K020706 · Pan Bio Pty. , Ltd. · GNP · Jun 13, 2002 · Microbiology
Device Facts
Record ID
K020706
Device Name
EBV-VCA IGG ELISA TEST, MODEL EBG-100
Applicant
Pan Bio Pty. , Ltd.
Product Code
GNP · Microbiology
Decision Date
Jun 13, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3235
Device Class
Class 1
Indications for Use
The Epstein Barr Virus Viral Capsid Antigen (EBV-VCA) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV-VCA in serum as an aid in the clinical laboratory diagnosis of EBV infection in patients with clinical symptoms consistent with infectious mononucleosis (IM). The PANBIO EBV-VCA IgG ELISA should be used in conjunction with other EBV serologies.
Device Story
The EBV-VCA IgG ELISA Kit is an enzyme-linked immunosorbent assay used in clinical laboratories to detect IgG antibodies to Epstein-Barr virus viral capsid antigen (EBV-VCA) in human serum. The device utilizes microwells coated with EBV-VCA gp125 antigen derived from P3HR1 cell lines. Patient serum is added to the wells; if specific IgG antibodies are present, they bind to the immobilized antigen. After washing, peroxidase-conjugated anti-human IgG is added, followed by a TMB/H2O2 substrate system. The enzymatic reaction produces a blue color, which turns yellow upon the addition of an acid stop solution. The intensity of the color, measured via spectrophotometry, indicates the presence of EBV-VCA IgG antibodies. Healthcare providers use these results in conjunction with other EBV serological tests to assist in the diagnosis of EBV infection in patients with symptoms of infectious mononucleosis.
Clinical Evidence
Clinical performance was evaluated using 508 total serum samples across two study sites. Site 1 (retrospective, n=156) showed 69.2% sensitivity for acute cases, 94.1% for past infection, and 96.4% specificity. Site 2 (prospective, n=352) showed 92.9% sensitivity for acute cases, 89.3% for past infection, and 91.7% specificity. Analytical specificity was confirmed with a 29-specimen cross-reactivity panel (CMV, VZV, HSV-1, HSV-2, ANA, Rheumatoid Factor), showing 0% cross-reactivity. Reproducibility was assessed via ANOVA across three sites, with total CVs ranging from 4.7% to 14.1%.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) using polystyrene microwells coated with immunopurified EBV-VCA gp125 antigen. Detection system utilizes peroxidase-conjugated anti-human IgG and TMB/H2O2 substrate. Results are determined by colorimetric change measured via spectrophotometry. The device is a manual or semi-automated in vitro diagnostic reagent kit.
Indications for Use
Indicated for qualitative detection of IgG antibodies to EBV-VCA in human serum as an aid in diagnosing EBV infection in patients presenting with clinical symptoms consistent with infectious mononucleosis (IM).
Regulatory Classification
Identification
Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).
Predicate Devices
DiaSorin Incstar EBV-VCA IgG ELISA
Related Devices
K981812 — EBV-VCA IGG ELISA TEST SYSTEM · Diamedix Corp. · Mar 4, 1999
K020717 — EBV V VCA IGM ELISA KIT MODEL# EBM-200 · Pan Bio Pty. , Ltd. · Jun 13, 2002
K020707 — EBV-EBNA IGG ELISA KIT, MODEL EBG-100 · Pan Bio Pty. , Ltd. · Jun 13, 2002
K980912 — VCA IGG ELISA TEST SYSTEM · Clark Laboratories, Inc. · Jul 22, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN 1 3 2002
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K020706
### Applicant Information:
| Date Prepared: | 07th June, 2002 |
|----------------|---------------------------------------------|
| Name: | PANBIO Limited |
| Address: | 116 Lutwyche Road<br>Windsor 4030 Australia |
| Contact Person: | Helen Jennings |
|-----------------|-------------------|
| Phone Number. | 61-(0)7-3357-1177 |
| Fax Number. | 61-(0)7-3357-1222 |
### Device Information:
| Trade Name: | EBV-VCA IgG ELISA Kit |
|----------------------|---------------------------------|
| Common Name. | EBV-VCA IgG EIA Test |
| Classification Name; | EBV-VCA IgG Serological Reagent |
#### Equivalent Device:
DiaSorin Incstar EBV-VCA IgG ELISA
#### Device Description:
The EBV-VCA IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to EBV-VCA antigen in human serum.
#### Intended Use:
The Epstein Barr Virus Viral Capsid Antigen (EBV-VCA) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV-VCA in serum as an aid in the clinical laboratory diagnosis of EBV infection in patients with clinical symptoms consistent with infectious mononucleosis (IM). The PANBIO EBV-VCA IgG ELISA should be used in conjunction with other EBV serologies.
