DIASORIN LIAISON TOXO IGG AND LIAISON TOXO IGM ASSAYS
Applicant
Diasorin S.P.A
Product Code
LGD · Microbiology
Decision Date
Feb 8, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3780
Device Class
Class 2
Indications for Use
The LIAISON® Toxo IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer (Catolog number 15970) for the the qualitative determination of specific IgG antibodies to Toxoplasma gondii in human serum. The results of this assay can be used as an aid in the assessment of the patient's serological status to infection with Toxoplasma gondii and in the determination of immune status of individuals including pregnant women. This assay has not been cleared/approved by the F.D.A for blood/plasma donor screening. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens, or infants. The DiaSorin LIAISON® Toxo IgM assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer for the presumptive qualitative determination of IgM antibodies to Toxoplasma gondii in human serum. The LIAISON® Toxo IgM can be used as an aid in the presumptive diagnosis of acute or recent Toxoplasma gondii infection. It is recommended that the LIAISON® Toxo IgM assay be performed in conjunction with a Toxoplasma gondii IgG assay. This assay has not been cleared/approved by the FDA for blood/plasma screening. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, cord blood, neonatal specimens, or infants.
Device Story
LIAISON® Toxo IgG and IgM assays are chemiluminescent immunoassays (CLIA) performed on the LIAISON® Analyzer. Input: human serum samples. Principle: antibody capture (IgM) or indirect (IgG) immunoassay using magnetic particles coated with T. gondii antigen and isoluminol-antibody conjugate. Analyzer performs incubation, wash cycles, and starter reagent addition; photomultiplier measures flash chemiluminescence as relative light units (RLU). Output: qualitative determination of anti-Toxo IgG/IgM antibodies. Used in clinical laboratories by trained personnel. Results aid clinicians in assessing immune status or diagnosing recent infection. Benefits: provides automated, standardized serological testing for Toxoplasma gondii.
Clinical Evidence
Performance evaluated via method comparison studies using 887 (IgG) and 892 (IgM) prospective samples. IgG assay showed 93.5% overall agreement with ELISA in US specimens and 96.5% in pregnant women. IgM assay showed 99.3% overall agreement with ELISA in US specimens. CDC Toxoplasma 1998 Human Serum Panel testing: IgG assay correctly identified 70/70 positive and 30/30 negative samples; IgM assay correctly identified 32/32 positive and 63/65 negative samples. Precision studies reported CVs of 5-8% (IgG) and 19-32% (IgM).
Technological Characteristics
Chemiluminescent immunoassay (CLIA) technology. Solid phase: magnetic particles coated with T. gondii antigen (IgG) or anti-human IgM (IgM). Conjugate: mouse monoclonal antibodies to T. gondii (IgM) or human IgG (IgG) linked to isoluminol derivative. Energy source: LIAISON® Analyzer (photomultiplier detection). Form factor: automated kit components. Connectivity: LIAISON® Analyzer platform. Sterilization: N/A (reagents).
Indications for Use
Indicated for qualitative determination of Toxoplasma gondii IgG and IgM antibodies in human serum. IgG assay aids assessment of serological status and immune status, including in pregnant women. IgM assay aids presumptive diagnosis of acute or recent infection. Not for blood/plasma donor screening.
Regulatory Classification
Identification
Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.
Predicate Devices
Is-Toxoplasma IgG ELISA Kit (k981498)
Toxoplasma IgM ELISA Kit (k963289)
Related Devices
K131441 — LIAISON TOXO IGM II, LIAISON CONTROL TOXO IGM II · DiaSorin, Inc. · Aug 9, 2013
K132234 — LIAISON(R) TOXO IGG II, LIAISON(R) CONTROL TOXO IGG II · DiaSorin, Inc. · Aug 28, 2013
K233932 — Alinity i Toxo IgM · Abbott Laboratories · Aug 30, 2024
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K052499
B. Purpose for Submission:
New device(s)
C. Measurand:
Toxoplasma IgG antibodies & Toxoplasma IgM antibodies
D. Type of Test:
Chemiluminescence immunoassay
E. Applicant:
DiaSorin, Inc.
F. Proprietary and Established Names:
LIAISON® Toxo IgG Assay & LIAISON® Toxo IgM Assay
G. Regulatory Information:
1. Regulation section: 21CFR 866.3780; Toxoplasma gondii Serological Reagents
2. Classification: Class: II
3. Product code: LGD; Enzyme Linked Immunosorbent Assay, Toxoplasma gondii
4. Panel: 83 Microbiology
H. Intended Use:
IgG Assay: The LIAISON® Toxo IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer (Catalog number 15970) for the qualitative determination of specific IgG antibodies to Toxoplasma gondii in human serum. The results of this assay can be used as an aid in the assessment of the patient's serological status to infection with Toxoplasma gondii and in the determination of immune status of individuals including pregnant women.
