← Product Code [LRG](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LRG) · K191918

# Thermo Scientific Sensititre ARIS HiQ System (K191918)

_Thermo Fisher Scientific · LRG · Nov 6, 2019 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LRG/K191918

## Device Facts

- **Applicant:** Thermo Fisher Scientific
- **Product Code:** [LRG](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LRG.md)
- **Decision Date:** Nov 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Thermo Scientific Sensititre ARIS HiQ System is part of the Sensititre AST system and is an automated plate management device containing an incubator and embedded OptiRead module. The Thermo Scientific Sensititre ARIS HiQ System is designed for use with the Thermo Scientific Sensititre SWIN Software System. The ARIS HiQ and the SWIN system work together to read Sensititre (18-24 hr) susceptibility plates, generating minimum inhibitory concentration (MIC) and interpreting breakpoint (BP) results for non-fastidious and fastidious microorganisms.

## Device Story

The Sensititre ARIS HiQ System is an automated plate management device for antimicrobial susceptibility testing (AST). It functions as an incubator and reader for Sensititre susceptibility plates. The device integrates with the Sensititre SWIN Software System. The system processes non-fastidious microorganisms on 18-24 hour plates. The embedded OptiRead module performs automated reading of the plates. The system generates MIC values and interprets breakpoint results. It is intended for use in clinical laboratory settings by trained laboratory personnel. The output is used by clinicians to guide antimicrobial therapy decisions for patients with bacterial infections.

## Clinical Evidence

No clinical studies performed. Evidence consists of analytical performance data: reproducibility study (25 isolates, 3 sites, 3 days) and method comparison study (75 challenge isolates of fastidious Gram-positive cocci). Method comparison against legacy ARIS/OptiRead showed >90% Essential Agreement (EA) and Category Agreement (CA) for all antimicrobial/organism combinations. QC testing with S. pneumoniae ATCC 49619 confirmed performance within expected ranges >95% of the time.

## Technological Characteristics

Automated plate management system; includes incubator and embedded OptiRead module; designed for use with Sensititre SWIN software; processes 18-24 hour susceptibility plates for non-fastidious microorganisms.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT

## I Background Information:

A 510(k) Number
K191918

B Applicant
Thermo Fisher Scientific

C Proprietary and Established Names
Thermo Scientific Sensititre ARIS HiQ System

D Regulatory Information
|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LRG | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology  |

## II Submission/Device Overview:

### A Purpose for Submission:
To demonstrate substantial equivalence of the Thermo Scientific Sensititre ARIS HiQ System to the predicate Sensititre ARIS Module, used in conjunction with the OptiRead, to read Sensititre (18-24 hour) susceptibility plates for fastidious Gram-positive microorganisms.

### B Measurand:
Used in conjunction with the embedded OptiRead module and Sensititre SWIN Software System to automatically read Sensititre MIC and Breakpoint (BP) Susceptibility test panels and interpret the antimicrobial susceptibility test results for fastidious Gram-positive microorganisms.

K191918 - Page 1 of 12

{1}

C Type of Test:
Automated plate management system containing an incubator and embedded OptiRead module, which is an automated fluorescence-based detection instrument used to read Sensititre Susceptibility plates.

III Intended Use/Indications for Use:

A Intended Use(s):
See Indications for Use below.

B Indication(s) for Use:
The Thermo Scientific Sensititre ARIS HiQ System is part of the Sensititre AST system and is an automated plate management device containing an incubator and embedded OptiRead module. The Thermo Scientific Sensititre ARIS HiQ System is designed for use with the Thermo Scientific Sensititre SWIN Software System. The ARIS HiQ and the SWIN system work together to read Sensititre (18-24 hr) susceptibility plates, generating minimum inhibitory concentration (MIC) and interpreting breakpoint (BP) results for non-fastidious and fastidious microorganisms.

