IS EBV-EBNA-1 IGG ELISA TEST SYSTEM

K981829 · Diamedix Corp. · GNP · Feb 16, 1999 · Microbiology

Device Facts

Record IDK981829
Device NameIS EBV-EBNA-1 IGG ELISA TEST SYSTEM
ApplicantDiamedix Corp.
Product CodeGNP · Microbiology
Decision DateFeb 16, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3235
Device ClassClass 1

Indications for Use

For the qualitative and semi-quantitative determination of IgG antibodies to Epstein-Barr Virus (recombinant) Nuclear Antigen-1 (EBV-EBNA-1 IgG) in human serum by indirect enzyme immunoassay. The Is-EBV-EBNA-1 IgG Test Kit may be used in combination with other Epstein-Barr serologies (Viral Capsid Antigen (VCA) IgG and IgM , Epstein-Barr Nuclear Antigen-1 (EBNA-1) IgM, Early Antigen-Diffuse (EA-D) IgG and IgM and heterophile antibody as an aid in the diagnosis of infectious mononucleosis (IM). The evaluation of paired sera, to determine a significant increase in EBNA-1 IgG antibody titer, can also aid in the diagnosis of acute infection. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated Processor.

Device Story

The Is-EBV-EBNA-1 IgG ELISA Kit is an in vitro diagnostic assay for detecting IgG antibodies to Epstein-Barr Nuclear Antigen-1 in human serum. The device uses recombinant EBNA-1 antigen bound to microwells. Patient serum is added; if anti-EBNA-1 IgG is present, it binds to the antigen. After washing, a horseradish peroxidase-labeled anti-human IgG conjugate is added, followed by a TMB substrate. The resulting color change is measured spectrophotometrically at 450/600 nm. The absorbance is directly proportional to the concentration of IgG antibodies. The assay can be performed manually or using the MAGO® Plus Automated Processor. Results are calculated as an index value (sample absorbance/cut-off absorbance). Clinicians use these results alongside other EBV serologies to diagnose infectious mononucleosis or acute infection.

Clinical Evidence

Clinical evaluation involved 175 characterized patient sera (102 convalescent, 34 seronegative, 39 current infection). Testing showed 95.4% overall agreement with EBV serological status. Sensitivity was 98.0% and specificity ranged from 87.2% to 97.1%. Precision studies (intra-assay and inter-assay) were conducted across three sites, with CVs generally ranging from 2.46% to 10.78%. Correlation between manual and MAGO Plus automated processing was high (Pearson r = 0.991). No cross-reactivity was observed with VZV, CMV, or HSV.

Technological Characteristics

In vitro diagnostic ELISA. Recombinant EBNA-1 antigen bound to microwells. HRP-labeled goat anti-human IgG conjugate. TMB substrate. Manual or automated (MAGO Plus) processing. Spectrophotometric readout at 450/600 nm. Storage at 2-8°C.

Indications for Use

Indicated for qualitative and semi-quantitative detection of IgG antibodies to EBV-EBNA-1 in human serum to aid in the diagnosis of infectious mononucleosis (IM) and acute infection. For use in clinical laboratory settings.

