← Product Code [NGZ](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ) · K151817

# VITEK 2  AST-YS Caspofungin <=0.125 - >=8 ug/mL (K151817)

_bioMerieux, Inc. · NGZ · Feb 18, 2016 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K151817

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [NGZ](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ.md)
- **Decision Date:** Feb 18, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** AI/ML

## Indications for Use

VITEK® 2 AST-YS Caspofungin is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Caspofungin is a quantitative test. Caspofungin has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifyingal. Candida albicans Candida glabrata Candida guilliermondii Candida krusei Candida parapsilosis Candida tropicalis The VITEK @ 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast.

## Device Story

VITEK® 2 AST-YS Caspofungin is an in vitro diagnostic test card used with VITEK® 2 and VITEK® 2 Compact systems. It performs automated quantitative susceptibility testing of Candida species. The process involves diluting isolated colonies in saline, rehydrating the antimicrobial medium in the card, and incubating. The system monitors well growth over 18 hours, calculating the Minimum Inhibitory Concentration (MIC) and interpretive category. Used in clinical laboratories by technicians; results assist physicians in selecting appropriate antifungal therapy for patients with Candida infections.

## Clinical Evidence

Bench testing only. Performance evaluated using 755 clinical and 111 challenge isolates. Combined clinical and challenge testing for automatic dilution showed 99.7% Essential Agreement (EA) and 99.8% Category Agreement (CA). Challenge testing for manual dilution on VITEK 2 and VITEK 2 Compact showed 100% EA and CA. Quality control testing (C. krusei ATCC 6258, C. parapsilosis ATCC 22019) met expected ranges 100% of the time.

## Technological Characteristics

Miniaturized broth microdilution card; antimicrobial medium rehydrated with saline-diluted isolate; automated incubation and optical monitoring; quantitative MIC determination; compatible with VITEK 2 and VITEK 2 Compact platforms.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- VITEK® 2 AST-YS Flucytosine ([K133952](/device/K133952.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K151817

B. Purpose for Submission:
To obtain a substantial equivalence determination for the addition of caspofungin to the VITEK® 2 and VITEK® Compact Systems Antimicrobial Susceptibility Test (AST) Systems

C. Measurand:
The VITEK 2 AST Yeast card contains the following concentrations of caspofungin: 0.125, 0.5, 2, and 8 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result reporting range for the card is 0.125 – 8 µg/mL.

D. Type of Test:
Automated quantitative or qualitative antifungal susceptibility test of Candida species to caspofungin.

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
VITEK® 2 AST-YST Caspofungin ≤0.12 - ≥8 µg/mL

G. Regulatory Information:
1. Regulation section:
21 CFR 866.1640, Antimicrobial Susceptibility Test Powder
2. Classification:
II
3. Product code:
NGZ – Susceptibility Test Plate, Antifungal

{1}

LRG – Instrument for Auto Reader and Interpretation of Overnight Susceptibility Systems

LTT – Panels, Test, Susceptibility, Antimicrobial

LTW – Susceptibility Test Cards, Antimicrobial

4. Panel:

83 Microbiology

H. Intended Use:

1. Intended use(s):

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

2. Indication(s) for use:

VITEK® 2 AST-YS Caspofungin is designed for antifungal susceptibility testing of Candida species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Caspofungin is a quantitative test. Caspofungin has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal.

Active in vitro and clinical infections:

- Candida albicans
- Candida glabrata
- Candida guilliermondii
- Candida krusei
- Candida parapsilosis
- Candida tropicalis

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

3. Special conditions for use statement(s):

For prescription use only.

4. Special instrument requirements:

{2}

For use with the VITEK® 2 and VITEK® 2 Compact Systems

The following limitations are included in the device labeling:

"One non-susceptible isolate of C. glabrata gave a susceptible MIC result, a potential very major error."

"The ability of VITEK 2 AST-YS to detect resistance to caspofungin is unknown because non-susceptible strains were not available for comparative testing. Isolates yielding caspofungin MIC results suggestive of a "non-susceptible" category (&gt;2 μg/mL) should be submitted to a reference laboratory for further testing."

