K023779 · Thermo Biostar, Inc. · GQG · Jan 14, 2003 · Microbiology
Device Facts
Record ID
K023779
Device Name
RSV OIA
Applicant
Thermo Biostar, Inc.
Product Code
GQG · Microbiology
Decision Date
Jan 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3480
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The Thermo BioStar® RSV OIA assay is an Optical ImmunoAssay test for the rapid and qualitative detection of respiratory syncytial virus (RSV) antigens (nucleoproteins and fusion proteins) from nasal wash and nasopharyngeal swab specimens. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in symptomatic neonatal and pediatric patients under the age of five. It is recommended that all negative test results be confirmed by cell culture.
Device Story
RSV OIA is an in vitro diagnostic assay for detecting RSV antigens in nasal wash and nasopharyngeal swab specimens. It utilizes Optical ImmunoAssay technology on a silicon wafer surface. When specimen is applied, immobilized antibodies capture RSV antigens; a substrate is added to increase the thickness of the molecular thin film via mass enhancement. This change in optical thickness alters the reflected light path, producing a visible color change (purple spot on gold background) indicating a positive result. The test is performed in clinical settings to aid diagnosis in symptomatic pediatric patients. Results are available in under 20 minutes. It provides a rapid alternative to traditional viral culture methods, which require >48 hours. Negative results should be confirmed by cell culture.
Clinical Evidence
Multicenter study of 491 patients (414 analyzed). Compared RSV OIA to cell culture with fluorescent antibody staining; PCR used for secondary confirmation of discordant results. For nasal wash: 86.8% sensitivity, 83.2% specificity. For nasopharyngeal swabs: 66.7% sensitivity, 96.4% specificity. Reproducibility study across 6 sites (4 hospital, 2 POL) showed 95% overall reproducibility. Analytical testing confirmed specificity and lack of interference from whole blood or OTC products.
Technological Characteristics
Optical ImmunoAssay technology using silicon wafer substrate. Principle: mass enhancement of molecular thin films via antigen-antibody binding, resulting in visible color change. Qualitative, manual test. No electronic components or software algorithms.
Indications for Use
Indicated for rapid, qualitative detection of RSV antigens (nucleoproteins and fusion proteins) in symptomatic neonatal and pediatric patients under age five using nasal wash or nasopharyngeal swab specimens.
Regulatory Classification
Identification
Respiratory syncytial virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to respiratory syncytial virus in serum. Additionally, some of these reagents consist of respiratory syncytial virus antisera conjugated with a fluorescent dye (immunofluorescent reagents) and used to identify respiratory syncytial viruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of respiratory syncytial virus infections and provides epidemiological information on diseases caused by these viruses. Respiratory syncytial viruses cause a number of respiratory tract infections, including the common cold, pharyngitis, and infantile bronchopneumonia.
K061008 — QUICKVUE RSV (20 TEST KIT), MODEL 20193; (2 TEST KIT), MODEL 20199 · Quidel Corp. · Sep 8, 2006
K022133 — BD DIRECTIGEN EZ RSV KIT · Becton, Dickinson & CO · Dec 10, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) SUMMARY (page 1 of 3) 8.0
# JAN 1 4 2003
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K023779
A. Safety and effectiveness information required per [§807.92(a)(1)]:
- SUBMITTER'S NAME: Thermo BioStar, Inc. ●
- ADDRESS: 331 South 104th Street .
- TELEPHONE: (303) 530-3888 ext. 612 ●
- FAX: (303) 581-6405 ●
- CONTACT PERSON: John G. Adams ●
- DATE 510(k) SUMMARY PREPARED: October, 2002 ●
B. Safety and effectiveness information required per [§807.92(a)(2)]:
- TRADE OR PROPRIETARY NAME: RSV OIA® .
- COMMON NAME: Respiratory Syncytial Virus assay ●
- CLASSIFICATION NAME: Antigen, , CF (including CF Controls), Respiratory Syncytial Virus ●
- C. Identification of legally marketed device to which we are comparing performance.
Device Technology:
Trade or Proprietary Name: Regulatory Class: Manufacturer: 510(k) Number:
RSV OIA test kit I Thermo BioStar K021172
Historical Reference Method:
Viral Culture
- Intended use of device [§807.92(a)(5)]: F.
The Thermo BioStar® RSV OLA assay is an Optical ImmunoAssay test for the rapid and qualitative detection of respiratory syncytial virus (RSV) antigens (nucleoproteins) from nasal wash and nasopharyngeal swab specimens. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in symptomatic neonatal and pediatric patients under the age of five. It is recommended that all negative test results be confirmed by cell culture.
