MICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS WITH TETRACYCLINE
K071268 · Dade Behring, Inc. · LON · Jun 26, 2007 · Microbiology
Device Facts
Record ID
K071268
Device Name
MICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS WITH TETRACYCLINE
Applicant
Dade Behring, Inc.
Product Code
LON · Microbiology
Decision Date
Jun 26, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Indications for Use
The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels - Tetracycline is intended for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive Enterococci and Staphylococci. The MicroScan® Synergies plus™ Panels are read on the WalkAway® -SI System (including upgraded WalkAway® -40 or WalkAway® -96 to meet WalkAway® SI equivalence).
Device Story
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels determine antimicrobial susceptibility for Enterococci and Staphylococci. Device uses dehydrated antibiotic panels rehydrated with inoculated broth (turbidity method). Panels incubated in WalkAway®-SI System; optical sensors measure growth via colorimetric detection. System performs rapid (4.5-16h) or overnight (16-20h) readings. Output provides quantitative MIC values and qualitative susceptibility interpretations. Used in clinical microbiology laboratories by technicians. Results assist clinicians in selecting appropriate antibiotic therapy for bacterial infections. Benefits include rapid turnaround time for susceptibility results compared to traditional overnight methods.
Clinical Evidence
Performance evaluated using 610 clinical isolates (564 fresh, 46 stock) and 75 challenge isolates across six sites. Compared against CLSI reference broth dilution method. Rapid method (4.5-16h) showed 94.9% Category Agreement (CA) and 94.8% Essential Agreement (EA). Overnight instrument read showed 98.0% CA and 98.2% EA. Overnight manual read showed 98.5% CA and 99.1% EA. E. faecium rapid read showed 5.4% very major error rate, leading to a specific reporting limitation. No clinical data; bench testing only.
Technological Characteristics
Growth-based antimicrobial susceptibility test system. Panels contain dehydrated antibiotics in water/broth. Inoculum prepared via turbidity method (0.5 McFarland). Detection via colorimetric optics system (lamp/photosensor). Connectivity: WalkAway®-SI instrument. Software: Automated growth-based detection algorithm. Sterilization: Not specified.
Indications for Use
Indicated for determining antimicrobial susceptibility of rapidly growing aerobic and facultative anaerobic Gram-positive enterococci and staphylococci (specifically Staphylococcus species and Enterococcus species) to tetracycline. For prescription use only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K062705 — SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH LINEZOLID · Dade Behring, Inc. · Oct 20, 2006
K051350 — SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS · Dade Behring, Inc. · Jul 25, 2005
K071316 — MICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS NEW ANTIMICROBIAL-STREPTOMYCIN SYNERGY SCREEN · Dade Behring, Inc. · Aug 20, 2007
K070860 — MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH CLINDAMYCIN · Dade Behring, Inc. · May 11, 2007
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K071268
B. Purpose for Submission:
To add tetracycline at concentrations of 0.25-16 µg/mL 7-Dilution Sequence, 0.5-8 µg/mL 5-Dilution Breakpoint Sequence and 2-8 µg/mL 3-Dilution Breakpoint Sequence for Enterococci and Staphylococci, to the Microscan® Synergies plus™ Gram-Positive MIC/Combo Panels.
C. Measurand:
Tetracycline 0.25-16 µg/mL 7-Dilution Sequence
Tetracycline 0.5-8 µg/mL 5-Dilution Breakpoint
Tetracycline 2-8 µg/mL 3-Dilution Breakpoint
D. Type of Test:
Quantitative and Qualitative growth-based detection algorithm using optics light detection
E. Applicant:
Dade Behring Inc, MicroScan®
F. Proprietary and Established Names:
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels - Tetracycline
G. Regulatory Information:
1. Regulation section:
866.1645 - Fully automated short-term incubation cycle antimicrobial susceptibility system
866.1640 - Antimicrobial Susceptibility Test Powder
2. Classification:
{1}
Class II
3. Product code:
LON – Automated AST system short incubation
LRG – Instrument for Auto Reader & Interpretation of Overnight Antimicrobial Susceptibility Systems
JWY – Manual Antimicrobial Susceptibility Test Systems
LTT – Panels, Test, Susceptibility, Antimicrobial
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels - Tetracycline is intended for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive Enterococci and Staphylococci. The MicroScan® Synergies plus™ Panels are read on the WalkAway® -SI System (including upgraded WalkAway® -40 or WalkAway® -96 to meet WalkAway® SI equivalence).
2. Indication(s) for use:
It is indicated for use for the addition of Tetracycline at concentrations 0.25-16 µg/mL (7-Dilution MIC Sequence), 0.5-8 µg/mL (5-Dilution BP Sequence), and 2-8 µg/mL (3-Dilution BP Sequence), to the gram-positive test panel for testing of Enterococci spp. and Staphylococci spp. at 4.5-16 hrs or 16-20 hours for an overnight reading.
3. Special conditions for use statement(s):
- Turbidity method of inoculum preparation only.
