CRAG LATERAL FLOW ASSAY (CRAG LFA)

K102286 · Immuno-Mycologics, Inc. · GMD · Jul 20, 2011 · Microbiology

Device Facts

Record IDK102286
Device NameCRAG LATERAL FLOW ASSAY (CRAG LFA)
ApplicantImmuno-Mycologics, Inc.
Product CodeGMD · Microbiology
Decision DateJul 20, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3165
Device ClassClass 2

Indications for Use

The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum.

Device Story

The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic test for detecting cryptococcal antigen in serum. The device uses specimen wicking to transport diluted serum across a membrane containing gold-conjugated anti-cryptococcal monoclonal antibodies. If antigen is present, it forms a sandwich complex with immobilized anti-cryptococcal antibodies at the Test Line (T), creating a visible line. A control line (C) confirms proper flow using gold-conjugated goat IgG and immobilized bovine anti-goat IgG. The assay is performed in a laboratory by trained personnel using test tubes or microtiter plates. Results are interpreted visually by the healthcare provider; a positive result shows two lines, while a negative result shows one. The test aids in the diagnosis of cryptococcosis, particularly in immunocompromised populations such as AIDS patients, organ transplant recipients, and those on steroid therapy. It provides a faster, simpler alternative to traditional latex agglutination methods by eliminating the need for serum pre-treatment (e.g., pronase).

Clinical Evidence

Performance evaluated via method comparison against gold standard (culture/India Ink) and EIA. Sensitivity 100% (96.0-100% CI) and specificity 100% (97.0-100% CI) observed against culture/India Ink. Comparison with EIA showed 100% positive agreement and 93% negative agreement. Analytical sensitivity (cut-off) is 1.25 ng/ml. Cross-reactivity testing performed on various fungal and non-fungal pathologies; interference testing performed on icteric, hemolyzed, and lipemic samples. High-dose hook effect identified at concentrations >200 ug/ml.

Technological Characteristics

Dipstick sandwich immunochromatographic assay. Components: lateral flow test strips with gold-conjugated anti-cryptococcal monoclonal antibodies, immobilized anti-cryptococcal monoclonal antibodies (test line), and immobilized bovine anti-goat IgG (control line). Specimen diluent included. Manual, non-instrumented. Storage at 20-25°C. No electronic components or software algorithms.

Indications for Use

Indicated for qualitative or semi-quantitative detection of Cryptococcus neoformans and Cryptococcus gattii capsular polysaccharide antigens in serum to aid in the diagnosis of cryptococcosis. For prescription-use in laboratory settings.

