K220607 · Becton, Dickinson and Company · PCH · Sep 21, 2022 · Microbiology
Device Facts
Record ID
K220607
Device Name
BD MAX Enteric Viral Panel
Applicant
Becton, Dickinson and Company
Product Code
PCH · Microbiology
Decision Date
Sep 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3990
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BD MAX™ Enteric Viral Panel performed on the BD MAX™ System is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric viral pathogens. The BD MAX™ Enteric Viral Panel detects nucleic acids from Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V), and Human Astrovirus (hAstro). Testing is performed on unpreserved soft to diarrheal or Cary-Blair preserved stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of relevant gene target DNA/RNA. The test utilizes fluorogenic gene-specific hybridization probes for detection of the amplified DNA. This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V) and hAstro infections. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test, and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
Device Story
The BD MAX Enteric Viral Panel is an automated in vitro diagnostic assay for detecting enteric viral pathogens (Norovirus, Rotavirus, Adenovirus, Sapovirus, Astrovirus) in stool specimens. The system uses real-time PCR to amplify target DNA/RNA and fluorogenic hybridization probes for detection. The device automates sample preparation (lysis, nucleic acid extraction/concentration), reagent rehydration, amplification, and detection. It is used in clinical laboratories; operators load specimens into the BD MAX System, which performs all subsequent steps. The system software interprets results (POS, NEG, UNR, IND, INC) based on target and Sample Processing Control (SPC) amplification. Results aid clinicians in differential diagnosis alongside clinical/epidemiological data. The device benefits patients by providing rapid, automated identification of viral causes of gastroenteritis, though it is not intended as the sole basis for patient management.
Clinical Evidence
Multicenter clinical study (8 sites) comparing FecalSwab to Cary-Blair preserved stool. Enrolled 802 specimens (591 prospective, 211 retrospective). Primary endpoints: Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for five viral targets. Results showed high concordance; e.g., Norovirus prospective PPA 87.0%, NPA 99.6%; retrospective PPA 99.0%, NPA 94.3%. Contrived specimens used for Adenovirus due to low clinical prevalence, showing 100% PPA/NPA. Study confirmed FecalSwab performance is substantially equivalent to the predicate.
Technological Characteristics
Multiplex nucleic acid-based assay; real-time PCR; fluorogenic hydrolysis (TaqMan) probes. Automated sample preparation (lysis, extraction, concentration) and detection on BD MAX System. Connectivity: standalone instrument. Software: automated interpretation of amplification curves (Ct.Score, EP fluorescence). Specimen types: unpreserved, Cary-Blair, or FecalSwab (nylon flocked swab in modified Cary-Blair).
Indications for Use
Indicated for symptomatic patients (pediatric and adult) with suspected acute gastroenteritis, enteritis, or colitis. Used for qualitative detection and differentiation of Norovirus GI/GII, Rotavirus A, Adenovirus F40/41, Sapovirus, and Human Astrovirus in unpreserved, Cary-Blair preserved, or FecalSwab preserved stool specimens.
Regulatory Classification
Identification
A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.
Special Controls
*Classification.* Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).
K181427 — BD MAX Enteric Viral Panel, BD MAX Instrument · Becton, Dickinson and Company · Nov 29, 2018
K170308 — BD MAX Extended Enteric Bacterial Panel, BD MAX System · Becton, Dickinson and Company · May 2, 2017
K143648 — BD MAX Enteric Parasite Panel, BD MAX Instrument · Becton, Dickinson and Company · Aug 25, 2015
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K220607
B Applicant
Becton, Dickinson and Company
C Proprietary and Established Names
BD MAX Enteric Viral Panel
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PCH | Class II | 21 CFR 866.3990 - Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assay | MI - Microbiology |
| OOI | Class II | 21 CFR 862.2570 - Instrumentation for clinical multiplex test systems | CH - Clinical Chemistry |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for BD MAX Enteric Viral Panel (BD MAX EVP) on the BD MAX System with stool specimens collected using the Copan FecalSwab Collection, Transport and Preservation System (Copan FecalSwab) and BD FecalSwab Collection, Transport and Preservation System (BD FecalSwab). The Copan FecalSwab is also co-branded as the BD FecalSwab Collection, Transport and Preservation System (BD FecalSwab) and has been FDA-cleared under K142094 (Copan Italia SpA, legal manufacturer); the terms Copan FecalSwab, BD FecalSwab, and FecalSwab may be used interchangeably.
