MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH CHLORAMPHENICOL (4-16 UG/ML)

K063564 · Dade Behring, Inc. · LON · Feb 8, 2007 · Microbiology

Device Facts

Record IDK063564
Device NameMICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH CHLORAMPHENICOL (4-16 UG/ML)
ApplicantDade Behring, Inc.
Product CodeLON · Microbiology
Decision DateFeb 8, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The testing of chloramphenicol is for use with MicroScan® Synergies plus™ Panels read on the WalkAway® -SI System (including upgraded WalkAway® - 40 or WalkAway® -96 to meet WalkAway® SI equivalence). MicroScan® panels are designed for use in determining identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive cocci and Listeria; the panels also provide quantitative and/or qualitative antimicrobial agent susceptibility for enterococci and staphylococci.

Device Story

MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels are used for antimicrobial susceptibility testing (AST) of gram-positive bacteria. Panels contain dehydrated antibiotics; rehydrated with bacterial inoculum prepared via turbidity method. Panels incubated in WalkAway®-SI system; optical system monitors growth periodically from 4.5 to 18 hours. System uses colorimetric optics (lamp/photosensor) to detect growth; determines MIC based on growth inhibition. Results provided to clinicians to guide antibiotic therapy. Benefits include rapid AST results compared to traditional overnight methods, facilitating timely clinical decision-making for bacterial infections.

Clinical Evidence

Bench testing only. Performance validated by comparing the subject panel against a frozen Reference Panel using fresh/stock efficacy isolates and challenge strains. Results demonstrated 96.1% overall categorical agreement for chloramphenicol. Reproducibility and precision were confirmed using the Turbidity inoculum method and WalkAway® SI system. Quality control testing met acceptable performance criteria.

Technological Characteristics

Miniaturized broth dilution panels; dehydrated antimicrobial agents in water/buffer/broth. Sensing via MicroScan® Rapid Fluorogenic Identification and AST technologies. Incubation in WalkAway® SI system (4.5-18 hours at 35°C). Connectivity via automated instrument reading or visual inspection. No specific material standards or software architecture details provided.