- For prescription use only.
- Hold all susceptible (<8 µg/mL) Enterococcus faecium isolates to overnight reporting (16/18 hours)
4. Special instrument requirements:
Not Applicable
I. Device Description:
{2}
Each panel contains two control wells: a negative control well, and a growth control well (contains test medium without antibiotic). Antibiotics are diluted in water, buffer, or minute concentrations of broth to selected concentrations prior to dehydration of the panels. The panel is rehydrated and inoculated at the same time with 0.1 ml of suspension prepared by the turbidity method (inoculum prepared in 0.4% saline with Pluronic®, then 0.1 ml transferred to 25 ml of inoculum Synergies plus Pos Broth with Pluronic®) for a final inoculum concentration of 3-7 X 10⁵ CFU/ml. Panels are incubated in a Walk-Away® System and read periodically starting at 4.5 hours until sufficient growth to determine the MIC. Alternately the panels may be incubated at 35°C in a non-CO₂ for 16-24 hours and read by visual observation of growth.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
MicroScan® Dried Gram-Positive and Gram-Negative MIC/Combo Panels
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels
2. Predicate 510(k) number(s):
k862140
k020185
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | MicroScan® panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic organisms | Same |
| Specimen | Isolated colonies from culture used | Same |
| Inoculum | Inoculum density to 0.5 McFarland standard | Same |
| Incubation | <16 hours
16 – 24 hours | Same |
| Results | Quantitative with qualitative interpretations | Same |
| Technology | Growth based | Same |
| Differences | | |
| Item | Device | Predicate |
{3}
| Panels | Dried tetracycline in water with 5% broth | Dried clindamycin or gentamicin in broth |
| --- | --- | --- |
| Reading | Uses both a ≤ 16 h read and overnight read method in the same system | Overnight system uses only the overnight reading method and <16 hour instruments use only the <16 hour read method. |
| Inoculum preparation | Turbidity method of inoculation only. | Inoculum prepared using either the Turbidity method or Prompt® system |
| Instrument | WalkAway® -SI System or equivalent | autoScan® -4 or WalkAway® |
| Antibiotic | Tetracycline at 0.25- 16 μg/ | Different concentrations depending on the antibiotic |
## K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; Clinical and Laboratory Standards Institute (CLSI) M7 (M100-S17) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”.
## L. Test Principle:
The WalkAway® SI uses a Colorimetric Optics System consisting of a color wheel/lamp assembly and a Photosensor. There is an initial read at 2.5 hours with a possible final read at 4.5, 5.5, 6.5, 8, 12, 16, or 18 hours (overnight instrument readings, manual readings) depending on the growth rate of the organism being tested. The time of final read is dependent on the growth rate of the organism and the sensitivity of the automatic reader since cell densities below $2 \times 10^{7}$ cells/ml are not detected.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Reproducibility was demonstrated using 10 isolates tested at six sites on three separate days in triplicate. The study included the testing on the Rapid (WalkAway® SI read at <16 hours), Overnight Instrument (WalkAway® 16-18 hour readings), and Overnight Manual (manual/visual read at 16-20 hours). There are three dilution sequences included in this submission using different reading algorithms namely 7-MIC Dilution sequence (i.e. $0.25 - 16\ \mu\mathrm{g/mL}$), 5-Dilution Breakpoint Sequence (i.e. $0.5 - 8\ \mu\mathrm{g/mL}$), and the 3-Dilution Breakpoint Sequence (i.e. $2 - 8\ \mu\mathrm{g/mL}$). All dilution sequences were done on the same 10 isolates. The BP dilutions (5-Dilution Breakpoint Sequence, and the 3-Dilution
{4}
Breakpoint Sequence) with reproducibility reading were based only on category agreement. There were no isolates that did not grow. All results were $>95\%$ reproducible.
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The recommended QC isolates, $E$ faecalis ATCC 29212 and $S$ aureus ATCC 29213 were tested a sufficient number of times with acceptable results most of the time with the reference method. Quality control results demonstrated the ability of the different reading parameters (manual and instrument) to produce acceptable results.
The following table provides the frequency of the results in each concentration with the expected range stated.
| Results | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Organism | Conc in μg/mL | # reference | MicroScan® | | | |
| Tetracycline | | | Manual overnight | Instrument overnight | <16 hrs instrument | |
| E. faecalisATCC 29212Expected Range8-32 μg/mL | <=0.06 | | | | | |
| | 0.12 | | | | | |
| | 0.25 | | | | | |
| | 0.5 | 1 | 1 | 1 | 1 | |
| | 1 | | | | | |
| | 2 | | | | | |
| | 4 | | | | | |
| | 8 | 39 | 2 | | 39 | |
| | 16 | 84 | 104 | 116 | 89 | |
| | 32 | 5 | 24 | 14 | 2 | |
| | >32 | | | | | |
| S. aureusATCC 29213Expected RangeCLSI:0.12 - 1 μg/mL | <=0.06 | | | | |
| --- | --- | --- | --- | --- | --- |
| | 0.12 | 5 | | | 14 |
| | 0.25 | 78 | 78 | 99 | 110 |
| | 0.5 | 50 | 55 | 36 | 6 |
| | 1 | 3 | 5 | | |
| | 2 | 1 | | | |
| | 4 | | 1 | | |
| | 8 | | | 2 | |
| | 16 | | | | |
| | 32 | | | | 2 |
| | >32 | | | 2 | 7 |
The 3 Dilution sequence, and 5 Dilution sequence breakpoint panels using separate reading algorithms also demonstrated acceptable quality control
{5}
results.