### B Measurand:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
The BD MAX Enteric Viral Panel detects nucleic acids from
- Norovirus GI & GII
- Rotavirus A
- Adenovirus F40/41
- Sapovirus (genogroups I, II, IV, V)
- Human Astrovirus (hAstro)
## C Type of Test:
The BD MAX EVP is a qualitative real-time polymerase chain reaction (PCR) assay for the amplification and detection of DNA and RNA from Norovirus, Rotavirus, Adenovirus, Sapovirus, and Astrovirus.
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The BD MAX Enteric Viral Panel performed on the BD MAX System, is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric viral pathogens. The BD MAX Enteric Viral Panel detects nucleic acids from
- Norovirus GI & GII
- Rotavirus A
- Adenovirus F40/41
- Sapovirus (genogroups I, II, IV, V)
- Human Astrovirus (hAstro)
Testing is performed on unpreserved soft to diarrheal or Cary-Blair preserved stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of relevant gene target DNA/RNA. The test utilizes fluorogenic gene-specific hybridization probes for detection of the amplified DNA.
This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V), and hAstro infections. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test, and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by
K220607 - Page 2 of 19
{2}
pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## D Special Instrument Requirements:
For use with the BD MAX System.
## IV Device/System Characteristics:
### A Device Description:
The BD MAX Enteric Viral Panel (BD MAX EVP) is a gastrointestinal bacterial panel multiplex nucleic acid-based assay system comprised of an instrument with associated hardware and accessories, disposable microfluidic cartridges, real-time PCR master mixes, unitized reagent strips, extraction reagents, and sample buffer tubes (SBT). The instrument automates sample preparation including target lysis, DNA extraction and concentration, reagent rehydration, and target nucleic acid amplification and detection using real-time PCR. The assays include a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors DNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX System software automatically interprets test results. The BD MAX Enteric Viral Panel performed on the BD MAX system is FDA cleared under K181427.
The BD MAX Enteric Viral Panel instructions for use include testing of unpreserved stool specimens collected and transported to the laboratory in a dry, clean container, Cary-Blair preserved stool specimens collected using a plastic paddle (scoop) to place a stool sample into 15 mL of Cary-Blair media for transport, or stool specimens collected and transported using the Copan FecalSwab Collection, Transport, and Preservation System (Copan FecalSwab). Stool specimens from rectal swabs or fixed stools have not been validated with the BD MAX EVP.
The Copan FecalSwab is comprised of a nylon flocked specimen collection swab co-packaged with a tube filled pre-filled with 2 mL of a modified Cary-Blair preservative. The Copan FecalSwab is also co-branded as the BD FecalSwab Collection, Transport and Preservation System (BD FecalSwab); the terms Copan FecalSwab, BD FecalSwab, and FecalSwab are used interchangeably.
For the BD MAX EVP a test result may be called as POS, NEG or UNR (Unresolved) based on the amplification status of the targets and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAX System failure.
### B Principle of Operation:
A stool specimen is collected and transported to the laboratory in a dry, clean container (for unpreserved specimens), in Cary-Blair transport media, or using the Copan FecalSwab Collection, Transport, and Preservation System. Unpreserved stool samples and Cary-Blair
K220607 - Page 3 of 19
{3}
preserved stool samples are placed in a BD MAX SBT using a 5 μL transfer loop for analysis on the BD Max System.
To use the Copan FecalSwab stool specimens, the operator transfers fecal material from an unpreserved stool specimen to the vial of FecalSwab transport medium using the nylon flocked specimen collection swab. The operator unscrews the cap of the FecalSwab transport medium tube and transfers the swab with sample into the tube. The operator breaks the swab shaft at the break point line molded into the shaft, discards the handle part of the swab shaft, and screws the cap onto the FecalSwab tube to close. Before analysis on the BD MAX system, FecalSwab stool samples are vortexed and then 25 μL of sample is pipetted into a BD MAX SBT.
The SBT is closed with a septum cap and vortexed. A worklist is created and the SBT, unitized reagent strips, master mix, extraction tubes, and PCR cartridges are loaded onto the BD MAX System. The BD MAX System automates sample preparation including cell lysis, DNA extraction and concentration, reagent rehydration, and target nucleic acid amplification and detection using real-time PCR. Amplified targets are detected with hydrolysis probes labeled with quenched fluorophores. The amplification, detection and interpretation of the signals are done automatically by the BD MAX System.