C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only

D Special Instrument Requirements:
The Thermo Scientific Sensititre ARIS HiQ System can only be used with version 3.4 of the Thermo Scientific Sensititre SWIN Software System.

IV Device/System Characteristics:

A Device Description:
The Thermo Scientific Sensititre ARIS HiQ System is an automated plate management device containing an incubator and an embedded OptiRead reader. In its current version, it is designed for use with Sensititre SWIN Software System (version 3.4). The ARIS HiQ system handles plate incubation and management. The ARIS HiQ and SWIN Software System work together to read and interpret up to 100 Sensititre plates, generating minimum inhibitory concentration (MIC) results and corresponding categorical interpretation based on breakpoints (BP) information within the SWIN software for non-fastidious Gram-negative, non-fastidious Gram-positive, and fastidious Gram-positive microorganisms.

B Principle of Operation:
The Thermo Scientific ARIS HiQ System is for use with Sensititre (18-24 hr) Susceptibility plates which are micro-versions of the classic broth dilution method for providing qualitative and quantitative susceptibility results in a dried plate format. Each plate (panel) contains a fluorogenic substrate to monitor growth and antimicrobial agents at specific two-fold dilutions to achieve drug concentrations over a specified range. When microbial metabolism and growth occurs, fluorescence is released from plate wells in amounts directly related to the activity of

K191918 - Page 2 of 12

{2}

bacterial growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism.

To use the ARIS HiQ, plates are inoculated with a standardized suspension of microorganism using the Sensititre AutoInoculator/AIM instrument, covered, and then loaded onto the ARIS HiQ for automated incubation and reading. After the incubation time indicated for each susceptibility panel, the microtiter plates are automatically moved by the robotic arm component of the ARIS HiQ to the embedded OptiRead module where the fluorescence count values are measured from each plate well. The fluorescence count values are transmitted to the SWIN software on an external PC for processing and interpretation of MIC and BP results. No result processing is performed within the ARIS HiQ.

C Instrument Description Information:

|  Modes of Operation | Yes | No  |
| --- | --- | --- |
|  Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐  |
|  Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑  |
|  Software  |   |   |
|  FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐  |

1. Instrument Name:

Thermo Scientific Sensititre ARIS HiQ System

2. Specimen Identification:

Specimen (Sensititre plate) identification information should be manually entered into the SWIN Software, followed by scanning the plate barcode. If plates have not yet been loaded in the ARIS HiQ, plate identification information can be sent from the SWIN to the ARIS HiQ. If plates have been loaded in the ARIS HiQ, the SWIN Software will automatically send plate identification information to the ARIS HiQ.

3. Specimen Sampling and Handling:

Sensititre plates inoculated with pure cultures can be loaded and unloaded from the ARIS HiQ in one of two ways: (1) individually, interacting with a rack mounted to the turntable at the front of the ARIS HiQ, or (2) in batches, using transportable microplate racks to simultaneously load, unload and handle up to ten microplates at a time.

4. Calibration:

All calibrations and adjustments to the ARIS HiQ System are performed by Sensititre system-trained service engineers during installation and regular servicing. No adjustments or calibrations are required by the end user during day-to-day operations.

K191918 - Page 3 of 12

{3}

5. Quality Control:
QC organisms were tested as appropriate for each drug and in accordance with CLSI recommendations. For results, see Section VII.A.5.

V Substantial Equivalence Information:

A Predicate Device Name(s):
Sensititre Aris Module

B Predicate 510(k) Number(s):
K911419

C Comparison with Predicate(s):

|  Similarities  |   |   |
| --- | --- | --- |
|  Device & Predicate Device(s): | ARIS HiQ
K191918
(Device) | Sensititre ARIS Module
K911419
(Predicate)  |
|  Intended Use | Automated plate handling and incubator instrument, used in conjunction with a plate reader and external software to read read Sensititre (18-24 hour) susceptibility plates, generating minimum inhibitory concentration (MIC) and interpreting breakpoint (BP) results. | Same  |
|  Plate Compatibility | FDA cleared Sensititre susceptibility plates designed for automated reading | Same  |
|  Barcode Reading | • Linear barcode reading capability on microtiter plates
• The entire plate inventory is read when plates are loaded/unloaded | Same  |
|  Reader Unit | OptiRead | Same  |
|  Incubation Temperature | 35°C ±1°C | Same  |
|  System Control & Software | Embedded processor running firmware | Same  |
|  Results Processing | Handled by established algorithms in SWIN system software | Same  |

