Disposable Sampler Inactivated Transport Media, Nest ITM
Applicant
Wuxi Nest Biotechnology Co., Ltd.
Product Code
QBD · Microbiology
Decision Date
Sep 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2950
Device Class
Class 2
Indications for Use
NEST ITM is an enclosed system intended for the collection, inactivation, stabilization and transportation of pharyngeal and nasal swabs suspected of containing adenovirus, influenza A virus or parainfluenza virus 2 from the collection site to the testing laboratory. The specimen transported in NEST ITM can be used for molecular detection in the laboratory.
Device Story
NEST ITM is an enclosed system for collecting, inactivating, and stabilizing viral specimens from pharyngeal and nasal swabs. The device consists of a polypropylene tube containing a chemical medium (guanidine isothiocyanate, TCEP, sodium acetate, PEG-8000, Tris HCl). Upon collection, the medium lyses cells, disrupts lipid membranes, denatures proteins, inactivates enzymes, and stabilizes viral RNA. The device is used at the point of collection (e.g., clinic) by healthcare providers and transported to a laboratory. The inactivation process provides a >4.0 log reduction in viral concentration within 10 seconds, reducing biohazard risk for laboratory personnel. Specimens remain stable for up to 15 days at 2-8°C or 25°C. Laboratory staff use the stabilized specimen for downstream molecular detection (e.g., PCR). The device benefits patients and staff by ensuring sample integrity during transport while mitigating infection risk.
Clinical Evidence
No clinical data. Bench testing only. Analytical performance included Limit of Detection (LoD) studies (5.0 x 10^2 copies/mL for all three viruses), viral stability studies over 15 days at 4°C and 25°C (maximum variation 0.9 Ct), and viral inactivation studies (>4.0 log reduction at 10 seconds).
Technological Characteristics
Polypropylene tube (5 mL/10 mL); flocked nylon fiber swab with ABS rod. Media components: Guanidine isothiocyanate, TCEP, sodium acetate, PEG-6000, Tris, HCl, purified water. Shelf life: 12 months. Single-use. Non-sterile tube; sterile swab. Complies with ASTM F1980-2016, ASTM D4169-2016, ISO 11607-1/2, ISO 15223-1, ISO 14971, and CLSI MM13-A.
Indications for Use
Indicated for collection, inactivation, stabilization, and transport of pharyngeal and nasal swabs suspected of containing adenovirus, influenza A virus, or parainfluenza virus 2 for molecular detection in a laboratory setting.
Regulatory Classification
Identification
A microbial nucleic acid storage and stabilization device is a device that consists of a container and reagents intended to stabilize microbial nucleic acids in human specimens for subsequent isolation and purification of nucleic acids for further molecular testing. The device is not intended for preserving morphology or viability of microorganisms.
Special Controls
The special controls for this device are: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of microorganisms and types of human specimens intended to be preserved. (2) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including all device components. (ii) Performance characteristics from applicable analytical studies, including but not limited to, nucleic acid stability and microorganism inactivation. (iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing. (iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies. (3) Design verification and validation must include the following: (i) Overall device design including all device components and all control elements incorporated into the analytical validation procedures. (ii) Thorough description of the microorganisms and methodology used in the validation of the device including, but not limited to, extraction platforms and assays used for the detection of preserved nucleic acids. (iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of microorganisms and types of human specimens intended to be preserved.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A detailed device description, including all device components;
(ii) Performance characteristics from applicable analytical studies, including nucleic acid stability and microorganism inactivation;
(iii) A limiting statement that erroneous results may occur when the transport device is not compatible with molecular testing; and
(iv) A limiting statement that the device has only been validated to preserve the representative microorganisms used in the analytical studies.
(3) Design verification and validation must include the following:
(i) Overall device design, including all device components and all control elements incorporated into the analytical validation procedures;
(ii) Thorough description of the microorganisms and methodology used in the validation of the device including, extraction platforms and assays used for the detection of preserved nucleic acids; and
(iii) The limit of detection (LoD) of the molecular test used to establish microorganism nucleic acid stability.
