Access Syphilis

K241427 · Beckman Coulter, Inc. · LIP · Sep 6, 2024 · Microbiology

Device Facts

Record IDK241427
Device NameAccess Syphilis
ApplicantBeckman Coulter, Inc.
Product CodeLIP · Microbiology
Decision DateSep 6, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3830
Device ClassClass 2

Indications for Use

The Access Syphilis assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative detection of total antibodies to Treponema pallidum in human serum and plasma using the Access Immunoassay Systems. It is intended to be used as an aid in the diagnosis of syphilis or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. The Access Syphilis assay is not intended for blood and tissue donor screening.

Device Story

Access Syphilis is a two-step enzyme immunoassay for qualitative detection of total antibodies to Treponema pallidum in human serum and plasma. Input samples are processed on Beckman Coulter Access Immunoassay Systems. The assay uses paramagnetic particles coated with recombinant Tp17 and Tp47 antigens and biotinylated Tp17/Tp47 antigens. After incubation, magnetic fields separate bound materials; alkaline phosphatase conjugates are added, followed by a chemiluminescent substrate. A luminometer measures light production, which is proportional to antibody concentration. Results are reported as Reactive, Non-reactive, or S/CO (Signal-to-Cutoff) based on instrument calibration. Used in clinical laboratories by technicians to assist physicians in diagnosing syphilis infection. The output provides diagnostic information to support clinical decision-making, aiding in patient management and treatment planning.

Clinical Evidence

Clinical study evaluated 1910 specimens (1104 prospective, 452 retrospective, 204 healthy, 150 medically diagnosed) across three sites. Prospective intended use population showed 100% positive percent agreement (184/184) and 96.7% negative percent agreement (890/920) against a comparator algorithm of three FDA-cleared syphilis assays. Medically diagnosed cohort showed 95.3% agreement. No high-dose hook effect observed.

Technological Characteristics

Two-step sandwich enzyme CMIA. Reagents: paramagnetic particles coated with recombinant Tp17/Tp47 antigens, biotinylated Tp17/Tp47 conjugates, alkaline phosphatase. Energy: chemiluminescence. Dimensions: automated analyzer-based. Connectivity: integrated with Access 2 and DxI 9000 systems. Software: automated result calculation based on S/CO ratio. Sterilization: N/A (reagents).

Indications for Use

Indicated for qualitative detection of total antibodies to Treponema pallidum in human serum and plasma to aid in syphilis diagnosis. For use in clinical laboratories on Access Immunoassay Systems. Not for blood or tissue donor screening.

