BD BACTEC Plus Aerobic/F Culture Vials (plastic)

K113558 · Becton, Dickinson and Company · MDB · Mar 16, 2012 · Microbiology

Device Facts

Record IDK113558
Device NameBD BACTEC Plus Aerobic/F Culture Vials (plastic)
ApplicantBecton, Dickinson and Company
Product CodeMDB · Microbiology
Decision DateMar 16, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2560
Device ClassClass 1

Indications for Use

The BD BACTEC Plus Aerobic/F medium is used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTEC fluorescent series instruments.

Device Story

Device is a plastic vial containing enriched soybean casein digest broth and resins for antimicrobial adsorption; used for aerobic blood culture. Sample (3-10 mL blood) is inoculated into vial; vial is inserted into BD BACTEC fluorescent-series instrument. Instrument incubates (35°C) and monitors vial every 10 minutes via chemical sensor detecting CO2-induced fluorescence. Increased fluorescence indicates microbial growth. Output is a positive/negative reading for presence of viable microorganisms. Used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing bloodstream infections; enables targeted antimicrobial therapy.

Clinical Evidence

Bench testing only. No clinical data. Analytical studies compared plastic vials to glass predicate using 738 paired sets across three BACTEC instruments (FX, 9240, 9050). 98.4% of sets were positive in both devices. Microbial detection limit study (270 sets) showed comparable recovery. False positive/negative rates were low and comparable to predicate. Antimicrobial neutralization capability confirmed for multiple drug classes. No statistically significant differences in recovery or time-to-detection for most organisms.

Technological Characteristics

Plastic vial containing 30 mL enriched soybean casein digest broth; includes resins for antimicrobial adsorption. Sensing principle: CO2-sensitive fluorescent sensor. Energy source: BACTEC fluorescent-series instrument (incubation/agitation/monitoring). Dimensions: 5.0 inches high. Connectivity: Standalone vial used within networked/integrated BACTEC instrument systems. Sterilization: Not specified.

Indications for Use

Indicated for aerobic culture and recovery of bacteria and yeast from human blood samples. Used with BD BACTEC fluorescent series instruments (BACTEC FX, 9240, 9050). Prescription use only.