K211675 — Dasky Disposable Sampling Tube (Model name: VSM02) · Ningbo Dasky Life Science Co., Ltd. · Aug 8, 2022
K233324 — Molecular Transport Media - MTM · Alb Luz · Jun 17, 2024
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K210440
B Applicant
Wuxi Nest Biotechnology Co., Ltd.
C Proprietary and Established Names
Disposable Sampler Inactivated Transport Media, Nest ITM
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QBD | Class II | 21 CFR 866.2950 - Microbial Nucleic Acid Storage And Stabilization Device | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To make a substantial equivalence determination for the NEST ITM for the collection, transport and storage of viral specimens to the laboratory for downstream testing.
B Measurand:
Storage and stability of nucleic acids from adenovirus, influenza A virus or parainfluenza virus 2
C Type of Test:
Microbial nucleic acid storage and stabilization device
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K210440 - Page 2 of 9
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
NEST ITM is an enclosed system intended for the collection, inactivation, stabilization and transportation of pharyngeal and nasal swabs suspected of containing adenovirus, influenza A virus or parainfluenza virus 2 from the collection site to the testing laboratory. The specimen transported in NEST ITM can be used for molecular detection in the laboratory.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
### D Special Instrument Requirements:
None
## IV Device/System Characteristics:
### A Device Description:
NEST ITM is an enclosed system intended for the collection, inactivation stabilization and transportation of swabs suspected of containing adenovirus, influenza A virus or parainfluenza virus 2 from the collection site to the testing laboratory. The specimen transported in NEST ITM can be used for molecular detection in the laboratory.
The preservation media tubes are made of polypropylene and are filled with Inactivated Transport Media. The tubes are packaged with or without sterile swabs. The specifications for the tubes and medium are listed as follows:
- A plastic screw-cap tubes filled with 1.0, 1.5, 2.0, 3.0 or 6.0 mL of medium
- A plastic screw-cap tube filled with a range of media and package with a nasopharyngeal swab
- A plastic screw-cap tube filled with a range of media and package with an oropharyngeal swab
### B Principle of Operation:
The device components are intended to inactivate influenza A virus, parainfluenza virus 2, lyse cells, disrupt/lyse lipid membranes, denatures proteins, inactivates enzymes, and stabilize viral RNA. The transport device is designed for storage of specimens between 2-8°C and 25 °C (for up to 15 days).
The media contains the following reagents:
- Guanidine Isothiocyanate
- TCEP
- Sodium Acetate
- PEG-8000
- Tris HCl
- purified water
{2}
V Substantial Equivalence Information:
A Predicate Device Name(s): PrimeStore MTM
B Predicate 510(k) Number(s): DEN170029
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K210440 | DEN170029 |
| --- | --- | --- |
| Device Trade Name | Disposable sampler inactivated transport media, Nest ITM | PrimeStore MTM |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | NEST ITM is an enclosed system intended for the collection, inactivation, stabilization and transportation of pharyngeal and nasal swabs suspected of containing adenovirus, influenza A virus or parainfluenza virus 2 from the collection site to the testing laboratory. The specimen transported in NEST ITM can be used for molecular detection in the laboratory. | PrimeStore MTM is intended for the stabilization, transportation and inactivation of infectious unprocessed nasal washes suspected of containing Influenza A virus RNA. PrimeStore MTM is also intended for the stabilization, transportation and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis DNA from human samples. |
| Inactivation tested | >4.0 log reduction in concentration at 10 seconds | Same |
| Storage temperatures | 2-8 °C up to 25°C | Same |
| General Device Characteristic Differences | | |
| Specimen stability | NEST ITM preserves Adenovirus, Influenza | Primestore MTM medium preserves influenza A |
K210440 - Page 3 of 9
{3}
| | A virus or Parainfluenza virus 2 for 15 days at 2-8 °C and 25°C | RNA for up to 8 days at 27°C and 29 days at 4°C |
| --- | --- | --- |
| Specimen Type | Nasopharyngeal or Oropharyngeal Swab | Nasal washes and sputum samples |
| Analyte | Nasopharyngeal or Oropharyngeal swab suspected of containing Adenovirus, Influenza A virus or Parainfluenza virus 2. | Nasal wash suspected of containing Influenza A virus. Sputum samples suspected of containing MTB. |
VI Standards/Guidance Documents Referenced:
Special controls that are applicable to regulation 21 CFR 866.2950.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
N/A
2. Linearity:
N/A
3. Analytical Specificity/Interference:
N/A
4. Assay Reportable Range:
N/A
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Shelf life: The shelf life for the NEST ITM is 12 months after the date of manufacture. The stability of the NEST ITM was performed using Realtime and Accelerated stability on a total of three (3) lots. Stability looked for bacterial and fungal growth in the media along with properties of the media, appearance, pH, and then confirmed with viral stabilization at room
K210440 - Page 4 of 9
{4}
temperature to the claimed 15 days demonstrating the stability of nucleic acids was not diminished with the age of media.