#### Principle of Procedure:
Serum containing antibodies to VCA antigen, when present, combine with EBV-VCA gp125 antigen attached to the polystyrene surface of the microwells. This antigen is grown in P3HR1 cell lines, and immunopurified utilising a monoclonal specific for gp125. Residual serum is removed by washing and peroxidase conjugated IgG is added. The microwells are washed and a colourless substrate system, anti-human tetramethylbenzidine/hydrogen peroxide (TMB/H2O2) is added. The substrate is hydrolysed by the enzyme and the chromogen changes to a blue colour. After stopping the reaction with acid, the TMB becomes yellow. Color development is indicative of the presence of EBV-VCA IgG antibodies in the test sample.
{1}------------------------------------------------
## PERFORMANCE CHARACTERISTICS
### Study Site 2:
156 frozen retrospective sera of various ages and genders were submitted to a state health lab in Maryland USA for EBV testing. The sera include samples from the following groups: 28 seronegative samples, 26 samples from patients with acute Infectious Mononucleosis, and 102 samples from patients with past exposure to EBV. These sera were tested on the PANBIO EBV-VCA IgG kit and EBV ELISA assays from an alternate manufacturer to determine the EBV status of the sera. The PANBIO EBV VCA IgG results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status. The data is summarized in the Table 1.
| PANBIO ELISA | | | | |
|--------------------------------------------------------------|----------|------------|----------|-------|
| EBV Status | Positive | Equivocal* | Negative | Total |
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-) | 1 | 0 | 27 | 28 |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-) | 18 | 4 | 4 | 26 |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 96 | 2 | 4 | 102 |
| Total | 115 | 6 | 35 | 156 |
# TABLE 1 EBV VCA IgG Serological Sensitivity and Specificity of PANBIO ELISA versus EBV Status
### 95% Confidence Interval
| Serological Sensitivity (Acute) | = 18/26 | = 69.2 % | 48.2 – 85.7 % |
|------------------------------------|-----------|----------|---------------|
| Serological Sensitivity (Past) | = 96/102 | = 94.1% | 87.6 – 97.8% |
| Serological Specificity (Negative) | = 27/28 | = 96.4 % | 81.6 – 99.9 % |
| Serological Agreement | = 141/156 | = 90.4 % | 84.6 – 94.5 % |
* Retesting of equivocal samples was not conducted, as the samples were unavailable.
Note: "Serological" sensitivity and specificity refers to the Comparison of the PANBIO assay results to that of other assays normally used to diagnose EBV associated IM. There was not an attempt to correlate the assay's results with disease presence or absence. No judgement can be made on the comparison's accuracy to predict disease. Since the above studies were performed on a pre-selected, retrospective, populations for the assay's positive and negative predictive value may be done or inferred.
{2}------------------------------------------------
### Study Site 3:
352 prospective sera of various ages and genders were tested at a private pathology laboratory in Queensland Australia for EBV testing. The sera include samples from the following groups: 48 seronegative samples, 42 samples from patients with acute Infectious Mononucleosis, and 262 samples from patients with past exposure to EBV. These sera were tested on the PANBIO EBV-VCA IgG kit and EBV ELISA assays from an alternate manufacturer to determine the EBV status of the sera. The PANBIO EBV-VCA IgG results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status. The data is summarized in Table 2.
| | PANBIO ELISA | | | |
|--------------------------------------------------------------|--------------|------------|----------|-----------------------------------------|
| EBV Status | Positive | Equivocal* | Negative | Total |
| Seronegative<br>VCA IgG (-)<br>VCA IgM (-)<br>EBNA IgG (-) | 4 | 0 | 44 | 48 |
| Acute<br>VCA IgM (+)<br>EBNA IgG (-) | 39 | 0 | 3 | 42 |
| Past Infection<br>VCA IgG (+)<br>VCA IgM (-)<br>EBNA IgG (+) | 234 | 1 | 27 | 262 |
| Total | 277 | 1 | 74 | 352 |
| Serological Sensitivity (Acute) | | = 39/42 | = 92.9% | 95% Confidence Interval<br>80.5 - 98-5% |
| Serological Sensitivity (Past) | | = 234/262 | = 89.3% | 85.6-93.1% |
| Serological Specificity (Negative) | = 44/48 | | = 91.7% | 80.0 - 97.7% |
| Serological Agreement | | = 317/352 | = 90.1% | 86.4-93.0% |
| TABLE 2 |
|--------------------------------------------------------|
| EBV VCA IgG Serological Sensitivity and Specificity of |
| PANBIO ELISA versus EBV Status |
*This equivocal sample was not tested on an alternative method due to insufficient sample. Collection of a further sample was not possible.