Regulatory Classification

Identification

A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k113558 B. Purpose for Submission: To obtain a substantial equivalent determination for a premarket notification for the BD BACTEC Plus Aerobic/F blood culture medium in plastic bottles. C. Measurand: Aerobic bacteria and yeast D. Type of Test: Liquid culture medium for recovery of microorganisms (bacteria and yeast) from blood using fluorescent instruments to detect increased CO₂ E. Applicant: Becton Dickinson and Company F. Proprietary and Established Names: BD BACTEC Plus Aerobic /F (plastic) G. Regulatory Information: 1. Regulation section: 21 CFR 866.2560 2. Classification: Class I 3. Product code: MDB {1} 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use: The BD BACTEC Plus Aerobic/F medium is used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTEC fluorescent series instruments. 2. Indications for use: The BD BACTEC Plus Aerobic/F medium is used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTEC fluorescent series instruments. 3. Special conditions for use statement(s): Prescription use 4. Special instrument requirements: BD BACTEC fluorescent series instrument (BACTEC FX, 9240, 9050) I. Device Description: The sample to be tested is inoculated into one or more vials which are inserted into the BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in $\mathrm{CO}_{2}$ produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of $\mathrm{CO}_{2}$ present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of medium. Resins have been described for the treatment of blood specimens both prior to and after their inoculation into culture media. Resins have been incorporated into BACTEC culture media to enhance recovery of organisms without a need for special processing. {2} J. Substantial Equivalence Information: 1. Predicate device name(s): BD BACTEC Plus Aerobic /F medium 2. Predicate K number(s): k921133 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood, with the BD BACTEC fluorescent series instrument | Same | | Specimen type | Human blood | Same | | Instrumentation | BD BACTEC fluorescent series | Same | | Detection Technology | Continuous monitoring; incorporate chemical sensor for detection of CO2 increases produced by the growth of aerobic bacteria and yeast | Same | | Incubation | 35°C (±1.5°C) up to 120 hrs | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Bottle | multilayer polycarbonate/nylon/polycarbonate plastic | Glass | | Ingredients | Sugar concentration 0.2% | Sugar concentration 0% | | Bottle weight | 20.9g | 113g | | Bottle height | 5.0 inches | 5.6 inches | | Sensor | 2.6g specific for the plastic bottle geometry | 1.75g | | Indicator | Increase concentrations of indicator and dye for the sensor changes | Different indicator and dye ratios | | Adhesive | Inert adhesion promoter for the adhesion of senor to the polycarbonate surface of the | Adhesion promoter not required | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | plastic bottles | | ## K. Standard/Guidance Document Referenced (if applicable): Not applicable ## L. Test Principle: If microorganisms are present in the test sample inoculated into the BACTEC vial, $\mathrm{CO}_{2}$ will be produced when the organisms metabolize the substrates present in the vial. Increases in the fluorescence of the vial sensor caused by the higher amount of $\mathrm{CO}_{2}$ are monitored by the BACTEC fluorescent series instrument. Analysis of the rate and amount of $\mathrm{CO}_{2}$ increase enables the BACTEC fluorescent series instrument to determine if the vial is positive, i.e., that the test sample contains viable organisms. ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The reproducibility study was evaluated by time to detection and recovery, in the Instrument Time to Detection and Recovery studies by using three lots. Both Lots 1 and 3 exhibited no statistically significant difference in time to detection between the new and predicate devices. Lot 2 exhibited a statistically significant difference in time to recovery due to the performance of BACTEC 9050 with Leuconostoc spp. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The range of time-to-detection in hours was $\leq 72$ hours for each of the organisms listed below. The QC study was performed on three lots and the QC specifications were met. ### Aerobic Medium Organisms Neisseria meningitidis Candida glabrata ATCC 13090 ATCC 66032 Haemophilus influenzae Staphylococcus aureus ATCC 19418 ATCC 25923 {4} Streptococcus pneumoniae* Escherichia coli ATCC 6305 ATCC 25922 Streptococcus pyogenes Alcaligenes faecalis ATCC 19615 ATCC 8750 Pseudomonas aeruginosa ATCC 27853 *CLSI recommended strain d. Detection limit: ## Microbial Detection Limit study The test includes 15 strains tested at two blood volumes, each with three low inoculum levels (0 to 1 and 1 to 10 CFU per bottle) over three lots for a total of 270 paired sets: 15 strains x 2 blood volumes x 3 inoculum levels x 3 lots = 270 Microbial Detection Limit Comparison summary | Condition | Number of Bottles | | --- | --- | | Growth and detection in both the new and predicate device | 200 per device | | Growth and detection in the predicate device (glass) only | 26 (9.6%) | | Growth and detection in the new device (plastic) only | 20 (7.4%) | | No growth and detection in the both new and predicate devices | 24 (8.9%) | There were 26 cultures positive only in the glass device, 20 of the negative cultures in the new plastic device were from the low inoculum (0 to 1 CFU per bottle) level. There were six failures at the 1 to 10 CFU per bottle inoculum level: Micrococcus luteus, Pseudomonas aeruginosa, Staphylococcus epidermidis, Streptococcus pneumoniae and 2 Candida glabrata. There were four false negative cultures that contributed to the overall difference. Two false negatives were with Leuconostoc species in the predicate device. Two false negatives were in the new device, one with Leuconostoc species and one with Staphylococcus epidermidis. The Staphylococcus epidermidis was a low inoculum and the inoculum plate count was culture negative. {5} The study demonstrated that the inoculum was 10- 100 CFU per bottle for the new BD BACTEC Plus Aerobic /F Plastic vial. e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: Performance of the BD BACTEC Plus Aerobic/F medium in plastic bottles was evaluated during internal analytical studies to demonstrate comparable performance to the predicate device – the BD BACTEC Plus Aerobic/F medium in glass bottles. Instrument Time to Detection (TTD) study The study included blood volumes, organism and BACTEC fluorescent-series blood culture instrument (BACTEC FX, 9240, 9050). Each organism was inoculated at 10 to 100 CFU per bottle, across two blood volumes (3 mL and 10 mL) over three lots of media for paired sets of the new and predicate devices. This testing was repeated in each BACTEC fluorescent-series blood culture instrument (BACTEC FX, 9240, 9050). There were 726 (98.4%) pair sets positive in both the new and predicate devices. 