K191918 - Page 4 of 12

{4}

|  Differences  |   |   |
| --- | --- | --- |
|  Device & Predicate Device(s): | ARIS HiQ
K191918
(Device) | Sensititre ARIS Module
K911419
(Predicate)  |
|  Plate Loading, Unloading & Handling | Plates are loaded, unloaded and handled in one of two ways:
(1) Individually, on a plate rack mounted to a turntable;
(2) In batches, using transportable plate racks to simultaneously load, unload and handle up to ten plates at a time. | Plates are loaded, unloaded, and handled individually on a fixed plate carousel.  |
|  Incubation Control | Five heater chimneys with a fan for each chimney to provide heated air circulation to the microplate racks. | Heated microplate carousel and a single heater chimney (no fan).  |
|  Reader Unit Implementation | OptiRead is fully integrated into the ARIS HiQ. | OptiRead is a separate instrument which plugs into the Sensititre ARIS module beneath the instrument.  |
|  User Interface | Touch screen display with icon driven graphical user interface. | Seven segment display with keypad.  |
|  SWIN interface/communication | USB | RS-232  |
|  Plate Capacity | 100 | 64  |
|  Automated Plate Organization | Shuffle function to automatically relocate plates within the instrument during incubation and to consolidate plates for removal | No automated plate relocation function  |

VI Standards/Guidance Documents Referenced:

- Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA, August 2009.
- CLSI M100-28th ed., Performance Standard for Antimicrobial Susceptibility Testing; Approved Standard, Twenty-eighth Edition, January 2018.
- CLSI M07, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard, Tenth Edition, January 2015.

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

K191918 - Page 5 of 12

{5}

A reproducibility study with fastidious Gram-positive cocci was conducted using 25 isolates of Streptococcus spp.: Streptococcus pneumoniae (ten isolates), Viridans Group Streptococcus spp. (five isolates), Streptococcus pyogenes (six isolates), and Streptococcus agalactiae (four isolates). Inocula were prepared using the Sensititre AutoInoculator.

The isolates were tested against the following antimicrobial agents: Amoxicillin/ Clavulanic acid, Cefepime, Cefotaxime, Ceftaroline, Chloramphenicol, Clindamycin, Daptomycin, Dalbavancin, Erythromycin, Levofloxacin, Linezolid, Meropenem, Penicillin, Trimethoprim/Sulfamethoxazole, Tedizolid, Telavancin, Tetracycline, and Vancomycin. The antimicrobial agents were selected as representatives of the major antimicrobial classes.

Testing was performed at one site over three days by three operators using Sensititre susceptibility plates on three separate ARIS HiQ instruments. Percent reproducibility was calculated as the number of isolates with MIC values falling within +/- one doubling dilution from the mode MIC value. Most of data points in the study were on-scale and within +/- one doubling dilution agreement as compared to the mode MIC.

In summary, the percent reproducibility for all antimicrobial/microorganism combinations tested met the acceptance criteria of greater than 95% for "best-case" as described in the AST Special Controls Guidance Document or "worst-case" calculations. The reproducibility results were acceptable.