I. Device Description:

The VITEK® 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK® 2 test card contains 64 microwells. A control well containing only culture medium is included on all cards, with the remaining wells containing premeasured amounts of a specific antimicrobial agent in a culture medium base. A suspension of organism from a pure culture is prepared in a tube containing 0.45-0.5% sterile saline and standardized to a McFarland 0.5 using the DensiCHEK Plus™. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader; manual methods can also be used for the inoculation of test cards for use in the VITEK 2 System. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK 2 AST-YST Caspofungin has the following concentrations in the card: 0.125, 0.5, 2 and 8 μg/mL (equivalent standard method concentration by efficacy in μg/mL). The MIC result range for the VITEK 2 AST-YST Caspofungin card is ≤ 0.12 - &gt;8μg/mL.

J. Substantial Equivalence Information:

1. Predicate device name(s): VITEK® 2 AST-YS Flucytosine
2. Predicate 510(k) number(s): K133952
3. Comparison with predicate:

{3}

4

Table 1. Comparison with the Predicate Device

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | VITEK® 2 AST-YS Caspofungin | VITEK® 2 AST-YS Flucytosine (K133952)  |
|  Intended Use | Quantitative and qualitative susceptibility for colonies of Candida spp. | Same  |
|  Test Methodology | Automated yeast antifungal susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems (VITEK 2 Systems) to determine the in vitro susceptibility of Candida species. | Same  |
|  Inoculum | Saline suspension of organism | Same  |
|  Test Card | VITEK 2 Test Card format | Same  |
|  Instrument | VITEK 2 and VITEK 2 Compact Systems | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | VITEK® 2 AST-YS Caspofungin | VITEK® 2 AST-YS Flucytosine (K133952)  |
|  Antimicrobial Agent | caspofungin | flucytosine  |
|  Antimicrobial Concentrations | Unique to caspofungin | Unique to flucytosine  |
|  Base broth | Unique to caspofungin | Unique to flucytosine  |
|  Analysis algorithms | Unique to caspofungin | Unique to flucytosine  |

K. Standard/Guidance Document Referenced:

CLSI Document CLSI M27-A3, Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts; Approved Standard - Third Edition

CLSI Document CLSI M27-S4, Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts; Fourth Informational Supplement

L. Test Principle:

The VITEK® 2 System optics use visible light to directly measure organism growth. The transmittance

{4}

optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. Several parameters based on the growth characteristics observed are used to provide appropriate input for the MIC calculations. Discriminate analysis is used to develop the algorithm that determines the susceptibility result for all antimicrobials on the VITEK® 2 System. The MIC result must be linked to organism identification in order to determine a category interpretation. A category interpretation (SIR) will be reported along with each MIC result.

## M. Performance Characteristics:

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Reproducibility studies were performed on the VITEK 2 and VITEK 2 Compact using both automatic dilution and manual dilution for panel inoculation. With each inoculation/reading methodology, the mode MIC was determined for each isolate and the reproducibility was calculated based on MIC values falling within ± 1 dilution of the mode MIC.

Testing using VITEK 2 and automatic dilution was performed at a total of four sites using 18 isolates representing species included in the intended use. Best case and worst case reproducibility for VITEK 2 with automatic dilution demonstrated acceptable performance at 96.7% and 95.3%, respectively.

Testing using VITEK 2 and manual dilution was performed using 10 Candida isolates tested at three sites. There were no off-scale results; best and worst case reproducibility was acceptable at 96.3%.

Reproducibility studies were also performed using manual dilution and the Compact option using ten Candida strains tested at three sites. There were no off-scale results; best case and worst case reproducibility was acceptable at 98.9%.

#### b. Linearity/assay reportable range:

Not applicable

#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Inoculum density control was monitored using the DensiCHEK Plus™ instrument. The DensiCHEK Plus™ was standardized weekly with all results recorded and in expected range.

A purity check of all organisms was performed at the time of VITEK 2 card inoculation. Only results obtained with pure cultures were evaluated.

Quality control testing was conducted throughout comparative testing at each site using two recommended quality control strains: Candida krusei (ATCC 6258) and Candida

5

{5}

parapsilosis (ATCC 22019). In those instances where the test result was out-of-range for all replicates of the reference method, all data from that day's testing was considered invalid and the testing for that day was repeated.

