MICROSCAN MICROSTREP PLUS PANEL TETRACYCLINE (0.006 TO 16 MCG/ML)
K062923 · Dade Behring, Inc. · LRG · Oct 13, 2006 · Microbiology
Device Facts
Record ID
K062923
Device Name
MICROSCAN MICROSTREP PLUS PANEL TETRACYCLINE (0.006 TO 16 MCG/ML)
Applicant
Dade Behring, Inc.
Product Code
LRG · Microbiology
Decision Date
Oct 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
To determine bacterial antimicrobial agent susceptibility
Device Story
MicroScan MICroSTREP plus Panel is a miniaturized broth dilution susceptibility test system. It uses panels rehydrated with Mueller-Hinton broth, 2-5% lysed horse blood, and 50 mM HEPES buffer. After inoculation with standardized organism suspensions, panels are incubated for 20-24 hours at 35°C in a non-CO2 incubator. Results are determined by visual inspection or by using the MicroScan WalkAway instrument. The instrument reads the lowest antimicrobial concentration (MIC) that inhibits bacterial growth. This submission specifically adds instrument-read capability for Tetracycline (0.06 to 16 mcg/ml). Healthcare providers use these MIC results to guide antimicrobial therapy decisions for patients with streptococcal infections. The device is intended for professional use in clinical laboratory settings.
Clinical Evidence
Bench testing only. Reproducibility study (10 isolates, 3 sites, 3 days) showed >95% reproducibility for both manual and automated methods. Comparison study of 70 streptococcal challenge isolates (including 53 S. pneumoniae) demonstrated 97.1% Essential Agreement (EA) for automated reading vs. reference broth dilution. No significant trending or performance differences observed between manual and automated reading methods.
Technological Characteristics
Miniaturized broth dilution susceptibility test. Components: panels rehydrated with Mueller-Hinton broth, 2-5% lysed horse blood (LHB), and 50 mM HEPES buffer. Incubation: 20-24 hours at 35°C +/- 1°C in non-CO2 incubator. Readout: visual or automated via MicroScan WalkAway instrument. Connectivity: integrated with WalkAway instrument system.
Indications for Use
Indicated for determining quantitative and/or qualitative antimicrobial susceptibility of aerobic streptococci, including Streptococcus pneumoniae, grown on solid media. Tetracycline testing is indicated for Streptococcus spp. and Streptococcus pneumoniae.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K091949 — MICROSCAN MICROSTREP PLUS, MODEL B1027-201 · Siemens Healthcare Diagnostics · Jan 26, 2010
K062773 — MICROSCAN MICROSTREP PLUS PANEL, PENICILLIN · Dade Behring, Inc. · Oct 13, 2006
K062375 — MICROSCAN MICROSTREP PLUS PANEL ERYTHROMYCIN (0.015 TO 2 MCG/ML) · Dade Behring, Inc. · Oct 26, 2006
K063053 — MICROSCAN MICROSTREP PLUS PANEL CEFUROXIME (0.12-8 MCG/ML) · Dade Behring, Inc. · Oct 26, 2006
K062585 — MICROSCAN MICROSTREP PLUS PANEL AZITHROMYCIN · Dade Behring, Inc. · Sep 19, 2006
Submission Summary (Full Text)
{0}
Page 1 of 6
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K062923
B. Purpose for Submission:
To add the option for automated reading of Tetracycline at 0.06 – 16 µg/mL to the MICroSTREP plus® Panel on the MicroScan® WalkAway System
C. Measurand:
Tetracycline at 0.06 – 16 µg/mL
D. Type of Test:
Quantitative and Qualitative growth based detection algorithm using optics light detection
E. Applicant:
Dade Behring Inc,
MicroScan®
F. Proprietary and Established Names:
MicroScan® MICroSTREP plus® Panel – Tetracycline at 0.06 – 16 µg/mL
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1640 – Antimicrobial Susceptibility Test Powder
2. Classification:
Class II
3. Product Code:
LRG – Instrument for Auto Reader & Interpretation of Overnight
Antimicrobial Susceptibility System
LTT – Panels, Test, Susceptibility, Antimicrobial
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
Tetracycline at 0.06 – 16 µg/mL is for use with MICroSTREP plus® Panels
MICroSTREP plus® Panels are designed for use in determining quantitative
{1}
Page 2 of 6
and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including *Streptococcus pneumoniae*.
2. **Indication(s) for use:**
This submission is for adding the option for automated reading of the Tetracycline at concentrations of 0.06 – 16 µg/mL to MICroSTREP plus® Panels on the MicroScan® WalkAway System.
3. **Special condition for use statement(s):**
Prescription Use Only
Turbidity method of inoculum preparation only
4. **Special instrument Requirements:**
Not Applicable
I. **Device Description:**
The MicroScan® MICroSTREP plus® Panel is a 96-well plastic dish which contains microdilutions of each antimicrobic in various concentrations dried in aqueous solutions. The panel is rehydrated and inoculated at the same time with a Mueller-Hinton broth supplemented with lysed horse blood (2 – 5%). The target inoculum concentration for each well should be approximately 5 × 10⁵ colony forming units (CFU)/mL. Panels are incubated in a 35°C non-CO₂ incubator for 20-24 hours. After incubation, the panels are read manually for growth. Additionally, panels may be incubated in and read by a MicroScan® WalkAway instrument. Each panel contains a “growth” but it does not contain a “no growth” control well.