Sterilization: The DNA/RNA Shield Collection tube with media are not sold as sterile nor are they intended to be sterilized by the user. These vials are single use devices that do not require cleaning by the operator. The Swabs are individually packages and sold as sterile.
# 6. Detection Limit:
a) LoD testing was conducted to determine the lowest concentration of analyte that can be detected with a greater than $95\%$ detection rate. The LoD studies for Adenovirus, Influenza A, and Parainfluenza virus 2 were designed using validated assays to establish a concentration of organisms used for additional testing noted below.
LoD testing was initially performed by spiking multiple concentrations of Adenovirus, Influenza A, and Parainfluenza virus 2 into contrived matrix and spiking it onto a swab. Adenovirus, Influenza A, and Parainfluenza virus 2 were spike at a final concertation range of $1.0 \times 10^{4}$ , $1.0 \times 10^{3}$ , $5.0 \times 10^{2}$ , and $1.0 \times 10^{2}$ copies/mL into the NEST ITM with a swab. A validated PCR assay was used to determine the LoD to be $5.0 \times 10^{2}$ copies/mL for each of the three viruses. Table 1 below shows the results of preliminary LoD for Adenovirus, influenza A, and Parainfluenza virus 2.
Table 1. Preliminary Limit of Detection
| Concentration (copies/mL) | Adenovirus, 5 Reps Average (Ct) | SD (Ct) | Influenza A, 5 Rep Average (Ct) | SD (Ct) | Parainfluenza virus 2, 5 Rep Average (Ct) | SD (Ct) |
| --- | --- | --- | --- | --- | --- | --- |
| 1.0x104 | 29.3 | 0.18 | 32.04 | 1.08% | 31.07 | 1.08% |
| 1.0x103 | 32.0 | 0.26 | 34.75 | 2.11% | 34.28 | 2.18% |
| 5.0x102 | 34.3 | 0.95 | 35.48 | 1.97% | 36.89 | 2.98% |
| 1.0x102 | >40 | - | >40 | - | >40 | - |
Confirmatory LoD testing was provided at a concentration of $5.0 \times 10^{2}$ copies/mL with 20 replicates. The validated assay had an LoD with an acceptance criteria of virus detection at a concentration range $5.0 \times 10^{2}$ copies/mL. The same detection range was replicated with the NEST ITM and further determine by the concentration that yielded at least a $95\%$ of the replicates were recoverable within this range. At a concentration of $5.0 \times 10^{2}$ copies/mL, 20 of 20 replicates had recoverable concentrations. Viral nucleic acids were extracted using a nucleic acid (DNA/RNA) extraction and purification kits (spin column) (SC903-50) (Wuxi TechstarTechnology Co., Ltd.) and amplified using the respective kits on the ABI 7500. The average $C_{t}$ values for each virus are listed below in Table 2.
