VIRAZYME CULTURE CONFIRMATION SCREEN FOR INFLUENZA AND PARAINFLUENZA VIRUSES
K950838 · Zymetx, Inc. · GNT · Jun 17, 1996 · Microbiology
Device Facts
Record ID
K950838
Device Name
VIRAZYME CULTURE CONFIRMATION SCREEN FOR INFLUENZA AND PARAINFLUENZA VIRUSES
Applicant
Zymetx, Inc.
Product Code
GNT · Microbiology
Decision Date
Jun 17, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3330
Device Class
Class 1
Indications for Use
The ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza is intended for use as a screening test for respiratory viral cultures infected with influenza type A and B and parainfluenza types 1, 2, 3, and 4. This test will screen culture fluids for the presence of these viruses, but it is not indented for the definitive typing of these viruses.
Device Story
ViraZyme® is a screening assay for respiratory viral cultures. Input: culture fluids from respiratory viral cultures. Principle: viral neuraminidase activity cleaves a modified N-acetylneuraminic acid (Neu5Ac) substrate coupled to a chromogen; released chromogen precipitates via diazonium salt to produce a red color; precipitate is concentrated on a filter device. Output: visual red color indicating presence of neuraminidase-producing viruses (influenza/parainfluenza). Used in clinical laboratories; performed by laboratory technicians. Results assist in confirming viral presence in culture. Benefits: rapid screening of culture fluids to identify positive samples for further investigation.
Clinical Evidence
Multi-site clinical evaluation (four US locations) comparing ViraZyme® to standard monoclonal antibody (IFA) culture confirmation. Studies used 177, 97, 123, and 403 patient specimens (frozen and fresh). Sensitivity varied by virus type: Influenza A/B (100%), Parainfluenza 1 (85.7-100%), Parainfluenza 2 (77.8-100%), and Parainfluenza 3 (1.9-63.4%). Reduced sensitivity for Parainfluenza 3 linked to bovine serum in culture media; procedure updated to exclude serum. Mumps virus also tested positive due to neuraminidase activity. Negative controls (uninoculated Primary Rhesus Monkey Kidney) showed no false positives.
Technological Characteristics
Chromogenic enzymatic assay. Components: modified Neu5Ac substrate, chromogen, diazonium salt, filter device. Principle: neuraminidase-mediated cleavage of substrate. Requires serum-free culture medium to prevent interference.
Indications for Use
Indicated for screening respiratory viral cultures for the presence of influenza A and B and parainfluenza 1, 2, 3, and 4. Not for definitive viral typing.
Regulatory Classification
Identification
Influenza virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to influenza in serum. The identification aids in the diagnosis of influenza (flu) and provides epidemiological information on influenza. Influenza is an acute respiratory tract disease, which is often epidemic.
Related Devices
K971494 — VIRAZYME INFLUENZA ID TEST FOR INFLUENZA TYPES A AND B VIRUSES · Zymetx, Inc. · Sep 10, 1997
K982429 — ZSTATFLU TEST FOR INFLUENZA TYPES A AND B VIRUSES · Zymetx, Inc. · Aug 25, 1998
K063689 — BD DIRECTIGEN EZ FLU A+B TEST · Becton, Dickinson & CO · Mar 7, 2007
K160164 — BD Veritor System for Rapid Detection of Flu A + B Labratory Kit · Becton, Dickinson & CO · Feb 25, 2016
K042472 — BD DIRECTIGEN EZ FLU A + B · Becton, Dickinson & CO · Jul 7, 2005
Submission Summary (Full Text)
{0}
K950838
JUN 17 1996
# Appendix C
## Safety and Effectiveness Summary for ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza
This information is provided as a summary of the safety and effectiveness of the ZymeTx, ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza Viruses. For more detailed information please refer to the product package insert.
The ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza is intended for use as a screening test for respiratory viral cultures infected with influenza type A and B and parainfluenza types 1, 2, 3, and 4. This test will screen culture fluids for the presence of these viruses, but it is not indented for the definitive typing of these viruses.
Influenza and parainfluenza viruses possess surface glycoproteins with neuraminidase activity, that hydrolyze substrates which contain alpha-ketosidically linked N-acetylneuraminic acid (Neu5Ac). A modified Neu5Ac molecule has been synthesized and coupled to a chromogen to produce the neuraminidase substrate. In the presence of influenza and parainfluenza the chromogenic substrate is then cleaved by the action of viral neuraminidase, releasing a free chromogen. This free chromogen is then precipitated by combining with a diazonium salt to produce a red color. The red precipitate is then concentrated and collected from the solution onto a filter device.
## Performance Characteristics
Viral studies were performed using the ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza at four separate locations throughout the United States. These studies compared the ViraZyme® Culture Confirmation Screen results to results obtained from standard culture confirmation with monoclonal antibodies.
