BINAX NOW MALARIA TEST; MODEL 660-000, 660-XXX

DEN070006 · Inverness Medical Professional Diagnostics-Binax · OAX · Jun 13, 2007 · Microbiology

Device Facts

Record IDDEN070006
Device NameBINAX NOW MALARIA TEST; MODEL 660-000, 660-XXX
ApplicantInverness Medical Professional Diagnostics-Binax
Product CodeOAX · Microbiology
Decision DateJun 13, 2007
DecisionDENG
Submission TypePost-NSE
Regulation21 CFR 866.3402
Device ClassClass 2

Indications for Use

The Binax NOW® Malaria Test is an in vitro immunochromatographic assay for the qualitative detection of Plasmodium antigens circulating in human venous and capillary EDTA whole blood of individuals with signs and symptoms of malarial infection. The test targets the histidine-rich protein II (HRPII) antigen specific to Plasmodium falciparum (P.f.) and a pan-malarial antigen, common to all four malaria species capable of infecting humans - P. falciparum, P. vivax (P.v.), P. ovale (P.o.), and P. malariae (P.m.). It is intended to aid in the rapid diagnosis of human malaria infections and to aid in the differential diagnosis of Plasmodium falciparum (P.f.) infections from other less virulent malarial infections. Negative results must be confirmed by thin / thick smear microscopy.

Device Story

In vitro immunochromatographic assay; detects Plasmodium falciparum HRPII antigen and pan-malarial antigen in human venous/capillary EDTA whole blood. Device consists of book-shaped hinged test strip with sample pad, wash pad, and absorbent pad. Sample applied to pad; antigen binds conjugated anti-malaria antibodies; complexes migrate along membrane; captured by immobilized antibodies forming pink-to-purple test lines. Control line confirms validity. Read visually at 15 minutes. Used in clinical laboratory settings by trained personnel. Aids rapid diagnosis and differential diagnosis of P. falciparum vs. other species. Negative results require confirmation by microscopy. Benefits include rapid identification of infection to guide clinical decision-making.

Clinical Evidence

Prospective multi-center study (n=4,122) in endemic regions compared device to Giemsa microscopy. Overall P.f. sensitivity 95.3% (93-97% CI); P.v. sensitivity 68.9% (66-72% CI). Sensitivity correlated with parasitemia levels. Specificity for P.f. 94.2%; P.v. 99.8%. Additional study (n=100) in low-incidence population showed 99% specificity. Clinical performance for P. ovale and P. malariae not adequately established.

Technological Characteristics

Immunochromatographic membrane assay; monoclonal antibodies immobilized on membrane; visualizing particles conjugated to control and anti-malaria antibodies. Form factor: book-shaped hinged test device. Sample type: EDTA whole blood. No electronic components or software.

Indications for Use

Indicated for laboratory diagnosis of malaria in individuals with signs and symptoms consistent with malaria infection. Prescription use only.

