MUELLER HINTON II BROTH (CATION-ADJUSTED) WITH LYSED HORSE BLOOD

K963036 · Becton Dickinson Microbiology Systems · JTZ · Oct 7, 1996 · Microbiology

Device Facts

Record IDK963036
Device NameMUELLER HINTON II BROTH (CATION-ADJUSTED) WITH LYSED HORSE BLOOD
ApplicantBecton Dickinson Microbiology Systems
Product CodeJTZ · Microbiology
Decision DateOct 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1700
Device ClassClass 2

Indications for Use

Mueller Hinton II Broth (Cation-Adjusted) with Lysed Horse Blood (MHLHB) is intended for use in broth dilution antimicrobial susceptibility testing of Streptococcus pneumoniae with 11 antimicrobial agents; i.e., cefaclor, cefotaxime, ceftriaxone, cefuroxime, chloramphenicol, erythromycin, imipenem, penicillin, tetracycline, trimethoprim/sulfamethoxazole, and vancomycin, according to the protocol described in the Approved Standard M7-A3, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically - Third Edition", dated 12/93, published by the National Committee for Clinical Laboratory Standards (NCCLS).

Device Story

Device is a specialized culture medium (Mueller Hinton II Broth with lysed horse blood) for antimicrobial susceptibility testing (AST) of fastidious Streptococcus pneumoniae. Clinical laboratory technicians perform broth dilution AST by inoculating serial dilutions of antimicrobial agents in the medium. Following incubation, technicians observe turbidity or pellets to determine the minimal inhibitory concentration (MIC). Quantitative MIC results allow clinicians to correlate in vitro drug inhibition with achievable patient drug concentrations in blood, urine, CSF, or bile, aiding in effective antimicrobial therapy selection. The medium supports growth of S. pneumoniae, which otherwise fails in standard cation-adjusted Mueller Hinton Broth.

Clinical Evidence

Bench testing only. Quality control testing using S. pneumoniae ATCC 49619 (n=220) showed 100% of MICs within expected NCCLS ranges. Reproducibility studies at two field sites using 10 strains (ATCC 49619 plus 9 well-characterized isolates) demonstrated 100% essential agreement for 9/11 antimicrobics, 99% for tetracycline, and 98% for trimethoprim/sulfamethoxazole compared to NCCLS reference methods.

Technological Characteristics

Cation-adjusted Mueller Hinton Broth supplemented with 2-5% lysed horse blood. Liquid medium form factor. Designed for use in standard broth dilution AST protocols (macrodilution or microtiter trays).

Indications for Use

Indicated for clinical laboratory use to determine susceptibility, intermediate, or resistance of pure culture Streptococcus pneumoniae isolates to 11 specific antimicrobial agents (cefaclor, cefotaxime, ceftriaxone, cefuroxime, chloramphenicol, erythromycin, imipenem, penicillin, tetracycline, trimethoprim/sulfamethoxazole, and vancomycin) using broth dilution methods.