Inoculum density control: A turbidity meter was used for the turbidity inoculation method. Turbidity inoculum verification provided.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Clinical testing was conducted at six sites using fresh isolates supplemented with stock isolates. A total of 610 gram-positive Staphylocci and Enterococci were tested of which 564 were fresh isolates and 46 were stock isolates. There were 75 challenge isolates tested at two sites and compared to the reference broth dilution result mode that was determined by previous testing of each isolate multiple times in the recommended reference panel. The Rapid method readings were obtained at times between 4.5 and 16 hours of incubation for >95% of the results. An additional comparison was done with readings on the instrument after overnight incubation and also read manually when incubated 16-18 hours. Performance by these alternate reading methods was also acceptable with no apparent differences or trends. The test device had a no growth rate of <10%.
The recommended CLSI reference method was followed with the exception of the use of a small amount (0.1%) Pluronic® in the final inoculum. A validation of the use of Pluronic® in the frozen reference panels was conducted. Similar calculations for the different reading methods were performed with very little difference. QC was also performed with no difference apparent in the results.
The charts below demonstrate the performance of all three reading methods (Rapid instrument readings at <16 hours, overnight on the WalkAway® and manually read at 18 hours using the touchScan®-SR) with the 7-Dilution sequence when compared to the reference method.
{6}
Summary Table for the Rapid Instrument Read (7-Dilution Sequence)
| | Total | EA | %EA | Total eval | EA of eval | %EA | CA | %CA | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Efficacy | 585 | 559 | 95.6 | 134 | 120 | 89.6 | 558 | 95.4 | 203 | 19 | 5 | 3 |
| Challenge | 66 | 58 | 87.9 | 12 | 9 | 75.0 | 60 | 90.9 | 29 | 2 | 4 | 0 |
| Combined | 651 | 617 | 94.8 | 146 | 129 | 88.4 | 618 | 94.9 | 232 | 21 | 9 | 3 |
Summary Table for the Overnight Instrument Read (7-Dilution Sequence)
| | Total | EA | %EA | Total eval | EA of eval | %EA | CA | %CA | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Efficacy | 610 | 598 | 98.0 | 86 | 76 | 88.4 | 597 | 97.9 | 217 | 11 | 2 | 0 |
| Challenge | 75 | 75 | 100 | 14 | 14 | 100 | 74 | 98.7 | 30 | 1 | 0 | 0 |
| Combined | 685 | 673 | 98.2 | 100 | 90 | 90.0 | 671 | 98.0 | 247 | 12 | 2 | 0 |
Summary Table for the Overnight Manual Read (7-Dilution Sequence)
| | Total | EA | %EA | Total eval | EA of eval | %EA | CA | %CA | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Efficacy | 610 | 604 | 99.0 | 105 | 99 | 94.3 | 601 | 98.5 | 217 | 8 | 1 | 0 |
| Challenge | 75 | 75 | 100 | 17 | 17 | 100 | 74 | 98.7 | 30 | 1 | 0 | 0 |
| Combined | 685 | 679 | 99.1 | 122 | 116 | 95.1 | 675 | 98.5 | 247 | 9 | 1 | 0 |
EA-Essential Agreement
CA-Category Agreement
R-resistant isolates
maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies
Essential agreement (EA) is when the MicroScan® Synergies plus panel agrees with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the MicroScan® Synergies plus panel interpretation agrees exactly with the reference panel interpretation. Evaluable (Eval) are results that are within the test range and on scale.
Similar results were obtained when the algorithm for reading the 3 Dilution sequence with 2, 4 and 8 µg/mL wells and algorithm for reading the 5 Dilution sequence with 0.5, 1, 2, 4 and 8 wells were applied with no difference in CA. EA was not calculated for these readings since insufficient numbers of concentrations were available for the evaluation.
The E. faecium group had a higher than expected vmj error rate of 5.4% with the Rapid read method. However, no vmj errors were observed with the overnight instrument or overnight manual read methods. Therefore, a limitation statement will be included for this drug/bug combination, e.g., "Hold all susceptible (<8 µg/mL) Enterococcus faecium isolates to overnight reporting (16/18 hours)."
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
{7}
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Staphylococcus spp. and Enterococcus spp. ≤4 (S), 8 (I), ≥16 (R)
N. Proposed Labeling:
The expected value range, interpretive criteria and QC for gram positive panels are included in the package insert. The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.