Regulatory Classification

Identification

A Plasmodium species antigen detection assay is a device that employs antibodies for the detection of specific malaria parasite antigens, including histidine-rich protein-2 (HRP2) specific antigens, and pan malarial antigens in human whole blood. These devices are used for testing specimens from individuals who have signs and symptoms consistent with malaria infection. The detection of these antigens aids in the clinical laboratory diagnosis of malaria caused by the four malaria species capable of infecting humans: Plasmodium falciparum, Plasmodium vivax, Plasmodium ovale, and Plasmodium malariae, and aids in the differential diagnosis of Plasmodium falciparum infections from other less virulent Plasmodium species. The device is intended for use in conjunction with other clinical laboratory findings.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document:*Plasmodium* species Antigen Detection Assays.” See § 866.1(e) for the availability of this guidance document.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE ### A. 510(k) Number: k061542 - B. Purpose for Submission: De Novo clearance - C. Measurand: Plasmodium antigens ### D. Type of Test: Qualitative, in vitro immunochromatographic assay E. Applicant: Binax, Inc., d/b/a Inverness Medical Professional Diagnostics - F. Proprietary and Established Names: Binax NOW® Malaria ## G. Regulatory Information: a) Regulation section: Plasmodium species antigen detection assay 21 CFR 866.3402. b) Classification: Class II Product Code: OAX c) Panel: 83 Microbiology ## H. Intended Use: - a) Intended use(s): The Binax NOW® Malaria Test is an in vitro immunochromatographic assay for the qualitative detection of Plasmodium antigens circulating in human venous and capillary EDTA whole blood of individuals with signs and symptoms of malarial infection. The test targets the histidine-rich protein II (HRPII) antigen specific to Plasmodium falciparum (P.f.) and a pan-malarial antigen, common to all four malaria species capable of infecting humans - P. falciparum, P. vivax (P.v.), P. ovale (P.o.), and P. malariae (P.m.). It is intended to aid in the rapid diagnosis of human malaria infections and to aid in the differential diagnosis of Plasmodium falciparum (P.f.) infections from other less virulent malarial infections. Negative results must be confirmed by thin / thick smear microscopy. # b) Indication(s) for use: The Binax NOW® Malaria Test is for the laboratory diagnosis of malaria in individuals with signs and symptoms consistent with malaria infection. - c) Special condition for use statement(s): The device is for prescription use only {1}------------------------------------------------ - d) Special instrument Requirements: NA - I. Device Description: - In vitro immunochromatographic immunoassay - J. Substantial Equivalence Information: - N/A De Novo ## K. Standard/Guidance Document Referenced (if applicable): ## L. Test Principle: The Binax NOW® Malaria Test is an immunochromatographic membrane assay that uses monoclonal antibodies to detect Plasmodium falciparum antigen and pan-malarial antigen (an antigen shared by all Plasmodium species causing human malaria) in venous and capillary whole blood specimens. These antibodies, and a control antibody, are immobilized on a membrane support as three distinct lines and are combined with a sample pad, which is impregnated with visualizing particles conjugated to control and anti-malaria antibodies, to create a test strip. This test strip is mounted in a book-shaped. hinged test device, along with wash and absorbent pads, intended to aid in the clearing of the membrane when the device is closed. To perform the test, whole blood is applied to the sample pad. Malarial antigen present in the sample reacts to bind the anti-malaria conjugated antibody. Reagent A is added to the bottom of the test strip and allows the antigen-conjugate complexes to migrate along the test strip, where they are captured by the immobilized antibodies, forming the Test Line(s). Immobilized control antibody captures control conjugate, forming the Control Line. Once the blood sample has migrated the