K071146 · bioMerieux, Inc. · NTM · Oct 11, 2007 · Microbiology
Device Facts
Record ID
K071146
Device Name
VIDAS B.R.A.H.M.S. PCT ASSAY
Applicant
bioMerieux, Inc.
Product Code
NTM · Microbiology
Decision Date
Oct 11, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3210
Device Class
Class 2
Indications for Use
VIDAS® BRAHMS PCT is an automated test for use on the VIDAS instruments for the determination of human procalcitonin in human serum or plasma (lithium heparin) using the ELFA (Enzyme-Linked Fluorescent Assay). VIDAS BRAHMS PCT is intended for use in conjunction with othe laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission, for progression to severe sepsis and septic shock.
Device Story
The VIDAS BRAHMS PCT is an automated enzyme-linked fluorescent immunoassay (ELFA) performed on VIDAS instruments. It measures procalcitonin (PCT) concentration in human serum or plasma. The device uses a Solid Phase Receptacle (SPR) as both a solid phase and pipetting device. The assay follows a one-step sandwich immunoassay principle: sample and alkaline phosphatase-labeled anti-PCT antibodies are cycled in the SPR; antigen binds to immunoglobulins on the SPR wall and conjugate; unbound compounds are washed away. A substrate (4-Methylumbelliferyl phosphate) is added; the enzyme catalyzes hydrolysis into a fluorescent product (4-Methyl-umbelliferone) measured at 450 nm. Fluorescence intensity is proportional to PCT concentration. Results are automatically calculated via calibration curves and printed. Used in clinical laboratory settings by trained personnel to assist clinicians in assessing the risk of severe sepsis or septic shock in critically ill patients upon ICU admission.
Clinical Evidence
Clinical study of 232 ICU patients (143 male, 89 female; age 18-92). Patients categorized by ACCP/SCCM criteria. Sensitivity/specificity evaluated at 0.5 ng/mL and 2.0 ng/mL cut-offs. At 0.5 ng/mL: 86/104 severe sepsis/septic shock patients correctly identified. At 2.0 ng/mL: 67/104 severe sepsis/septic shock patients correctly identified. Method comparison with predicate (n=204) showed 97.1% concordance at 0.5 ng/mL and 94.1% at 2.0 ng/mL.
Technological Characteristics
Enzyme-Linked Fluorescent Assay (ELFA). Materials: mouse monoclonal anti-human PCT immunoglobulins coated on SPR. Energy: automated instrument-based fluorescence detection at 450nm. Connectivity: standalone VIDAS/miniVIDAS instruments. Calibration: Master Lot Entry (MLE) card for lot-specific master curve, plus two-point recalibration. Sample volume: 200 μL. Assay time: ~20 minutes.
Indications for Use
Indicated for the determination of human procalcitonin in human serum or plasma (lithium heparin) to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.
Regulatory Classification
Identification
An endotoxin assay is a device that uses serological techniques in whole blood. The device is intended for use in conjunction with other laboratory findings and clinical assessment of the patient to aid in the risk assessment of critically ill patients for progression to severe sepsis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance entitled “Class II Special Controls Guidance Document: Endotoxin Assay.” See § 866.1(e) for the availability of this guidance document.
Predicate Devices
BRAHMS PCT LIA Assay
Related Devices
K160911 — VIDAS B.R.A.H.M.S. PCT (PCT) · bioMerieux, Inc. · Jun 28, 2016
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K071146
B. Purpose for Submission:
Substantial equivalence for the VIDAS BRAHMS PCT which is an automated test for use on the VIDAS instruments for the determination of human procalcitonin (PCT) in human serum or plasma
C. Measurand:
Procalcitonin
D. Type of Test:
Enzyme - linked Fluorescent Assay
E. Applicant:
bioMerieux Inc.
F. Proprietary and Established Names:
VIDAS BRAHMS PCT Assay
G. Regulatory Information:
1. Regulation section:
21CFR 866.3610, Endotoxin activity
2. Classification:
Class II
3. Product code:
NTM – Antigen, inflammatory response marker, sepsis
{1}
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use:
The VIDAS BRAHMS PCT is an automated test for use on the VIDAS instruments for the determination of human procalcitonin in human serum or plasma (lithium heparin) using the Enzyme-Linked Fluorescent Assay (ELFA) technique. The VIDAS BRAHMS PCT is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.
