VITEK 2 AST-Gram Positive Linezolid (<=0.5 – >=8 µg/mL)

K212849 · Biomérieux, Inc. · LON · Feb 4, 2022 · Microbiology

Device Facts

Record IDK212849
Device NameVITEK 2 AST-Gram Positive Linezolid (<=0.5 – >=8 µg/mL)
ApplicantBiomérieux, Inc.
Product CodeLON · Microbiology
Decision DateFeb 4, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

VITEK® 2 AST-Gram Positive Linezolid is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Positive Linezolid is a quantitative test. Linezolid has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Enterococcus faecium (vancomycin-resistant isolates only) Staphylococcus aureus (including methicillin-resistant isolates) Streptococcus agalactiae In vitro data are available, but clinical significance is unknown: Enterococcus faecalis (including vancomycin-resistant isolates) Enterococcus faecium (vancomycin-susceptible isolates) Staphylococcus epidermidis (including methicillin-resistant isolates) Staphylococcus haemolyticus The VITEK® 2 Gram-positive Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed.

Device Story

VITEK 2 AST-Gram Positive Linezolid is an in vitro diagnostic test for determining antimicrobial susceptibility of gram-positive bacteria. The device uses VITEK 2 or VITEK 2 Compact instruments to perform automated dilution of antimicrobial agents. Input consists of bacterial isolates prepared as direct colony suspensions. The system measures growth inhibition to determine Minimum Inhibitory Concentration (MIC) values. Results are provided to clinicians to guide antibiotic therapy selection. The modification involves reanalysis of MIC results for Staphylococcus spp. using updated FDA-recognized interpretative criteria. The device is used in clinical microbiology laboratories by trained technicians. The output assists in identifying resistant strains and informing clinical decision-making for linezolid treatment.

Clinical Evidence

Performance evaluated using clinical and challenge isolates compared to CLSI agar dilution reference method. Results: Enterococcus spp. (EA 99.8%, CA 98.0%), Staphylococcus spp. (EA 97.2%, CA 99.7%), S. agalactiae (EA 100%, CA 100%). Staphylococcus haemolyticus subset showed EA 82.9% and CA 97.1% with one major error. Reproducibility was 100%.

Technological Characteristics

Microdilution-based 64-well card; contains premeasured linezolid (0.5, 1, 2 ug/mL) and culture media. Automated incubation and optical growth monitoring. Compatible with VITEK 2 and VITEK 2 Compact systems. Uses Discriminant Analysis algorithm.