K210440 - Page 5 of 9
{5}
Table 2. Limit of detection ${\mathrm{C}}_{\mathrm{t}}$ values for each virus
| | Ct Value | | |
| --- | --- | --- | --- |
| Replicates | Adenovirus | Influenza A | Parainfluenza virus 2 |
| 1 | 36.2 | 32.0 | 38.0 |
| 2 | 37.0 | 32.4 | 38.2 |
| 3 | 38.1 | 35.4 | 37.8 |
| 4 | 38.1 | 35.2 | 36.7 |
| 5 | 36.8 | 35.4 | 36.4 |
| 6 | 36.9 | 37.0 | 38.4 |
| 7 | 35.5 | 35.8 | 37.3 |
| 8 | 36.1 | 35.0 | 35.2 |
| 9 | 37.2 | 35.4 | 35.9 |
| 10 | 35.8 | 35.7 | 36.5 |
| 11 | 37.7 | 35.8 | 36.2 |
| 12 | 36.3 | 35.8 | 36.2 |
| 13 | 36.4 | 35.3 | 34.2 |
| 14 | 37.5 | 36.2 | 35.3 |
| 15 | 36.7 | 34.1 | 36.1 |
| 16 | 35.7 | 35.6 | 35.1 |
| 17 | 37.0 | 35.9 | 36.6 |
| 18 | 37.0 | 35.7 | 35.8 |
| 19 | 36.8 | 36.2 | 37.5 |
| 20 | 37.1 | 35.0 | 36.3 |
| AVG: | 36.8 | 35.2 | 36.3 |
| SD: | 0.73 | 1.2 | 1.1 |
LoD testing at $5.0 \times 10^{2}$ copies/mL resulted in all 20 replicates for the concentration meeting the pre-defined acceptance criteria.
# b) Viral Stability
The stability of Adenovirus, Influenza A, and Parainfluenza virus 2 at $1 \times$ LoD ( $5.0 \times 10^{2}$ copies/mL) was evaluated by spiking virus into simulated matrix incubated in the NEST ITM at refridgerated temperature ( $4^{\circ} \mathrm{C}, 39^{\circ} \mathrm{F}$ ) for 15 days (see Table 3), and ambient temperature ( $25^{\circ} \mathrm{C}, 77^{\circ} \mathrm{F}$ ) for 15 days (see Table 4). Validated PCR assays was used to determine stability of Adenovirus, Influenza A, and Parainfluenza virus 2 in the NEST ITM. The stability study analyzed a total of three lots near the manufacturer claimed 12-month stability. Testing used at least 20 replicates for each virus, time point and storage condition. An initial time point designated as Day 0 was included as the initial $C_{t}$ average for each of the two temperature ranges tested. Testing at three time points was performed at Day 0, 9 and 15 for refrigerated temperature ( $2-8^{\circ} \mathrm{C}, 36-39^{\circ} \mathrm{F}$ ), and three time points, Day 0, 9, and 15, for ambient temperature ( $25^{\circ} \mathrm{C}, 77^{\circ} \mathrm{F}$ ).
The validated PCR assays were run with all applicable controls to valid and confirm the detection of the target virus, Adenovirus, Influenza A, and Parainfluenza virus 2. A predefined acceptance criteria of $(+/-)$ $3.0\mathrm{C_t}$ from the initial time zero value was the acceptance criteria.
K210440 - Page 6 of 9
{6}
Table 3. Adenovirus, Influenza A, and Parainfluenza 2 virus (5.0x10² copies/mL) stability at 4°C
| | Day 0 | Day 9 | Day 15 |
| --- | --- | --- | --- |
| Adenovirus AVG (Ct): | 36.0 | 36.7 | 36.9 |
| CV (Ct): | 2.15% | 1.8% | 2.0% |
| Influenza A AVG (Ct): | 35.4 | 35.4 | 35.7 |
| CV (Ct): | 2.0% | 2.3% | 2.5% |
| Parainfluenza 2 AVG (Ct): | 36.1 | 37.0 | 37.0 |
| CV (Ct): | 2.0% | 2.0% | 2.4% |
Table 4. Adenovirus, Influenza A, and Parainfluenza 2 virus (5.0x10² copies/mL) stability 25°C
| | Day 0 | Day 9 | Day 15 |
| --- | --- | --- | --- |
| Adenovirus AVG (Ct): | 36.1 | 37.0 | 36.9 |
| CV (Ct): | 2.0% | 2.4% | 2.0% |
| Influenza A AVG (Ct): | 35.3 | 36.1 | 35.3 |
| CV (Ct): | 2.4% | 2.6% | 2.1% |
| Parainfluenza 2 AVG (Ct): | 36.2 | 36.9 | 37.0 |
| CV (Ct): | 2.3% | 2.6% | 2.1% |
Stability testing of RNA from whole Adenovirus, Influenza A, and Parainfluenza virus 2 were spiked into matrix and stored in NEST ITM, resulted in a maximum average variation of 0.8 Cₜ over 15 days at 25°C and a maximum variation of 0.9 Cₜ over 15 days at 2-8°C.