length of the test strip, the device is closed, allowing Reagent A that has been added to the wash pad to clear the test strip of excess blood. Test results are interpreted by the presence or absence of visually detectable pink-topurple colored lines. A positive test result, read in 15 minutes, will include the detection of both a Test Line (or Test Lines) and a Control Line. A negative test result, read in 15 minutes, will produce only a Control Line, indicating that malarial antigens were not detected in the sample. Failure of the Control Line to appear, whether the Test Line(s) is present or not, indicates an invalid result. ## M. Performance Characteristics (if/when applicable): ### Analytical performance: ### a) Precision/Reproducibility: A blind study of the Binax NOW® Malaria Test was conducted at 3 separate sites using panels of blind coded specimens containing negative, limit of detection, and low positive P.f. and P.v. samples. Participants tested each sample multiple times on 3 different days. There was 97% (140/144) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (3 different days), between sites (3 sites), or between {2}------------------------------------------------ operators (6 operators). The overall percent detection of each sample type is summarized below. | Sample Type | % Detection | |-------------------|--------------| | P.f. Low Positive | 94% (17/18) | | P.f. LOD | 97% (35/36) | | P.v. Low Positive | 94% (17/18) | | P.v. LOD | 100% (36/36) | | Negative | 3% (1/36)* | #### Overall Percent Detection of P.f. and P.v. Samples * One operator called a negative sample a P.f. positive. - a. Linearity/assay reportable range: NA - Traceability, Stability, Expected values (controls, calibrators, or method): b. NA - c. Detection limit: #### P.f. and P.v. Limits of Detection: In the study described above. Binax NOW® test clinical limit of detection (LOD) for P.f., defined as the parasitemia level in infected blood that produces positive Binax NOW® test results approximately 95% of the time, was determined to be 1001-1500 parasites per ul and the clinical LOD for P.v. was determined to be 5001-5500 parasites per ul. #### d. Analytical specificity: To determine the analytical specificity of the Binax NOW® Malaria Test, 28 pathogenic microorganisms (7 bacteria, 5 protists and 16 viruses) that may be present in whole blood were tested. All were negative when tested at the concentrations listed below. {3}------------------------------------------------ | TYPE | PATHOGEN TESTED | CONCENTRATION<br>TESTED | |----------|-------------------------------------------------|---------------------------| | Bacteria | Borrelia burgdorferi<br>(N40 strain) | $2.3 x 10^6$ organisms/ml | | | Leptospira interrogans<br>(icterohaemorrhagiae) | $1.0 x 10^7$ organisms/ml | | | Leptospira biflexa<br>(andamana) | $1.0 x 10^7$ organisms/ml | | | Treponema pallidum | $1.0 x 10^5$ organisms/ml | | | Rickettsia conorii<br>(Malish 7) | $1.0 x 10^7$ organisms/ml | | | Rickettsia typhi<br>(Wilmington) | $1.0 x 10^7$ organisms/ml | | | Orientia tsutsugamushi -<br>Rickettsia (Karp) | $1.0 x 10^7$ organisms/ml | | | Babesia microti<br>(RMNS strain) | $4.4 x 10^7$ parasites/ml | | Protists | Trypanosoma cruzi<br>(Y strain) | $1.3 x 10^6$ parasites/ml | | | Leishmania donovani | $1.0 x 10^6$ parasites/ml | | | Leishmania infantum | $1.0 x 10^6$ parasites/ml | | | Leishmania chagasi | $1.0 x 10^6$ parasites/ml | | | Cytomegalovirus (CMV)<br>(AD169) | $1.2 x 10^5$ PFU/ml | | | Epstein-Barr virus (EBV) | $1.1 x 10^4$ copies/ml | | | Dengue virus - West Pac 74 | $1.2 x 10^5$ PFU/ml | | | Dengue virus - S16803 | $3.9 x 10^4$ PFU/ml | | | Dengue virus - CH53489 | $1.3 x 10^4$ PFU/ml | | | Dengue virus - TVP360 | $1.4 x 10^5$ PFU/ml | | | Yellow Fever virus | $7.9 x 10^6$ PFU/ml | | Viruses | West Nile virus | $1.6 x 10^5$ PFU/ml | | | Chikungunya virus | $4.0 x 10^5$ PFU/ml | | | Ross-River virus | $1.0 x 10^6$ PFU/ml | | | Influenza A - Bayern/7/95 | $2.5 x 10^7$ TCID50/ml | | | Influenza B - Victoria/2/87 | $1.0 x 10^7$ TCID50/ml | | | HIV-1 (Subtype B) | $1.4 x 10^5$ copies/ml | | | Hepatitis B | $2.0 x 10^5$ IU/ml | | | Hepatitis