## C Interpretation of Results
Targets amplified by both mastermixes D6 and D5 are detected with hydrolysis probes (TaqMan probes) labelled at one end with a fluorescent reporter dye and at the other end with a quencher moiety. Six separate probes labelled with different reporter dyes are used to detect, in four different BD MAX System optical channels, the amplicons generated by their respective primers for MM D6 Enteric Viral Panel targets (Astrovirus, Sapovirus, and a sample processing control) and eight probes for MM D5 Enteric Viral Panel targets (Adenovirus, Norovirus, Rotavirus, and a sample processing control). The BD MAX System monitors several amplification curve metrics including: the fluorescence height at the end of the amplification curve (EP fluorescence), the location and value of the curve's first derivative, the location and value of the curve's second derivative, the threshold cycle value with an additional quality control applied to prevent false positives due to signal drift (Ct.Score), and measurements of system noise along the PCR amplification curve. Curve metrics are transformed into results via comparison of the metrics to cutoffs and the use of logic statements. Specifically, cutoffs are applied to the Ct.Score and EP values to determine amplification status with other metrics serving as quality controls.
Results are available on the "Results" tab in the "Results" window on the BD MAX System monitor. The BD MAX System software automatically interprets test results. Results are reported for each of the analytes and for the Sample Processing Control. A test result may be called as NEG (negative), POS (positive) or UNR (unresolved) based on the amplification status of the target and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAX System failure and require sample repeat testing. A sample can be re-tested directly from the already prepared Sample Buffer Tube or following the preparation of a new Sample Buffer Tube inoculation. In the case of a partial UNR, where one or more targets have a POS result and all other targets have a UNR result, the targets with a UNR result will not be called NEG. This will be reported on a per Master Mix basis.
K220607 - Page 4 of 19
{4}
Erroneous results may occur from improper specimen collection, handling, storage, technical error, specimen mix-up or because the number of organisms in the specimen is below the analytical sensitivity of the test.
The Sample Processing Control has been added to the test to aid in the identification of samples that contain inhibitors to PCR amplification and as a control for reagent integrity and of the assay system as a whole. The Sample Processing Control does not indicate if nucleic acid has been lost due to inadequate collection, transport or storage of samples, or whether viral capsids have been adequately lysed.
V Substantial Equivalence Information:
A Predicate Device Name(s): BD MAX Enteric Viral Panel
B Predicate 510(k) Number(s): K181427
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K220607 | K181427 |
| --- | --- | --- |
| Device Trade Name | BD MAX Enteric Viral Panel | BD MAX Enteric Viral Panel |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The BD MAX Enteric Viral Panel performed on the BD MAX System, is an automated in vitro diagnostic test for the direct qualitative detection and differentiation of enteric viral pathogens. The BD MAX™ Enteric Viral Panel detects nucleic acids from
• Norovirus GI & GII
• Rotavirus A
• Adenovirus F40/41
• Sapovirus (genogroups I, II, IV, V)
• Human Astrovirus (hAstro) | Same |
K220607 - Page 5 of 19
{5}
| | Testing is performed on unpreserved soft to diarrheal or Cary-Blair preserved stool specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of relevant gene target DNA/RNA. The test utilizes fluorogenic gene-specific hybridization probes for the detection of the amplified DNA.
This test is intended for use, in conjunction with clinical presentation, laboratory findings, and epidemiological information, as an aid in the differential diagnosis of Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V), and hAstro infections. Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test, and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be |
| --- | --- |
K220607 - Page 6 of 19
{6}
K220607 - Page 7 of 19
| | due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn’s disease. | |
| --- | --- | --- |
| Organisms Detected | • Norovirus GI & GII
• Rotavirus A
• Adenovirus F40/41
• Sapovirus (genogroups I, II, IV, V)
• Human Astrovirus (hAstro) | Same |
| Assay Format | Amplification: PCR Detection: Fluorogenic target-specific hybridization | Same |
| Assay Format | Amplification: PCR Detection: Fluorogenic target-specific hybridization | Same |
| Assay Targets | Presence of RdRp and VP1 genes specific for Norovirus GI & GII
• non-coding sequence after nsp-3 gene specific for Rotavirus A
• hexon gene specific for Adenovirus F40/41
• RdRp and VP1 genes specific to Sapovirus (genogroups I, II, IV, V)
• RdRp gene specific to Human Astrovirus (hAstro) | Same |
| Interpretation of Test Results | Automated (BD MAX System diagnostic software) | Same |
| Analysis Platform | BD MAX System | Same |
| PCR Sample Preparation | Automated by the BD MAX System | Same |
| Detection probes | TaqMan Probe | Same |
| Assay Controls | Sample Processing Control (SPC) | Same |
{7}
| General Device Characteristic Differences | | |
| --- | --- | --- |
| Specimen Type | • Unpreserved stool
• Cary-Blair preserved stool
• FecalSwab (modified Cary-Blair) preserved stool | • Unpreserved stool
• Cary-Blair preserved stool |
| Sample Volume Tested | • 25 μL via Pipette from the FecalSwab collection tube
• 5 μL Transport Loop from the unpreserved or Cary-Blair preserved stool | • 5 μL Transport Loop from the unpreserved or Cary-Blair preserved stool |
VI Standards/Guidance Documents Referenced:
Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens, November 2, 2015.
