K092454 · bioMerieux, Inc. · NGZ · May 21, 2010 · Microbiology
Device Facts
Record ID
K092454
Device Name
VITEK 2 YEAST VORICONAZOLE
Applicant
bioMerieux, Inc.
Product Code
NGZ · Microbiology
Decision Date
May 21, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
VITEK® 2 Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Voriconazole is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Voriconazole has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal. Candida albicans Candida tropicalis Candida krusei Candida lusitaniae Candida parapsilosis Candida guilliermondii The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.
Device Story
VITEK 2 Yeast Voriconazole is an in vitro diagnostic test for determining minimum inhibitory concentrations (MIC) of voriconazole against Candida species. The device consists of AST cards containing antifungal concentrations. Clinical laboratory technicians prepare isolate samples by diluting them in 0.45% saline to rehydrate the card medium. The VITEK 2 system automates filling, sealing, and incubation; the VITEK 2 Compact system requires manual filling/sealing. The system monitors well growth via optical sensors over 36 hours. The device outputs MIC values and interpretive categories. Results assist clinicians in selecting appropriate antifungal therapy for patients with Candida infections.
Clinical Evidence
External evaluation conducted using fresh and stock clinical isolates and challenge strains. Performance compared to CLSI reference method at 24 and 48 hours. Results: 24-hour incubation showed 99.2% Essential Agreement and 99.2% category agreement. 48-hour incubation showed 96.9% Essential Agreement and 98.7% category agreement. Reproducibility and Quality Control testing met acceptance criteria.
Technological Characteristics
Miniaturized broth microdilution AST card. Sensing principle: automated optical growth monitoring. Energy source: electrical (VITEK 2/Compact systems). Connectivity: networked system integration. Software: automated analysis of growth kinetics to determine MIC.
Indications for Use
Indicated for antifungal susceptibility testing of Candida species (C. albicans, C. krusei, C. parapsilosis, C. tropicalis, C. lusitaniae, C. guilliermondii) in clinical laboratory settings. For prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of a stylized circle that is split into two halves. The left half of the circle has horizontal lines, while the right half is solid black. The text "BIOMERIEUX" is written in a stylized font below the circle.
X092454
## 510(k) SUMMARY
### VITEK® 2 Yeast Voriconazole
### 510(k) Submission Information:
## MAY 2 1 2010
| Submitter's Name: | bioMérieux, Inc. |
|----------------------|--------------------------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Nancy Weaver<br>Associate Director, Regulatory Affairs |
| Phone Number: | 314 -731-8695 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | August 06, 2009 |
| B. Device Name: | |
| Formal/Trade Name: | VITEK® 2 Yeast Voriconazole |
| Classification Name: | 21 CFR 866.1640<br>Antimicrobial Susceptibility Test<br>Product Code NGZ |
| Common Name: | VITEK® 2 AST-YS Voriconazole |
| C. Predicate Device: | VITEK® 2 Yeast Fluconazole (K061945) |
### D. 510(k) Summary:
VITEK® 2 Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Voriconazole is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a Íaboratory aid in the determination of in vitro susceptibility to antifungal agents. Voriconazole has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.
| Candida albicans | Candida tropicalis |
|----------------------|------------------------|
| Candida krusei | Candida lusitaniae |
| Candida parapsilosis | Candida guilliermondii |
The antifungal presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antifungal medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours for
### bioMérieux, Inc. 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 http://www.biomerieux-usa.com
{1}------------------------------------------------
2
yeast). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antifungal contained on the card.
VITEK 2 Yeast Voriconazole demonstrated substantially equivalent performance when compared with the CLSI reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Feb. 5, 2003.
The Premarket Notification (510(k)) presents data in support of VITEK® 2 Yeast Voriconazole. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 Yeast Voriconazole by comparing its performance with the CLSI reference method incubated at 24 and at 48 hrs. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK® 2 Yeast Voriconazole demonstrated acceptable performance of 99.2% overall Essential Agreement and 99.2% overall category agreement with the reference method incubated for 24 hrs. With the reference method incubated 48 hrs the overall Essential Agreement was 96.9% and overall category agreement was 98.7%. Reproducibility and Quality Control demonstrated acceptable results.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Biomerieux, Inc. c/o Nancy Weaver Associate Director, Regulatory Affairs 595 Anglum Rd. Hazelwood, Missouri 63042
MAY 2 1 2010
| Re: | k092454 |
|--------------------|--------------------------------------------------------------|
| Trade/Device Name: | Vitek ®2 Yeast Voriconazole |
| Regulation Number: | 21CFR §866.1640 |
| Regulation Name: | Antifungal susceptibility of Candida species to Voriconazole |
| Regulatory Class: | Class II |
| Product Code: | NGZ |
| Dated: | May 10, 2010 |
| Received: | May 11, 2010 |
Dear Ms. Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section
{3}------------------------------------------------
Page 2 - Nancy Weaver
510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Saly aath
Sally A. Hojvat, M.Sc., Ph.I Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):_ KO9 2 459
Device Name: VITEK 2 Yeast Voriconazole (0.12 - 8 µg/ml)
Indications For Use:
VITEK® 2 Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species. VITEK 2 Yeast Voriconazole is a quantitative test intended for use with the VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Voriconazole has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.
Active in vitro and in clinical infections Candida albicans Candida krusei Candida parapsilosis Candida tropicalis
Active in vitro but their clinical significance is uknown: Candida lusitaniae Candida guilliermondii
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD)
Freddie M. Cooke
Luvision Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) KD92454
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.