Regulatory Classification

Identification

Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K241427 B Applicant Beckman Coulter, Inc C Proprietary and Established Names Access Syphilis D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LIP | Class II | 21 CFR 866.3830 - Treponema Pallidum Treponemal Test Reagents | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To establish substantial equivalence to a predicate device and to obtain market clearance for a new assay designed to detect antibodies to Treponema pallidum in human serum and plasma. B Measurand: Total antibodies to Treponema pallidum C Type of Test: Chemiluminescent microparticle immunoassay (CMIA) ## III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} B Indication(s) for Use: The Access Syphilis assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative detection of total antibodies to Treponema pallidum in human serum and plasma using the Access Immunoassay Systems. It is intended to be used as an aid in the diagnosis of syphilis or in conjunction with a non-treponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. The Access Syphilis assay is not intended for blood and tissue donor screening. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: Access Immunoassay Systems - Access 2 Immunoassay System - DxI 9000 Access Immunoassay Analyzer IV Device/System Characteristics: A Device Description: The Access Syphilis assay is a two-step enzyme chemiluminescent microparticle immunoassay. The test is performed by adding a sample to the reaction vessel, along with buffer, paramagnetic particles coated with recombinant Treponema pallidum antigens Tp17 and Tp47, and biotinylated Treponema Tp17 and Tp47 antigens. The sample mixture is incubated in the reaction vessel, allowing materials to bind to the solid phase where they are held in a magnetic field. Any unbound materials are washed away. B Principle of Operation: The Access Syphilis assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative detection of total antibodies to Treponema pallidum in human serum and plasma using the Access Immunoassay Systems. The Access Immunoassay System is fully automated to allow assay-specific reagents to form immune complexes, immobilize the bound immune complex by magnetic field, and allow the unbound components to be washed away. The Access Syphilis assay uses the following reagents: - Treponema pallidum antigens labeled with alkaline phosphatase - Antigen-coated micron-sized paramagnetic particles - TRIS-buffered saline with biotinylated Tp17 and Tp47 conjugates The assay is a two-step immunoenzymic "sandwich" assay. The sample is added to the reaction vessel along with buffer, recombinant Tp17- and Tp47-coated paramagnetic particles, and biotinylated Treponema Tp17 and Tp47 antigens. The mixture is incubated to allow Treponemal antibodies that may be present to be captured by the antigens. Following incubation, any unbound materials are washed away, and the materials bound to the solid phase are held in a magnetic field. Alkaline phosphatase conjugates are added that bind to immunoglobulin captured on the particles. A chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The measured light allows for determination of the K241427 - Page 2 of 18 {2} presence of the analyte by comparison with a cut-off value defined during the assay calibration on the instrument. Test results are determined automatically by the system software. Results expressed as Signal/Cutoff (S/CO) are reported to be "reactive" or "nonreactive" as a function of their relationship with the "cut-off". The result interpretation for the Access Syphilis assay is as follows: S/CO &lt; 1.00 Nonreactive S/CO ≥ 1.00 Reactive **Interpretation of results:** Reactive: Treponemal antibodies detected; indicates active or previous infection with *Treponema pallidum*. Nonreactive: Treponemal antibodies not detected; indicates no active or previous infection with *Treponema pallidum*. Test results are intended to aid in diagnosis only. As with all serological tests for syphilis, results should always be interpreted in conjunction with additional treponemal or nontreponemal serologic test results (as appropriate), the patient's clinical symptoms, medical history, and other clinical and/or laboratory findings to produce a diagnosis of syphilis by disease stage. **Access Syphilis Calibrator** Intended for the calibration of the Access Syphilis assay. Provided as one negative and one positive calibrator for *Treponema pallidum* antibody. **Access Syphilis Controls** Intended for the estimation of test precision and the detection of systematic analytical deviations of the Access Immunoassay System when used for the qualitative detection of antibodies to *Treponema pallidum* (TP) in human serum and plasma. Provided as three vials each of *Treponema pallidum* antibody positive control and *Treponema pallidum* negative control. V Substantial Equivalence Information: A Predicate Device Name(s): ARCHITECT Syphilis TP Reagent, ARCHITECT Syphilis TP Calibrator, ARCHITECT Syphilis TP Control B Predicate 510(k) Number(s): K153730 C Comparison with Predicate(s): | Device & Predicate Device(s): | K241427 | K153730 | | --- | --- | --- | | Device Trade Name | Access Syphilis | ARCHITECT Syphilis TP | K241427 - Page 3 of 18 {3} K241427 - Page 4 of 18 | General Device Characteristic Similarities | | | | --- | --- | --- | | Intended Use/Indications for Use | The Access Syphilis assay is a paramagnetic particle, chemiluminescent immunoassay for the qualitative detection of total antibodies to Treponema pallidum in human serum and plasma using the Access Immunoassay Systems. It is intended to be used as an aid in the diagnosis of syphilis or in conjunction with a non-treponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. The Access Syphilis assay is not intended for blood and tissue donor screening. | The ARCHITECT Syphilis TP assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of antibodies (IgG and IgM) directed against Treponema pallidum (TP) in human serum and plasma. The ARCHITECT Syphilis TP assay is intended to be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection. | | Environment of Use | Clinical laboratory | Same | | Methodology | Chemiluminescent microparticle immunoassay (CMIA) | Same | | Cut-off Index | 1.00 S/CO | Same | | Assay Type | Two-step sandwich enzyme immunoassay | Same | | Detection Method | Automated, chemiluminescence | Same | | Controls | 2 (Negative and Positive) | Same | | General Device Characteristic Differences | | | | Antigens | Recombinant Treponema pallidum antigens Tp17 and Tp47, and biotinylated Treponema Tp17 and Tp47 antigens | Recombinant TP antigens: TpN15, TpN17, and TpN47 (obtained in E. coli) | | Instrument | Access Immunoassay | ARCHITECT iSystem | {4} K241427 - Page 5 of 18 | | Systems: • Access 2 Immunoassay System • DxI 9000 Access Immunoassay Analyzer | | | --- | --- | --- | | Components | 1. Microparticles – paramagnetic particles coated with recombinant Ag Tp17 and Tp47 in TRIS-buffered saline with surfactant, protein, sodium azide, and ProClin 300. 2. Conjugate – TRIS-buffered saline with biotinylated Tp17 and Tp47 conjugates, surfactant, protein, and sodium azide. 3. Assay Diluent – Phosphate buffer with alkaline phosphatase conjugates, protein, sodium azide, and ProClin 300. | 1. Microparticles – TP antigen (E. coli, recombinant) coated microparticles in HEPES buffer with detergent. 2. Conjugate – Murine anti-IgG/anti-IgM acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer. 3. Assay Diluent – Syphilis TP Assay Diluent containing MES buffer with detergent. Preservatives: ProClin 950 and other antimicrobial agents. | | Sample Volume | ~45 μL | 30 μL | | Calibrators | 1 negative and 1 positive antibody calibrator. | 1 positive calibrator, stable up to 30 days. | | Compatible Anticoagulants | Human Serum: Serum and serum separator tube Human Plasma: Lithium Heparin Lithium Heparin separator tube Dipotassium (K_{2}) EDTA | Human Serum: Serum and serum separator tube Human Plasma: Lithium Heparin Lithium Heparin separator tube Dipotassium (K_{2}) EDTA | {5} | | Tripotassium (K_{3}) EDTA Sodium Citrate Acid Citrate Dextrose (ACD) Citrate Phosphate Dextrose (CPD) Citrate Phosphate Dextrose with Adenine (CPDA) | Tripotassium (K_{3}) EDTA Sodium Heparin | | --- | --- | --- | VI Standards/Guidance Documents Referenced: CLSI EP05-A3 7-251 Evaluation of Precision of Qualitative Measurement Procedures; Approved Guideline – Third Edition CLSI EP07 3rd Edition 7-275 Interference Testing in Clinical Chemistry CLSI EP12-A2 (2008) User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline – Second Edition CLSI EP24-A2 7-234 Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Approved Guideline – Second Edition CLSI EP25-A 7-235 Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline CLSI EP37 1st Edition 7-284 Supplemental Tables for Interference Testing in Clinical Chemistry CLSI GP44-A4 7-213 Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline – Fourth Edition CLSI GP41 7th Edition 7-277 Collection of Diagnostic Venous Blood Specimens ISO 7000 Sixth edition 2019-07 5-124 Graphical symbols for use on equipment – Registered symbols ISO 7010 Third edition 2019-07 5-130 Graphical symbols – Safety colors and safety signs – Registered safety signs ISO 15223-1 Fourth Edition 2021-07 5-134 Medical devices – Symbols to be used with information to be supplied by the manufacturer – Part 1: General requirements ISO 13485:2016 Medical devices – Quality management systems – Requirements for regulatory purposes K241427 - Page 6 of 18 {6} VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Within-lab Precision The precision of the Access Syphilis assay was evaluated in an internal study using three reagent lots and three calibrator lots. The panel consisted of four samples contrived in undiluted, pooled human serum containing azide and ProClin. Positive specimens were prepared by spiking negative serum with high positive CPD plasma specimens. Table 1. Precision panel members | Panel Member | Target ratio (S/CO) | | --- | --- | | Low nonreactive | < 0.1 | | High nonreactive | 0.5 to 1.0 | | Low reactive | 1.0 to 2.0 | | Reactive | > 5.0 | Negative and positive quality controls were also included in each run, with mean S/CO values of 0.10 and 1.56, respectively. Samples were tested on both Access 2 and DxI 9000 instruments in 2 replicates per run, 2 runs per day for a 20-day period. A total of 80 replicates were tested for each panel member for each lot. Table 2. Precision study data – DxI 9000 instrument | Precision – DxI 9000 Instrument | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean S/CO | Between Lot | | Between Day | | Between Run | | Within Run | | Overall | | | | | | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | | Neg Control | 238* | 0.166 | 0.064 | N/A | 0.007 | N/A | 0.007 | N/A | 0.008 | N/A | 0.065 | N/A | | Pos Control | 238* | 1.786 | 0.082 | 4.6% | 0.059 | 3.3% | 0.088 | 4.9% | 0.041 | 2.3% | 0.140 | 7.8% | | Low Negative | 240 | 0.078 | 0.018 | N/A | 0.003 | N/A | 0.003 | N/A | 0.005 | N/A | 0.019 | N/A | | High Negative | 239* | 0.779 | 0.065 | 8.4% | 0.021 | 2.8% | 0.023 | 3.0% | 0.019 | 2.4% | 0.075 | 9.6% | | Low Positive | 240 | 1.994 | 0.078 | 3.9% | 0.048 | 2.4% | 0.072 | 3.6% | 0.046 | 2.3% | 0.125 | 6.3% | | Positive | 240 | 7.598 | 0.294 | 3.9% | 0.241 | 3.2% | 0.240 | 3.2% | 0.163 | 2.1% | 0.478 | 6.3% | *A total of five results were flagged during the study due to instrument errors. These specimens were not retested and were excluded from result calculations. K241427 - Page 7 of 18 {7} Table 3. Precision study data – Access 2 instrument | Precision – Access 2 Instrument | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | N | Mean S/CO | Between Lot | | Between Day | | Between Run | | Within Run | | Overall | | | | | | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | SD (S/CO) | CV (%) | | Neg Control | 240 | 0.107 | 0.029 | N/A | 0.009 | N/A | 0.007 | N/A | 0.007 | N/A | 0.032 | N/A | | Pos Control | 240 | 1.704 | 0.083 | 4.9% | 0.057 | 3.4% | 0.074 | 4.3% | 0.048 | 2.8% | 0.134 | 7.9% | | Low Negative | 240 | 0.064 | 0.004 | N/A | 0.003 | N/A | 0.004 | N/A | 0.003 | N/A | 0.007 | N/A | | High Negative | 240 | 0.708 | 0.018 | 2.6% | 0.023 | 3.2% | 0.022 | 3.2% | 0.017 | 2.5% | 0.041 | 5.8% | | Low Positive | 240 | 1.819 | 0.116 | 6.4% | 0.061 | 3.4% | 0.051 | 2.8% | 0.039 | 2.2% | 0.146 | 8.0% | | Positive | 240 | 6.721 | 0.491 | 7.3% | 0.216 | 3.2% | 0.203 | 3.0% | 0.179 | 2.7% | 0.601 | 8.9% | ## Multi-site Reproducibility A multicenter reproducibility study was performed at three clinical sites using three lots each of the Access Syphilis assay, three lots of calibrators, and three lots of controls. Each panel consisted of one low negative, one high negative, one low positive, and one positive, as well as two assay controls. Panel members were tested using four replicates per run, two runs per day, for a total of five days to generate 120 data points at each site per platform. Data was analyzed for intra-assay and inter-assay reproducibility. Table 4. Reproducibility study data – Access 2 instrument | Reproducibility – Access 2 Instrument | | | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | Mean Value | N | Repeatability | | Between-Run | | Between-Day | | Between-Lot | | Between-Site | | Reproducibility | | | | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Neg Control | 0.10 | 360 | 0.004 | N/A | 0.005 | N/A | 0.006 | N/A | 0.025 | N/A | 0.005 | N/A | 0.027 | N/A | | Pos Control | 1.56 | 360 | 0.043 | 2.8% | 0.058 | 3.7% | 0.016 | 1.0% | 0.010 | 0.6% | 0.054 | 3.4% | 0.092 | 5.9% | | Low Negative | 0.06 | 360 | 0.003 | N/A | 0.003 | N/A | 0.003 | N/A | 0.004 | N/A | 0.005 | N/A | 0.009 | N/A | | High Negative | 0.75 | 360 | 0.019 | 2.6% | 0.018 | 2.4% | 0.018 | 2.5% | 0.013 | 1.8% | 0.005 | 0.6% | 0.035 | 4.7% | | Low Positive | 1.89 | 360 | 0.042 | 2.2% | 0.065 | 3.5% | 0.037 | 2.0% | 0.058 | 3.1% | 0.000 | N/A | 0.104 | 5.5% | | Positive | 7.19 | 360 | 0.223 | 3.1% | 0.197 | 2.7% | 0.000 | 0.0% | 0.247 | 3.4% | 0.068 | 0.9% | 0.393 | 5.5% | K241427 - Page 8 of 18 {8} Table 5. Reproducibility study data – DxI 9000 instrument | Reproducibility – DxI 9000 Instrument | | | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sample | Mean Value | N | Repeatability | | Between-Run | | Between-Day | | Between-Lot | | Between-Site | | Reproducibility | | | | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Neg Control | 0.10 | 360 | 0.005 | N/A | 0.003 | N/A | 0.009 | N/A | 0.033 | N/A | 0.010 | N/A | 0.036 | N/A | | Pos Control | 1.59 | 360 | 0.037 | 2.3 | 0.050 | 3.2 | 0.059 | 3.7 | 0.000 | N/A | 0.019 | 1.2 | 0.088 | 5.5 | | Low Negative | 0.05 | 360 | 0.003 | N/A | 0.001 | N/A | 0.004 | N/A | 0.007 | N/A | 0.005 | N/A | 0.010 | N/A | | High Negative | 0.75 | 360 | 0.017 | 2.3 | 0.018 | 2.4 | 0.029 | 3.8 | 0.018 | 2.4 | 0.030 | 4.0 | 0.052 | 6.9 | | Low Positive | 1.94 | 360 | 0.044 | 2.3 | 0.060 | 3.1 | 0.070 | 3.6 | 0.053 | 2.7 | 0.052 | 2.7 | 0.126 | 6.5 | | Positive | 7.48 | 360 | 0.271 | 3.6 | 0.271 | 3.6 | 0.213 | 2.8 | 0.297 | 4.0 | 0.387 | 5.2 | 0.655 | 8.8 | 2. Linearity: Not applicable. 