Regulatory Classification

Identification

Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Sent bv: IMMUNO PROBE 3016957824 Page 3 / 8 2 | 16/99 K981829 Job 39 ## 510k Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K981829 #### Applicant Information; | Date Prepared: | May 18, 1998 | |----------------|------------------------------------------------------| | Name: | Columbia Bioscience, Inc. | | Address: | 8775 M Centre Park Drive, #559<br>Columbia, MD 21045 | | Contact Person: | Norman Jenkins | |-----------------|----------------| | Phone Number: | 410-995-1278 | | Fax Number: | 410-995-0508 | #### Device Information: | Trade Name: | ZEBV EBNA-1 IgG ELISA Kit | |----------------------|----------------------------------------| | Common Name. | EBV Nuclear Antigen-1 EIA Test | | Classification Name; | Epstein Barr Virus Serological Reagent | #### Equivalent Device Description: Wampole EBNA-1 IgG ELISA. Wampole EBNA-1 IgG ELISA kit contains instructions and materials for the qualitative and semi-quantitative detection of IgG antibodies to EBV-EBNA-1 IgG in human serum by indirect ELISA Device Description: The CEBV-EBNA-1 IgG ELISA Kit is an enzyme-linked inımunosorbent assay (ELISA) for the detection of IgG to Epstein Barr Nuclear antigen-1 in human serum. Intended Use: For the qualitative and semi-quantitative determination of IgG antibodies to Epstein-Barr Virus (recombinant) Nuclear Antigen-1 (EBV-EBNA-1 IgG) in human serum by indirect enzyme immunoassay. The Is-EBV-EBNA-1 IgG Test Kit may be used in combination with other Epstein-Barr serologies (Viral Capsid Antigen (VCA) IgG and IgM , Epstein-Barr Nuclear Antigen-1 (EBNA-1) IgM, Early Anigen-Diffise (EA-D) IgG and IgM and heterophile antibody as an aid in the diagnosis of infectious mononucleosis (IM). The evaluation of paired sera, to determine a significant increase in EBNA-1 IgG antibody titer, can also and in the diagnosis of acute infection. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated Processor. #### Principle of Procedure: Recombinant EBNA-1 antigen is bound to microwells. Diluted patient sera. Cut-Off Calibrator and controls are placed inthe microwells and incubated. Anti-EBNA-1 IgG antibodes, if present, will bind to the antigen forming antigen-antibody complexes. Residual sample is eliminated by aspirating and washing. Conjugate (horseradish peroxidase-labeled anti-human IgG) is added and will bind to these complexes. Unbound conjugate is removed by aspiration and washing. Substrate is then added and incubated. In the presence of bound enzyme the substrate is converted to an end product. The absorbance of this end product can be read spectrophotometrically at 4.0 {1}------------------------------------------------ nm (reference 600-630 nm) and is directly proportional to the concentration of lgG antibodies to EBNA-1 present in the sample. The Is-EBV-EBNA-1 IgG ELISA kit and the Wampole EBNA-1 IgG ELISA are substantially equivalent in that. - Both are in vitro immunologic methods. ﻟﺴﻨﺔ - Both are intended for use in the detection of IgG antibody to EBV-EBNA-1 in human serum 2. - Both are based on the formation of a complex between EBNA-1 antigens and antibody నా - 4. Both use antigen coated microtiter plates. - Both are qualitative/semi-quantitative assays. ળ - Both use goat anti-human IgG conjugated to horseradish peroxidase. 