COPALIS EBV-M ANTIBODY ASSAY

K991459 · DiaSorin, Inc. · LJN · May 14, 1999 · Microbiology

Device Facts

Record IDK991459
Device NameCOPALIS EBV-M ANTIBODY ASSAY
ApplicantDiaSorin, Inc.
Product CodeLJN · Microbiology
Decision DateMay 14, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3235
Device ClassClass 1
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The Copalis® EBV-M Antibody Assay uses Coupled Particle Light Scattering technology in a microparticle agglutination-based assay for the qualitative detection of IgM antibodies to the EBV VCA antigen. The assay is designed for human serum using the Copalis I® Immunoassay System. The presence of VCA IgM antibodies is used as an aid in the diagnosis of EBV associated mononucleosis when used in coniunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations.

Device Story

The Copalis® EBV-M Antibody Assay is an in vitro diagnostic test for human serum. It utilizes Coupled Particle Light Scattering (Copalis®) technology on the Copalis I® Immunoassay System. Polystyrene microparticles coated with synthetic VCA antigen are contained in a reaction well; dried reagent is reconstituted with buffer. Patient sample is added and incubated for 10 minutes. If IgM antibodies to VCA are present, microparticles agglutinate, reducing the monomer population. The reaction mixture passes through a flow cell; the instrument measures light scattering to determine residual monomer concentration. This concentration is compared to a cutoff value to determine reactivity. The assay is performed in clinical laboratories. Results aid clinicians in diagnosing EBV-associated mononucleosis by identifying primary infection or confirming serological status in conjunction with other markers (IgG, EBNA, EA, heterophile).

Clinical Evidence

Clinical evaluation included prospective and retrospective analysis of fresh (12%) and frozen (88%) samples from pediatric, adult, transplant donor, and recipient populations. Primary infection sensitivity was 100% (12/12 pediatric; 59/59 adult). Specificity in seronegative populations was 95.2% (pediatric) and 100% (adult). Reactivated disease specificity was 74.4%. Serial sample studies (n=7) showed Copalis® closely matched expected marker patterns and, in 3/7 cases, detected IgM for a longer duration than IFA. Reproducibility was assessed across 3 sites with 6 samples; %CV ranged from 1.3% to 26.7%.

Technological Characteristics

Uses Coupled Particle Light Scattering technology. Components: Polystyrene microparticles coated with synthetic VCA antigen. Instrument: Copalis I® Immunoassay System. Principle: Microparticle agglutination measured via light scattering in a flow cell. Form factor: Test cup with reaction well. Connectivity: Standalone instrument. Software: Automated analysis of monomer concentration vs. cutoff values.

Indications for Use

Indicated for the qualitative detection of IgM antibodies to the Epstein-Barr Virus (EBV) VCA antigen in human serum. Used as an aid in the diagnosis of EBV-associated mononucleosis in pediatric, adult, transplant donor, and transplant recipient populations, in conjunction with other EBV serologies.