Regulatory Classification

Identification

Cryptococcus neoformans serological reagents are devices that consist of antigens used in serological tests to identify antibodies to Cryptococcus neoformans in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) and are used to identify Cryptococcus neoformans directly from clinical specimens or from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of cryptococcosis and provides epidemiological information on this type of disease. Cryptococcosis infections are found most often as chronic meningitis (inflammation of brain membranes) and, if not treated, are usually fatal.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K102286 B. Purpose for Submission: To obtain substantial equivalence for an original 510(k) for a device which detects Cryptococcal Antigen. C. Measurand: Capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) D. Type of Test: Qualitative and semi-quantitative dipstick sandwich lateral flow immunochromatographic assay E. Applicant: Immuno-Mycologics, Inc. F. Proprietary and Established Names: CrAg Lateral Flow Assay G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | GMD | II | 866.3165 | 83-Microbiology | H. Intended Use: 1. Intended use: The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) in serum. {1} 2. Indication for use: The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of *Cryptococcus* species complex (*Cryptococcus neoformans* and *Cryptococcus gattii*) in serum. The CrAg Lateral Flow Assay is a prescription use laboratory assay which can aid in the diagnosis of cryptococcosis 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Not applicable I. Device Description: The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic assay which detects cryptococcal antigen in serum. The assay consists of CrAg Lateral Flow test strips which have a gold-conjugated antibody and a gold-conjugated, anti-cryptococcal antibody deposited onto a sample membrane and anti-Crypto antibody and control-line capture antibody striped onto a membrane. Also in the kit is a specimen diluent. J. Substantial Equivalence Information: 1. Predicate Device names Latex-*Cryptococcus* Antigen Detection System 2. Predicate K number: K791382 {2} Comparison with predicate: Table 1: Comparison Between New Device and Predicate Device | SIMILARITIES | | | | --- | --- | --- | | Feature | Cryptococcal Antigen Lateral Flow Assay | Latex-Cryptococcus Antigen Detection System (K791382) | | Intended Use | | | | Intended Use | Test for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus in serum | Test for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus in serum and CSF | | Indication For Use | Prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis | Prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis | | Device Description | | | | Sample Matrix | Serum | Serum and CSF | | Instruments | None | None | | Detection Antibody | Anti-cryptococcal monoclonal antibody | Anti-cryptococcal antibody | | | | | | DIFFERENCES | | | | Feature | Cryptococcal Antigen Lateral Flow Assay | Latex-Cryptococcus Antigen Detection System (K791382) | | Intended Use | | | | Intended Use | No difference | No difference | | Indication For Use | No difference | No differences | | Device Description | | | | Assay Principle | Lateral flow assay | Latex agglutination | | Assay Components | Specimen diluent, lateral flow strips, built-in control, gold conjugated antibodies | Positive control, negative control, latex cards, latex conjugated anti-cryptococcal polyclonal antibodies | | Serum Pre-Treatment | Serum dilution | Pronase with pronase inhibitor | | CSF Pre-Treatment | none | Boil | | Detection Antibodies | Gold-conjugated anti-cryptococcal monoclonal antibodies | Latex-conjugated anti-cryptococcal polyclonal antibodies | | Storage Requirements | 20-25°C | 4°C ± 2° | | | | | # K. Standard/Guidance Document Referenced (if applicable): Not Applicable {3} L. Test Principle: The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic assay, which detects cryptococcal antigen in serum. For the qualitative procedure, specimens are diluted 1:2 in 1x Specimen Diluent and analyzed. For the semi-quantitative procedure, specimens are diluted 1:5 in 1x Specimen Diluent followed by 1:2 serial dilutions. All dilutions are then analyzed as in the qualitative procedure. Specimens are placed into an appropriate reservoir, such as a test tube or microtiter plate, and the lateral flow device is then placed into the reservoir allowing the specimen to come into contact with the test membrane. The test uses specimen wicking to capture gold-conjugated, anti-cryptococcal monoclonal antibodies and gold-conjugated control antibodies that are deposited onto a membrane. If cryptococcal antigen is present in the specimen, it binds to the gold-conjugated, anti-cryptococcal antibodies. The gold-labeled antibody-antigen