2. Linearity:
Not applicable

3. Analytical Specificity/Interference:
Not applicable

4. Assay Reportable Range:
Not applicable

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

## Quality Control (QC) Testing

The FDA/CLSI recommended QC strain for Gram-positive organisms (S. pneumoniae ATCC 49619) was tested using Sensititre susceptibility plates at two sites. Inocula were prepared using the Sensititre AutoInoculator. The QC organism was tested a minimum of 20 times per site for each antimicrobial using both the ARIS HiQ and the legacy ARIS (ARIS 2X, used with the OptiRead instrument). All QC results were within the expected range &gt;95% of the time. Overall, QC performance was acceptable. Results are summarized in Table 1. Shading in the table indicate the expected concentration range.

K191918 - Page 6 of 12

{6}

Table 1. Quality Control Results Summary

|  QC Organism
S. pneumoniae ATCC 49619 | Conc.
(μg/mL) | ARIS HiQ | Legacy ARIS  |
| --- | --- | --- | --- |
|  Amoxicillin/ Clavulanic acid
Expected Range
0.03/0.015-0.12/0.06 μg/mL | 0.015/0.008 |  |   |
|   |  0.03/0.015 |  |   |
|   |  0.06/0.03 | 31 | 36  |
|   |  0.12/0.06 | 8 | 3  |
|   |  0.25/0.12 |  |   |
|  Cefepime
Expected Range
0.03-0.25 μg/mL | 0.015 |  |   |
|   |  0.03 |  | 1  |
|   |  0.06 | 29 | 32  |
|   |  0.12 | 10 | 6  |
|   |  0.25 |  |   |
|   |  0.5 |  |   |
|  Cefotaxime
Expected Range
0.03-0.12 μg/mL | 0.015 |  |   |
|   |  0.03 |  |   |
|   |  0.06 | 37 | 38  |
|   |  0.12 | 2 | 1  |
|   |  0.25 |  |   |
|  Chloramphenicol
Expected Range
2-8 μg/mL | 1 |  |   |
|   |  2 | 2 | 4  |
|   |  4 | 37 | 35  |
|   |  8 |  |   |
|   |  >8 |  |   |
|  Clindamycin
Expected Range
0.03-0.12 μg/mL | 0.015 |  |   |
|   |  0.03 | 3 | 3  |
|   |  0.06 | 36 | 34  |
|   |  0.12 |  | 2  |
|   |  0.25 |  |   |
|  Daptomycin
Expected Range
0.06-0.5 μg/mL | 0.03 |  |   |
|   |  0.06 |  |   |
|   |  0.12 | 12 | 16  |
|   |  0.25 | 23 | 17  |
|   |  0.5 | 2 | 1  |
|   |  1 | 2 | 5  |
|  Erythromycin
Expected Range
0.03-0.12 μg/mL | 0.015 |  | 1  |
|   |  0.03 | 30 | 28  |
|   |  0.06 | 9 | 10  |
|   |  0.12 |  |   |
|   |  0.25 |  |   |
|  Levofloxacin
Expected Range
0.5-2 μg/mL | 0.25 |  |   |
|   |  0.5 |  |   |
|   |  1 | 39 | 39  |
|   |  2 |  |   |
|   |  4 |  |   |
|  Linezolid
Expected Range
0.5-2 μg/mL | 0.25 |  |   |
|   |  0.5 |  | 1  |
|   |  1 | 39 | 36  |
|   |  2 |  | 2  |
|   |  >2 |  |   |
|   | 0.03 |  | 1  |