Regulatory Classification

Identification

A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} August 2, 1996 # SUMMARY OF SAFETY AND EFFECTIVENESS SUBMITTED BY: Virginia C. Weinknecht Regulatory Affairs Specialist Becton Dickinson Microbiology Systems 250 Schilling Circle Cockeysville, MD 21030-0243 K963036 OCT - 7 1996 ## NAME OF DEVICE: Trade Name: Mueller Hinton II Broth (Cation-Adjusted) with Lysed Horse Blood Catalog Number 4320500 Common Name/Description: Antimicrobial Susceptibility Culture Medium Classification Name: Culture Medium for Antimicrobial Susceptibility Tests ## PREDICATE DEVICE: AB Biodisk Etest® for Susceptibility Testing of Pneumococci (K913459) ## DEVICE DESCRIPTION: **INTENDED USE:** Mueller Hinton II Broth (Cation-Adjusted) with Lysed Horse Blood (MHLHB) is intended for use in broth dilution antimicrobial susceptibility testing of *Streptococcus pneumoniae* with 11 antimicrobial agents; i.e., cefaclor, cefotaxime, ceftriaxone, cefuroxime, chloramphenicol, erythromycin, imipenem, penicillin, tetracycline, trimethoprim/sulfamethoxazole, and vancomycin, according to the protocol described in the Approved Standard M7-A3, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically - Third Edition", dated 12/93, published by the National Committee for Clinical Laboratory Standards (NCCLS). **INDICATIONS FOR USE:** Use of BBL® Mueller Hinton II Broth (Cation-Adjusted) with Lysed Horse Blood is indicated when *Streptococcus pneumoniae* has been isolated in pure culture in the clinical laboratory, and a C-1 {1} determination is desired as to whether the organism is susceptible, intermediate, or resistant to selected antimicrobial agents. In cases such as these, broth dilution antimicrobial susceptibility testing may be performed on BBL® Mueller Hinton II Broth (Cation-Adjusted) with Lysed Horse Blood, using cefaclor, cefotaxime, ceftrixone, cefuroxime, chloramphenicol, erythromycin, imipenem, penicillin, tetracycline, trimethoprim/sulfamethoxazole, and vancomycin. ## PRODUCT DESCRIPTION: The development of laboratory tests to determine the activity of antimicrobial agents has paralleled the development of these agents. Fleming used a serial dilution technique to measure the lowest concentration of penicillin that prevented growth of a test organisms in broth. Ericsson and Sherris have published an excellent review of the various methods for susceptibility testing and the relationship of dilution and diffusion methods. Broth dilution antimicrobial susceptibility tests (AST) are performed by inoculating serial dilutions (usually 2-fold) of the drug in a suitable liquid medium. The test may be performed in tubes (macrodilution), usually in 1.0 ml volumes, or in microtiter trays in volumes of 0.05 to 0.1 ml. Following incubation, the tubes or wells are examined for the presence of growth (turbidity or pellet). The lowest concentration of an antimicrobial agent at which no visible growth occurs is defined as the minimal inhibitory concentration, or MIC. The rationale for an MIC susceptibility test rather than the disc diffusion test is that it gives quantitative information. It allows the clinician to correlate the amount of antimicrobial agent required to inhibit the growth of an organism in vitro and the achievable concentrations in the blood, urine, cerebrospinal fluid or bile. Effective antimicrobial therapy, however, also depends on other factors. Broth dilution ASTs are usually performed in cation-adjusted Mueller Hinton Broth (CAMHB). However, this medium is not satisfactory for fastidious organisms such as S. pneumoniae. CAMHB supplemented with 2 to 5% lysed horse blood is the medium recommended for susceptibility testing of S. pneumoniae. Interpretive criteria for the antimicrobial susceptibility testing of S. pneumoniae are provided in the NCCLS Standard, M7-A3. This document and NCCLS Informational Supplement M100-S6 should be consulted for further details. C-2 {2} C-3 # PERFORMANCE DATA: Antimicrobial broth dilution susceptibility testing using the quality control strain recommended by NCCLS Document M7-A3, *Streptococcus pneumoniae* ATCC 49619, was performed in-house with cefaclor, cefotaxime, ceftriaxone, cefuroxime, chloramphenicol, erythromycin, imipenem, penicillin, tetracycline, trimethoprim/sulfamethoxazole, and vancomycin. Following the test procedures described in M7-A3, twenty tests with the quality control strain and microtitre panels containing the eleven antimicrobics were performed over a period of ten test days. For all eleven antimicrobics, 100% (220/220) of the MIC’s fell within the expected MIC ranges published in Table 3C of NCCLS Supplement M100-S6. The standard deviation for all antimicrobics was less than 1 µg/ml. Reproducibility studies (3x/day for 3 days) were done at two field sites with the antimicrobics listed above against *S. pneumoniae* ATCC 49619 and nine (9) additional well-characterized *S. pneumoniae* strains. MIC interpretive standards from Table 2C of NCCLS Supplement M100-S6 were followed for each antimicrobic. Testing with cefaclor, cefotaxime, ceftriaxone, cefuroxime, chloramphenicol, erythromycin, imipenem, penicillin, and vancomycin resulted in 100% Essential Agreement with the NCCLS reference method. Testing with tetracycline and trimethoprim/sulfamethoxazole resulted in 99% and 98% Essential Agreement respectively with the reference method.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...