J. **Substantial Equivalence Information:**
1. **Predicate device name(s):**
MICroSTREP plus®
2. **Predicate K number(s):**
K020939
3. **Comparison with predicate:**
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Determination of susceptibility to antimicrobics with aerobic streptococci including *Streptococcus pneumoniae* | Same |
| Isolates | For use with aerobic streptococci including *Streptococcus* isolated colonies from culture | Same |
| Results | Quantitative with qualitative interpretations | Same |
{2}
| Incubation | 20 – 24 hours | | Same |
| --- | --- | --- | --- |
| Panels | Tetracycline dried in aqueous solution | | Same |
| Differences | | | |
| Item | | Device | Predicate |
| Technology | | Growth based using algorithm with optics light detection | Growth based |
| Reading | | Overnight method Manual or automated | Overnight method Manual read only |
| Instrument | | MicroScan® WalkAway System or Microdilution viewer | Microdilution viewer |
# K. Standard/Guidance Document Referenced (if applicable):
"Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA"; CLSI M7 (M100-S16) "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard."
# L. Test Principle:
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with $115~\mu \mathrm{L}$ Mueller-Hinton broth supplemented with $2 - 5\%$ lysed horse blood (LHB), after inoculation of the broth with a standardized suspension of the organism. The target inoculum concentration for each well should be approximately $5\times 10^{5}$ colony forming units (CFU)/mL. After incubation in a non- $\mathrm{CO}_{2}$ incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organisms is determined by observing the lowest antimicrobial concentration showing inhibition of growth. Panels can be read manually using indirect light or the panels can be read on the MicroScan® WalkAway instrument using optics light detection.
# M. Performance Characteristics (if/when applicable):
# This submission is for the AST Panel only. The ID System was not reviewed.
The Reproducibility studies, QC performance data, and Challenge isolates evaluated by the manual and automated reading methods are contained in this submission to demonstrate that there is no difference between manual reading and automated reading in the MicroScan® WalkAway System. The clinical efficacy performance was previously established using the manual read method and was therefore not required for this submission.
{3}
Page 4 of 6
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility was demonstrated using 10 isolates tested at 3 sites on 3 separate days in triplicate. The study included testing on the MicroScan® WalkAway System with automated reading at 20-24 hours, and manual readings at 20-24 hours incubation. Both reading methods demonstrated >95% reproducibility, and no differences were observed.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or method):
The recommended QC isolate S. pneumoniae ATCC 49619 was tested a sufficient number of times with acceptable results on all testing days with the reference method. Quality control results demonstrated the ability of the different reading parameters (manual and instrument) to produce acceptable results. The following table provides the frequency of results in each concentration with the expected range stated. Both reading methods produced the same mode.
| Organism | Concentration μg/mL | Reference results | MicroScan® WalkAway results | |
| --- | --- | --- | --- | --- |
| | | | Manual Overnight | Instrument Overnight |
| S. pneumoniae ATCC 49619
Expected range 0.12 – 0.5 μg/mL | <=0.06 | | | |
| | 0.12 | | | |
| | 0.25 | 52 | 56 | 52 |
| | 0.5 | 10 | 12 | 16 |
| | 1 | | | |
| | 2 | | | |
| | 4 | | | |
Inoculum density control: A turbidity meter, which was verified each day of testing, was used for the turbidity inoculation method. Colony counts were performed weekly, on the ATCC S. pneumoniae with all results in the expected range of approximately 5 x 10⁵.
No trending was observed.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
{4}
Page 5 of 6
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Clinical efficacy testing with manual result reading was conducted in the previous submission (K020939). In this submission, Challenge isolates were evaluated by the manual and automated reading methods to demonstrate that there is no difference between manual reading and instrument reading on the MicroScan® WalkAway System. There were 70 streptococcal challenge isolates including 53 S. pneumoniae CDC Challenge Set strains tested at one site and compared to the reference broth dilution result. A comparison was done with readings on the instrument after 20 hours incubation, and also read manually when incubated for 20-24 hours. Performance by the automated reading method was acceptable with no differences or trends.
The recommended CLSI reference method was followed with the exception of the use of a small amount (0.1%) of Pluronic (a wetting agent) in the final inoculum. A validation of the use of Pluronic in the frozen reference panel was conducted. QC was also performed with no difference apparent in the results.
Read method comparison of Streptococcus species and Tetracycline
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Challenge Manual | 70 | 69 | 98.6 | 41 | 40 | 97.6 | 35 | 0 | 1 | 0 |
| Challenge Automated | 70 | 68 | 97.1 | 41 | 39 | 95.1 | 35 | 1 | 1 | 0 |
EA-Essential Agreement
R-resistant isolates
maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies
Essential agreement (EA) is when the Microscan® MICroSTREP plus® panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Evaluable (Eval) are results that are within the test range and on scale.
Automated reading results were the same as the manual reading results with no trending. There were no vmj, 1 maj error produced by both reading methods, and there was 1 additional minor error generated by the automated reading method. The overall EA% of 98.6% and Eval EA% of 97.6% for the manual read and overall EA% of 97.1% and Eval EA% of 95.1% for the automated reading were both very good.
The test device had a growth rate of >95% for both the manual reading and the automated reading methods.
{5}
Page 6 of 6
The comparison of the reading methods demonstrates that the manual reading method and the automated reading on the MicroScan® WalkAway System are no different. The efficacy data performed with the manual reading method would therefore be expected to have no differences.
The performance data currently documented in the package insert will not change.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Interpretive criteria:
<=2 (Susceptible), 4 (Intermediate), >=8 (Resistant)
The expected value range, interpretive criteria and QC are included in the package insert.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.