K191918 - Page 7 of 12

{7}

|  QC Organism
S. pneumoniae ATCC 49619 | Conc.
(μg/mL) | ARIS HiQ | Legacy ARIS  |
| --- | --- | --- | --- |
|  Meropenem
Expected Range
0.06-0.25 μg/mL | 0.06 | 39 | 36  |
|   |  0.12 |  | 2  |
|   |  0.25 |  |   |
|   |  0.5 |  |   |
|  Penicillin
Expected Range
0.25-1 μg/mL | 0.12 |  |   |
|   |  0.25 | 15 | 14  |
|   |  0.5 | 24 | 25  |
|   |  1 |  |   |
|   |  2 |  |   |
|  Tetracycline
Expected Range
0.12-0.5 μg/mL | 0.06 |  |   |
|   |  0.12 | 25 | 24  |
|   |  0.25 | 13 | 13  |
|   |  0.5 |  |   |
|   |  ≥1 | 1 | 2  |
|  Trimethoprim/
sulfamethoxazole
Expected Range
0.12/2.38-1/19 μg/mL | 0.06/1.19 |  |   |
|   |  0.12/2.38 | 1 | 2  |
|   |  0.25/4.75 | 37 | 36  |
|   |  0.5/9.5 | 1 | 1  |
|   |  1/19 |  |   |
|   |  2/38 |  |   |
|  Vancomycin
Expected Range
0.12-0.5 μg/mL | 0.06 |  |   |
|   |  0.12 |  | 2  |
|   |  0.25 | 38 | 36  |
|   |  0.5 | 1 |   |
|   |  1 |  | 1  |
|  Ceftaroline
Expected Range
0.008-0.03 μg/mL | 0.004 |  |   |
|   |  0.008 |  |   |
|   |  0.015 | 14 | 13  |
|   |  0.03 | 24 | 26  |
|   |  0.06 | 1 |   |
|  Dalbavancin
Expected Range
0.008-0.03 μg/mL | 0.004 |  |   |
|   |  0.008 | 10 | 10  |
|   |  0.015 | 22 | 27  |
|   |  0.03 | 7 | 2  |
|   |  0.06 |  |   |
|  Telavancin
Expected Range
0.004-0.015 μg/mL | 0.002 |  |   |
|   |  0.004 | 4 | 3  |
|   |  0.008 | 19 | 25  |
|   |  0.015 | 14 | 8  |
|   |  0.03 | 2 | 3  |
|  Tedizolid
Expected Range
0.12-0.5μg/mL | 0.06 |  | 1  |
|   |  0.12 | 13 | 13  |
|   |  0.25 | 25 | 23  |
|   |  0.5 |  |   |
|   |  ≥1 | 1 | 2  |

6. Detection Limit:

Not applicable

K191918 - Page 8 of 12

{8}

7. Assay Cut-Off:

Not applicable

8. Accuracy (Instrument):

Not applicable

9. Carry-Over:

Not applicable

## B Comparison Studies:

1. Method Comparison with Predicate Device:

The performance of the ARIS HiQ was evaluated in a method comparison study conducted at one internal site. The study compared the performance of the ARIS HiQ to legacy ARIS and OptiRead instruments using Sensititre (18-24 hour) susceptibility plates. Microorganism suspensions were prepared using the Sensititre Nephelometer, and each Sensititre susceptibility plate was inoculated with the prepared microorganism suspension using the Sensititre AutoInoculator. A purity check and colony count check of all organisms was performed on the final inocula. Only those cultures that were pure were evaluated.

The performance criteria described in the "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems" were used to evaluate the performance of the ARIS HiQ. Performance evaluations used current FDA-recognized susceptibility interpretative criteria (STIC) as described on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/antibacterial-susceptibility-test-interpretive-criteria).

Testing was conducted on 75 challenge isolates of fastidious Gram-positive cocci. These included Streptococcus pneumoniae (25), Viridans Group Streptococcus spp. (18), and the following β-hemolytic Group Streptococcus spp.: Streptococcus agalactiae (11), Streptococcus dysgalactiae (7), and Streptococcus pyogenes (14). One isolate each of S. dysgalactiae and S. pneumoniae failed to grow for a growth rate of 97.3% (73/75).

The EA and CA were greater than 90% for all microorganism group/antimicrobial combinations. The overall performance was acceptable for all antimicrobials tested. Results are summarized in Table 2 below.

