The REMEL RPR Card Test is an in vitro diagnostic assay used to detect reagin antibodies in human serum or plasma. The test utilizes an antigen suspension that reacts with reagin antibodies present in the patient sample; the presence of antibodies is indicated by macroscopic flocculation. The test is performed by laboratory personnel in a clinical setting. Results are interpreted visually by the technician to determine reactive or nonreactive status, with reactive samples often titered to determine the endpoint. The test aids in the screening for syphilis infection. Clinical utility is supported by comparison to established RPR and USR antigens, with results correlated against patient history of Treponema pallidum infection and treatment status.
Clinical Evidence
Bench-only study. Two batches of REMEL RPR antigen (in-house and LEE Laboratories) were evaluated against BD RPR and Difco USR antigens. Testing included reactive and nonreactive samples with documented Treponema pallidum infection stages and treatment histories. Results were compared against MHA-TP. No statistically significant differences in performance were observed between REMEL antigens and predicate antigens. Discrepant samples were resolved via blinded re-testing at a secondary site, showing consistency with reference methods. No clinical prospective/retrospective patient outcome data provided.
Technological Characteristics
In vitro diagnostic card test. Employs macroscopic flocculation principle using RPR antigen suspension. Manual visual interpretation. No electronic components, software, or energy sources. Sterilization not applicable.
Indications for Use
Indicated for the qualitative and semi-quantitative detection of reagin antibodies in human serum or plasma as a screening test for syphilis.
Regulatory Classification
Identification
Treponema pallidum nontreponemal test reagents are devices that consist of antigens derived from nontreponemal sources (sources not directly associated with treponemal organisms) and control sera (standardized sera with which test results are compared) used in serological tests to identify reagin, an antibody-like agent, which is produced from the reaction of treponema microorganisms with body tissues. The identification aids in the diagnosis of syphilis caused by microorganisms belonging to the genus Treponema and provides epidemiological information on syphilis.
Predicate Devices
BD RPR antigen
Difco USR antigen
Reference Devices
MHA-TP test
Related Devices
K992124 — AVANTI POLAR LIPIDS VDRL ANTIGEN SLIDE TEST KIT WITH 10 X 0.5 ML AMPULES, AVANTI POLAR LIPIDS BULK BDRL ANTIGEN, AVANTI · Avanti Polar Lipids, Inc. · Sep 20, 1999
K993633 — SYNTHETIC VDRL ANTIGEN · Centers for Disease Control and Prevention · Feb 23, 2000
K952802 — VISUWELL REAGIN II · Biomira Diagnostics, Inc. · May 23, 1996
Submission Summary (Full Text)
{0}
K955136
JUL 29 1996
Food and Drug Administration
Center for Devices and Radiological Health
Document Mail Center (HFZ-401)
9200 Corporate Blvd.
Rockville, MD 20850
RE: 510(k) Submission for RPR Card Test Kit Summary of Safety and Effectiveness
November 8, 1995
Dear Sir:
In accordance with 21 CFR Chapter 1, Subpart E, Sec. 807.93, and as part of the 510(k) submission for the RPR Liquid Controls, the following information is provided.
Reasonable assurance that this device is safe and effective was determined through the use of valid scientific nonclinical investigations including in vitro studies. The results of these studies are summarized in the attached Technical Insert.
This device is effective when tested under the conditions for its intended use. Each test kit contains a technical insert which provides adequate directions and precautions for use. Lastly, each insert contains information on the limitations and the performance characteristics of the test (including relative sensitivity and specificity data) when used according to the directions.
SUMMARY OF SAFETY AND EFFECTIVENESS DATA:
Raw performance data
Intended use of the product
Limitations of the product
Relative sensitivity and specificity data
Directions for procedure
510(k) Submission
Should there be any questions concerning this matter, please contact me at (706)736-6011.
Sincerely,
David A. Wall
Manager, Immunodiagnostics
{1}
# REMEL RPR CARD TEST
## PERFORMANCE CHARACTERISTICS
### University of Texas
Dr. Beth Hartwell
&
Michigan Department of Public Health
Mr. Harlan Stiefel
Two batches of RPR antigen were tested at the Univ. of Texas in comparison to the BD RPR antigen. Antigen #1 was made by REMEL and evaluated by the CDC before use in determining its performance characteristics. This material was considered "Satisfactory" by the CDC (copy of report enclosed). Antigen #2 was made for REMEL at LEE Laboratories in Grayson, GA and was also tested and deemed "Satisfactory" by the CDC. Both materials were evaluated since REMEL would like the option to make or purchase this antigen as needed.
The REMEL manufactured RPR antigen (Ag#1) was tested at U. of Texas without any significant discrepancies. It was tested vs the BD RPR antigen, and MHA-TP (for all reactives). Patient genders were documented for all specimens. Patient gender did not affect the outcome of the tests. Many of the reactive samples were titered for both RPR antigens and there was no statistically significant difference in performance between the two. Additionally, data is provided on several of the reactive samples which includes stage of infection with Treponema pallidum and history of treatment. Reactive results are consistent with the history of treatment and staging of illness.
The LEE Labs RPR antigen (Ag#2) was tested at both the Univ. of Texas vs the BD RPR antigen and Michigan Dept. of Health. vs the Difco USR antigen. Again, patient genders were recorded. Gender had no affect on the outcome of the tests. Also, stage of infection with Treponema pallidum and historical treatment on many of the reactive samples were recorded. All reactive samples were also tested vs the MHA-TP test with results consistent with the history of treatment and staging of illness. There were a number of the reactive samples tested at the U. of Texas that were read as nonreactive with the Ag#2 and reactive (undiluted endpoint titer) with the BD Ag. Repeat testing of the same samples with the same material at the U. of Texas yielded the same results. Because the same antigen (#2) was used at the Michigan site without any significant # of discrepancies, the remainder of the U. of Texas discrepant samples along with 12 known reactive samples and 12 known nonreactive samples were sent blinded to Michigan. Michigan reported results which were consistent between the known results at Texas and between RPR and USR. Additionally, except with only a couple of samples, the RPR results were consistent with the USR test in the discrepant samples. The raw data is enclosed and a summary of the testing follows.
81/10
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.