At a Southern medical center, 177 previously identified patient specimens from freeze were examined in the ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza. A total of 98/177 specimens were positive in the ViraZyme® Culture Confirmation Screen. Of these 98, ViraZyme® was positive in 100% (23/23) of influenza A IFA confirmed positives, 100% (22/22) of influenza B IFA confirmed positives, 94.4% (17/18) of parainfluenza 1 IFA confirmed positives, 100% (8/8) parainfluenza 2 IFA confirmed positives, and 63.4% (26/41) of parainfluenza 3 IFA confirmed positives. The test site reported two virus-negative specimens by IFA as appearing weakly positive in ViraZyme®. The reminder of the specimens were IFA-negative specimens and these were all negative in ViraZyme®.
Page 114
{1}
Appendix C - Safety and Effectiveness Summary for ViraZyme® Culture Confirmation Screen for Influenza and Parafinfluenza
At a Southwestern medical center, 97 previously identified patient specimens from freeze were examined in the ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza. A total of 37/97 specimens were positive in the ViraZyme® Culture Confirmation Screen. Of these 37, ViraZyme® was positive in 100% (36/36) of influenza A IFA confirmed positive and 1.9% (1/53) of Parainfluenza 3 IFA confirmed positives. The remainder of the specimens at this site were virus-negative by IFA and likewise negative in ViraZyme®.
The low ViraZyme® sensitivity for parainfluenza 3 at this site is believed to be related to the presence of bovine serum in the feed medium used at this site. This site routinely used a commercial medium containing 2% bovine serum to feed the inoculated PRMK cultures. It has been shown, in our in-house studies that presence of serum at this concentration will result in a reduced ViraZyme® result. It has been observed that parainfluenza 3 typically gives a lower ViraZyme® result relative to a typical influenza result. Due to the results of our own in-house evaluation with serum and those seen at this test site, we have defined the ViraZyme® culture procedure to exclude serum from virus culture medium.
At a Midwestern medical center, 123 previously identified patient specimens from freeze were examined in the ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza. A total of 62/123 specimens were positive in the ViraZyme® Culture Confirmation Screen. Of these 62, ViraZyme® was positive in 100% (11/11) of influenza A IFA confirmed positives, 100% (16/16) of influenza B IFA confirmed positives, 100% (5/5) of parainfluenza 1 IFA confirmed positives, 77.8% (7/9) of parainfluenza 2 IFA confirmed positives and 55.6% (5/9) of parainfluenza 3 IFA confirmed positives. Eighteen specimens, which were originally identified as virus positive by IFA at the time of isolation from the patient as being influenza A (five), parainfluenza 1 (four) and parainfluenza 3 (nine), were ViraZyme®-positive, but IFA negative, when tested for these same viruses. The remainder of the specimens were virus-negative by IFA and appropriately negative in ViraZyme®.
In an in-house evaluation, in the Southwest, a total of 403 previously identified frozen as well as fresh patient specimens were examined in the ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza. A total of 185/403 specimens were positive in the ViraZyme® Culture Confirmation Screen. Of these 185, ViraZyme® was positive in 100% (49/49) of influenza A IFA confirmed positives.
100% (99/99) of influenza B IFA confirmed positives, 85.7% (12/14) of parainfluenza 1 IFA confirmed positives, 100% (7/7) of parainfluenza 2 IFA confirmed positives, and 77.8% (14/18) of parainfluenza 3 IFA confirmed positives. Two other ViraZyme®-positives were IFA confirmed for parainfluenza 4 along with two additional ViraZyme®-positives which were
Page 115
{2}
Appendix C - Safety and Effectiveness Summary for ViraZyme® Culture Confirmation Screen for Influenza and Parafinfluenza
mumps virus strains. Mumps, being the other human virus, besides the influenza and parainfluenza to produce neuraminidase, gave predictably positive in ViraZyme® as well as by IFA confirming MAb. The reminder of the specimens were IFA-negative for influenza and parainfluenza and these were also all negative in ViraZyme®.
Numerous lots of uninoculated Primary Rhesus Monkey Kidney were also included as negative culture controls to establish that there was no indication of false positive ViraZyme® results associated with either the cells or breakthrough endogenous simian virus. ViraZyme® was shown to be negative in all of these culture controls.
## Technical Information
For technical information and comments regarding this product, you may contact the ViraZyme® Product Manager at (405) 271-1383.
This Safety and Effectiveness Summary has been provided as a part of the 510(k) notification for the ViraZyme® Culture Confirmation Screen for Influenza and Parainfluenza.
Craig D. Shimasaki
Executive Director of Research
2/23/95
Date
Page 116
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.