C | $1.9 x 10^5$ IU/ml | | | Rubella virus | > $2.0 x 10^2$ TCID50/ml | {4}------------------------------------------------ ## Interference from Exogenous Blood Components: The following substances that may be artificially introduced into whole blood were evaluated in the Binax NOW® Malaria Test at the concentrations listed and were found not to affect test performance. Note: The analytical effects of these drugs on the Binax NOW® test were studied by taking whole blood and spiking it with quantities at high therapeutic concentrations and then testing these samples. | Substance Type | Substance | Concentration | |-------------------------------------------|----------------------------------------------|---------------| | Anti-malarial drugs<br>(prevention) | Mefloquine (Lariam®) | 1 mg/ml | | | Doxycycline* (Vibramycin®) | 1 mg/ml | | | Chloroquine | 1 mg/ml | | | Hydroxychloroquine sulfate | 1 mg/ml | | | Paludrine (Proguanil®) | 1 mg/ml | | | Primaquine | 1 mg/ml | | | Quinine | 1 mg/ml | | | Sulfadoxine and<br>Pyrimethamine (Fansidar®) | 1 mg/ml | | Antibiotic<br>(treatment) | Amoxicillin (Trimox®) | 0.1 mg/ml | | | Cephalexin | 0.1 mg/ml | | | Ciprofloxacin | 0.1 mg/ml | | | Erythromycin | 0.1 mg/ml | | Anti-Inflammatory<br>Drugs<br>(treatment) | Aspirin | 1 mg/ml | | | Acetaminophen | 1 mg/ml | | | Ibuprofen (NSAID) | 1 mg/ml | * Doxcycline is also used as an antibiotic, typically at a lower dose than that tested in this study. ### Interference from Endogenous Blood Components: The Binax NOW® Malaria test was evaluated for possible interference from high levels of endogenous blood components, based on guidelines described in CLSI EP7. EDTA whole blood samples were tested that contained hemoglobin, protein, bilirubin (conjugated and unconjugated) or triglycerides at concentrations above physiological levels. None of the endogenous blood components affected test performance. ## Interference from Unrelated Medical Conditions: To assess the impact of unrelated medical conditions on the specificity of the Binax NOW® Malaria Test, 116 specimens from subjects with a variety of medical conditions unrelated to malaria were tested. Only five (5) of the 116 specimens tested produced a false positive result on the Binax NOW® Test, four (4) from subjects known to be positive for rheumatoid factor and one (1) from a subject with a positive human antimouse antibody (HAMA) titer. Comparison studies: ## a. Method comparison with predicate device: N/A - b. Matrix comparison: {5}------------------------------------------------ ## Clinical studies: #### a. Clinical sensitivity: Clinical Sample Performance – Binax NOW® Malaria Test Sensitivity & Specificity - Endemic Population: The performance of the Binax NOW® test was compared to Giemsa malaria microscopy in a multi-center prospective study conducted in 2001 outside the U.S., in regions considered endemic for malaria. A total of 4,122 whole blood specimens collected from patients presenting with malaria-like symptoms were evaluated on the Binax NOW® test. Microscopy was considered positive only when asexual malaria forms were detected, since asexual forms (not gametocytes) are indicative of active infection. Forty-four percent (1.796/4.122) of the tested population was microscopy positive for malaria, including 557 patients with P.f., 1,187 with P.v., 16 with P.m., 2 with P.o., and 34 with mixed P.f./P.v. infections. Fifty-nine percent of patients were male, 41% female, 19% pediatric (<18 years) and 81% adult (≥18 years). Binax NOW® test performance for detection of the individual malaria species and for mixed P.f./P.v. infections is summarized below. No differences in Binax NOW® Malaria Test performance were observed based on patient age or gender. Binax NOW® test specificity for P.f. trends slightly lower (89.4%) in the 5% of patients who were on anti-malarial drug therapy, than in patients not receiving therapy (94.4%), but does not achieve statistical significance. ### Detection of P.f. Infection Binax NOW® test sensitivity and specificity for detection of P.f. vs. microscopy is presented below. Sensitivity was evaluated based on the levels of parasitemia (parasites