3. Analytical Specificity/Interference: Interference by Endogenous Substances Potential interference with hemoglobin, bilirubin, triolein, protein (albumin), and gamma globulin were evaluated in an analytical study by preparing the panel members at treponemal concentrations targeted at the levels described in the table below. Table 6. Interference panel members | Panel Member | Target Ratio (S/CO) | | --- | --- | | Low negative | < 0.1 | | High negative | 0.5 – 1.0 | | Low positive | 1.0 – 2.0 | | Positive | 5.0 – 10.0 | Each of the four panel members was split into seven aliquots (for each of the six conditions plus a control). Each aliquot was spiked directly with the substance at the target final concentration. If an effect was observed at a given analyte concentration, a dose effect study was performed to determine the minimum interfering concentration. Table 7. Interferent concentration levels | Interferent | Interferent Level | | --- | --- | | Hemoglobin | 1000 mg/dL | | Conjugated Bilirubin | 40 mg/dL | | Unconjugated Bilirubin | 40 mg/dL | | Triolein | 36 g/L | | Total Protein (Human Albumin) | 15 g/dL | | Gamma globulin | 47.5 g/L* | *Interference was noted at a concentration of 60 g/L. No interference was observed ≤ 47.5 g/L K241427 - Page 9 of 18 {9} Interference was observed with the Access Syphilis test in the presence of gamma globulin at a concentration of 60 mg/mL in negative and positive specimens. The highest concentration for which no interference was observed across three lots of the Access Syphilis assay was 47.5 mg/mL. No interference was observed for the other listed substances up to the levels indicated in the table above. ## Drug Interference Potential interference due to the presence of common pharmaceutical compounds was tested in a drug interference study. Panels were prepared as described for the endogenous interference study. Each panel member was divided into ten aliquots (for seven exogenous substances and three control conditions). The aliquots were spiked directly with the drug of interest at the final target concentration listed in the table below. For the control specimens, the same amount of drug diluent (deionized water or ethanol) was spiked-in to avoid bias between test and control specimens. No interference was observed at the levels tested. Table 8. Interfering substances and concentrations | Substance | Drug class | Target | | --- | --- | --- | | Acetylsalicylic acid | Anti-inflammatory | 3 mg/dL | | Ibuprofen | Anti-inflammatory | 21.9 mg/dL | | Acetaminophen | Antipyretic | 15.6 mg/dL | | Azithromycin | Antibiotic (macrolides) | 1.11 mg/dL | | Ceftriaxone sodium | Antibiotic (cephalosporins) | 84 mg/dL | | Doxycycline hyclate | Antibiotic (tetracyclines) | 1.8 mg/dL | | Biotin | Vitamin | 0.351 mg/dL | ## Cross-reactivity in Specimens with Medical Conditions not Related to Syphilis Specimens from individuals diagnosed with other diseases that could cause false positive results were tested in duplicate by the Access Syphilis assay. Discordant results were investigated by additional testing. For each specimen, one replicate was tested on the Access 2 instrument and one replicate was tested with the DxI 9000 instrument. Table 9. Cross-reactivity panel | Potential Cross-reactant | N | Number of Reactive Samples | | --- | --- | --- | | Pregnant Multipara | 14 | 0 | | Hemodialysis | 10 | 0 | | Transplant | 10 | 0 | | Rheumatoid Factor (RF) | 10 | 0 | | Human anti-mouse antibody (HAMA) | 10 | 0 | | Anti-nuclear antibody (ANA) | 10 | 0 | | Myeloma | 11 | 1a | | Flu-vaccinated patients | 10 | 0 | | Anti-phospholipid | 10 | 0 | | Lyme Disease (Borrelia garinii, Borrelia afzelii & Borrelia burgdorferi s.s.) | 10 | 0 | K241427 - Page 10 of 18 {10} | Leptospirosis | 10 | 0 | | --- | --- | --- | | Systemic Lupus Erythematosus (SLE) | 10 | 0 | | Hepatitis A Virus (HAV) Total Ab | 5 | 0 | | Hepatitis A Virus (HAV) IgM | 5 | 0 | | Hepatitis B Virus (HBV) HBs Ag positive | 10 | 0 | | Hepatitis B Virus (HBV) anti-HBs positive | 22 | 0 | | Hepatitis B Virus (HBV) HBc IgM | 10 | 1^{a} | | Hepatitis B Virus (HBV) HBc Total | 20 | 0 | | Hepatitis C Virus (HCV) Ig Total Ab | 10 | 0 | | HTLV-1 | 10 | 0 | | HTLV-2 | 14 | 1^{b} | | Human Immunodeficiency Virus (HIV)-1 | 10 | 2^{b} | | Human Immunodeficiency Virus (HIV)-2 | 10 | 1^{a} | | Herpes Simplex Virus (HSV) 1/2 IgM | 15 | 0 | | Herpes Simplex Virus (HSV) 1 IgG | 10 | 0 | | Herpes Simplex Virus (HSV) 2 IgG | 2 | 0 | | Cytomegalovirus (CMV) IgG | 5 | 0 | | Cytomegalovirus (CMV) IgM | 5 | 0 | | Rubella IgG | 5 | 0 | | Rubella IgM | 10 | 0 | | Toxoplasma gondii IgG | 5 | 0 | | Toxoplasma gondii IgM | 5 | 0 | | Epstein Barr Virus (EBV) IgG | 5 | 0 | | Epstein Barr Virus (EBV) IgM | 5 | 0 | | Epstein Barr Virus (EBV) anti-EBNA IgG and anti-VCA IgG | 5 | 0 | | Epstein Barr Virus (EBV) anti-VCA IgM | 5 | 0 | | Anti-Escherichia coli (E. coli)^{c} | 10 | 0 | a Volume insufficient for duplicate testing. b Specimens were determined to be true syphilis positives. c Anti-E. coli specimens were prepared by spiking 10 negative specimens with anti-E. coli rabbit antibodies. 4. Assay Reportable Range: Not applicable. 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Calibrator The Access Syphilis Calibrators are not provided with the reagent kit. The calibrators are provided as one negative for TP antibody and one positive for TP antibody. External Controls The Access Syphilis assay controls are not provided with the reagent kit. The negative control is made from human serum shown to be non-reactive for antibodies to TP. The positive control is made from TP-non-reactive human serum spiked with a pool of anti-TP reactive human serum or plasma. Sample Stability Studies This study was conducted using serum and plasma specimens. The stability of TP antibodies in various storage matrices was evaluated for: - Refrigerated Stability (at 2 to 8°C) at day 0 and daily at days 5 – 8. - Room temperature stability (at 20 to 25°C) at 0 h, 24 h, 48 h, 72 h, and 96 h. - Frozen