Indications for Use

Indicated for antimicrobial susceptibility testing of Staphylococcus spp., Enterococcus spp., and Streptococcus agalactiae to linezolid in clinical settings using VITEK 2 or VITEK 2 Compact systems.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K212849 B Applicant bioMérieux, Inc C Proprietary and Established Names VITEK 2 AST-Gram Positive Linezolid (≤0.5 - ≥8 μg/mL) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring 510(k). The following items are present and acceptable. 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. Trade Name: VITEK 2 AST-GP Linezolid (≤0.5 - ≥8μg/mL) 510(k) #: K032766 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). The Intended Use/Instructions for Use have not changed, but language has been harmonized to align with recent VITEK 2 antimicrobials. In addition, The VITEK 2 Compact instrument is now included in the Intended Use since the VITEK 2 Compact performance was similar to VITEK 2 System as demonstrated in K050002. 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. The change was for reanalysis of MIC results for Staphylococcus spp. with linezolid using currently recognized interpretative criteria. bioMérieux also tested 11 S. epidermidis clinical stock linezolid-resistant isolates and 20 S. haemolyticus clinical stock isolates to assess performance. All testing was performed with the auto dilution method at a single external clinical site. The methods used to provide data to support this change were in conformity to those outlined in CLSI M07, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – 11th Edition; January 2018. 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The device performance for Streptococcus agalactiae and Enterococcus spp. was evaluated in K032766. Please refer to K032766 for results obtained with the bioMérieux VITEK 2 AST-Gram Positive Linezolid when compared to results obtained with the CLSI reference agar dilution method. ## Staphylococcus spp. Performance Study: A supplemental clinical study was conducted in accordance with the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, issued August 28, 2009 (AST Guidance document). Supplemental testing was performed at one external clinical site. All reference method testing follows CLSI approved agar dilution testing conditions, which consisted of: - Medium: Mueller Hinton agar with the appropriate dilutions of antimicrobial solution added - Inoculum: Direct colony suspension - Incubation: 35°C; 16-20 hours K212849 - Page 2 of 5 {2} Previously collected clinical and challenge data and newly collected clinical data of 390 isolates of Staphylococcus spp. (337 clinical and 53 challenge isolates) were reanalyzed using the current FDA-recognized breakpoints using auto dilution method. Results obtained with the auto dilution method were acceptable (Table 1). Table 1. Staphylococcus spp. of Clinical and Challenge Isolates for Linezolid: VITEK 2 Auto-Dilution | Organism Type | EA Tot | EA N | EA % | Eval. EA Tot | Eval. EA N | Eval. EA % | CA N | CA % | #R | #S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Staphylococcus^{1} spp. [Breakpoints (μg/mL): ≤4 (S). - (I), ≥8 (R)] | | | | | | | | | | | | | | | Clinical | 337 | 328 | 97.3 | 324 | 315 | 97.2 | 336 | 99.7 | 11 | 326 | 0 | 1 | 0 | | Challenge | 53 | 51 | 96.2 | 53 | 51 | 96.2 | 53 | 100 | 0 | 53 | 0 | 0 | 0 | | Combined | 390 | 379 | 97.2 | 377 | 366 | 97.1 | 389 | 99.7 | 11 | 379 | 0 | 1 | 0 | EA – Essential Agreement CA – Category Agreement EVAL – Evaluable isolates R – Resistant isolates min – minor errors maj – major errors vmj – very major errors S – Susceptible isolates 1When evaluating performance of the clinical and challenge isolates, testing with Staphylococcus haemolyticus isolates yielded an EA of 82.9% (29/35) and an CA of 97.1% (34/35). There was one major error (2.9%, 1/35). For supplemental testing, DensiChek was used for the preparation of inoculum for the VITEK 2 AST cards in the trial and standardized daily. Two quality control organisms, Enterococcus faecalis ATCC 29212 and Staphylococcus aureus ATCC 29213, were tested throughout comparative testing by both the VITEK 2 AST-GP Linezolid and reference agar dilution methods. The QC organisms were tested a minimum of twenty times by both the VITEK 2 card and reference method. QC results for the VITEK 2 AST-GP Linezolid were within expected results' range &gt;95% of the time for automatic dilution mode of the VITEK 2. Performance of all clinical