{3}------------------------------------------------
# REPRODUCIBILITY
The reproducibility of the PANBIO EBV-VCA IgG ELISA kit was determined by testing 8 sera 3 times each on three different days at three Australian study sites. Two sites were private pathology laboratories and the third site was PANBIO. Within-run, between day, between site and total precision were estimated by analysis of variance (ANOVA Type II) and are presented in table 3 below.
| Sample | n | *Mean | Within | | Between Day | | Between Site | | Total | |
|----------|----|-------|--------|-------|-------------|------|--------------|------|-------|-------|
| | | | *S.D | CV | *S.D | CV | *S.D | CV | *S.D | CV |
| Positive | 27 | 2.71 | 0.24 | 9.0% | 0.12 | 4.5% | 0.12 | 4.5% | 0.28 | 10.4% |
| Cut-off | 27 | 1.00 | 0.06 | 5.7% | 0.00 | 0.0% | 0.00 | 0.0% | 0.05 | 4.7% |
| Negative | 27 | 0.19 | 0.02 | 8.5% | 0.00 | 2.6% | 0.00 | 0.0% | 0.02 | 8.7% |
| #1 | 27 | 2.08 | 0.21 | 10.2% | 0.07 | 3.6% | 0.11 | 5.4% | 0.24 | 11.5% |
| #2 | 27 | 3.29 | 0.24 | 7.3% | 0.00 | 0.0% | 0.16 | 5.0% | 0.27 | 8.3% |
| #3 | 27 | 1.47 | 0.12 | 8.5% | 0.07 | 5.0% | 0.07 | 4.6% | 0.15 | 10.2% |
| #4 | 27 | 1.24 | 0.08 | 6.8% | 0.01 | 1.0% | 0.05 | 4.0% | 0.09 | 7.6% |
| #5 | 27 | 0.89 | 0.08 | 9.5% | 0.05 | 6.1% | 0.04 | 4.4% | 0.10 | 11.3% |
| #6 | 27 | 1.13 | 0.08 | 7.0% | 0.00 | 0.0% | 0.04 | 3.7% | 0.08 | 7.5% |
| #7 | 27 | 2.24 | 0.27 | 11.9% | 0.07 | 3.0% | 0.19 | 8.6% | 0.32 | 14.1% |
| #8 | 27 | 3.74 | 0.27 | 7.2% | 0.01 | 0.4% | 0.18 | 4.9% | 0.31 | 8.3% |
## TABLE 3 - REPRODUCIBILITY DATA PANBIO EBV-VCA IgG Study Site 1, 2 & 3 Precision Measures (Using Cut-Off Ratio)
All values are calculated from Ratios (Cut-Off using O.D) SD = Standard Deviation; CV = Coefficient of Variation
Note: Standard Deviation results have been rounded to two decimal places for tabulation purposes
{4}------------------------------------------------
# POTENTIAL CROSS-REACTIVITY
### Study Site 5:
A panel of 29 specimens from patients with confirmed diseases other than Epstein Barr Virus was tested to establish the analytical specificity of the EBV-VCA IgG ELISA Test. The specimens were from patients with diseases that have the potential for cross-reactivity. Each of the specimens included in the study was characterized with respect to disease diagnosis prior to analysis with the EBV-VCA IgG ELISA Test. No crossreactivity was observed for any of the 29 specimens tested from the disease panel. Refer to Table 4 for a summary of the results.
### TABLE 4
| Disease Type | Total Specimens | Positive Result |
|------------------------|-----------------|-----------------|
| Cytomegalovirus | 7 | (0/7) |
| Varicella zoster | 9 | (0/9) |
| Herpes Simplex virus 1 | 6 | (0/6) |
| Herpes Simplex virus 2 | 1 | (0/1) |
| Anti-Nuclear Antigen | 3 | (0/3) |
| Rheumatoid Factor | 3 | (0/3) |
| Total | 29 | (0/29) |
### PANBIO EBV-VCA IgG CROSS-REACTIVITY SPECIMEN PANEL
Results indicate that no specimens (0/29) were positive when analysed with the EBV-VCA IgG ELISA Kit. Refer to 'Study Document -- Site 5' for raw data and section 2.3.4.1 for the summary table.
The true negative result of 100% for the above disease panel is consistent with good analytical specificity for the EBV-VCA IgG ELISA Test.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUN 1 3 2002
Ms. Helen Jennings Quality and Regulatory Affairs Manager PANBIO Limited 116 Lutwyche Road Windsor. Brisbane Queensland, 4030 Australia
Re: k020706
> Trade/Device Name: Epstein Barr Virus VCA IgG ELISA Test Regulation Number: 21 CFR 866.3235 Regulation Name: Epstein - Barr virus Serological Reagents Regulatory Class: Class I Product Code: GNP Dated: May 9, 2002 Received: May 15, 2002
Dear Ms. Jennings:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and v additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page 1 of of 1
510(k) Number (if known): K020706
Device Name:
Indications For Use:
The Epstein Barr Virus Viral Capsid Antigen (EBV-VCA) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV-VCA in serum as an aid in the clinical laboratory diagnosis of EBV infection in patients with clinical symptoms consistent with infectious mononucleosis (IM). The PANBIO EBV-VCA IgG ELISA should be used in conjunction with other EBV serologies
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Deibres
Unical Laboratory Devices
PRESCRIPTION USE X
(Optional Format 3-10-98)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.