Caution: This assay has not been cleared/approved by the F.D.A for blood/plasma donor screening.
{1}
IgM Assay: The DiaSorin LIAISON® Toxo IgM assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer for the presumptive qualitative determination of IgM antibodies to Toxoplasma gondii in human serum. The LIAISON® Toxo IgM can be used as an aid in the presumptive diagnosis of acute or recent Toxoplasma gondii infection. It is recommended that the LIAISON® Toxo IgM assay be performed in conjunction with a Toxoplasma gondii IgG assay.
Caution: This assay has not been cleared/approved by the F.D.A for blood/plasma donor screening.
2. Indication(s) for use:
IgG Assay: The LIAISON® Toxo IgG assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer (Catalog number 15970) for the the qualitative determination of specific IgG antibodies to Toxoplasma gondii in human serum. The results of this assay can be used as an aid in the assessment of the patient's serological status to infection with Toxoplasma gondii and in the determination of immune status of individuals including pregnant women.
This assay has not been cleared/approved by the FDA for blood/plasma donor screening.
IgM Assay: The DiaSorin LIAISON® Toxo IgM assay uses chemiluminescent immunoassay (CLIA) technology on the LIAISON® Analyzer for the presumptive qualitative determination of IgM antibodies to Toxoplasma gondii in human serum. The LIAISON® Toxo IgM can be used as an aid in the presumptive diagnosis of acute or recent Toxoplasma gondii infection. It is recommended that the LIAISON® Toxo IgM assay be performed in conjunction with a Toxoplasma gondii IgG assay.
This assay has not been cleared/approved by the FDA for blood/plasma donor screening.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
None
I. Device Description:
IgG Assay: The method for qualitative determination of IgG antibodies to Toxoplasma gondii (anti-Toxo IgG) is an indirect chemiluminescence
{2}
immunoassay (CLIA). All assay steps (with the exception of magnetic particle resuspension) and incubations are performed by the LIAISON® Analyzer. The principal components of the test are magnetic particles (solid phase) coated with Toxoplasma gondii and a conjugate of mouse monoclonal antibodies to human IgG linked to an isoluminol derivative (isoluminol-antibody conjugate). During the first incubation, Toxoplasma gondii antibodies present in diluted calibrators, samples or controls bind to the solid phase. During the second incubation, the monoclonal antibody conjugate reacts with anti-Toxo IgG that is already bound to the solid phase. After each incubation, unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal and therefore, the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of the presence of anti-Toxo IgG in calibrators, samples or controls.
**IgM Assay:** The method for qualitative determination of IgM antibodies to Toxoplasma gondii (anti-Toxo IgM) is an antibody capture chemiluminescence immunoassay (CLIA). All assay steps (with the exception of magnetic particle resuspension) and incubations are performed by the LIAISON® Analyzer. The principal components of the test are magnetic particles (solid phase) coated with IgG to human IgM (mouse, monoclonal), Toxoplasma gondii antigen, and a conjugate of mouse monoclonal antibodies to Toxoplasma gondii linked to an isoluminol derivative (isoluminol-antibody conjugate). During the first incubation, IgM antibodies present in diluted calibrators, samples or controls bind to the solid phase. During the second incubation, the mouse monoclonal antibody conjugate reacts with Toxoplasma gondii antigen previously added and the immune complex thus formed reacts with IgM already bound to the solid phase. After each incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and therefore the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of the presence of anti-Toxo IgM in calibrators, samples or controls.
**J. Substantial Equivalence Information:**
1. Predicate device name(s): Is-Toxoplasma IgG ELISA Kit and Toxoplasma IgM ELISA Kit
2. Predicate 510(k) number(s): k981498 (IgG Device) and k963289 (IgM Device)
3. Comparison with predicate:
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| IgG Assay: | | |
| Specimen Type | Human serum | Human serum |
| IgM Assay: | | |
| Method and type | Qualitative | Qualitative |
| Specimen Type | Human serum | Human serum |
| Differences | | |
| Item | Device | Predicate |
| IgG Assay: | | |
| Method and Type | Chemiluminescent | ELISA |
| | Immunoassay | |
| | Qualitative | Qualitative & quantitative |
| Instrumentation | Automated | Manual and automated |
| IgM Assay: | | |
| Method and Type | Chemiluminescent | ELISA |
| | Immunoassay | |
| Instrumentation | Automated | Manual |
# K. Standard/Guidance Document Referenced (if applicable):
Not applicable
# L. Test Principle:
Chemiluminescence immunoassay
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
IgG Assay: An assay reproducibility study was conducted at two external US laboratories and at DiaSorin. A coded panel comprised of 9 frozen repository samples was prepared by DiaSorin and provided to each site for testing by the LIAISON® Toxo IgG assay. The panel contained samples prepared to represent low to mid positive analyte levels. All panel members were divided into aliquots and stored frozen prior to testing. The same coded panel was tested at all three sites, in three replicates per run for ten runs. The overall CV for the positive specimens ranged from $5 - 8\%$ .