6. - 7. Both use TMB as the enzyme substrate A detailed comparison between the proposed devise and the predicate device is shown in Table 1 Conclusions: The Diamedix Is-EBV-EBNA-1 IgG is substantially equivalent to the Wampole EBNA-1 ELISA for the detection of IgG antibodies to EBV-EBNA-1 in human serum to aid in the diagnosis of infectious mononucleosis. The device is as safe, as effective, and performs as well as the legally marketed device described. {2}------------------------------------------------ | Table 1 | | | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | PROPOSED DEVICE<br>Diamedix Is-EBV-EBNA IgG ELISA Kit | PREDICATE DEVICE<br>Wampole EBNA-1 IgG ELISA | | Intended Use | For the qualitative and semi-quantitative determination of<br>IgG antibodies to Epstein-Barr Virus (recombinant)<br>Nuclear Antigen-1 (EBV-EBNA-1 IgG) in human serum<br>by indirect enzyme immunoassay. The Is-EBV-EBNA-1<br>IgG Test Kit may be used in combination with other<br>Epstein-Barr serologies (Viral Capsid Antigen (VCA)<br>IgG and IgM, Epstein-Barr Nuclear Antigen-1 (EBNA-1)<br>IgM, Early Antigen-Difluse (EA-D) IgG and IgM and<br>heterophile antibody as an aid in the diagnosis of<br>infectious mononucleosis (IM). The evaluation of paired<br>sera, to determine a significant increase in EBNA-1 IgG<br>antibody titer, can also aid in the diagnosis of acute<br>infection. These reagents can be used either manually or<br>in conjunction with the MAGO® Plus Automated<br>Processor. | The Wampole Laboratories (Wampole) Epstein-Barr<br>Virus Nuclear Antigen-1 (EBNA-1) IgG Enzyme-Linked<br>Immunosorbent Assay (ELISA) is intended for the<br>qualitative and semi-quantitative determination of IgG<br>antibody in human serum to EBNA-1 recombinant<br>antigen. The Wampole EBNA-1 IgG assay may be used<br>in conjunction with other Epstein-Barr serologies (VCA<br>IgG, VCA IgM, EA (R&D), and heterophile) as an aid in<br>the diagnosis of infectious mononucleosis. | | Methodology | Enzyme immunoassay (EIA) | Enzyme Linked Immunosorbent Assay (ELISA) | | Specifications | For in vitro diagnostic use.<br>For use with fresh or frozen human serum.<br>Avoid lipemic, hemolyzed, contaminated, or icteric sera.<br>Assay performed on 1:21 dilution of serum at 18-30°C.<br>Store at 2-8°C. | For in vitro diagnostic use. For use with fresh or frozen<br>human serum. Assay performed on 1:21 dilution of<br>serum at 21-25°C. Store at 2-8°C. | | Design | Is-EBV-EBNA IgG Test Kit. 96 determinations. Un-<br>diluted Calibrator, Positive, and Negative controls. | EBNA-1 IgG ELISA. 96 determinations. Undiluted<br>Calibrator, High positive, Low positive, and Negative<br>controls. | | Principles of<br>Operation | Purified, recombinant EBNA-1 antigen is bound to<br>microwells (solid phase). Diluted human serum is asses<br>to the microwell which binds human anti-EBNA IgG, if<br>present. Solid phase is washed and exposed to anti-<br>human IgG conjugate. Solid phase is washed and<br>exposed to enzyme substrate to develop color. Strong<br>acid is added to stop reaction. The color is read at<br>450/600 nm on an EIA reader. | Diluted patient serum is incubated with purified,<br>recombinant EBNA-1 antigen bound to the solid surface<br>of a microtiter well. If IgG antibodies against EBV-<br>EBNA-1 are present in the serum, antigen-antibody<br>complexes are formed. These complexes bind with HRP<br>labeled anti-human IgG which react with the addition of<br>chromogen, resulting in a color development. The<br>absorbance is measured at 450/630 nm. | | Performance<br>Characteristics | Relative Sensitivity: 98.0%<br>Relative Specificity (Current Infection): 87.2% Relative<br>Specificity (Seronegative): 97.1<br>Agreement: 95.4%<br>Intra-assay Precision (Positive samples, all sites)<br>Overall Manual- 2.46-8.69 MAGO Plus- 3.33-6.55<br>Interassay Precision (Positive