stability at -20°C and -80°C - Non-reactive specimens were evaluated at days 0, 29, 36, 90, and 105. K241427 - Page 11 of 18 {11} o Reactive serum specimens were evaluated at days 0, 29, and 36. o Reactive plasma specimens were evaluated at days 0, 29, 36, 90, and 111. - Freeze/thaw stability for 0 - 6 cycles (at -20 and -80°C) Each sample was tested in duplicate with the Access 2 and DxI 9000 instruments at each time point. The evaluation of results was performed separately for each concentration, storage condition, and time point. The mean S/CO was calculated for each subsequent time point for each panel member, concentration level, storage condition, and tube type. The difference between the mean S/CO at each time point and the corresponding baseline mean S/CO, taken at day zero, was calculated. For negative panel members (S/CO &lt; 1.00), the difference in measured concentration was within ± 0.10 compared to the baseline and the negative agreement was 100% among the different time points. For positive panel members (S/CO ≥ 1.00), the median S/CO percentage bias was within ± 15% versus the baseline and the positive agreement was at least 95% for the low positive specimens (&lt; 2.5× S/CO) and 100% for the positive specimens (≥ 2.5× S/CO). The study data supported the following storage conditions: - 2 to 8°C for 7 days - 20 to 25°C for 72 hours - -20°C or colder for 30 days - Samples may be frozen and thawed up to 5 times. ## Reagent Stability To test reagent stability, three lots of Access Syphilis reagent packs were tested in triplicate within a month after production. After the initial time point, testing was conducted in duplicate at 3, 6, 9, 10, 11, and 12 months approximately for each lot, with additional testing for lots one and two at 13 months and testing for lot one through 15 months. The positive calibrator and negative calibrators were tested at each time point to validate the run and calibrate the reagent pack stored at 2-10°C. Four panel members plus a positive and negative control were used to assess real-time reagent stability. The study data from three lots supported stability up to 10.9 months when stored at 2-10°C unopened. ## On-board Reagent Stability To assess the on-board reagent stability of Access Syphilis, ten open kits were tested multiple times between day 0 and day 62. The positive calibrator and negative calibrators were tested at each time point to validate the run and calibrate the reagent pack stored at 2-10°C. Four panel members plus a positive and negative controls were used to assess real-time reagent stability. The study data supported stability up to 56 days after opening when stored at 2-10°C. ## Hook Effect The sponsor conducted a study to demonstrate that the Access Syphilis assay does not produce false non-reactive results when testing samples with high levels of anti-TP antibody. The study utilized five CPD plasma specimens with medium to high levels of anti-TP antibody which were serially diluted with negative serum. Each of the diluted specimens was tested with the Access Syphilis assay. The data from this study showed that there was no high dose hook effect, K241427 - Page 12 of 18 {12} suggesting that the Access Syphilis assay is not susceptible to interference from specimens with high levels of anti-syphilis TP antibodies. 6. Detection Limit: Not applicable. 7. Assay Cut-Off: The Access Syphilis cut-off was evaluated by ROC analysis using 301 antibody-positive and 3047 negative specimens. Specimens were tested in one replicate on the DxI 9000 and Access 2 instruments. To determine the cut-off value, a normalized ratio was determined for each specimen by dividing the specimen signal by the signal obtained with the positive calibrator tested in triplicate during the same testing runs. The normalized ratio values were used to build the ROC curve, and the optimal normalized ratio value that gave the best compromise between sensitivity and specificity was determined. Result classification: S/CO &lt; 1.00 sample is interpreted as "non-reactive" S/CO ≥ 1.00 sample is interpreted as "reactive" B Comparison Studies: 1. Method Comparison with Predicate Device: See the "Clinical Studies" section. 2. Matrix Comparison: The suitability of various blood collection tube types for use with the Access Syphilis assay was evaluated in an analytical study. Samples were obtained from multiple donors and venous blood was collected in the following tube types: - Serum (collected in dry tube) – used as the control - Serum separator tubes - Citrate phosphate dextrose with adenine (CPDA) - Lithium heparin - Lithium heparin with gel separator tube - Dipotassium (K2) EDTA - Tripotassium (K3) EDTA - Sodium citrate - Citrate phosphate dextrose (CPD) - Acid citrate dextrose (ACD) A total of 65 paired specimens, 25 non-reactive and 40 reactive, were tested for each matrix. Positive specimens were prepared by spiking negative specimens with highly reactive syphilis samples to targeted concentrations. A total of four levels of reactivity were evaluated: K241427 - Page 13 of 18 {13} Table 10. Matrix comparison panel members | Analyte Target Level | Number of Expected Specimens | | --- | --- | | 2.5 – 5x cut-off | 10 | | < 2.5x cut-off* | 30 | | Negative | 35 | | Total | 65 | *About half of the samples were prepared in