complex will continue to wick up the membrane where it will interact with the Test Line (T). The Test Line is immobilized anti-cryptococcal monoclonal antibodies. If the specimen contains cryptococcal antigen, a sandwich is created with the gold-labeled antibodies and the immobilized antibodies, causing a visible line to develop at the test line site (T). If proper flow occurs and the reagents are reactive at the time of use, the wicking of any specimen, positive or negative, will cause the gold-conjugated control goat IgG antibody to move to the Control Line (C) which is immobilized bovine anti-goat IgG antibody. The immobilized anti-goat antibody will bind to the gold-conjugated goat IgG Control antibody and will cause a visible line to develop (C). A positive test result will create two lines, while a negative test result will create one line (Figure 1). If the control line fails to develop a line, then the test is not valid. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Serum: Repeatability and reproducibility with serum specimens were determined by spiking a serum specimen pool that was negative by the IMMY Latex-Cryptococcus Antigen Detection System with cryptococcal antigen at four concentrations: Negative, high negative (C₅), low positive (near C₉₅), and medium positive. The samples were analyzed on the CrAg Lateral Flow Assay in triplicate on five different days, at three different sites with a total of five different operators, on one lot, according to EP5-A2. One site was internal (Site 1) and two were a South African reference laboratory (Site 2) and a US hospital laboratory (Site 3). For repeatability, percent positive and percent negative detected were calculated for each site (Table 2). For reproducibility, overall percent positive and percent {4} negative detected were calculated by combining the data from all three sites (last two rows of Table 2). | SERUM PANEL | Site 1 % Pos | Site 2 % Pos | Site 3 % Pos | Overall % Pos | | --- | --- | --- | --- | --- | | Negative | 0% (0/30) | 0% (0/30) | 0% (0/15) | 0% (0/75) | | High Negative | 7% (2/30) | 0% (0/30) | 0% (0/15) | 3% (2/75) | | Low Positive | 100% (30/30) | 100% (30/30) | 100% (15/15) | 100% (75/75) | | Moderate Positive | 100% (30/30) | 100% (30/30) | 100% (15/15) | 100% (75/75) | As expected, the high negative tested positive nearly 5% of the time and 3% of the time in serum. All other samples performed 100% as expected across all sites, all runs, and all operators. ## b. Linearity/assay reportable range: Not Applicable ## c. Traceability, Stability, Expected values (controls, calibrators, or methods): The CrAg Lateral Flow Assay has a three-month expiration date when stored at room temperature. Real-time stability testing is on-going. Initial accelerated stability was determined by testing three lots of kits stored at 25°C and 37°C for three months. Buffer was spiked with cryptococcal antigen at four concentrations: 1.5, 3.0, 6.0 and 9.0ng/ml. Non-spiked buffer was used as a negative control. Each sample was tested in triplicate by one operator. Each spiked sample replicate tested positive and each negative control tested negative on all lots of the CrAg Lateral Flow Assay, stored at both 25°C and 37°C. This initial data indicate that the CrAg Lateral Flow Assay is stable at room temperature for three months. Stability studies are on-going to demonstrate a one-year expiration dating. ## d. Detection limit: ### Detection Limit/Analytical Cut-off Analytical sensitivity for the CrAg Lateral Flow Assay was estimated by running varying concentrations of cryptococcal antigen diluted in LF Specimen Diluent. Test results determined that the analytical cut-off is 1.25ng/ml (Table 3). {5} Table 3. Analytical Cut-Off | Sample Concentration (ng/ml) | No. Positive | No. Tested | % Positive | | --- | --- | --- | --- | | 0.50 | 0 | 24 | 0% | | 0.75 | 0 | 24 | 0% | | 1.00 | 4 | 24 | 17% | | 1.25 | 12 | 24 | 50% | | 1.50 | 21 | 24 | 88% | | 1.75 | 24 | 24 | 100% | | 2.00 | 24 | 24 | 100% | | 2.50 | 24 | 24 | 100% | | 3.00 | 24 | 24 | 100% | | 3.50 | 24 | 24 | 100% | | 4.00 | 24 | 24 | 100% | Analytical specificity: Table 4. Analytical Specificity | Pathology | No. of Specimens | No. of Replicates Per Specimen | Total Tests | Total Number of Positives | % Positive (No Pos/Total) | | --- | --- | --- | --- | --- | --- | | Non-Fungal Pathologies | | | | | | | HAMA (Human anti-Mouse antibody) Positive | 5 | 3 | 15 | 0 | 0% | | Syphilis | 10 | 3 | 30 | 0 | 0% | | Rubella | 5 | 3 | 15 | 0 | 0% | | Mycoplasma | 10 | 3 | 30 | 0 | 0% | | Toxoplasmosis | 7 | 3 | 21 | 0 | 0% | | CMV Infection | 10 | 3 | 30 | 0 | 0% | | Rheumatoid factor* | 10 | 3 | 30 | 0 | 0% | | Total | 57 | 3 | 171 | 0 | 0% | | Fungal Pathologies | | | | | | | Penicilliosis | 5 | 3 | 15 | 0 | 0% | | Sporotrichosis | 6 | 3 | 18 | 0 | 0% | {6} | Blastomycosis | 10 | 3 | 30 | 0 | 