MicroScan Dried Gram-Negative MIC/Combo Panels with Meropenem (Mer) (0.004-32 µg/mL)
K201423 · Beckman Coulter, Inc. · LTT · Jun 29, 2020 · Microbiology
Device Facts
Record ID
K201423
Device Name
MicroScan Dried Gram-Negative MIC/Combo Panels with Meropenem (Mer) (0.004-32 µg/mL)
Applicant
Beckman Coulter, Inc.
Product Code
LTT · Microbiology
Decision Date
Jun 29, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
To determine antimicrobial agent susceptibility
Device Story
MicroScan Dried Gram-Negative MIC/Combo Panels utilize broth microdilution to determine minimum inhibitory concentrations (MIC) of meropenem for gram-negative bacteria. Panels contain dehydrated antimicrobial agents; clinical samples are inoculated into the panel using either Prompt or Turbidity methods. Following incubation, panels are read via automated systems (WalkAway, autoSCAN 4) or manual visual inspection. The device provides quantitative MIC results, which clinicians use to guide antibiotic therapy selection for patients with gram-negative infections. The modification adds Acinetobacter spp. to the existing indications for meropenem testing. The device is used in clinical microbiology laboratories by trained laboratory personnel.
Clinical Evidence
Bench testing only. Performance evaluated using fresh and stock efficacy and challenge isolates of Acinetobacter spp. compared to CLSI frozen reference panels. Results showed 91.1% Essential Agreement and 98.2% Categorical Agreement for Meropenem. Reproducibility and precision testing confirmed acceptable performance across both Prompt and Turbidity inoculum methods and both WalkAway and autoSCAN-4 instrument platforms. Quality control testing met acceptance criteria.
Technological Characteristics
Miniaturized broth microdilution panels; dehydrated antimicrobial agents. Inoculation via standardized suspension (Turbidity or Prompt method). Incubation 16-20 hours at non-CO2 conditions. Readout via automated instrumentation (WalkAway or autoSCAN-4).
Indications for Use
Indicated for determining quantitative and/or qualitative antimicrobial agent susceptibility of rapidly growing aerobic and facultative anaerobic gram-negative bacilli, including Acinetobacter spp., using colonies grown on solid media.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K172255 — MicroScan Dried Gram Negative MIC/Combo Panels with Ceftolozane/tazobactam (0.25/4 - 16/4 ug/mL) · Beckman Coulter, Inc. · Dec 1, 2017
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K201423
B Applicant
Beckman Coulter, Inc.
C Proprietary and Established Names
MicroScan Dried Gram-Negative MIC/Combo Panels with Meropenem (Mer) (0.004-32 µg/mL)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LTT, JWY, LRG, LTW | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
## II Review Summary:
This 510(k) submission contains information/data on modifications made to the submitter's own Class II device requiring 510(k). The following items are present and acceptable:
1. The name and 510(k) number of the SUBMITTER'S previously cleared device.
2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling.
3. A description of the device MODIFICATION was provided in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for the addition of Acinetobacter spp. as an indicated species for testing with meropenem.
4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.
Table 1. Performance of MicroScan Dried Gram Negative MIC/Combo Panel with Meropenem for Acinetobacter spp.
| Inoculation Method | Reading Method | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Prompt | WalkAway | 56 | 51 | 91.1 | 36 | 31 | 86.1 | 55 | 98.2 | 37 | 19 | 1 | 0 | 0 |
| | autoSCAN 4 | 56 | 52 | 92.9 | 37 | 33 | 89.2 | 55 | 98.2 | 37 | 19 | 1 | 0 | 0 |
| | Manual | 56 | 52 | 92.9 | 36 | 32 | 88.9 | 55 | 98.2 | 37 | 19 | 1 | 0 | 0 |
| Turbidity | WalkAway | 56 | 54 | 96.4 | 38 | 35 | 92.1 | 52 | 92.9 | 37 | 18 | 4 | 0 | 0 |
| | autoSCAN 4 | 56 | 54 | 96.4 | 37 | 35 | 94.6 | 52 | 92.9 | 37 | 19 | 4 | 0 | 0 |
| | Manual | 56 | 54 | 96.4 | 37 | 35 | 94.6 | 51 | 91.1 | 37 | 19 | 5 | 0 | 0 |
Table 2. Observed Trending for Acinetobacter spp. with Meropenem
| Inoculation Method | Read Method | Total Evaluable for Trending | ≥1 Dilution lower No. (%) | Exact No. (%) | ≥1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Prompt | WalkAway | 37 | 10 (27.0) | 17 (46.0) | 10 (27.0) | 0 | No |
| | autoSCAN4 | 40 | 10 (25.0) | 20 (50.0) | 10 (25.0) | 0 | No |
| | Manual | 37 | 11 (29.7) | 20 (54.0) | 6 (16.2) | -13.5 | No |
| Turbidity | WalkAway | 42 | 17 (40.5) | 17 (40.5) | 8 (19.1) | -21.4 | No |
| | autoSCAN4 | 42 | 19 (45.2) | 16 (38.1) | 7 (16.7) | -28.6 | No |
| | Manual | 41 | 19 (46.3) | 18 (43.9) | 4 (9.8) | -36.6 | Yes |
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission referred to the breakpoint change protocol that was reviewed and accepted by FDA in K192355. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that Beckman Coulter intends to use to evaluate the MicroScan Dried Gram-Negative MIC/Combo Panels with Meropenem (Mer) (0.004 - 32 µg/mL) when revised breakpoints for meropenem are published on the FDA STIC webpage. The breakpoint change protocol indicated that if specific criteria are met, Beckman Coulter will update the meropenem device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K201423 - Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.