← Product Code [NGZ](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ) · K082534

# ETEST VORICONAZOLE 0.002-32 UG/ML (K082534)

_Ab Biomerieux · NGZ · Nov 2, 2009 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K082534

## Device Facts

- **Applicant:** Ab Biomerieux
- **Product Code:** [NGZ](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ.md)
- **Decision Date:** Nov 2, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

This submission is for Etest Voriconazole for MIC determinations across 0.002-32 µg/mL with C. albicans, C. glabrata, C. parapsilosis and C. tropicalis. Etest is an agar-based gradient technique for quantitative antifungal susceptibility testing of Candida species. It uses a predefined concentration gradient of the specific antifungal agent to determine the Minimum Inhibitory Concentration (MIC) in ug/mL inhibiting the growth of the test organism under defined conditions.

## Device Story

Etest® is an agar-based gradient strip for quantitative antifungal susceptibility testing. Input: isolated Candida colonies inoculated on agar plates. Operation: plastic strip with predefined exponential gradient of dried Voriconazole applied to inoculated agar; antifungal agent diffuses into agar creating concentration gradient; incubation at 35°C. Output: symmetrical inhibition ellipse intersecting strip at MIC value (µg/mL). Used in clinical laboratories; manual reading by technicians. Benefits: provides discrete MIC values for antifungal susceptibility, aiding clinical decision-making regarding treatment.

## Clinical Evidence

No clinical data provided. Performance is based on bench testing for MIC determination of Voriconazole against specified Candida species.

## Technological Characteristics

Thin, inert, non-porous plastic strip (5mm x 60mm). Contains predefined exponential gradient of dried antifungal agent. Principle: combination of dilution and diffusion. Manual reading of inhibition ellipse. No electronic components or software.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 5

# 10(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K082534

B. Purpose for Submission:
AST test for *Candida albicans*, *C. glabrata*, *C. krusei*, *C. parapsilosis* and *C. tropicalis*.

C. Measurand:
Voriconazole 0.002 – 32 µg/mL

D. Type of Test:
Quantitative AST growth based detection

E. Applicant:
AB bioMerieux (former AB BIODISK)

F. Proprietary and Established Names:
Etest® for Antimicrobial Susceptibility Testing

G. Regulatory Information:

1. Regulation section:
866.1640 Antimicrobial Susceptibility Test

2. Classification:
II

3. Product code:
NGZ

4. Panel:
83 Microbiology

H. Intended Use:

1. Intended use(s):
Etest® is an agar-based gradient technique for quantitative antifungal susceptibility testing of *Candida*. It uses a predefined concentration gradient of the specific antifungal agent to determine the Minimum Inhibitory Concentration (MIC) in ug/mL inhibiting the growth of the test organism under defined conditions.

{1}

2. Indication(s) for use:
This submission is for Etest® Voriconazole for MIC determination across 0.002 – 32 µg/mL with Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis.

3. Special conditions for use statement(s):
For prescription use

4. Special instrument requirements:
Manual readings only

I. Device Description:
Etest® consists of a thin, inert and non-porous plastic strip, 5mm wide and 60 mm long. One side of the strip carries a two-letter code designating the identity of the antifungal agent and is calibrated with MIC values in terms of ug/mL, a predefined and exponential gradient of the dried and established antifungal agent which covers a continuous concentration range across two-fold dilutions.

J. Substantial Equivalence Information:

1. Predicate device name(s):
Etest

2. Predicate 510(k) number(s):
K040560

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | Quantitative susceptibility to antifungal agents for Candida species | Same  |
|  Incubation | 35° | Same  |
|  Inoculation | Isolated colonies of Candida species | Same  |
|  Result | MIC | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Antibiotic | Voriconazole | Fluconazole
Itraconazole
Flucytosine  |

Page 2 of 5

{2}

Page 3 of 5

K. Standard/Guidance Document Referenced (if applicable):
“Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; Reference Method for Broth Dilution Antifungal Susceptibility of Yeasts: Third Informational Supplement (M27-S3)