2. Indications for use:
The VIDAS BRAHMS PCT is an automated test for use on the VIDAS instruments for the determination of human procalcitonin in human serum or plasma (lithium heparin) using the Enzyme-Linked Fluorescent Assay (ELFA) technique. The VIDAS BRAHMS PCT is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission for progression to severe sepsis and septic shock.
3. Special conditions for use statement:
For prescription use
4. Special instrument requirements:
VIDAS and mini VIDAS instruments
I. Device Description:
The VIDAS BRAHMS PCT kit consists of 60 tests. The contents of the kit are 60 PCT strips (each strip consists of 10 wells), 60 ready to use solid phase receptacles (SPR) coated with mouse monoclonal anti-human procalcitonin immunoglobulins, 2 lyophilized PCT controls C1 and C2, TRIS NaCl buffer, recombinant human PCT and preservatives, 2 lyophilized PCT calibrators S1 and S2 and a master lot entry card (MLE).
J. Substantial Equivalence Information:
1. Predicate device name:
BRAHMS PCT LIA
2. Predicate 510(k) number:
{2}
K040887
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Determination of PCT in human serum or plasma | Same |
| Specimen | Human serum or plasma | Same |
| Analyte | PCT | Same |
| Antibody | Anti-PCT antibody (monoclonal mouse) | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Principle | Immunoassay based on sandwich principle – automated assay | Immunoluminometric assay based on sandwich principle – non automated assay |
| Detection method | Fluorescence of 4 methyl umbelliferyl measured at 450nm | Luminescence signal measurement via luminometer |
| Instrument requirements | VIDAS & miniVIDAS | Luminometer |
| Sample volume | 200 μl | 20 μl |
| Assay Time | Approx 20 mins | Approx 90 mins |
| Measurement range | 0.05 to 200 ng/ml | 0.3 to 500 ng/ml |
K. Standard/Guidance Document Referenced (if applicable):
N/A
L. Test Principle:
The VIDAS BRAHMS PCT assay combines a one-step immunoassay sandwich method with a final fluorescent detection. The solid phase receptacle whose interior is coated with mouse monoclonal anti procalcitonin immunoglobulins, serves as the solid phase as well as the pipetting device. Assay reagents are ready to use and pre dispensed in the sealed reagent strips. The instrument performs all assay steps automatically. The sample is transferred into the wells containing the anti procalcitonin antibodies labeled with alkaline phosphatase
{3}
(conjugate). The sample/conjugate mixture is cycled in and out of the SPR several times. The antigen binds with the immunoglobulins to form a sandwich. Washing steps eliminate unbound compounds. Two detection steps are then performed successively. During each step, the substrate is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product, the fluorescence of which is measured at 450nm. The intensity of the fluorescence is proportional to the concentration of antigen present in the sample. Results are automatically calculated by the instrument in relation to two calibration curves corresponding to the two detection steps. A fluorescence threshold value determines the calibration curve to be used for each sample
## M. Performance Characteristics:
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Six serum samples were tested in duplicate in 20 different runs (2 runs per day) with 2 reagent lots using the same instrument at 3 sites. The repeatability (intra-run precision), inter-run reproducibility (inter-run precision), inter-site reproducibility (inter-site precision) and inter-lot reproducibility (total precision = intra-run, inter-run, inter-day, inter-site, inter-lot) were calculated based on recommendations from CLSI EP 5-A2 document. Total precision ranged from 6.17 – 15.31%CV; intra-run precision from 1.93 – 4.61% CV; inter-run precision from 3.57 – 7.04% CV and inter-site precision from 4.21 – 11.40%CV.
#### b. Linearity/assay reportable range:
A dilution study was performed on 2 verification lots to validate the dilution claimed in the package insert. With the diluent provided in the VIDAS BRAHMS PCT kit, dilutions up to 1/20 are acceptable. Recovery percentages up to 1/20 are within the acceptable range of 80 to 120%. On the studied range [0.11-229ng/ml], a PCT negative sample can be used as a diluent for samples.