{4}
IgM Assay: An assay reproducibility study was conducted at two external US laboratories and at DiaSorin. A coded panel comprised of 9 frozen repository serum samples was prepared by DiaSorin and provided to each site for testing by the LIAISON® Toxo IgM assay. The panel contained 3 sets of serum samples. The sets were prepared to represent low to mid positive analyte levels. All panel members were divided into aliquots and stored frozen prior to testing. The same coded panel was tested at all three sites, in three replicates per run for ten runs. The overall CV for the positive specimens ranged from 19 – 32%.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable
d. Detection limit:
Not applicable
e. Analytical specificity:
IgG Assay: Cross-reactivity studies for the LIAISON® Toxo IgG assay were designed to evaluate potential interference from IgG immunoglobulins directed against closely-related members of the herpes virus family, organisms that may cause symptoms similar to Toxoplasmosis and other conditions that may result from atypical immune system activity. No positive result was found for the samples when tested by LIAISON® Toxoplasma IgG.
IgM Assay: Cross-reactivity studies for the LIAISON® Toxo IgM assay were designed to evaluate potential interference from IgM immunoglobulins directed against closely-related members of the herpes virus family, organisms that may cause symptoms similar to Toxoplasmosis and other conditions that may result from atypical immune system activity. No positive result was found for the samples when tested by LIAISON® Toxoplasma IgM
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
IgG Assay
A total of 887 prospectively collected samples were tested – 613 collected from the U.S. and 274 collected from Europe. The U.S. collection included 200 samples from pregnant women. In addition, 200 archived samples from acute, chronic,
{5}
and normal patients were tested. The testing was performed at four sites. All samples were tested with the LIAISON® Toxo IgG Assay and an enzyme immunoassay, ELISA or Sabin Feldman Dye test.
U.S. specimens – agreement with comparison IgG ELISA
Positive 91.8% (78/85)
Negative 95.3% (304/319)
Overall 93.5% (386/413)
European specimens - agreement with comparison IgG ELISA
Positive 98.6% (139/141)
Negative 91.3% (116/127)
Overall 94.5% (255/268)
Pregnant women (U.S.) - agreement with comparison IgG ELISA
Positive 96% (48/50)
Negative 97.3% (143/147)
Overall 96.5% (190/197)
Retrospective specimens: Two-hundred specimens were compared against the reference method Sabin Feldman Dye test, out of which 199 samples agreed with the DiaSorin test.
CDC Panel Study: The CDC (Centers for Disease Control and Prevention) Toxoplasma 1998 Human Serum Panel was tested by LIAISON® Toxo IgG assay. The panel is comprised of 100 frozen blinded specimens, 70 Toxoplasma positive samples and 30 toxoplasma negative samples. The data obtained were submitted to the CDC for analysis. The LIAISON® Toxo IgG Assay correctly detected the 70 positive and the 30 negative specimens.
**IgM Assay**
A total of 892 prospectively collected samples were tested – 613 collected from the U.S and 279 collected from Europe. The U.S. collection included 200 samples from pregnant women. The testing was performed at three sites – a hospital, a physician’s laboratory, and at DiaSorin. All samples were tested with the LIAISON® Toxo IgG Assay and an enzyme immunoassay, ELISA.
U.S. specimens – agreement with comparison IgM ELISA
Positive 33.3% (1/3)
Negative 99.8% (408/409)
{6}
Overall 99.3% (409/412)
European specimens - agreement with comparison IgM ELISA
Positive 91.1% (41/45)
Negative 98.7% (223/226)
Overall 94.9% (265/279)
Pregnant women (U.S.) - agreement with comparison IgM ELISA
Positive 33.3% (1/3)
Negative 99.5% (194/195)
Overall 98.5% (195/198)
CDC Panel Study: The CDC (Centers for Disease Control and Prevention) Toxoplasma 1998 Human Serum Panel was tested by LIAISON® Toxo IgM assay. The panel is comprised of 100 frozen blinded specimens: 32 Toxoplasma IgM positive, 3 dilutions of three true Toxoplasma IgM positive, and 65 Toxoplasma IgM negative samples. Of the 65 IgM negative samples 30 were Toxoplasma IgG negative and 35 IgG positive. The data obtained were submitted to the CDC for analysis. The LIAISON® Toxo IgM Assay correctly detected the 32 out of the 32 IgM positives, 1 out of the three IgM positive dilutions, and 63 out of the 65 IgM negatives.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
{7}
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.