per ul) observed in microscopy. | SENSITIVITY FOR P.f. | | | |-----------------------|----------------------|-----------| | Parasitem<br>ia Level | % Sensitivity | 95%CI | | > 5000 | 99.7% (326 /<br>327) | 98 - 100% | | 1000 -<br>5000 | 99.2% (126 /<br>127) | 96 - 100% | | 500 -<br>1000 | 92.6% (25 / 27) | 76 - 99% | | 100 - 500 | 89.2% (33 / 37) | 75 - 97% | | 0-100 | 53.9% (21 / 39) | 37 - 70% | | Overall | 95.3% (531 /<br>557) | 93 - 97% | ## Binax NOW® Malaria Test Sensitivity and Specificity for P.f. vs. Microscopy {6}------------------------------------------------ | SPECIFICITY FOR P.f. | | |----------------------|--------| | % Specificity | 95% CI | | 94.2% (3297 / 3500) | 93-95% | ### Detection of P.v. Infection Binax NOW® test sensitivity and specificity for detection of P.v. vs. microscopy is presented below. Sensitivity was evaluated based on the levels of parasitemia (parasites per ul) observed in microscopy. There were 68 samples generating two Binax NOW® test lines that were microscopy positive for P.v. only. When these samples are included in the true positive calculation, Binax NOW® test sensitivity for overall detection of P.v. increases from 68.9% to 74.6% (886/1,187). ## Binax NOW® Malaria Test Sensitivity and Specificity for P.v. vs. Microscopy | SENSITIVITY FOR P.v. | | | |-----------------------|--------------------|-------------| | Parasitem<br>ia Level | % Sensitivity | 95%CI | | > 5000 | 93.5% (462 / 494) | 91 -<br>96% | | 1000 -<br>5000 | 81.0% (277 / 342) | 76 -<br>85% | | 500 -<br>1000 | 47.4% (37 / 78) | 36 -<br>59% | | 100 - 500 | 23.6% (34 / 144) | 17 -<br>31% | | 0 - 100 | 6.2% (8 / 129) | 3 -<br>12% | | Overall | 68.9% (818 / 1187) | 66 -<br>72% | | SPECIFICITY FOR P.v. | | | | | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--| | % Specificity | તે જેન્જી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો | | | | | | CI | | | | | 99.8% (2863 / | 99- | | | | | 2870) | 100% | | | | ### Detection of P.m. and P.o. Infection Binax NOW® test sensitivity was 43.8% (7/16) for detection of P.m. and 50% (1/2) for detection of P.o. When five P.m. microscopy positive samples that generated two test {7}------------------------------------------------ lines in the Binax NOW® test are included in the true positive calculation, Binax NOW® test sensitivity for P.m. increases from 43.8% to 75.0% (12/16). #### Detection of Mixed P.f./P.v. Infection Thirty four samples were both P.f. and P.v. positive by microscopy, based on the detection of asexual forms of both species. The Binax NOW® test detected 32 of these samples by generating both test lines, for a sensitivity of 94.1% (95% CI of 81-98%). #### b. Clinical specificity: The performance of the Binax NOW® test was compared to Giemsa malaria microscopy in a prospective study conducted in the eastern US in 2006-2007. One hundred (100) whole blood specimens collected from febrile patients were evaluated on the Binax NOW® test and on microscopy. All 100 samples were negative for malaria on microscopy, and 99 of these samples generated negative Binax NOW® test results, yielding a specificity of 99% (99/100) in this low incidence population. Binax NOW® test specificity versus microscopy is presented below. ## Binax NOW® Malaria Test Specificity vs. Microscopy | | - / - | + / - | % Spec | 95% CI | |------------------|-------|-------|--------|---------| | P.f. | 100 | 0 | 100% | 96-100% | | P.v., P.o., P.m. | 99 | 1 | 99% | 95-100% | #### N. Proposed labeling: The labeling is sufficient and it satisfies the requirement of 21 CFR Part 809.10. #### WARNING This test should only be used by laboratories that have or can acquire blood samples containing Plasmodium falciparum for use as a positive control. It is recommended that the level of the positive control used challenge the assay cutoff. Clinical performance has not been adequately established for P. ovale (P.o.) and P. malariae (P.m.). The user must establish performance characteristics of this test with these Plasmodium species. The test is not intended for use in screening asymptomatic populations. #### O. Conclusion: The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...