and challenge isolates for linezolid using the automatic dilution method were stratified by organism and the results were acceptable (Table 2). Performance of all challenge isolates for linezolid using the manual dilution method stratified by organism remains unchanged from K032766. A trending analysis was conducted using the combined clinical and challenge data for all organisms. Evaluation of the results for S. haemolyticus showed a trend toward higher MIC values using the auto-dilution method (Table 3). K212849 - Page 3 of 5 {3} Table 2. Performance of All Clinical and Challenge Isolates for Linezolid Stratified by Organism: VITEK 2 Auto-Dilution | Organism Type | EA Tot | EA N | EA % | Eval. EA Tot | Eval. EA N | Eval. EA % | CA N | CA % | #R | #S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Staphylococcus1 spp. [Breakpoints (μg/mL): ≤4 (S), - (I), ≥8 (R)] | | | | | | | | | | | | | | | Clinical | 337 | 328 | 97.3 | 324 | 315 | 97.2 | 336 | 99.7 | 11 | 326 | 0 | 1 | 0 | | Challenge | 53 | 51 | 96.2 | 53 | 51 | 96.2 | 53 | 100 | 0 | 53 | 0 | 0 | 0 | | Combined | 390 | 379 | 97.2 | 377 | 366 | 97.1 | 389 | 99.7 | 11 | 379 | 0 | 1 | 0 | | Enterococcus spp. [Breakpoints (μg/mL): ≤2 (S), 4 (I), ≥8 (R)] | | | | | | | | | | | | | | | Clinical | 368 | 367 | 99.7 | 359 | 358 | 99.7 | 360 | 97.8 | 8 | 354 | 8 | 0 | 0 | | Challenge | 35 | 35 | 100 | 35 | 35 | 100 | 35 | 100 | 0 | 35 | 0 | 0 | 0 | | Combined | 403 | 402 | 99.8 | 394 | 393 | 99.7 | 395 | 98.0 | 8 | 389 | 8 | 0 | 0 | | Streptococcus agalactiae [Breakpoints (μg/mL): ≤2 (S), - (I), - (R)] | | | | | | | | | | | | | | | Clinical | 54 | 54 | 100 | 53 | 53 | 100 | 54 | 100 | 0 | 54 | 0 | 0 | 0 | | Challenge | 10 | 10 | 100 | 10 | 10 | 100 | 10 | 100 | 0 | 10 | 0 | 0 | 0 | | Combined | 64 | 64 | 100 | 63 | 63 | 100 | 64 | 100 | 0 | 64 | 0 | 0 | 0 | EA - Essential Agreement CA - Category Agreement EVAL - Evaluable isolates R - Resistant isolates min - minor errors maj - major errors vmj - very major errors S - Susceptible isolates 1When evaluating performance of the clinical and challenge isolates, testing with Staphylococcus haemolyticus isolates yielded an EA of 82.9% (29/35) and an CA of 97.1% (34/35). There was one major error (2.9%, 1/35). Table 3. Trending by S. haemolyticus (clinical and challenge isolates) | VITEK 2 Auto-Dilution | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Organism | Total Evaluable for Trending | ≥1 dil. Lower # (%) | Exact # (%) | ≥1 dil. Higher # (%) | Percent Difference (95% CI) | Trending Noted | | Staphylococcus haemolyticus | 35 | 1 (2.86) | 7 (20.00) | 27 (77.14) | 74.29% (54.35 to 85.33) | Yes¹ High | ¹Trending is addressed in a footnote to the performance table in the device labeling. To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the VITEK 2 AST-GP Linezolid when revised breakpoints for linezolid are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the linezolid device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation. The data included in this submission supported the sponsor's proposal to revise the following from the device instructions for use: 1. Updated performance for Staphylococcus spp. using the currently-recognized interpretive criteria for linezolid and Staphylococcus spp. (as noted on the FDA Susceptibility Test Interpretive Criteria website (STIC)). K212849 - Page 4 of 5 {4} 2. Added the following footnotes under the table Performance Characteristics Antimicrobial Susceptibility Testing in the labeling: "VITEK 2 AST-Gram Positive Linezolid MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Staphylococcus haemolyticus compared to the CLSI reference agar dilution method." "When evaluating performance of the clinical and challenge isolates, testing with Staphylococcus haemolyticus isolates yielded an EA of 82.9% (29/35) and an CA of 97.1% (34/35). There was one major error (2.9%%, 1/35)." 3. Removed S. epidermidis from the limitation regarding lack of resistant isolate availability in the labeling. The modified limitation is shown below: "The ability of the AST card to detect resistance with the following combination(s) is unknown because resistant strains were not available at the time of comparative testing: - [lzn02n] Linezolid: Enterococcus spp., Staphylococcus aureus, Staphylococcus haemolyticus, Streptococcus agalactiae" The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their pre-amendment) device. K212849 - Page 5 of 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...