the 1.00 – 1.50 S/CO range. Each sample was tested in duplicate. Percent agreement and bias analysis were calculated in comparison to the control condition (serum collected in a dry tube). The resulting data demonstrates that serum, serum with gel separator, lithium heparin, lithium heparin with gel separator, dipotassium EDTA, tripotassium EDTA, sodium citrate, citrate phosphate dextrose, acid citrate dextrose, and citrate phosphate dextrose with adenine tubes are acceptable for use with the Access Syphilis assay. # C Clinical Studies: A multicenter study was conducted on the Access 2 and DxI 9000 instruments between June and September 2023 to evaluate the ability of the Access Syphilis assay to detect antibodies directed against Treponema pallidum (TP). A total of 1923 specimens were tested in the Access Syphilis clinical study on both the Access 2 and the DxI 9000 instruments. Of these 1923, 13 were excluded from analysis. The remaining 1910 included 1104 prospectively collected from the intended use population; 50 retrospective samples from a population at high-risk for STIs; 402 retrospective specimens including 22 pregnant women; 204 apparently healthy individuals; and 150 from medically diagnosed individuals with primary, secondary, or latent syphilis. All cohorts were tested at three testing sites in a randomized and blinded fashion. The clinical performance of the Access Syphilis assay was evaluated by calculating positive percent agreement and negative percent agreement of the assay with the final comparator result based on an algorithm of results from three FDA-cleared syphilis assays: a treponemal chemiluminescent immunoassay (TP-CLIA), a nontreponemal assay (RPR), and a second treponemal assay (TP-PA). Clinical Performance in Prospectively Collected Specimens in the Intended Use Population The 1104 prospectively collected specimens in the intended use population, analyzed in the Access Syphilis clinical study consisted of 399 specimens sent for routine syphilis testing (201 females and 198 males, 12 to $&gt;89$ years old), 405 pregnant females (16 to 46 years old), and 300 HIV-positive specimens (98 females and 202 males, 23 to 77 years old). Of the 1104 prospectively collected specimens, 55 were from patients $\leq 21$ years old. Thirteen (13) prospective specimens were excluded from analysis due to system error, duplicate testing, or missing comparator result. The comparison between the Access Syphilis result and the final comparator result for the prospectively collected specimens in the intended use population is shown in the following table. Table 11. Percent agreement in the prospective intended use population K241427 - Page 14 of 18 {14} | Access Syphilis Result Interpretation | Final Comparator Result | | | --- | --- | --- | | | Positive | Negative | | Reactive | 184 | 30 | | Non-reactive | 0 | 890 | Positive percent agreement was 100% (184/184) with a 95% confidence interval of 98.0% to 100.0%. Negative percent agreement was 96.7% (890/920) with a 95% confidence interval of 95.4% to 97.7%. Table 12. Prospective percent agreement by category | Category | Positive Percent Agreement % (x/n) | 95% Confidence Interval (%) | Negative Percent Agreement % (x/n) | 95% Confidence Interval (%) | | --- | --- | --- | --- | --- | | Routine Syphilis | 100 (60/60) | 94.0 - 100 | 99.7 (338/339) | 98.3 - 99.9 | | Pregnant | 100 (6/6) | 61.0 - 100 | 99.7 (398/399) | 98.6 - 100 | | HIV positive | 100 (118/118) | 96.8 - 100 | 84.6^{a} (154/182) | 78.7 - 89.1 | a The 28 comparator-negative specimens were tested with an additional TP-CLIA test. Of these, 25 (89.3%) were also reactive. ## Clinical Performance in Retrospective Samples Of the total 1910 specimens analyzed in the Access Syphilis clinical study, 452 were retrospective specimens. The 402 retrospective specimens included 22 specimens from pregnant individuals. An additional 50 retrospective specimens from a population at high risk for contracting STIs were also tested. Each of these specimens was analyzed with the Access Syphilis assay on the Access 2 and DxI 9000 instruments, and comparator assays, employing an algorithm of three syphilis tests to determine the comparator result. The agreements between the Access Syphilis result and the final comparator result are presented below. Table 13. Percent agreement in retrospective specimens | Category | Positive Percent Agreement % (x/n) | 95% Confidence Interval (%) | Negative Percent Agreement % (x/n) | 95% Confidence Interval (%) | | --- | --- | --- | --- | --- | | Retrospective Specimens | 100 (378/378) | 99.0 - 100.0 | 0 (2/0) | NA | | Retrospective Specimens - Pregnant | 100 (20/20) | 83.9 - 100.0 | 50.0 (1/2)^{a} | 9.4 - 90.6 | | Retrospective High-risk STI population | 100 (20/20) | 83.9 - 100.0 | 80.0 (24/30)^{b} | 62.7 - 90.5 | a Specimen reactive by Access Syphilis was also reactive with another FDA-cleared TP-CLIA test. b Of the six specimens reactive by Access Syphilis, one was also reactive with another FDA-cleared TP-CLIA test. NA = Not applicable ## Clinical Performance in Pregnant Females A total of 427 pregnant female samples were analyzed in the Access Syphilis clinical study; 405 samples were prospectively collected from women coming to the health care facilities at the collection sites and 22 specimens were