samples, all sites)<br>Overall Manual- 3.73-8.86 MAGO Plus- 7.80-9.23<br>No Cross-reactivity | Relative Sensitivity: 97.8%<br>Relative Specificity: 100%<br>Agreement: 98.3%<br>Inter-Site Precision (Positive samples, all sites)<br>Overall: 7.70-10.78%<br>No Cross-reactivity | | Enzyme Used | Horesradish Peroxidase | Horesradish Peroxidase | | Substrate | TMB | TMB | | Specimen | Serum<br>Sample Absorbance/Cut-off Absorbance = Index Value | Serum<br>Sample Absorbance/Cut-off Absorbance = ISR | | Calculation of Results | | | | Interpretation | <0.90 Negative for EBNA IgG<br>0.90-1.09 Equivocal for EBNA IgG<br>> 1.10 Positive for EBNA IgG | ≤0.90 Negative for EBNA-1 IgG<br>0.91-1.09 Equivocal for EBNA-1 IgG<br>> 1.10 Positive for EBNA-1 IgG | | Materials | 96 microwells in 12x8 strips, Wash concentrate, Sample<br>Diluent, Conjugate, Calibrator, Controls, Substrate, Stop<br>Solution | 96 microwells in 12x8 strips, Wash Buffer, Serum<br>Diluent, HRP Conjugate, Calibrator, Controls,<br>Chromogen, Stop Solution | {3}------------------------------------------------ # Performance Characteristics ## A. Clinical Sensitivity and Specificity Using Characterized Sera Scra from one hundred and seventy-five patients were characterized using commercially available kits for VCA IgG, VCA IgM, EBNA IgG and heterophile antibodies. Based on the results of this testing, the patient sera were characterized as follows : - · 102 sera were characterized as convalescent (past infection). These were positive for VCA IgG and/or EBNA IgG antibodies and negative for VCA IgM and heterophile antibody. - · 34 sera were characterized as seronegative. These were negative for VCA IgG, VCA IgM, EBNA IgG and heterophile antibody. - · 39 sers were characterized as having a current (recent) infection. These were positive for VCA IgM and/or heterophile antibody and were negative for EBNA IgG. All 175 sera were then tested by an independent clinical commercial laboratory in the 18-15BNA-1 IgG Test Kit. The results obtained are shown in Table 2: | | TABLE 2 | Convaloscont | Current Infection | Seronegative | |-------|------------|--------------|-------------------|--------------| | 1 IgG | POSITIVE | 100 | 5 | 1 | | | NEGATIVE | 2 | 34 | 33 | | | *EQUIVOCAL | 0 | 0 | 0 | EBV Serological Status 15-1:BNA-1 - · Of the 102 convalescent sera tested, 100 were positive for anti-EBNA IgG and two were negative. - · Of the thirty-mine current (recent) infection samples tested, thirty-four were negative for anti-EBNA IgG and five were positive. - · Of the thirty-four seronegative sera tested, thirty-three were negative for anti-EBNA IgG and one serum icsted positive. - · The overall agreement of the Is-EBNA IgG test kit compared to EBV serological status was 161/175 = 95.4%. {4}------------------------------------------------ #### B. Precision To decemine the precision of the Is-EBNA-1 IgG Test Kit, four positive shares and site is site its is site it in site it it site it it site it it sit site it it sit site it i To deternine the precision of the IS-EBNA-T ISO Text Kits in 1981 your read interassay precision obtained at each site is shown in Tables 3, 4 and 5. TABLE 3 : Site #1 - Intra-Assay and Interassay Precision | SERUM | INTRA-ASSAY RUN 1 | | INTRA-ASSAY RUN 2 | | INTRA-ASSAY RUN 3 | | INTERASSAY | | |---------|-------------------|-------|-------------------|-------|-------------------|-------|---------------|-------| | | MEAN<br>INDEX | CV% | MEAN<br>INDEX | CV% | MEAN<br>INDEX | CV% | MEAN<br>INDEX | CV% | | A (POS) | 1.61 | 4.44 | 1.62 | 6.57 | 1.58 | 8.23 | 1.60 | 6.47 | | B (POS) | 2.28 | 4.27 | 