0% | | --- | --- | --- | --- | --- | --- | | Coccidioidomycosis | 10 | 3 | 30 | 0 | 0% | | Histoplasmosis | 10 | 3 | 30 | 0 | 0% | | Candidiasis | 10 | 3 | 30 | 0 | 0% | | Aspergillosis ** | 10 | 3 | 30 | 3 | 10% | | Aspergillus terreus Culture Filtrate | 3 | 3 | 9 | 0 | 0% | | Aspergillus niger Culture Filtrate | 3 | 3 | 9 | 0 | 0% | | Aspergillus flavus Culture Filtrate | 3 | 3 | 9 | 0 | 0% | | Aspergillus fumigatus Culture Filtrate | 3 | 3 | 9 | 0 | 0% | | Paracoccidioides brasiliensis Culture Filtrate | 3 | 3 | 9 | 6 | 67% | | Total | 76 | 3 | 228 | 9 | 3.9% | * Rheumatoid factor concentrations tested ranged from 112IU/ml to 6479IU/mls. ** The three positives were the result of three replicates of the same specimen, which was positive. In addition to the cross-reactivity study, interference testing was also performed on five icteric, five hemolyzed, and five lipemic serum specimens. Each specimen was spiked with cryptococcal antigen at three times the C95 concentration. All specimens were then tested at IMMY, on one lot of CrAg Lateral Flow assay in triplicate: spiked and unspiked. Percent positivity was determined for each condition. All of the unspiked specimens had negative results on the CrAg Lateral Flow Assay. All spiked specimens were positive, thus, these types of serum specimens do not interfere with the CrAg Lateral Flow Assay. However, it is possible that hemolyzed samples could lead to false negatives due to the high background color on the strip. As such, a statement is included in the Limitations of the Procedure section of the package insert: "Hemolyzed serum samples could lead to false negative results due to the high background color on the strip." Due to specimen availability, the following conditions were not tested in the CrAg Lateral Flow Assay: Candida dubliniensis, Candida tropicalis, Candida parapsilosis, Candida krusei, Candida glabrata, Cladosporium trichoides, Neisseria meningitidis, Salmonella typhi, Pneumocystis carinii, Trichosporon beigelii, {7} Zygomycetes, ANA+, HAV, HCV, Staph, and Strep. # f. Assay cut-off: High dose hook effect concentrations with serum specimens were determined by spiking negative serum that were negative by the IMMY Latex-Cryptococcus Antigen Detection System and CrAg Lateral Flow Assay, with cryptococcal antigen at various concentrations between 20 and $500\mathrm{ug / ml}$ . Each concentration was tested in triplicate at IMMY on one lot of CrAg Lateral Flow Assay, according to the package insert. It was determined that serum specimens with a cryptococcal antigen concentration higher than $200\mathrm{ug / ml}$ can produce a high dose hook effect and therefore may produce a false negative result. # 2. Comparison studies: # a. Method comparison with predicate device: The CrAg Lateral Flow Assay was compared to the gold standard diagnoses of cryptococcosis (culture and/or India Ink) to evaluate the sensitivity and specificity of the assay. These studies contained a mix of both prospective and retrospective specimens. A summary table of the data collected is included below. | Serum | | Culture/India Ink | | | --- | --- | --- | --- | | | | Positive | Negative | | CrAg LFA Assay | Positive | 91 | 0 | | | Negative | 0 | 123 | | Serum | Calculated | 95% CI | | --- | --- | --- | | Sensitivity | 100% | 96.0% - 100% | | Specificity | 100% | 97.0% - 100% | # b. Matrix comparison: # Serum Specimens A panel of 197 serum specimens that were submitted to a US reference laboratory for cryptococcal antigen testing, using a Cryptococcal Antigen EIA, were collected and stored frozen until method comparison studies were performed. Of the 197 specimens, 96 were positive and 101 were negative by the EIA. All specimens were analyzed, according to EP12-A2, concurrently in the CrAg Lateral Flow Assay and in the IMMY Latex-Cryptococcus Antigen Detection System to ensure the data was not affected by the freeze- {8} thaw cycle. Each specimen was tested at IMMY in duplicate in both tests according to each test's package insert. The data is presented in a 2x2 contingency table (Table 8). The percent agreement positive, percent agreement negative, and 95% confidence interval are also presented (Table 9). Table 8. Serum 2x2 Contingency Table | | EIA (+) | EIA (-) | | --- | --- | --- | | CrAg Lat Flow (+) | 96 | 7 | | CrAg Lat Flow (-) | 0 | 94 | Table 9. Serum Statistical Analysis | | | 95% CI | | --- | --- | --- | | % Agreement Positive | 100% | 96-100% | | % Agreement Negative | 93% | 86-97% | Semi-Quantitative Method Comparison In addition, 62 of these specimens were tested using the semi-quantitative titration procedure in both the CrAg Lateral Flow Assay and the IMMY Late Cryptococcal Antigen Detection System (REF CR1003). Linear regression analysis of the data yielded an R² value of 0.890 3. Clinical studies: Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Not Applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...