VII Performance Characteristics (if/when applicable):
A. Analytical Performance:
1. Precision/Reproducibility:
Within-lab precision and reproducibility studies were evaluated previously using unpreserved stool specimens. This represents the most challenging specimen type and therefore additional studies were not conducted. Please see K181427.
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Analytical Specificity/Cross-Reactivity: Analytical specificity/cross-reactivity was evaluated previously in the presence of unpreserved stool that represents the most challenging specimen type. Please refer to K181427. No specificity issues were identified from additional studies conducted to support testing FecalSwab stool specimens.
K220607 - Page 8 of 19
{8}
Interference Substances (exogenous/endogenous): Interfering substances were previously evaluated in the presence of unpreserved stool which contains the highest concentration of potential inhibitors. Please refer to K181427.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Please refer to K181427.
An additional specimen stability study was conducted to evaluate specimen stability in FecalSwab collection tube and corresponding SBT meet the following BD MAX EVP assay specimen stability claims:
A. FecalSwab stool specimens before testing: 25 ± 2 °C up to 48 hours (2 days), or 2-8°C (5 °C ± 3 °C) up to 120 hours (5 days)
B. Specimen added to the BD MAX Sample Buffer Tube (SBT): 25 ± 2°C up to 48 hours (2 days) or 2-8°C up to 120 hours (5 days)
FecalSwab collection tubes were prepared with the positive and negative matrices for testing. Four positive FecalSwab tubes and one negative FecalSwab tube from each of three lots were prepared for each time point. Two BD MAX Sample Buffer Tubes were prepared from each FecalSwab tube. A minimum of 24 test results for positive specimens and six test results for negative specimens were performed at each test point. To determine stability, contrived samples containing Rotavirus A strain Va70, Adenovirus Type F41, and Astrovirus Type 4 were prepared from viral stocks. Stability testing for Norovirus GII and Sapovirus GI utilized quantitated clinical specimens. All positive samples contained analyte levels at 3X LoD.
Table 1 below describes the stability conditions and timepoints.
Table 1: Nested FecalSwab Specimen Stability Conditions and Timepoints
| Test Point | FecalSwab | | SBT | | Total Storag e Time |
| --- | --- | --- | --- | --- | --- |
| | Storage Condition 1 | | Storage Condition 2 | | |
| 1 (Baseline) | N/A | | | | |
| 2 | 25 °C ± 2 °C | 2 days | N/A | | 2 days |
| 3 | | | 25 °C ± 2 °C | 2 days | 4 days |
| 4 | | | | 3 days | 5 days |
| 5 | | | 5 °C ± 3 °C | 5 days | 7 days |
| 6 | | | | 6 days | 8 days |
| 7 | | 3 days | N/A | | 3 days |
| 8 | | 5 days | N/A | | 5 days |
| 9 | | | | 2 days | 7 days |
K220607 - Page 9 of 19
{9}
| 10 | 5 °C ± 3 °C | | 25 °C ± 2 °C | 3 days | 8 days |
| --- | --- | --- | --- | --- | --- |
| 11 | | | 5 °C ± 3 °C | 5 days | 10 days |
| 12 | | | | 6 days | 11 days |
| 13 | | 6 days | N/A | | 6 days |
The acceptance criteria for the study were 100% of spiked positive samples at baseline should report "POS" as the result and 100% of un-inoculated negative samples at baseline should report "NEG" as the result. For all subsequent time points, ≥95% of spiked positive samples should report "POS" as the result and 100% of un-inoculated negative samples should report "NEG" as the result. A minimum of 24 valid results for positive and 6 valid results for negative samples at each test point was required.