c) Inactivation
Adenovirus, Influenza A, and Parainfluenza virus 2 at a concentration of 1.0 x 10⁷ TCID₅₀/ml was incubated with NEST ITM for 10 seconds. Each virus only and NEST ITM were also incubated accordingly to serve as controls. Adenovirus, Influenza A, and Parainfluenza 2 virus (virus alone), Virus and NEST ITM, or NEST ITM alone was then incubated on to cell cultures after incubation. Four days after inoculation, the cells were fixed and stained with 0.06% crystal violet in 1% glutaraldehyde. Cells that did not take up the stain were considered evidence of a viral cytopathic effect (CPE) and as a result were considered a measure of viral viability. The titer of the virus CPE was calculated and recorded as the TCID₅₀.
Inactivation time:
The NEST ITM showed no cytotoxicity on MDCK cells at a 1:1,000 dilution factor and greater; therefore at least a 1:1,000 dilution factor is needed to avoid a direct cytotoxic effect of the NEST ITM. Adenovirus, Influenza A, and Parainfluenza virus 2, were then exposed to NEST ITM for 10 seconds prior to serial 10 fold dilutions and incubations
K210440 - Page 7 of 9
{7}
(final concentration $\leq 10^{3}\mathrm{TCID}_{50} / \mathrm{mL}$ ), while Influenza Adenovirus, Influenza A, and Parainfluenza virus 2 samples had viral loads of greater than $1.0 \times 10^{6} \mathrm{TCID}_{50}$ of virus and NEST ITM alone was diluted 1:1000 to see no CPE. The NEST ITM rapidly inactivated all viruses tested with a $>4.0$ log reduction at a 1:10 specimen to media concentration at 10 seconds. Viral CPE could not be observed at $< 3.0$ logs due to cellular destruction by NEST ITM See Table 5 below.
Table 5 Adenovirus, Influenza A, and Parainfluenza virus 2 inactivation in NEST ITM
| 10s incubation | Adenovirus TCID50(log) | Influenza A TCID50(log) | Influenza A TCID50(log) |
| --- | --- | --- | --- |
| Virus only | 6.14 | 6.75 | 6.63 |
| Virus and NEST ITM | < 3.0 | < 3.0 | < 3.0 |
| NEST ITM only* | ≤ 3.0 | ≤ 3.0 | ≤ 3.0 |
* NEST ITM shows cytotoxicity on MDCK cells when diluted to 1:1,000.
NEST ITM must be used at a ratio of at least 1:10 at a minimum of 10 seconds exposure time to demonstrate inactivation of Adenovirus, Influenza A, and Parainfluenza virus 2. Measuring Adenovirus, Influenza A, and Parainfluenza virus 2 inactivation below $1 \times 10^{3}$ $\mathrm{TCID}_{50}$ was not possible because of the cytotoxic affects NEST ITM has on the cell culture based assay.
7. Assay Cut-Off:
N/A
B Comparison Studies:
1. Method Comparison with Predicate Device:
N/A
2. Matrix Comparison:
N/A
C Clinical Studies:
1. Clinical Sensitivity:
N/A
2. Clinical Specificity:
N/A
K210440 - Page 8 of 9
{8}
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
N/A
D Clinical Cut-Off:
N/A
E Expected Values/Reference Range:
N/A
F Other Supportive Instrument Performance Characteristics Data:
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K210440 - Page 9 of 9
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.