2.30 | 5.84 | 2.19 | 7.04 | 2.26 | 5.98 | | C (POS) | 1.54 | 6.26 | 1.43 | 7.15 | 1.48 | 8.69 | 1.49 | 7.82 | | D (POS) | 1.91 | 4.43 | 1.82 | 4.74 | 1.82 | 4.18 | 1.85 | 4.92 | | E (NEG) | 0.08 | 20.41 | 0.07 | 62.39 | 0.04 | 42.61 | 0.06 | 53.16 | | F (NEG) | 0.49 | 12.12 | 0.39 | 12.06 | 0.36 | 10.72 | 0.42 | 17.68 | | | | | | | | CAL | 1.00 | 9.56 | | | | | | | | PC | 1.44 | 6.84 | | | | | | | | NC | 0.04 | 110.2 | | TABLE 4 : Site #2- Intra-Assay and Interassay Precision | |---------------------------------------------------------| |---------------------------------------------------------| | SERUM | INTRA-ASSAY RUN 1<br>MEAN<br>INDEX | CV% | INTRA-ASSAY RUN 2<br>MEAN<br>INDEX | CV% | INTRA-ASSAY RUN 3<br>MEAN<br>INDEX | CV% | INTERASSAY<br>MEAN<br>INDEX | CV% | | |---------|------------------------------------|-------|------------------------------------|------|------------------------------------|-------|-----------------------------|-------|--------| | A (POS) | 1.54 | 2.97 | 1.52 | 4.47 | 1.50 | 4.41 | 1.55 | 4.19 | | | B (POS) | 2.30 | 2.96 | 2.25 | 3.13 | 2.35 | 3.80 | 2.30 | 3.73 | | | C (POS) | 1.42 | 5.56 | 1.40 | 3.10 | 1.51 | 4.60 | 1.44 | 5.45 | | | D (POS) | 1.77 | 2.46 | 1.73 | 2.82 | 1.88 | 2.88 | 1.79 | 4.49 | | | E (NEG) | 0.08 | 43.69 | 0.08 | 8.34 | 0.10 | 12.57 | 0.08 | 27.49 | | | F (NEG) | 0.42 | 39.98 | 0.34 | 5.60 | 0.53 | 7.67 | 0.43 | 28.98 | | | | | | | | | CAL | 1.00 | 1.74 | n = 18 | | | | | | | | PC | 1.54 | 3.48 | n = 12 | | | | | | | | NC | 0.05 | 28.39 | n = 12 | | | TABLE 5 : Site #3 - Intra-assay and Interassay Precision | | |--|----------------------------------------------------------|--| |--|----------------------------------------------------------|--| | SERUM | INTRA-ASSAY RUN 1<br>MEAN<br>INDEX | INTRA-ASSAY RUN 1<br>CV% | INTRA-ASSAY RUN 2<br>MEAN<br>INDEX | INTRA-ASSAY RUN 2<br>CV% | INTRA-ASSAY RUN 3<br>MEAN<br>INDEX | INTRA-ASSAY RUN 3<br>CV% | INTERASSAY<br>MEAN<br>INDEX | INTERASSAY<br>CV% | |---------|------------------------------------|--------------------------|------------------------------------|--------------------------|------------------------------------|--------------------------|-----------------------------|-------------------| | A (POS) | 1.72 | 6.19 | 1.51 | 4.50 | 1.63 | 7.41 | 1.62 | 8.20 | | B (POS) | 2.53 | 5.89 | 2.33 | 4.95 | 2.42 | 6.25 | 2.43 | 6.51 | | C (POS) | 1.57 | 4.41 | 1.39 | 5.22 | 1.51 | 7.10 | 1.49 | 7.45 | | D (POS) | 2.04 | 5.41 | 1.78 | 6.14 | 2.00 | 8.52 | 1.94 | 8.86 | | E (NEG) | 0.06 | 26.22 | 0.06 | 23.80 | 0.04 | 38.91 | 0.05 | 32.41 | | F (NEG) | 0.45 | 10.04 | 0.31 | 6.19 | 0.49 | 7.99 | 0.42 | 20.80 | | | | | | | | CAL | 1.00 | 5.38 | | | | | | | | PC | 1.63 | 5.21 | | | | | | | | NC | 0.05 | 10.83 | #### C. Specificity with Potentially Cross-Reactive Sera Sixteen sera, non-reactive) for IgG antibudies to EBNA-1 in the Is-EBNA-1 IgG Test Kit, were tested by IIIA for IgG antibody to varietla zoster, cytomegalovirus and herpes simplex virus. 15/15 anti-VZV IgC passive sers were non-reactive for anti-EBNA-1 IgG; 3/3 anti-CMV IgG positive sera were non-reactive for anti-EBNA-1 IgG and 3/3 and-HSV positive sera were non-reactive for anti-EDNA-1 IgG. "This suggests that no specific crass-tenchyly should be expected with the Is-EBNA-1 IgG Test Kit from these analytes. ## D. Correlation of Manual and MAGO Plus Results The Is-EBNA-1 IgG Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence of the manual and MAGO Plus procedures, the results of 197 serum samples tested by hoth methods were plotted. A seattergram and regression line of the results obtained with 