All target organisms passed the prespecified acceptance criteria, therefore the data were considered acceptable for the proposed stability claim. Stool preserved in FecalSwab can be stored for up to 120 hours (5 days) at 2 - 8 °C or for up to 48 hours at 2 - 25 °C. Inoculated Sample Buffer Tube can be stored at 2 - 8 °C for a maximum of 120 hours (5 days) or at 2 - 25 °C for a maximum of 48 hours (2 days).
## 6. Limit of Detection - Equivalency Study
The LoD for each target on the BD MAX EVP panel was previously determined in unpreserved stool specimens and Cary-Blair preserved stool specimens (see K181427). A LoD equivalency study was performed to demonstrate comparable analytical sensitivity for the BD MAX EVP panel testing Cary-Blair preserved stool samples and FecalSwab preserved stool samples.
Viral stocks were used for LoD serial dilution testing of Rotavirus A strain Va70, Adenovirus Type F41, and Astrovirus Type 4. Quantitated positive stools diluted to a working solution were used to generate the LoD dilutions for Norovirus GII and Sapovirus GI. One viral stock or positive stool was tested for each of the above viruses. Testing was performed with BD MAX EVP reagents, three lots of FecalSwab (4 FecalSwab replicates per reagent lot, 2 SBTs per FecalSwab), and one lot of Cary-Blair preserved specimen medium. A five-fold serial titration resulting in a total of five dilutions, was performed for each of the 5 assay targets and tested in both sample types (Cary-Blair and FecalSwab). Sample Buffer Tubes (SBT) were created by pipetting 25 μL from FecalSwab or looping 5 μL loop from Cary-Blair, altogether 24 SBTs each. The SBTs were tested on the BD MAX System.
Equivalence between the specimen collection methods was confirmed when LoDs for each target were within one five-fold dilution of each other. LoD is defined as when positivity is greater than 95% (23/24 or more POS or NEG results).
The LoDs using Cary-Blair preserved stool specimens and FecalSwab stool specimens were within one five-fold dilution of each other for all EVP targets. All analytes met the acceptance criteria and the results demonstrated comparable analytical sensitivity when FecalSwab preserved stool specimens and Cary-Blair preserved stool specimens were tested with BD MAX EVP targets on the BD MAX System.
K220607 - Page 10 of 19
{10}
7. Carry-Over:
Carry-over was established previously using unpreserved stool in EBP sample buffer which represents the worst-case scenario for carry-over contamination; therefore, additional studies were not necessary. Please refer to K181427.
8. Assay Cut-Off:
Assay cut-offs remain unchanged from the previously cleared version of the multiplex panel (see K181427).
9. User Variability:
The objective of this study was to determine whether the preparation of the FecalSwab collection tube by different users introduces variability in the expected results for the BD MAX EVP assay on the BD MAX System.
Two different FecalSwab collection tubes were prepared by six different operators for each of the five panel members. The test panel included contrived specimens co-spiked with Norovirus and Astrovirus at the following levels: one negative sample, three samples at 2x LoD, and one sample at 4x LoD. These targets were selected to represent each of the two PCR Master Mix formulations of the BD MAX Enteric Viral Panel. All steps subsequent to the FecalSwab preparation (i.e., pipetting to the BD MAX SBT, application of the septum cap, final vertexing, and initiating the run on the BD MAX) were performed by a single experienced BD MAX user.
For acceptance criteria, 100% negative results for the 12 negative samples, ≥95% positive results for the 36 samples tested at 2X LoD, and 100% positive for the 12 samples tested at 4X LoD.
Table 2 below presents results for user variability testing with Norovirus and Astrovirus.
Table 2: User Variability Assay Results
| Organism | Load | Assay Results (POS/NEG/Total) | Pass/Fail Acceptance Criteria |
| --- | --- | --- | --- |
| Norovirus | 2x LoD | 36 / 0 / 36 | Pass |
| | 4x LoD | 12 / 0 / 12 | |
| | NEG | 0 / 12 / 12 | |
| Astrovirus | 2x LoD | 36 / 0 / 36 | |
| | 4x LoD | 12 / 0 / 12 | |
| | NEG | 0 / 12 / 12 | |
Tables 3 and 4 present results for user variability testing with Norovirus and Astrovirus, respectively.
Table 3: User Variability of the FecalSwab Collection Tube with Norovirus
| EVP Target: Norovirus | | | | |
| --- | --- | --- | --- | --- |
| Panel | 2X LOD | 4X LOD | Negative | Grand Total |
K220607 - Page 11 of 19
{11}
| Result | NEG | POS | NEG | POS | NEG | POS | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| User 1 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 2 | 0 | 7* | 0 | 2 | 2 | 0 | 11 |
| User 3 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 4 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 5 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 6 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| Grand Total | 0 | 37 | 0 | 12 | 12 | 0 | 61 |
*One sample had an invalid result and was retested. This sample tested positive for that target in the repeat run.