A second linearity study was also performed as per CLSI EP 6-A. Evaluation of the linearity of Quantitative Measurement Procedures, Vol. 23, No. 16. Two verification lots were used and high and low PCT level samples were diluted in various proportions to obtain nine dilutions (1/10 to 9/10). Each sample and each dilution were tested once in 3 separate runs. The assay was linear over the studied range [0.07 - 218 ng/ml]
No Hook Effect was found up to procalcitonin concentrations of 2,600 ng/ml
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Studies were performed to test kit stability; calibrator and control stability as well as specimen stability (serum and plasma tubes).
4
{4}
Kit stability results provided were within the expected ranges for up to 9 months for 2 verification lots at 2-8°C. Testing is on going to obtain 12 month stability data.
Stability studies were also conducted using calibrators and controls. Study results show that after reconstitution, calibrators and controls are stable at 11 months of storage at -25±6°C. At 2-8°C calibrators and controls are stable at 2 months of storage. At 18-25 °C, calibrators and controls are stable at 2 weeks of storage and are also stable after 6 freeze/thaw cycles.
For the different sampling tubes (serum and plasma) stability studies were performed to determine the storage conditions at various temperatures, storage times and after multiple freezing/thawing cycles. At 2-8°C, storage at 2 and 3 days did not impact the results. Results were also acceptable for storage at 7 months at -25 ± 6°C and for 2 months at less than -60°C. Samples were also stable after 4 freeze/thaw cycles.
d. Detection limit:
The analytical detection limit is stated as <0.05 ng/ml. Recombinant human PCT was diluted at various concentrations in a pool of human serum. 2 verification lots were tested. The analytical detection limit was defined as the lowest concentration of PCT that can be distinguished from standard zero. The limit for Lot 1 was 0.012 ng/ml and for Lot 2 was 0.011 ng/ml
e. Analytical specificity:
No significant interference was noted with bilirubin at 574μmol/l, hemoglobin at 347 μmol/l and triglycerides at 30g/l.
No significant interference was noted with protein(albumin) at 4g/dl, human calcitonin at 60 ng/ml, human katacalcin at 10 ng/ml, human alpha-CGRP at 10μg/ml and human beta-CGRP at 10μg/ml.
No significant interference was noted with imipenem at 0.5mg/ml, cefotaxime at 180 mg/dl, vancomycin at 3 mg/ml, dopamine at 26mg/dl, noradrenalin at 4μg/ml, dobutamine at 22.4 μg/ml, heparin at 16,000U/L and furosemide at 4mg/dl.
f. Assay cut-off:
The assay cut offs are:
(a) >2 ng/ml = high risk of severe sepsis and/or septic shock
(b) <0.5 ng/ml = low risk of severe sepsis and septic shock
{5}
6
2. Comparison studies:
a. Method comparison with predicate device:
A comparison study was performed between the VIDAS BRAHMS PCT and the BRAHMS PCT LIA using 204 samples. The percentages of concordance between the two assays at cut offs 0.5 ng/ml and 2 ng/ml are 97.1% and 94.1%
| VIDAS BRAHMS PCT | BRAHMS PCT LIA | → | → |
| --- | --- | --- | --- |
| | ≤ 0.5 ng/ml | >0.5ng/ml | Total |
| ≤ 0.5ng/ml | 74 | 1 | 75 |
| > 0.5 ng/ml | 5 | 124 | 129 |
| Total | 79 | 125 | 204 |
| VIDAS BRAHMS PCT | BRAHMS PCT LIA | → | → |
| --- | --- | --- | --- |
| | ≤ 2 ng/ml | >2ng/ml | Total |
| ≤ 2 ng/ml | 109 | 4 | 113 |
| > 2 ng/ml | 8 | 83 | 91 |
| Total | 117 | 87 | 204 |
b. Matrix comparison:
A matrix comparison study was performed using serum and plasma samples. Serum reference tube, serum with gel or silica, heparin plasma and EDTA plasma tubes were used. Thirty two blood donors were tested using each tube type and each sampling tube was spiked with PCT. No significant difference was noted between the PCT concentrations of the serum samples or plasma samples except for the EDTA tubes. A decrease of PCT concentration was detected with plasma samples from EDTA tubes. A limitation statement to this effect is included in the package insert.