Table 2. Fastidious Challenge Isolate Performance (ARIS HiQ vs. Legacy ARIS/OptiRead)

|  Antimicrobial | Organism Group | Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R or NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Amoxicillin/clavulanic acid | S. pneumoniae | 24 | 24 | 100 | 22 | 22 | 100 | 23 | 95.8 | 2 | 1 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Cefepime | S. pneumoniae | 24 | 24 | 100 | 24 | 24 | 100 | 23 | 95.8 | 1 | 1 | 0 | 0  |

K191918 - Page 9 of 12

{9}

|  Antimicrobial | Organism Group | Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R or NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 30 | 30 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Total | 55 | 55 | 100 | 54 | 54 | 100 | 54 | 98.2 | 1 | 1 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Cefotaxime | S. pneumoniae | 24 | 24 | 100 | 23 | 23 | 100 | 22 | 91.7 | 4 | 2 | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 18 | 18 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 15 | 15 | 100 | 17 | 94.4 | 3 | 1 | 0 | 0  |
|   |  Total | 73 | 73 | 100 | 56 | 56 | 100 | 70 | 95.9 | 7 | 3 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Ceftaroline | S. pneumoniae | 24 | 24 | 100 | 24 | 24 | 100 | 23 | 95.8 | 1 | N/A | 1* | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 25 | 25 | 100 | 25 | 25 | 100 | 25 | 100 | 0 | N/A | 0 | 0  |
|   |  Total | 49 | 49 | 100 | 49 | 49 | 100 | 48 | 98.0 | 1 | N/A | 1* | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Chloramphenicol | S. pneumoniae | 24 | 24 | 100 | 12 | 12 | 100 | 24 | 100 | 14 | N/A | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 31 | 31 | 100 | 31 | 100 | 0 | 0 | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 18 | 18 | 100 | 18 | 100 | 2 | 0 | 0 | 0  |
|   |  Total | 73 | 73 | 100 | 61 | 61 | 100 | 73 | 100 | 16 | 0 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clindamycin | S. pneumoniae | 24 | 24 | 100 | 14 | 14 | 100 | 24 | 100 | 10 | 0 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Daptomycin | β-hemolytic Group Streptococcus spp. | 31 | 30 | 96.8 | 31 | 30 | 96.8 | 31 | 100 | 0 | N/A | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Dalbavancin | β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 31 | 31 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 16 | 16 | 100 | 18 | 100 | 2 | N/A | 0 | 0  |
|   |  Total | 49 | 49 | 100 | 47 | 47 | 100 | 49 | 100 | 2 | N/A | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Erythromycin | S. pneumoniae | 24 | 24 | 100 | 8 | 8 | 100 | 24 | 100 | 15 | 0 | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 31 | 30 | 96.8 | 26 | 25 | 96.2 | 31 | 100 | 5 | 0 | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 7 | 7 | 100 | 18 | 100 | 9 | 0 | 0 | 0  |
|   |  Total | 73 | 72 | 98.6 | 41 | 40 | 97.6 | 73 | 100 | 29 | 0 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Levofloxacin | S. pneumoniae | 24 | 24 | 100 | 24 | 24 | 100 | 24 | 100 | 0 | 0 | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 27 | 27 | 100 | 31 | 100 | 1 | 0 | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 18 | 18 | 100 | 18 | 100 | 0 | 0 | 0 | 0  |
|   |  Total | 73 | 73 | 100 | 69 | 69 | 100 | 73 | 100 | 1 | 0 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Linezolid | S. pneumoniae | 24 | 24 | 100 | 21 | 21 | 100 | 24 | 100 | 3 | N/A | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 31 | 31 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Total | 55 | 55 | 100 | 52 | 52 | 100 | 55 | 100 | 3 | N/A | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Meropenem | S. pneumoniae | 24 | 24 | 100 | 17 | 17 | 100 | 24 | 100 | 18 | N/A | 0 | 0  |