Table 4: User Variability of the FecalSwab Collection Tube with Astrovirus
| EVP Target: Astrovirus | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Panel | 2X LOD | | 4X LOD | | Negative | | Grand Total |
| Result | NEG | POS | NEG | POS | NEG | POS | |
| User 1 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 2 | 0 | 7* | 0 | 2 | 2 | 0 | 11 |
| User 3 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 4 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 5 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| User 6 | 0 | 6 | 0 | 2 | 2 | 0 | 10 |
| Grand Total | 0 | 37 | 0 | 12 | 12 | 0 | 61 |
*One sample had an invalid result and was retested. This sample tested positive for that target in the repeat run.
Results for both targets met the acceptance criteria. Differences in specimen workflow between the FecalSwab and Cary-Blair collection procedures did not demonstrate an observable effect on expected results with the BD MAX EVP on the BD MAX System.
## B. Comparison Studies:
### 1. Method Comparison with Predicate Device:
A method comparison study was performed to demonstrate equivalent performance between testing FecalSwab stool specimens and Cary-Blair preserved stool specimens with the BD MAX EVP on the BD MAX System.
The study was conducted in accordance with the Declaration of Helsinki and in compliance with applicable regulations and ICH Good Clinical Practice (GCP). Informed consent was not needed because only de-identified remnant specimens were enrolled and tested. Fresh prospective specimens were collected at six external sites. Previously characterized retrospective specimen were collected at two external sites and one internal site at BD. Samples were also contrived, where applicable, at an internal BD site. The following inclusion and exclusion criteria were used for prospective and retrospective specimen collection:
- Prospective inclusion criteria: unpreserved soft to diarrheal stool specimen from a pediatric or adult patient admitted to a healthcare facility (e.g., hospital, outpatient
K220607 - Page 12 of 19
{12}
clinica, or long term care facility) and suspected of having acute hasteroenteritis, enteritis, or colitis for which diagnostic tests were indicated and/or ordered
- Prospective exclusion criteria: unlabeled or mislabeled solid/formed stools or rectal swabs and specimen from patients suspected and/or confirmed Clostridioides (Clostridium) difficile diarrheal disease
- Retrospective inclusion criteria: unpreserved stool specimen confirmed positive for one of the EVP
- Retrospective exclusion criteria: unlabeled or mislabeled specimens
Adenovirus contrived specimens were prepared such that a minimum of 50% of the specimens contained analyte concentrations corresponding to 1.5x - 2x LoD in a unique specimen matrix consisting of negative stool and collected using the FecalSwab. A total of 53 positive and 53 negative contrived samples were prepared using 53 unique negative specimen matrices, all of which had been previously determined to be negative for all targeted analytes.
Table 5: Adenovirus Contrived Samples Panel Target Level Description
| xLoD | Adenovirus |
| --- | --- |
| 2 | 27 |
| 4 | 7 |
| 10 | 7 |
| 25 | 6 |
| 1000 | 6 |
| Negative | 53 |
| Total | 106 |
A total of 594 prospective specimens and 211 retrospective specimens were enrolled in the clinical evaluation. Three prospective specimens were excluded from the data analysis due to specimen exclusion criteria.
Table 6 presents the BD MAX positivity rate for BD MAX EVP assay, by specimen origin, device type and site. The NYH and TGH sites were the only sites actively collecting retrospective specimens during study enrollment. All other retrospective specimens had been collected during previous studies.
The 802 (591 prospective and 211 retrospective) compliant specimens enrolled by patient age, sex, and specimen type. Ten (10) additional prospective samples were excluded from the data analysis due to Sample Buffer Tube or instrument level exclusion criteria. The final data analysis included 581 compliant prospective and 211 compliant retrospective subjects for Norovirus GI & GII, Rotavirus A, Adenovirus F40/41, Sapovirus (genogroups I, II, IV, V), and Human Astrovirus (hAstro) targets.
Table 5 presents the BD MAX positivity rate for BD MAX EVP assay, by specimen origin, device type and site. The NYH and TGH sites were the only sites actively collecting retrospective specimens during study enrollment. All other retrospective specimens had been collected during previous studies.