L. Test Principle:
The Etest® gradient technology is based on a combination of the concepts of dilution and diffusion test methods for susceptibility testing. Etest® is processed like disc diffusion test (i.e. 0.5 McFarland suspension inoculum), but directly quantifies antimicrobial susceptibility in terms of discrete MIC values. When the Etest® strip containing the antibiotic is applied to an inoculated agar plate, the antifungal agent is immediately released from the plastic surface into the agar. A predefined, continuous gradient of antifungal concentrations is created and maintained directly underneath the strip. After incubation whereby growth becomes visible, a symmetrical inhibition ellipse centered along the strip will be seen. The inhibition zone edge intersects the strip of the MIC value in $\mu \mathrm{g} / \mathrm{mL}$ of the agent.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:
Precision/Reproducibility was determined by testing 26 Candida species one time at each of three sites. The mode was determined and then the reproducibility was calculated based on plus or minus one well of the mode. The reproducibility for 24 hours and 48 hours was $&gt;95\%$ (77/78).

b. Linearity/assay reportable range:
Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Quality control was performed each day of testing on both the reference method and the test method at four testing sites.

Voriconazole Quality Control

|  Organism | Concentration (μg/mL) | Reference | Etest®  |   |
| --- | --- | --- | --- | --- |
|   |   |  48 hr | 24 hr | 48 hr  |
|  C. krusei | 0.064 |  |  |   |
|  ATCC 6258 | 0.125 |  | 9 |   |
|  FDA/CLSI (48 hrs) | 0.25 | 42 | 46 | 25  |
|  0.12-1 μg/mL | 0.5 | 23 | 21 | 54  |
|   |  1 |  |  | 1  |

{3}

A 0.5 McFarland (1-5 x 10⁶ yeast cells per mL) is used to determine the correct inoculum. Colony counts were performed periodically at each site to demonstrate that the inoculum procedure results were in the expected CFU/ml.

*Candida krusei* ATCC 6258

The reference method and Etest were both within range. The mode is the same as that of the 48 hours reference method at 24 hours but one dilution higher at 48 hours.

The no growth rate was &lt;10%.

d. Detection limit:
Not Applicable

e. Analytical specificity:
Not Applicable

f. Assay cut-off:
Not Applicable

2. Comparison studies:

a. Method comparison with predicate device:

Clinical laboratory studies were performed at three or more clinical laboratories and compared to the CLSI broth micro-dilution method as described in M27 A2. Seventy four challenge strains were tested at one site and were included in the evaluation. Both tests were incubated in ambient air at 35°C. The reference method was read at 48 hours and the Etest® was read at 24 hours and confirmed at 48 hours. A two concentration allowance was used for this comparison due to the difficulty in reading the Essential Agreement (EA) for both the reference method and the Etest®. Category Agreement (CA) was determined based on the interpretation of each test.

Summary Table for Etest® 24 hr vs. Reference 48 hrs (≤1, 2, ≥4)

|   | Total | EA # | % EA (±two 2-fold dil) | CA | % CA | NS | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinical | 633 | 615 | 97.2 | 616 | 98.2 | 23 | 11 | 0 | 0  |
|  Challenge | 74 | 73 | 98.6 | 69 | 98.6 | 5 | 1 | 0 | 0  |
|  Combined | 707 | 688 | 97.3 | 685 | 98.3 | 28 | 12 | 0 | 0  |

Summary Table for Etest® 48 hr vs Reference 48 hrs (≤1, 2, ≥4)

|   | Total | EA # | % EA (±two 2-fold dil) | CA | % CA | NS | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Clinical | 633 | 613 | 96.8 | 608 | 97.1 | 23 | 15 | 3 | 0  |
|  Challenge | 74 | 74 | 100 | 70 | 100 | 5 | 0 | 0 | 0  |
|  Combined | 707 | 687 | 97.2 | 678 | 97.4 | 28 | 15 | 3 | 0  |

{4}

EA-Essential Agreement
CA- Category Agreement
NS- Non-susceptible

maj- major discrepancies
vmj- very major discrepancies
min- minor discrepancies

The three maj errors at 48 hours were due to C. glabrata, resulting in a maj error rate of 2.5% (Etest: 1; reference: 0.25, 0.5, to 1) and a combined overall EA of 93.2%, 91.8% CA. The overall maj error rate was 0.5% (3/610) which was acceptable. There were no maj or vmj at 24 hours.

b. Matrix comparison:
Not Applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable

4. Clinical cut-off:
Not Applicable

5. Expected values/Reference range:
$\leq 1, 2, \geq 4 \ \mu\mathrm{g}/\mathrm{mL}$

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

Page 5 of 5

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K082534](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K082534)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