K191918 - Page 10 of 12

{10}

|  Antimicrobial | Organism Group | Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R or NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | S. pyogenes | 14 | 14 | 100 | 2 | 2 | 100 | 14 | 100 | 0 | N/A | 0 | 0  |
|   |  Total | 38 | 38 | 100 | 19 | 19 | 100 | 38 | 100 | 18 | N/A | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Penicillin | S. pneumoniae | 24 | 24 | 100 | 19 | 19 | 100 | 24 | 100 | 16 | 0 | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 14 | 14 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 13 | 13 | 100 | 18 | 100 | 5 | 0 | 0 | 0  |
|   |  Total | 73 | 73 | 100 | 46 | 46 | 100 | 73 | 100 | 21 | 0 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Trimethoprim/sulfamethoxazole | S. pneumoniae | 24 | 24 | 100 | 4 | 4 | 100 | 23 | 95.8 | 20 | 1 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Tedizolid | β-hemolytic Group Streptococcus spp. | 31 | 31 | 100 | 31 | 31 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 16 | 16 | 100 | 18 | 100 | 2 | N/A | 0 | 0  |
|   |  Total | 49 | 49 | 100 | 47 | 47 | 100 | 49 | 100 | 2 | N/A | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Telavancin | β-hemolytic Group Streptococcus spp | 31 | 31 | 100 | 31 | 31 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 16 | 16 | 100 | 18 | 100 | 2 | N/A | 0 | 0  |
|   |  Total | 49 | 49 | 100 | 47 | 47 | 100 | 49 | 100 | 2 | N/A | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Tetracycline | S. pneumoniae | 24 | 24 | 100 | 10 | 10 | 100 | 24 | 100 | 14 | 0 | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp | 31 | 30 | 96.8 | 18 | 17 | 94.4 | 31 | 100 | 13 | 0 | 0 | 0  |
|   |  Viridans Group Streptococcus spp. | 18 | 18 | 100 | 11 | 11 | 100 | 17 | 94.4 | 7 | 1 | 0 | 0  |
|   |  Total | 73 | 72 | 98.6 | 39 | 38 | 97.4 | 72 | 98.6 | 34 | 1 | 0 | 0  |
|  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Vancomycin | S. pneumoniae | 24 | 24 | 100 | 21 | 21 | 100 | 24 | 100 | 3 | N/A | 0 | 0  |
|   |  β-hemolytic Group Streptococcus spp | 31 | 31 | 100 | 31 | 31 | 100 | 31 | 100 | 0 | N/A | 0 | 0  |
|   |  Total | 55 | 55 | 100 | 52 | 52 | 100 | 55 | 100 | 3 | N/A | 0 | 0  |

a. A single categorical very major error was observed for ceftaroline when testing S. pneumoniae (1/24). Based on the essential agreement and lack of an intermediate breakpoint for ceftaroline, the overall adjusted very major error rate for S. pneumoniae isolates was 0% (0/24).
N/A - Not applicable, as there are no intermediate breakpoints.

EA – Essential Agreement
Min – minor discrepancies
CA – Category Agreement
Maj – major discrepancies
Eval – Evaluable isolates
Vmj – very major discrepancies
R – Resistant isolates
NS - Non-susceptible isolates

Essential agreement (EA) is when the ARIS HiQ result is within plus or minus one doubling dilution of the legacy ARIS result. Category Agreement (CA) is when the ARIS HiQ result interpretation agrees exactly with the legacy ARIS result interpretation. Evaluable results are those that are on-scale for both the ARIS HiQ and legacy ARIS instruments.

K191918 - Page 11 of 12

{11}

2. Matrix Comparison:
Not applicable

C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
See comparison studies in Section VII.B above.

D Clinical Cut-Off:
Not applicable

E Expected Values/Reference Range:
Not applicable

F Other Supportive Instrument Performance Characteristics Data:
Not applicable

VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K191918 - Page 12 of 12

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LRG/K191918](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LRG/K191918)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
