELITe MGB Assay and one of the samples was confirmed as dual positive by the composite reference method.
{20}
Cutaneous and Mucocutaneous HSV 1&2 Prevalence by Age and Gender
| Gender | Age Group | Total | HSV 1&2 ELITe MGB Assay HSV-1 results | | HSV 1&2 ELITe MGB Assay HSV-2 results | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Positive | Prevalence | Positive | Prevalence |
| Female | <20 | 42 | 18 | 42.9% | 12 | 28.6% |
| | 20-29 | 244 | 68 | 27.9% | 70 | 28.7% |
| | 30-39 | 143 | 24 | 16.8% | 45 | 31.5% |
| | 40-49 | 97 | 14 | 14.4% | 25 | 25.8% |
| | 50-59 | 88 | 18 | 20.5% | 24 | 27.3% |
| | ≥60 | 123 | 21 | 17.1% | 30 | 24.4% |
| | All | 737 | 163 | 22.1% | 206 | 28.0% |
| Male | <20 | 20 | 4 | 20.0% | 2 | 10.0% |
| | 20-29 | 144 | 25 | 17.4% | 33 | 22.9% |
| | 30-39 | 117 | 15 | 12.8% | 25 | 21.4% |
| | 40-49 | 48 | 5 | 10.4% | 15 | 31.3% |
| | 50-59 | 44 | 6 | 13.6% | 9 | 20.5% |
| | ≥60 | 61 | 5 | 8.2% | 13 | 21.3% |
| | All | 434 | 60 | 13.8% | 97 | 22.4% |
| | Gender not identified | 1 | 0 | 0% | 0 | 0% |
| ALL | | 1172 | 223 | 19.0% | 303 | 25.9% |
Cutaneous HSV 1&2 Expected Values by Lesion Source
| Lesion Source | Total | HSV 1&2 ELITe MGB Assay HSV-1 results | | HSV 1&2 ELITe MGB Assay HSV-2 results | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Prevalence | Positive | Prevalence |
| Genital/Anogenital | 248 | 38 | 15.3% | 78 | 31.5% |
| Skin lesion | 297 | 47 | 15.8% | 53 | 17.8% |
| Overall | 545 | 85 | 15.6% | 131 | 24.0% |
{21}
22
Mucocutaneous HSV 1&2 Expected Values by Lesion Source
| Lesion Source | Total | HSV 1&2 ELITe MGB
Assay HSV-1 results | | HSV 1&2 ELITe MGB
Assay HSV-2 results | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Prevalence | Positive | Prevalence |
| Genital/Vaginal/Cervical | 501 | 109 | 21.8% | 163 | 32.5% |
| Oral | 74 | 21 | 28.4% | 2 | 2.7% |
| Other | 27 | 5 | 18.5% | 2 | 7.4% |
| Anorectal | 12 | 2 | 16.7% | 5 | 41.7% |
| Urethral | 6 | 0 | 0% | 0 | 0% |
| Ocular | 5 | 0 | 0% | 0 | 0% |
| Nasal | 2 | 1 | 50.0% | 0 | 0% |
| Overall | 627 | 138 | 22.0% | 172 | 27.4% |
N. Instrument Name:
ELITe InGenius system
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ or No ☑
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☑
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☑ or No ☐
3. Specimen Identification: Not Applicable
4. Specimen Sampling and Handling: Not Applicable
{22}
5. Calibration: Not Applicable
6. Quality Control: Not Applicable
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
The HSV 1&2 ELITe MGB Assay is used on the ELITe InGenius system. The ELITe InGenius system is a bench top instrument integrating all required hardware, reagent and software components to perform nucleic acid sample preparation and real-time PCR operations.
The ELITe InGenius system can process from 1 to 12 samples in 12 parallel tracks, and samples may be loaded in primary tubes or in secondary tubes provided. The system utilizes a universal, cassette-based process for sample extraction, and allows for multiple and independent PCRs to be performed from a single nucleic acid eluate. The system can operate in three different modes: nucleic acid extraction only, PCR amplification only, or nucleic acid extraction with PCR amplification.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
23