K220607 - Page 13 of 19
{13}
Table 6: Compliant Clinical Trial Enrollment Summary by Age, Sex, and Specimen Type
| Specimen Type | Mean Age in years (SD) | Median Age in years | Min Age in years | Max Age in years | Sex of Total N |
| --- | --- | --- | --- | --- | --- |
| Prospective Total N = 591 | 47.0 (22.7) | 49.0 | <1 | 95 | Male: 44.8% |
| Unknown Age: 0 | | | | | Female: 55.2% |
| Known Age: 591 | | | | | Unknown: 0.0% |
| Retrospective Total N = 211 | 42.0 (24.4) | 46.0 | <1 | 88 | Male: 41.2% |
| Unknown Age: 59 | | | | | Female: 53.1% |
| Known Age: 152 | | | | | Unknown: 5.7% |
| Overall Total N = 802 | 46.0 (23.1) | 49.0 | <1 | 95 | Male: 43.9% |
| Unknown Age: 59 | | | | | Female: 54.6% |
| Known Age: 743 | | | | | Unknown: 1.5% |
The primary study endpoint was to demonstrate equivalent performance of testing FecalSwab stool specimens to the performance of testing Cary-Blair stool specimen with the BD MAX EVP on the BD MAX System. Acceptance criteria were established as a PPA of $\geq 95\%$ for each target and an NPA of $\geq 90\%$ for each target with $95\%$ lower bound of the confidence interval of $\geq 90\%$ for each target for both PPA and NPA.
Percent agreement comparing the FecalSwab specimen performance to the Cary-Blair specimen performance with the BD MAX EVP targets is presented in Tables 7-13.
Table 7: Norovirus PPA and NPA of the BD MAX Enteric Viral Panel – FecalSwab Compared to Cary-Blair Preserved
| Norovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| Prospective | Positive | 20 | 2 | 22 |
| | Negative | 3 | 554 | 557 |
| | Total | 23 | 556 | 579 |
| PPA: 87.0% (67.9%, 95.5%)
NPA: 99.6% (98.7%, 99.9%) | | | | |
| Retrospective | Positive | 104 | 6 | 110 |
| | Negative | 1 | 100 | 101 |
K220607 - Page 14 of 19
{14}
| Norovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| | Total | 105 | 106 | 211 |
| PPA: 99.0% (94.8%, 99.8%) NPA: 94.3% (88.2%, 97.4%) | | | | |
Table 8: Rotavirus PPA and NPA of the BD MAX Enteric Viral Panel – FecalSwab Compared to Cary-Blair Preserved
| Rotavirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| Prospective | Positive | 3 | 0 | 3 |
| | Negative | 0 | 576 | 576 |
| | Total | 3 | 576 | 579 |
| PPA: 100.0% (43.9%, 100.0%) NPA: 100.0% (99.3%, 100.0%) | | | | |
| Retrospective | Positive | 39 | 3a | 42 |
| | Negative | 7a | 162 | 169 |
| | Total | 46 | 165 | 211 |
| PPA: 84.8% (71.8%, 92.4%) NPA: 98.2% (94.8%, 99.4%) | | | | |
$^{\mathrm{a}}$ 9/10 specimens with discrepant results have target at or near the LoD.
For the BD FecalSwab Collection, Transport and Preservation System, the BD MAX™ Enteric Viral Panel identified $100.0\%$ and $99.5\%$ of the prospectively collected Adenovirus positive and negative specimens, respectively, and $100.0\%$ and $99.0\%$ of the retrospectively collected Adenovirus positive and negative specimens, respectively (refer to Table 9).
Table 9: Adenovirus PPA and NPA of the BD MAX EVP FecalSwab Compared to Cary Blair Preserved
| Adenovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| Prospective | Positive | 1 | 3 | 4 |
| | Negative | 0 | 575 | 575 |
K220607 - Page 15 of 19
{15}
| Adenovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| | Total | 1 | 578 | 579 |
| PPA: 100.0% (20.7%, 100.0%) NPA: 99.5% (98.5%, 99.8%) | | | | |
| Retrospective | Positive | 9 | 2 | 11 |
| | Negative | 0 | 200 | 200 |
| | Total | 9 | 202 | 211 |
| PPA: 100.0% (70.1%, 100.0%) NPA: 99.0% (96.5%, 99.7%) | | | | |
Due to the small number of Adenovirus positive specimens in the study, contrived specimens were also evaluated. The BD FecalSwab Collection, Transport and Preservation System on the BD MAX Enteric Viral Panel identified 100.0% of the Adenovirus contrived positive and negative specimens, when compared to expected results (refer to Table 10).