Indications for Use

Indicated for determining antimicrobial susceptibility of rapidly growing aerobic and facultative anaerobic gram-positive enterococci and staphylococci (specifically Staphylococcus aureus) isolated from solid media. For prescription use in clinical laboratory settings.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K063564 B. Purpose for Submission: To add chloramphenicol to the MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels C. Measurand: Chloramphenicol (4-16 µg/mL) D. Type of Test: Qualitative growth based detection algorithm using optics light detection E. Applicant: Dade Behring Inc, MicroScan® F. Proprietary and Established Names: MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels with Chloramphenicol (4-16 µg/mL) G. Regulatory Information: 1. Regulation section: 866.1645 - Fully automated short-term incubation cycle antimicrobial susceptibility system 866.1640 - Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: LON – Automated AST system short incubation LRG-Instrument for Auto Reader &amp; Interpretation of Overnight Antimicrobial {1} Susceptibility Systems JWY - Manual Antimicrobial Susceptibility Test Systems 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): The testing of chloramphenicol is for use with MicroScan® Synergies plus™ Panels read on the WalkAway® -SI System (including upgraded WalkAway® - 40 or WalkAway® -96 to meet WalkAway® SI equivalence). MicroScan® panels are designed for use in determining identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive cocci and Listeria; the panels also provide quantitative and/or qualitative antimicrobial agent susceptibility for enterococci and staphylococci. 2. Indication(s) for use: The testing of chloramphenicol at concentrations of 4-16 µg/mL to the gram-positive test panel for 4.5-16 hours or 16-20 hours for a overnight reading. 3. Special conditions for use statement(s): - Turbidity method of inoculum preparation only. - For prescription use only. 4. Special instrument requirements: Not Applicable I. Device Description: This submission is for the AST system only. The ID system was not reviewed. Each panel contains two control wells: a negative control well, and a growth control well (contains test medium without antibiotic). Antibiotics are diluted in water, buffer, or minute concentrations of broth to selected concentrations prior to dehydration of the panels. The panel is rehydrated and inoculated at the same time with 0.1 ml of suspension prepared by the turbidity method (inoculum prepared in 0.4% saline with PLURONIC®, then 0.1 ml transferred to 25 ml of inoculum Synergies plus Pos Broth with PLURONIC®) for a final inoculum concentration of 3-7 X 10⁵ CFU/ml. Panels are incubated in a Walk-Away® System and read periodically starting at 4.5 hours until sufficient growth to determine the MIC. Alternately the panels may be incubated at 35°C in a non-CO₂ for 16-18 hours and read by visual observation of growth. J. Substantial Equivalence Information: {2} 1. Predicate device name(s): MicroScan® Dried Gram-Positive and Gram-Negative MIC/Combo Panels 2. Predicate 510(k) number(s): k862140 k020185 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | MicroScan® panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic organisms | Same | | Specimen | Isolated colonies from culture used | Same | | Inoculum | Inoculum density to 0.5 McFarland standard | Same | | Incubation | <16 hours 16 – 18 hours | Same | | Results | Qualitative interpretations | Same | | Technology | Growth based | Same | | Differences | | | | Item | Device | Predicate | | Panels | Dried chloramphenicol in water | Dried clindamycin or gentamicin in broth | | Reading | Uses both an early read and overnight read methods in the same system | Overnight system uses only the overnight reading methods and <16 hour instruments use only the <16 hour read methods. | | Inoculum preparation | Turbidity method of inoculation only | Inoculum prepared from isolated colonies using either the Turbidity method or Prompt® system | | Instrument | WalkAway® -SI System or equivalent | autoScan® -4 or WalkAway® | | Antibiotic | chloramphenicol at 4-16 μg/mL | Different concentrations depending on the antibiotic | | | | infection | | Inoculum preparation | Turbidity method of inoculation only | Inoculum prepared from isolated colonies using either the Turbidity method or Prompt® system | | Inoculum preparation | WalkAway® -SI System or equivalent | AutoScan® -4 or WalkAway® | {3} 4 K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; Clinical and Laboratory Standards Institute (CLSI) M7 (M100-S16) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”. L. Test Principle: The WalkAway® SI uses a Colorimetric Optics System consisting of a color wheel/lamp assembly and a Photosensor. There is an initial read at 2.5 hours with a possible final read at 4.5, 5.5, 6.5, 8, 12, 16, or 18 hours (overnight instrument readings, manual readings) depending on the growth rate of the organism being tested. The time of final read is dependent on the growth rate of the organism and the sensitivity of the automatic reader since cell densities below $2 \times 10^{7}$ cells/ml are not detected. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Reproducibility was demonstrated using ten isolates tested at 3 sites on 3 separate days in triplicate. The study included the testing on the WalkAway® SI read at &lt;16 hours, WalkAway® 16-18 hour readings and manual readings at 16-20 hours incubation. Reproducibility was determined using only category results since less than 5 dilutions will be the panel format. All results were &gt;95% reproducible. Observations for trending demonstrated a trend for the ≤ 16 hour readings to provide a more susceptible result when there was a difference at all. b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): The recommended QC isolate was tested a sufficient number of times with acceptable results on all testing days with the reference method. Quality control results demonstrated the ability of the different reading parameters (manual and instrument) to produce acceptable results. The following table provides the frequency of the results in each concentration with the expected range stated. {4} | Results | | | | | | | --- | --- | --- | --- | --- | --- | | Organism | Conc in μg/ml | # reference | MicroScan® | | | | | | | Manual overnight | Instrument overnight | 4.5-18h instrument | | E. faecalis ATCC 29212 Range 4-16 μg/mL | ≤ 4 | 30 | 65 | 76 | 110 | | | 8 | 75 | 47 | 35 | 1 | | | 16 | | | | | | | >16 | | | | | | | | | | | | | S. aureus ATCC 29213 Range 2-16 μg/mL | ≤ 4 | 1 | 10 | 10 | 72 | | | 8 | 93 | 89 | 86 | 28 | | | 16 | 1 | 1 | | | | | >16 | | | 3 | | There appears to be a trend for the $\leq 16$ hour readings to be more susceptible than the reference method by one dilution for both the ATCC S. aureus and the E. faecalis. The E. faecalis also appears to trend to a more susceptible result for all MicroScan® readings when compared to the mode of the reference method. Inoculum density control: A turbidity meter was used for the turbidity inoculation method. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: A total of 575 gram-positive isolates were tested of which 495 were fresh isolates. A comparison was made to the standard reference method result performed as described in the CLSI document M7 except for the modification described in the submission. The reference panels were inoculated with a very small amount of PLURONIC® in water and added to the frozen reference panels containing the antibiotic in broth. The use of PLURONIC® water is not part of the reference method but MicroScan® uses it in very small amounts as a wetting solution for {5} inoculation of the panels with the small volume of inocula used. A validation of the use of chloramphenicol with PLURONIC® was conducted demonstrating that there was no difference in the results with and without PLURONIC®. The Synergies plus™ readings were obtained at times between 4.5 and 16 hours of incubation for &gt; 95% of the results. Performance of the different read methods was based on category agreement only since only 3 dilutions of chloramphenicol are tested and a category result only will be provided. Essential Agreement (EA) can't be determined with less than 5 dilutions of antibiotic to evaluate. An additional comparison was done with readings on the instrument after overnight incubation and also read manually when incubated 16- 18 hours. Performance by these alternate reading methods was also acceptable but when differences occurred the &lt;16 hour readings were more susceptible. The chart below demonstrates the performance of all three reading methods, (Synergies plus™ readings at &lt;16 hours, overnight on the WalkAway® and manually read at 18 hours using the touchScan®-SR) when compared to the reference method result of clinical and challenge isolates. | | total | CA | %CA | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | | <16 h | 644 | 619 | 96.1 | 46 | 21 | 2 | 2 | | Overnight Instrument | 650 | 621 | 95.5 | 46 | 22 | 7 | 0 | | Overnight manual | 650 | 627 | 96.5 | 46 | 17 | 4 | 0 | CA-Category Agreement R-resistant isolates vmj-very major discrepancies min- minor discrepancies maj-major discrepancies CA is when the interpretation of the reference method agrees exactly with the interpretation of the MicroScan® result. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable {6} c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: Staphylococcus spp. and Enterococcus spp. interpretive criteria: ≤ 8 (susceptible) 16 (intermediate) ≥ 32 (resistant) All values are included in the package insert. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...