3 lots x 2 blood volumes x 3 instruments x 41 isolates = 738 pair sets Organisms with median TTD greater than one hour were summarized in the table below: | Organism | Median TTD Difference (hours) | | --- | --- | | Candida glabrata | -2.83 Earlier in new device (plastic) | | Cryptococcus neoformans | -1.67 Earlier in new device (plastic) | | Haemophilus parainfluenzae biotype I | 1.33 Earlier in predicate device (glass) | | Micrococcus luteus | 2.83 Earlier in predicate device (glass) | | Leuconostoc spp. | 8.00 Earlier in predicate device (glass) | {6} # Percent Recovery (Sensitivity) study A total of 738 paired sets were evaluated in the Percent Recovery comparison. There were nine paired sets failed to recover organisms in both the new and predicate device (1.6%) in the percent recovery study. One paired set recovered in the predicate device only (the paired new device was negative) and two paired sets recovered in the new device only (the predicate device was negative). | Condition | Number of Bottles | | --- | --- | | Detected in both the new and predicate devices | 726 per device | | Detected in the predicate device only | 1 | | Detected in the new device only | 2 | | Not detected in either device | 9 per device | The recovery failure included S. pneumoniae, Eikenella corrodens, and Leuconostoc spp. The observed failures of S. pneumoniae and Eikenella corrodens were replicates in a single instrument type. An additional study was conducted using the failed recovery strains; there were no recovery failure. # False Positive Rate Study False positivity was assessed with bottles inoculated with freshly drawn human blood (i.e. 2, 4, 6, and 8 mL). No organisms were added to the bottle. There were 240 paired sets comprised of 80 bottles from each of the three lots. There were two false positive plastic bottles observed but they were inoculated outside of the recommended usage range for the device (<3 mL). # False Negative Rate Study The data for this study was generated from the Instrument Time to Detection, Percent Recovery (Sensitivity), and Microbial Detection Limit studies. Bottles from that study that are expected to be positive (i.e., those inoculated with viable organisms) will form the dataset for the False Negative Rate. A total of 82 paired sets were evaluated to determine the false negatives, with the following results: Detected in predicate device only 27 Detected in new device only 22 Detected in neither devices (Negatives) 33 paired sets There was one false negative result with the new device: Leuconostoc spp. with 3 mL of blood (plate count 35 CFU). {7} # BACTEC Instrument Platform Compatibility Data for this study were from a subset of organisms included in the Instrument Time to Detection study. A total of 246 paired sets (new and predicate devices) were tested in each the BACTEC FX, BACTEC 9240 and BACTEC 9050 fluorescent-series blood culture instruments. There were nine paired sets that failed to recover in both the new and predicate devices in the BACTEC FX instrument. Recovery Failures in the BACTEC FX | Device | Organism | Blood Volume | Replicates | | --- | --- | --- | --- | | New (Plastic) and Predicate Devices (Glass) | S. pneumoniae | 3 mL | 3 | | New (Plastic) and Predicate Devices (Glass) | S. pneumoniae | 10 mL | 3 | | New (Plastic) and Predicate Devices (Glass) | E. corrodens | 10 mL | 3 | A total of three recovery failures were observed with the BACTEC 9050 instrument. All three recovery failures were associated with Leuconostoc spp. The BACTEC 9050 exhibited a statistically significant difference in time to detection between the new and predicate devices. In the BACTEC 9050, the Wilcoxon median time to detection difference estimate is 0.417 hours (25 minutes) in favor of the predicate device. Recovery Failures in the BACTEC 9050 | Device | Organism | Blood Volume | Replicates | | --- | --- | --- | --- | | Predicate Device (Glass) | Leuconostoc spp. | 3 mL | 1 | | Predicate Device (Glass) | Leuconostoc spp. | 10 mL | 1 | | New Device (Plastic) | Leuconostoc spp. | 3 mL | 1 | There was no recovery failures observed with the BACTEC 9240 instrument. # Antimicrobial Neutralization Capability The amount of antimicrobial added to each bottle represents the amount found in 7 mL of blood at or near the peak serum level. The antimicrobials were: | Amoxicillin/Clavulanate | Aztreonam | Ceftazidime | | --- | --- | --- | | Ciprofloxacin | Ceftriaxone | Cefotaxime | | Ertapenem | Fluconazole | Cefepime | | Gentamicin | Imipenem | Levofloxacin | | Meropenem | Tetracycline | Tigecycline | | Ticarcillin/Clavulanate | Piperacillin/Tazobactam | Vancomycin | {8} A supplemental antimicrobial neutralization study, at a lower concentration was conducted with a representative drug in each antimicrobial class: Aztreonam, Ciprofloxacin, Cefotaxime, Cefepime, Gentamicin, Meropenem, Piperacillin, and Tazobactam. No statistically significant difference was observed in the studies. a. Method comparison with predicate device: Performance of the new BD BACTEC Plus Aerobic /F Plastic was compare to that of the BACTEC Plus Aerobic/F medium in glass bottles. b. Matrix comparison: BD BACTEC Plus culture media, human blood volume, common bloodstream pathogens 3. Clinical studies: Not Applicable, analytical seeded studies for comparison between the new BD BACTEC Plus Aerobic /F Plastic and the BACTEC Plus Aerobic/F medium in glass bottles a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Not Applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. {9} O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...