The QC organisms were tested a minimum of 20 times with the reference method and with VITEK 2 (manual and automatic dilution) and with the Compact option using manual dilution. QC results were interpreted after 24 hours of incubation. QC results for the VITEK 2 AST-YS Caspofungin were within the expected results range 100% of the time for both instrument platforms and both dilution methods. A summary of the QC performance is provided in Table 2 below.

Table 2. Quality Control Results VITEK 2 Interpreted after 24 Hours of Incubation

|   |  | VITEK 2 Automatic Dilution |   | VITEK 2 Manual Dilution |   | VITEK 2 Compact Manual Dilution  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  QC Organism | Conc. (μg/mL) | Test | Ref. | Test | Ref. | Test | Ref.  |
|  C. krusei ATCC 6258
FDA and CLSI
Expected QC
Range at 24 hours:
0.12 - 1μg/mL | 0.015625 |  |  |  |  |  |   |
|   |  0.03125 |  |  |  |  |  |   |
|   |  0.0625 |  |  |  |  |  |   |
|   |  0.125 | 51 |  | 20 |  | 11 |   |
|   |  0.25 | 10 | 42 | 41 | 42 | 50 | 42  |
|   |  0.5 |  | 19 |  | 19 |  | 19  |
|   |  1 |  |  |  |  |  |   |
|   |  2 |  |  |  |  |  |   |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  ≥16 |  |  |  |  |  |   |
|  |   |   |   |   |   |   |   |
|  C. parapsilosis ATCC 22019
FDA and CLSI
Expected QC
Range at 24 hours:
0.25 - 1 μg/mL | 0.15625 |  |  |  |  |  |   |
|   |  0.03125 |  |  |  |  |  |   |
|   |  0.0625 |  |  |  |  |  |   |
|   |  0.125 |  |  |  |  |  |   |
|   |  0.25 | 1 | 1 | 1 | 1 | 1 | 1  |
|   |  0.5 | 60 | 48 | 60 | 48 | 60 | 48  |
|   |  1 |  | 12 |  | 12 |  | 12  |
|   |  2 |  |  |  |  |  |   |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  ≥16 |  |  |  |  |  |   |

{6}

d. Detection limit:

Not applicable

e. Analytical specificity:

Not applicable

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Results obtained with the bioMérieux VITEK2 AST-YS card with caspofungin were compared to results obtained with the reference frozen broth microdilution panel. The VITEK 2 AST-YS Caspofungin card contains the following concentrations: 0.125, 0.5, 2, and 8 µg/mL (Equivalent standard method concentration by efficacy in µg/mL). The MIC result range for the VITEK 2 card is 0.125 – 8 µg/mL. The frozen reference panel contained two-fold serial dilutions with a range of 0.0156 – 32 µg/mL.

Test inocula were standardized using the DensiCHEK Plus instrument. VITEK 2 AST-YS cards were inoculated using automatic dilution (for reading on the VITEK 2 instrument) or using a manual method (for reading on the VITEK 2 instrument or the VITEK 2 COMPACT instrument). Reference panels were inoculated as outlined in CLSI M27-A3.

A total of 755 clinical isolates were evaluated at three sites with VITEK 2 cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. The majority of isolates were fresh and recent isolates; 40 clinical isolates (5.3%) were stock isolates.

A total of 111 challenge isolates were tested at one site. The challenge set was tested with both of the card inoculation options, automatic dilution and manual dilution, on the VITEK 2 System and with the manual dilution on the VITEK 2 Compact System.