95% confidence interests in shown in Figure 3. The data indicate good correlation with a Pearson Correlation Coefficient of 0.991. . {5}------------------------------------------------ Image /page/5/Figure/3 description: The image is a scatter plot that shows the relationship between MAGO PLUS INDEX VALUES and MANUAL INDEX VALUES. The x-axis represents MANUAL INDEX VALUES, ranging from 0 to 6, while the y-axis represents MAGO PLUS INDEX VALUES, ranging from -1 to 7. A regression line is plotted on the scatter plot, with the equation Y = -0.00022 + 1.1060 X, and the correlation coefficient (r) is 0.991. ## FIGURE 3 : Manual and MAGO Plus Result Correlation #### D. MAGO Plus Precision The precision of the assay when performed on the MAGO Plus Autonated EIA Processor was determined by assaying six sera ten times each in three different runs. Table 6 shows the intra-and interassay precision obtained using the MAGO Plus. | SERUM | INTRA-ASSAY RUN 1 | | INTRA-ASSAY RUN 2 | | INTRA-ASSAY RUN 3 | | INTERASSAY | | | |---------|-------------------|-------|-------------------|-------|-------------------|--------|------------|-------|-------| | | MEAN INDEX | CV% | MEAN INDEX | CV% | MEAN INDEX | CV% | MEAN INDEX | CV% | | | A (POS) | 1.49 | 3.81 | 1.65 | 6.55 | 1.38 | 4.58 | 1.51 | 9.05 | | | B (POS) | 2.26 | 5.60 | 2.55 | 3.33 | 2.15 | 5.02 | 2.32 | 8.65 | | | C (POS) | 1.52 | 5.19 | 1.66 | 5.82 | 1.39 | 6.30 | 1.52 | 9.23 | | | D (POS) | 1.76 | 6.11 | 1.94 | 4.35 | 1.69 | 5.18 | 1.80 | 7.00 | | | E (NEG) | 0.09 | 35.14 | 0.10 | 0.00 | 0.04 | 129.10 | 0.08 | 56.11 | | | F (NEG) | 0.37 | 42.35 | 0.38 | 11.10 | 0.34 | 15.19 | 0.36 | 26.54 | | | | | | | | | | CAL | 1.00 | 4.95 | | | | | | | | | PC | 1.40 | 14.29 | | | | | | | | | NC | 0.07 | 86.60 | TABLE 6 : Sile #2- InIra-Assay and Interassay Precision - MAGO Plus {6}------------------------------------------------ Image /page/6/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB 16 1999 Diamedix Corporation c/o Norman Jenkins Columbia Bioscience, Inc. 8775 M Centre Park Drive, #559 Columbia, MD 21045 K981829 Re: Trade Name: Is EBV-EBNA-1 IgG ELISA Test System Regulatory Class: I Product Code: GNP Dated: December 14, 1998 Received: December 14, 1998 Dear Mr. Jenkins: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {7}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours. Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ 510(k) Number: K981829 Device Name: EBV- EBNA-1 IgG ELISA Indications For Use: The The EBNA-1 IgG kit is an Enzyme-Linked Immunosorbent Assay (ELISA) For the qualitative and semi-quantitative determination of IgG antibodies to Epstein-Barr Virus (recombinant) Nuclear Antigen-1 (EBV-EBNA-1 IgG) in human serum by indirect enzyme immunoassay. The Is-EBV-EBNA-1 IgG Test Kit may be used in combination with other Epstein-Barr serologies (Viral Capsid Antigen (VCA) IgG and IgM , Epstein-Barr Nuclear Antigen-1 (EBNA-1) IgM, Early Antigen-Diffuse (EA-D) IgG and IgM and heterophile antibody as an aid in the diagnosis of infectious mononucleosis The evaluation of paired sera, to determine a significant increase in EBNA-(IM). 1 IgG antibody titer, can also aid in the diagnosis of acute infection. These reagents can be used either manually or in conjunction with the MAGO® Plus Automated Processor. ### PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Y (Per 21 CFR 801.109) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ OR Over-The Counter Use (Optional Format 1-2-96) Division of Clinical Laboratory Devices K981829 510(k) Number_ (
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...