{1}------------------------------------------------
### 8.0 510(k) SUMMARY (page 2 of 3)
### D. Description of device [§807.92(a)(4)]:
#### Principle of the Test:
The RSV OIA test involves the qualitative extraction and detection of protein antigens unique to RSV (nucleoprotein and fusion protein). The Optical ImmunoAssay technology enables the direct visual detection of a physical change in the optical thickness of molecular thin films. This change is a result of antigen-antibody binding on an optical surface (silicon wafer). When an extracted specimen is placed directly on the optical surface, the immobilized specific antibodies capture the antigen. After washing, the substrate is added, increasing the thickness (mass enhancement) of
the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color change. Slight changes in optical thickness produce a distinct, visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the original gold color indicating a negative result .
#### DEVICE COMPARISON:
#### Device Technology:
The RSV OIA assay addressed in this submission is identical to the RSV OIA assay previously cleared in that:
- Both assays are rapid diagnostic tests that utilize Optical ImmunoAssay technology ●
- Both assays are used to detect and identify antigen proteins specific to RSV infections. ●
- Both assays can provide results in less than 20 minutes. ●
- Both assays are qualitative. ●
The RSV OIA assay differs from the currently marketed RSV OIA assay in that:
- The original RSV OIA assay (K021172) specimen types are limited to nasal wash specimens, and do not include nasopharyngeal swab specimens.
#### Comparison to historical standard method:
The RSV OIA assay is similar to culture methods in that:
- Both assays are used to detect and identify respiratory syncytial virus. .
- Both assays can detect the RSV virus from nasal wash and nasopharyngeal swab specimens. .
- Both assays are qualitative .
The RSV OIA assay differs with traditional culture methods in that:
- RSV OIA assay detects an antigen unique to the respiratory syncytial virus while the traditional culture . methods detect the whole living organism.
- . RSV OIA assay can provide results in less than 20 minutes, in contrast to culture methods that can take in excess of 48 hours.
{2}------------------------------------------------
#### 8.0 510(k) SUMMARY (page 3 of 3)
#### SUMMARY OF PERFORMANCE DATA:
#### CLINICAL STUDIES
Performance characteristics for the RSV OIA assay were initially established in a multicenter study with geographically diverse clinical sites.
H. Summary of clinical testing [§807.92(b)(2)]:
#### Reproducibility
Reproducibility testing was conducted at four hospital laboratories and two physician office laboratories (POL). Nine blinded samples were tested at each site at three separate times. The samples consisted of negative, low and moderate levels of RSV antigen, and the Kit positive and negative controls. These samples were prepared in buffered protein solution, and were spiked onto swabs for analysis. The reproducibility study included both the swab extraction and OIA testing of the antigen associated with RSV.
There were no significant differences in performance among the sites. Overall reproducibility for the sample panel was 95% across all sites (151/159).
#### Clinical Sensitivity and Specificity
A study was conducted comparing the RSV OIA test to commercially available cell culture, with confirmation and typing by fluorescent antibody staining. Secondary confirmation testing of specimens positive to OIA and negative to culture was done by specific nucleic acid detection by PCR . A total of 491 patients were enrolled into the multicenter study. A total of 414 nasalpharyngeal specimens were included in the data analysis, with 77 specimens excluded.
The RSV OIA assay was evaluated versus cell culture with confirmation and typing by fluorescent antibody staining. Sensitivity and specificity for symptomatic patients was 86.8% and 83.2% respectively for nasal wash specimens. Sensitivity and specificity for symptomatic patients and nasopharyngeal swas 66.7% and 96.4% respectively.
1. Conclusions from nonclinical / clinical testing [§807.92(b)(3)]:
Analytical testing was performed on both microbial and viral panels to assess specificity and cross reactivity.
Whole blood and several types of over the counter (OTC) products were evaluated to assess the potential for interference.
The results of the above described internal and external studies demonstrated that the RSV OIA test is as safe and effective as the comparative devices.
#### J. Additional information [§807.92(d)]:
No additional information has been requested by FDA at this time.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 1 4 2003
Mr. John G. Adams Regulatory Affairs Thermo BioStar 331 South 104th Street Louisville, CO 80027
k023779 Re: Trade/Device Name: RSV OIA® Regulation Number: 21 CFR 866.3480 Regulation Name: Respiratory Syncytial Virus Serological Reagents Regulatory Class: Class I Product Code: GQG Dated: November 8, 2002 Received: November 12, 2002
Dear Mr. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
## Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT 12.0
### INDICATIONS FOR USE STATEMENT
Koa3779 510(k) Number (if known): RSV OIA® Device Name:
Indications For Use:
The Thermo BioStar® RSV OIA assay is an Optical ImmunoAssay test for the rapid and qualitative detection of respiratory syncytial virus (RSV) antigens (nucleoproteins and fusion proteins) from nasal wash and nasopharyngeal swab specimens. This test is intended for in vitro diagnostic use to aid in the diagnosis of RSV infections in symptomatic neonatal and pediatric patients under the age of five. It is recommended that all negative test results be confirmed by cell culture.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Freddie W. Poole
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number _K 02
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.