3. Clinical studies:
a. Clinical Sensitivity:
Clinical studies were conducted on 232 patients on their first day of admission to the ICU. Patients admitted for trauma, surgery, burns or prolonged or severe cardiogenic shock were excluded. Patients were classified into 5 categories based on the ACCP/SCCM consensus criteria namely no infection, SIRS, sepsis, severe sepsis and septic shock. There were 143 males and 89 females with age ranges from 18 to 92 yrs. Results at both cut offs were as follows:
Results with a cut-off at 0.5ng/ml
{6}
7
| | No infection/SIRS/sepsis | Severe sepsis/septic shock | Total |
| --- | --- | --- | --- |
| PCT≤ 0.5ng/ml | 74 | 18 | 92 |
| PCT> 0.5ng/ml | 54 | 86 | 140 |
| Total | 128 | 104 | 232 |
Results with a cut off at 2.0 ng/ml
| | No infection/SIRS/sepsis | Severe sepsis/septic shock | Total |
| --- | --- | --- | --- |
| PCT≤ 2ng/ml | 98 | 37 | 135 |
| PCT>2ng/ml | 30 | 67 | 97 |
| Total | 128 | 104 | 232 |
b. Clinical specificity:
See 3(a) above
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
4. Clinical cut-off:
A total of 204 samples from patients admitted to the ICU were tested on both the VIDAS BRAHMS PCT and the BRAHMS PCT LIA. Concordance between both methods at the cut off of 0.5 and 2.0 ng/ml was excellent. The Passing & Bablok slopes showed some variability but the total concordance at the relevant clinical cut-offs was good.
5. Expected values/Reference range:
In agreement with the literature, results obtained with the VIDAS PCT kit during a study performed on patients admitted to ICU showed that:
<0.5 ng/ml represents a low risk of severe sepsis and/or septic shock
>2 ng/ml represents a high risk of severe sepsis and/or septic shock
Concentrations <0.5 ng/ml do not exclude an infection or a systemic infection in its initial stages. Increased procalcitonin can occur without infection. PCT concentrations between 0.5 and 2.0 ng/ml can be interpreted taking into account the patient's history
N. Instruments Name:
The VIDAS PC and the miniVIDAS
{7}
8
O. System Descriptions:
1. Modes of Operation:
The VIDAS PC instrument cleared in 1989 under K891385 is attached to a computer and a printer. Each instrument has 5 independent sections allowing 5 different assays to be run simultaneously. Each section can process up to six samples. When fully loaded the device can process up to 30 samples. The miniVIDAS is a smaller compact version of the VIDAS PC and was cleared under K923579 in 1993. It has a built in computer, keyboard and printer and is used most frequently in physician's office laboratories. Two independent sections each accept six tests so twelve samples can be processed simultaneously. The VIDAS assay can be run on either instrument.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ or No ☑
3. Specimen Identification:
All assay steps are controlled automatically by the instrument. The sample is transferred into the wells. The PCT strip consists of ten wells covered with labeled foil seal. The label comprises a bar code which indicates the assay code (PCT), kit lot number and expiration date.
4. Specimen Sampling and Handling:
The solid phase receptor (SPR) serves as both the solid phase and the pipetting device. The foil of the 1st well is perforated to allow introduction of the sample into well 1. The last well (well 10) of each strip is a cuvette in which the fluorometric reading is performed. The center wells of the strip contain the various reagents required for the assay
5. Calibration:
The kit contains 2 PCT controls and 2 PCT calibrators See Sec. I and M 1(c).
6. Quality Control:
The Master curve is established at time of manufacture for each lot of reagents. It is provided with each test kit and is entered into the VIDAS instrument using the Master Lot Entry card (MLE) included in the kit. Data from the MLE card is entered once for each lot of reagents. Each lab establishes its own calibration curve (recalibration) based on the mathematical master curve data and the test results of two calibrators tested in
{8}
duplicate by the lab. Recalibration serves to control for minor variations in assay signal from one VIDAS instrument to another and is therefore specific for each instrument.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
N/A
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
9
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.