retrospective. The percent agreement between the Access Syphilis results and the final comparator results for prospectively collected and retrospective specimens from pregnant females is shown below, stratified by pregnancy trimester. K241427 - Page 15 of 18 {15} Table 14. Percent agreement in pregnant women | Category | Positive Percent Agreement % (x/n) | 95% Confidence Interval (%) | Negative Percent Agreement % (x/n) | 95% Confidence Interval (%) | | --- | --- | --- | --- | --- | | Prospectively Collected | 100.0 (6/6) | 61.0 – 100.0 | 99.7 (398/399) | 98.6 – 100.0 | | First Trimester | 100.0 (1/1) | 20.7 – 100.0 | 100.0 (59/59) | 93.9 – 100.0 | | Second Trimester | 100.0 (1/1) | 20.7 – 100.0 | 100.0 (37/37) | 90.6 – 100.0 | | Third Trimester | 100.0 (4/4) | 51.0 – 100.0 | 99.7 (302/303) | 98.2 – 99.9 | | Retrospective | 100.0 (20/20) | 83.9 – 100.0 | 50% (1/2) | NA | | First Trimester | 100.0 (6/6) | 61.0 – 100.0 | NA | NA | | Second Trimester | 100.0 (9/9) | 70.1 – 100.0 | NA | NA | | Third Trimester | 100.0 (5/5) | 56.6 – 100.0 | 100.0 (1/1) | 20.7 – 100.0 | | Trimester Unknown | NA | NA | 0% (0/1)a | NA | NA = Not applicable a Specimen tested non-reactive by Access Syphilis and another FDA-cleared TP-CLIA test. ## Clinical Performance in Medically Diagnosed Individuals Samples were tested from 150 individuals medically diagnosed with primary, secondary, or latent syphilis based on standard of care testing and/or physician diagnosis. These specimens included 49 females and 101 males. Results of the Access Syphilis assay for this cohort are summarized below. Table 15. Reactivity of the Access Syphilis assay in subjects medically diagnosed with syphilis | Medically Diagnosed Individuals | | | Access Syphilis Result | | Percent Agreement with Expected Outcome | | --- | --- | --- | --- | --- | --- | | Syphilis Stage | Treatment Status | N | Reactive | Non-reactive | | | Primary | Treated | 27 | 26 | 1 | 96.3 | | | Untreated | 49 | 49 | 0 | 100.0 | | Secondary | Treated | 23 | 23 | 0 | 100.0 | | | Untreated | 13 | 13 | 0 | 100.0 | | Latent | Treated | 25 | 25 | 0 | 100.0 | | | Untreated | 13 | 7 | 6a | 53.8 | | Total | | 150 | 143 | 7 | 95.3 | a Specimens tested non-reactive with Access Syphilis and another FDA cleared TP-CLIA. ## Clinical Performance in Apparently Healthy Individuals Of the total 1910 specimens analyzed in the Access Syphilis clinical study, 204 specimens were collected from apparently healthy individuals. This included 87 females, 116 males, and one from a person of unknown sex. Of these, 10 specimens from subjects younger than 21 years old were included. The reactivity of the Access Syphilis assay in the apparently healthy population is shown below. K241427 - Page 16 of 18 {16} Table 16. Reactivity of the Access Syphilis assay in the apparently healthy population | Category | Access Syphilis Result | | | | --- | --- | --- | --- | | | Number of Reactive (%) | Number of Nonreactive (%) | Total | | Adults | 18 (9.3%) | 176 (90.7%) | 194 | | < 21 yrs. old | 0 (0.0%) | 10 (100.0%) | 10 | | Total | 18 (8.8%) | 186 (91.2%) | 204 | 1. Clinical Sensitivity: The clinical sensitivity of the assay was expressed as positive percent agreement of the Access Syphilis assay results with the final comparator result, as explained above. 2. Clinical Specificity: The clinical specificity of the assay was expressed as negative percent agreement of the Access Syphilis assay results with the final comparator result, as explained above. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable. D Clinical Cut-Off: Not applicable. E Expected Values: In this clinical study, there were a total of 1104 specimens prospectively collected from the intended use population and tested with the Access Syphilis assay. The Access Syphilis assay was reactive in 214 subjects, for a 19.4% positivity rate of treponemal antibodies in the studied population. The distribution of the Access Syphilis reactive and non-reactive results by age and gender is summarized below. Table 17. Distribution of the Access Syphilis reactive and non-reactive results in the intended use population by age and gender. | Age Range (Yr) | Sex | N | Reactive | | Non-reactive | | | --- | --- | --- | --- | --- | --- | --- | | | | | # Reactive | % Reactive | # Non-reactive | % Non-reactive | | 12-20 | Male | 7 | 0 | 0.0% | 7 | 100.0% | | | Female | 48 | 2 | 4.2% | 46 | 95.8% | | 21-30 | Male | 44 | 9 | 20.4% | 35 | 79.6% | | | Female | 256 | 8 | 3.1% | 248 | 96.9% | | 31-40 | Male | 72 | 36 | 50.0% | 36 | 50.0% | | | Female | 211 | 12 | 5.7% | 199 | 94.3% | | 41-50 | Male | 87 | 38 | 43.7% | 49 | 56.3% | | | Female | 68 | 9 | 13.2% | 59 | 86.8% | | 51-60 | Male | 123 | 54 | 43.9% | 69 | 56.1% | | | Female | 68 | 17 | 25.0% | 51 | 75.0% | | 61-70 | Male | 54 | 16 | 29.6% | 38 | 70.4% | | | Female | 38 | 10 | 26.3% | 28 | 73.7% | K241427 - Page 17 of 18 {17} | 71-89a | Male | 13 | 2 | 15.4% | 11 | 84.6% | | --- | --- | --- | --- | --- | --- | --- | | | Female | 15 | 1 | 6.7% | 14 | 93.3% | | Overall | Male | 400 | 155 | 38.8% | 245 | 61.3% | | | Female | 704 | 59 | 8.4% | 645 | 91.6% | | Total | | 1104 | 214 | 19.4% | 890 | 80.6% | a One (1) subject was &gt;89 Yr. ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K241427 - Page 18 of 18
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...