Table 10: Adenovirus Contrived FecalSwab Specimen Results
| Adenovirus | Expected Result | | |
| --- | --- | --- | --- |
| FecalSwab | Positive | Negative | Total |
| Positive | 52 | 0 | 52 |
| Negative | 0 | 53 | 53 |
| Total | 52 | 53 | 105 |
| PPA: 100.0% (93.1%, 100.0%) NPA: 100.0% (93.2%, 100.0%) | | | |
Table 11: Adenovirus Contrived FecalSwab Specimen Results Compared to Cary-Blair contrived Specimen Results
| Adenovirus | Cary-Blair | | |
| --- | --- | --- | --- |
| FecalSwab | Positive | Negative | Total |
| Positive | 50 | 2 | 52 |
| Negative | 0 | 53 | 53 |
K220607 - Page 16 of 19
{16}
| Adenovirus | Cary-Blair | | |
| --- | --- | --- | --- |
| FecalSwab | Positive | Negative | Total |
| Total | 50 | 55 | 105 |
| PPA: 100.0% (92.9%, 100.0%) NPA: 96.4% (87.7%, 99.9%) | | | |
For the BD FecalSwab Collection, Transport and Preservation System, the BD MAX Enteric Viral Panel identified $50.0\%$ and $99.3\%$ of the prospectively collected Sapovirus positive and negative specimens, respectively (refer to Table 12).
Table 12: Sapovirus PPA and NPA of the BD MAX Enteric Viral Panel – FecalSwab Compared to Cary-Blair Preserved
| Sapovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| Prospective | Positive | 1 | 4 | 5 |
| | Negative | \( 1^a \) | 574 | 575 |
| | Total | 2 | 578 | 580 |
| PPA: 50.0% (9.5%, 90.5%) NPA: 99.3% (98.2%, 99.7%) | | | | |
| Retrospective | Positive | 22 | 4 | 26 |
| | Negative | 0 | 185 | 185 |
| | Total | 22 | 189 | 211 |
| PPA: 100.0% (85.1%, 100.0%) NPA: 97.9% (94.7%, 99.2%) | | | | |
${}^{a}$ One missed sample by the FecalSwab was at or near the LoD.
Table 13: Astrovirus PPA and NPA of the BD MAX Enteric Viral Panel – FecalSwab Compared to Cary-Blair Preserved
| Astrovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| Prospective | Positive | 1 | 4a | 5 |
| | Negative | 0 | 575 | 575 |
| | Total | 1 | 579 | 580 |
K220607 - Page 17 of 19
{17}
| Astrovirus | | Cary-Blair | | Total |
| --- | --- | --- | --- | --- |
| Specimen Origin | FecalSwab | Positive | Negative | |
| PPA: 100.0% (20.7%, 100.0%)
NPA: 99.3% (98.2%, 99.7%) | | | | |
| Retrospective | Positive | 26 | 6^{a} | 32 |
| | Negative | 1^{a} | 178 | 179 |
| | Total | 27 | 184 | 211 |
| PPA: 96.3% (81.7%, 99.3%)
NPA: 96.7% (93.1%, 98.5%) | | | | |
aSeven out of eleven of the discrepant samples were at or near the LoD.
These results demonstrate the FecalSwab specimen collection type has comparable and acceptable performance when compared to the predicate Cary-Blair specimen type with all targets in the BD MAX EVP on the BD MAX System.
## Non-Reportable Results:
The rates of unresolved (UNR) results due to sample processing control (SPC) failure and the rates of indeterminate (IND) results due to a BD MAX system failure were also estimated for FecalSwab stool specimens and Cary-Blair stool specimens tested with the EVP assay. The total non-reportable rates (NRR= UNR+IND) observed in the clinical study was 0% (0/581) and 0% (0/581) for FecalSwab and Cary-Blair stool specimens, respectively.
2. Matrix Comparison:
Not applicable.
## C. Clinical Studies:
1. Clinical Sensitivity:
Clinical performance of the BD MAX EVP was established previously with prospective clinical studies. Please see K181427 for additional details.
2. Clinical Specificity:
Clinical performance of the BD MAX EVP was established previously with prospective clinical studies. Please see K181427 for additional details.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
K220607 - Page 18 of 19
{18}
D. Clinical Cut-Off:
Not applicable
E. Expected Values/Reference Range:
The expected values/reference range for analytes on the BD MAX EVP was established previously. See K181427.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220607 - Page 19 of 19
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.