Regulatory Classification

Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K991459 MAY 1 4 1999 May 5, 1999 ## 510 (k) SUMMARY Judith J. Smith SUBMITTED BY: DiaSorin, Inc. 9175 Guilford Rd. Suite 100 Columbia, MD 21046 NAME OF DEVICES: Copalis® EBV-M Antibody Assay Trade Name: Immunoassay for the Detection of Common Names/Descriptions: lqM Antibodies to VCA antigen of Epstein Barr Virus EBV Serology Test Classification Names: PREDICATE DEVICES: Gull Laboratories EBV IgM IFA ## DEVICE DESCRIPTION: INTENDED USE: The Copalis® EBV-M Antibody Assay uses Coupled Particle Light Scattering technology in a microparticle agglutination-based assay for the qualitative detection of IgM antibodies to the EBV VCA antigen. The assay is designed for human serum using the Copalis I® Immunoassay System. The presence of VCA IgM antibodies is used as an aid in the diagnosis of EBV associated mononucleosis when used in coniunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. KIT DESCRIPTION: Coupled Particle Light Scattering (Copalis®) technology provides a rapid method for the measurement of antibodies to specific pathogens. The Copalis® EBV-M Antibody Assay is a microparticle agglutination test using the Copalis® light scattering technology. Polystyrene microparticles are coated with svnthetic VCA antigen and are contained within a special covered reaction well in the test cup. The dried reagent is reconstituted with a reaction buffer on the instrument at the start of the assay. Patient sample is added to the reaction mixture and incubated for 10 minutes. The presence of IgM antibodies specific to the VCA antigen in the patient sample results in agglutination of the monomer microparticles to form aggregates. The reaction mixture is passed through a flow cell and the instrument uses light scattering technology to measure the monomer concentration. The decrease in the monomer population resulting from agglutination is related to the amount of antibody in the sample. The residual monomer concentration in each reaction mixture is compared to a cutoff values to determine sample reactivity and nonreactivity. ## PERFORMANCE DATA: Clinical Correlation: Fresh (12%) and frozen (88%) samples were analyzed at two clinical laboratories and at DiaSorin. Patients from the disease states defined below and representing the eastern, midwestern and western United States were tested. {1}------------------------------------------------ The screening population is a group of samples from patients suspected of disease. The samples were chosen for each disease state population based upon comparison to expected patterns of serological testing results for the 4 EBV markers (EBV VCA IgG and IgM, EBNA, and EA), and for primary infection, heterophile test . For primary infection, the first level of inclusion/exclusion was based on IFA results for VCA IgM, VCA IgG and EBNA. Samples with positive results for the VCA antigen and negative results for the EBNA were included and further examined. The next level of inclusion/exclusion criteria was based upon the heterophile results. A positive result is expected for this test in acute EBV infection. Samples with positive VCA IgM, VCA IgG, and heterophile test and negative EBNA result were included in the primary infection population. If a negative result was obtained on the heterophile test, the results of the EA IFA test were examined. A positive EA IFA resulted in inclusion of the sample. A negative EA IFA result resulted in exclusion of the sample. A summary of the expected marker patterns for the EBV markers is summarized below. | Antibody | Seronegative | Primary | Reactivated | Past | |--------------|--------------|---------|-------------|------| | IgM anti-VCA | - | + | - | - | | IgG anti-VCA | - | + | + | + | | Anti EBNA | - | - | + | -/+ | | Anti-EA | - | + | + | + | | Heterophile | - | + | N/A | N/A | ## Summary of Expected Patterns for EBV Antigen Reactivity The age of the primary disease population varied from 1 to 48 years of age (mean and median age: 22 years old). The age of the patients with reactivated disease varied from 3 to 68 (mean: 27, median 23). The age range of the seronegative group was 1 to 74 (mean: 14, median: 12). The Copalis EBV-M Antibody Assay was compared to expected marker patterns in the defined disease states and separated into adult and pediatric populations. The results of these studies are summarized below with the 95% confidence intervals (95% CI). Equivocal results by Copalis or IFA or non-specific staining by IFA were not included in the calculations. | Expected Pattern | Heterophile +; VCA IgM +; VCA IgG +; EBNA –; EA + | | |----------------------|---------------------------------------------------|-----------------| | | Pediatric | Adult | | Copalis®/IFA Result* | EBV VCA IgM | EBV VCA IgM | | Sensitivity | 12/12 = 100% | 59/59* = 100.0% | | (95% CI) | (73.5 - 100.0%) | (93.9 - 100.0%) | | Specificity | NA | NA | | Agreement | 12/12 = 100% | 59/59 = 100.0% | | Prevalence Copalis® | 12/12 = 100% | 59/61 = 96.7% | | Prevalence IFA | 12/12 = 100% | 61/61 = 100.0% | #### Primary Disease State Population *(2 Copalis equivocal/IFA positive) {2}------------------------------------------------ ### Reactivated Disease State Population | Expected Pattern | VCA IgM a; VCA IgG +; EBNA +; EA + | |----------------------|------------------------------------| | Copalis®/IFA Resultb | EBV VCA IgM | | Sensitivity | NA | | Specificity | 29/39 = 74.4% | | (95% CI) | (57.8 – 87.0%) | | Agreement | 29/39 = 74.4% | | Prevalence Copalis® | 10/39 = 25.6% | | Prevalence IFAc | 0/39 = 0% | 3 Literature reports show that IgM can be present in reactivated disease ' °2 Copalis equivocal/IFA negative, 1 Copalis equivocal/IFA non-specific staining (nss); 1 Copalis negative/IFA nss °Prevalence of IFA before exclusion was 4/50 = 8% #### EBV Screening Population | Expected Pattern | VCA IgM -/+<br>(Primary, Reactiviated, Seronegative, or Past | | |-------------------------|--------------------------------------------------------------|-------------------------------| | | Pediatric | Adult | | Copalis®/IFA* Result | EBV VCA IgM | EBV VCA IgM | | Sensitivity<br>(95% CI) | N/A | N/A | | Specificity<br>(95% CI) | 13/13 = 100.0%<br>(75. - 100.0%) | 25/28* = 89.3%<br>(71.8-97.7) | | Agreement | 13/13 = 100.0% | 25/28 = 89.3% | | Prevalence Copalis® | 0/14 = 0.0% | 3/29 = 10.3% | | Prevalence IFA | 0/14 = 0.0% | 0/29 = 0.0% | (1 Copalis equivocal/IFA negative) ## SeroNegative Population | Expected Pattern | VCA IgM -; VCA IgG -; EBNA -; EA - | | |-------------------------|------------------------------------|-------------------------------| | | Pediatric | Adult | | Copalis®/IFA* Result | EBV VCA IgM | EBV VCA IgM | | Sensitivity<br>(95% Cl) | NA | NA | | Specificity<br>(95% Cl) | 40/42 = 95.2%<br>(83.8 - 99.4%) | 8/8 = 100%<br>(63.1 - 100.0%) | | Agreement | 40/42 = 95.2% | 8/8 = 100.0% | | Prevalence Copalis® | 2/42 = 4.8% | 0/8 = 0% | | Prevalence IFA | 0/42 = 0% | 0/8 = 0% | *(no equivocal results) ## Apparently Healthy Adult Population | Expected Pattern | - | |----------------------|----------------| | Copalis®/IFA* Result | EBV VCA IgM | | Sensitivity | NA | | Specificity | 91/97 = 93.8% | | (95% CI) | (82.0 - 97.8%) | | Agreement | 91/97 = 93.8% | | Prevalence Copalis® | 6/97 = 6.2% | | Prevalence IFA | 0/97 = 0% | *(5 Copalis equivocal/IFA negative) {3}------------------------------------------------ #### Transplant Recipients (Reactivated) | Copalis®/IFA* Result | | |----------------------|----------------| | Sensitivity | NA | | Specificity | 30/41 = 73.2% | | (95% CI) | (57.1 - 85.8%) | | Agreement | 3/42 = 71.4% | | Prevalence Copalis® | 10/42 = 23.8% | | Prevalence IFA | 4/46 = 8.7% | *(3 Copalis equivocal/IFA positive; 3 Copalis equivocal/IFA negative; 1 Copalis negative/IFA nss; 1 Copalis equivocal/IFA nss) #### Transplant Donors | Copalis®/IFA* Result | | |----------------------|----------------| | Sensitivity | NA | | Specificity | 46/50 = 92.0% | | (95% CI) | (80.8 - 97.8%) | | Agreement | 56/50 = 92.0% | | Prevalence Copalis® | 4/46 = 8.7% | | Prevalence IFA | 0/53 = 0% | *(4 Copalis equivocal/IFA negative) The Copalis® EBV-M Antibody Assay identified all patients with primary disease. It gave equivalent or better prevalence results than IFA in defined disease states when compared to the literature. Review of the data from serial samples (next section) shows that the Copalis® assay closely matches the expected pattern for the markers in primary infection (acute and convalescent). In the reactivated disease state population, the IFA result was negative in 10 samples in which the Copalis® assay results were positive. Although most tables of marker patterns in defined diseases state that the lgM should be negative in reactivated disease, literature reports show that IgM can be present in reactivated disease. Thus, the presence of IgM in this state is not unexpected. ### Serial Sample Studies Seven serial sample panels were evaluated by Copalis® EBV-M Antibody Assay and by IFA. The panels were also tested for EBV VCA, EBNA and EA antibodies. The samples ranged from 2 to 960 days following onset of illness. The Copalis® assay was equivalent to IFA in the detection of marker patterns. In 3 of 7 patients, the Copalis® EBV-M assay detected the presence of IgM for more days than detected by IFA. In one sample, the Copalis® EBV-M assay and IFA detected IgM for the same number of days. The assay CI's show changes in IgM antibody titers more clearly than IFA. When reviewing all 4 markers together, in 6/7 patients the Copalis® assay pattern was more consistent with disease than IFA. {4}------------------------------------------------ Reproducibility: Reproducibility studies were performed at the 3 sites using one lot of tests. Assay reproducibility was determined by testing 6 samples that spanned the range of the assay components CTRs. Samples were tested in duplicate once a day for 5 days. The results are summarized below. | Sample | Mean<br>mg/dL | Within Run<br>%CV | %CV | |-------------|---------------|-------------------|-------| | Neg Control | 0.90 | -- | 1.3% | | Pos Control | 2.25 | -- | 10.7% | | RP 1 | 1.87 | 7.8% | 12.6% | | RP 2 | 0.89 | 2.0% | 1.7% | | RP 3 | 5.06 | 8.4% | 19.3% | | RP 4 | 0.90 | 0.9% | 1.4% | | RP 5 | 5.53 | 19.4% | 26.7% | | RP 6 | 1.35 | 7.6% | 13.2% | | N for RP | 30 | 30 | 30 | REPRODUCIBILITY RESULTS FOR COPALIS® EBV-M ANTIBODY ASSAY - COMBINED SITES {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 1 4 1999 DiaSorin c/o Carole Stamp TUV Product Service Inc. 1775 Old Highway 8 NW Suite 104 New Brighton, MN 55112 K991459 Re: Trade Name: Copalis® EBV-M Assay Regulatory Class: I Product Code: LJN Dated: May 7, 1999 Received: May 10, 1999 Dear Ms. Stamp: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {6}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a Iegally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number (if known): K991459 Copalis® EBV-M Assay Device Name: Indications For Use: The Copalis® EBV-M Assay uses Coupled Particle Light Scattering technology in a microparticle agglutination-based assay for the qualitative detection of IgM antibodies to the EBV VCA antigen. The assay is designed for human serum using the Copalis® I Immunoassay System. The presence of VCA IgM antibodies is used as an aid in the diagnosis of EBV associated mononucleosis when used in conjunction with other EBV serologies in pediatric, adult, transplant donor and transplant recipient populations. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ### Concurrence of CDRH, Office of Device Evaluation (ODE) Woody Dubois k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use _ //////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////// (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...