For isolates interpreted using the VITEK 2 System and inoculated using the automatic dilution option, the combined results from clinical and challenge testing demonstrated a combined EA of 99.7% and CA of 99.8% (Table 3). A total of 145 isolates were determined to be evaluable; the EA of evaluable isolates was determined to 98.6%. Two isolates of C. glabrata were determined to be non-susceptible by the reference method but susceptible by VITEK 2. Because the FDA Reference List Drug Label for caspofungin lists only a susceptible breakpoint (susceptible ≤ 2 µg/mL and non-susceptible &gt; 2 µg/mL) these two isolates would be considered potential very major errors. Analysis of results from one of the two C. glabrata isolates indicated a contaminated culture; the sponsor chose not to remove results of this isolate from the performance determination; however, according to FDA's

{7}

current policy this isolate was not considered to be a potential very major error due to the technical error. The sponsor was asked to add the following limitation to the device labeling:

"One non-susceptible isolate of C. glabrata gave a susceptible MIC result, a potential very major error."

For challenge isolates tested with the VITEK 2 System and inoculated using the manual dilution method, the results from challenge testing demonstrated an EA and CA of 100% (Table 4).

For challenge isolates tested with the VITEK 2 Compact System and inoculated using the manual dilution method, the results from challenge testing demonstrated an EA and CA of 100% (Table 5).

Because a limited number of non-susceptible isolates were evaluated, the sponsor was asked to include the following limitation in their device labeling:

"The ability of VITEK 2 AST-YS to detect resistance to caspofungin is unknown because non-susceptible strains were not available for comparative testing. Isolates yielding caspofungin MIC results suggestive of a "non-susceptible" category (&gt;2 µg/mL) should be submitted to a reference laboratory for further testing."

8

{8}

Table 3: Performance of Clinical and Challenge Isolates VITEK 2 System, Automatic Dilution Method

|   | Total | No. EAa | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. NS | No. S | minb | majb | vmjb  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Overall  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 755 | 752 | 99.6 | 119 | 117 | 98.3 | 753 | 99.7 | 2 | 755 | N/A | N/A | N/A (2)c  |
|  Challenge | 111 | 111 | 100.0 | 26 | 26 | 100.0 | 111 | 100.0 | 1 | 110 | N/A | N/A | 0  |
|  Combined | 866 | 863 | 99.7 | 145 | 143 | 98.6 | 864 | 99.8 | 3 | 863 | N/A | N/A | N/A (2)c  |

a For antifungal agents, essential agreement is  $\pm$  two 2-fold dilutions
b There are no intermediate or resistant interpretive criteria for caspofungin. The current absence of resistant isolates precludes defining any results other than "susceptible."
These isolates are considered potential very major errors (a susceptible result obtained for an organism that tested non-susceptible by the reference method).

EA - Essential Agreement (+/- 2 dilutions)

CA - Category Agreement

EVAL - Evaluable isolates

NS - non-susceptible isolates

S - susceptible isolates

min - minor discrepancies

maj - major discrepancies

vmj - very major discrepancies

Table 4. Performance of Challenge Isolates, VITEK 2 Manual Dilution Method

|   | Total | No. EAa | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. NS | No. S | minb | majb | vmjb  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Overall  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 111 | 111 | 100.0 | 33 | 33 | 100.0 | 111 | 100.0 | 1 | 110 | N/A | N/A | N/A  |

a For antifungal agents, essential agreement is  $\pm$  two 2-fold dilutions
b There are no intermediate or resistant interpretive criteria for caspofungin. The current absence of resistant isolates precludes defining any results other than "susceptible."

{9}

Table 5. Performance of Challenge Isolates, VITEK 2 Compact, Manual Dilution Method

|   | Tot | No. EAa | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. NS | No. S | minb | majb | vmjb  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Overall  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 111 | 111 | 100.0 | 32 | 32 | 100.0 | 111 | 100.0 | 1 | 110 | N/A | N/A | N/A  |

a For antifungal agents, essential agreement is  $\pm$  two 2-fold dilutions
b There are no intermediate or resistant interpretive criteria for caspofungin. The current absence of resistant isolates precludes defining any results other than "susceptible."

{10}

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

Table 6. Susceptibility Interpretive Criteria

|  VITEK® 2 AST-YST | FDA Interpretive Categories* (MIC in μg/mL) Broth Microdilution MIC (mcg/mL) at 24 hours  |   |
| --- | --- | --- |
|   |  Candida sp.  |   |
|  Caspofungin | Susceptible | Non-Susceptible  |
|   |  ≤2 | >2  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K151817](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K151817)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
