VRESelect™ is a selective and differential chromogenic medium, containing 8 µg/mL of vancomycin, for the qualitative detection of gastrointestinal colonization of vancomycin-resistant Enterococcus faecium (VREfm) and vancomycin-resistant Enterococcus faecalis (VREfs) and to aid in the prevention and control of vancomycin-resistant Enterococcus (VRE) in healthcare settings. The test is performed on rectal swabs, or fecal specimens from patients to be screened for VRE colonization. VRESelect™ is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Results can be interpreted after 24 to 28 hours incubation. Subculture to non-selective media (e.g., trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.
Device Story
VRESelect™ is a selective/differential chromogenic culture medium containing 8 µg/mL vancomycin; used in healthcare settings for VRE screening. Input: rectal swabs or fecal specimens. Principle: antifungal/antibiotic mixture inhibits non-VRE organisms; chromogenic substrates cleaved by specific enzymes produce color-coded colonies (pink for VREfm, blue for VREfs). Output: visual colony identification after 24-28 hours incubation. Interpretation: manual/visual by laboratory personnel. Blue colonies require confirmation via catalase test and antimicrobial susceptibility testing. Subculture to non-selective media required for further identification/typing. Benefits: aids in VRE prevention and control programs.
Clinical Evidence
Performance evaluated at three sites using 946 fecal samples. Compared against Bile Esculin Azide Agar (BEAV) plus conventional identification (Gram stain, catalase, PYR, VITEK 2, E-Test). At 28 hours, positive agreement was 98% (186/189) and negative agreement was 95% (721/757) vs BEAV. VREfm positive agreement vs Vitek 2 was 97% at 28 hours; VREfs positive agreement was 94%. Analytical sensitivity (LOD) is 10³ CFU/mL. No cross-reactivity observed with 119 strains.
Technological Characteristics
Selective/differential chromogenic culture medium. Contains 8µg/mL vancomycin and antifungal/antibiotic mixture. Principle: enzymatic cleavage of chromogenic substrates. Form factor: plated medium. Incubation: 24-28 hours at 35-37°C. No software or electronic components.
Indications for Use
Indicated for qualitative detection of gastrointestinal colonization of vancomycin-resistant Enterococcus faecium (VREfm) and Enterococcus faecalis (VREfs) in patients via rectal swabs or fecal specimens. Aids in VRE prevention/control in healthcare settings. Not for diagnosing active infection or guiding treatment.
Regulatory Classification
Identification
A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.
K103684 — VRESELECT CULTURE MEDIUM · Bio-Rad · Oct 21, 2011
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k122187
B. Purpose for Submission:
Premarket notification in support of clearance of Bio-Rad VRESelect™ for the qualitative detection of vancomycin-resistant Enterococcus faecium and E. faecalis in stool samples.
C. Measurand:
Vancomycin resistant E. faecium and E. faecalis
D. Type of Test:
Detection of vancomycin resistant Enterococcus faecium and Enterococcus faecalis using a selective and differential chromogenic medium.
E. Applicant:
Bio-Rad
F. Proprietary and Established Names:
VRESelect™ Culture Medium
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1700
2. Classification:
Class II
3. Product code:
JSO Culture media, Antimicrobial susceptibility test, excluding Mueller Hinton Agar
{1}
4. Panel:
83; Microbiology
H. Intended Use:
1. Intended use:
VRESelect™ is a selective and differential chromogenic medium, containing 8 µg/mL of vancomycin, for the qualitative detection of gastrointestinal colonization of vancomycin-resistant Enterococcus faecium (VREfm) and vancomycin-resistant Enterococcus faecalis (VREfs) and to aid in the prevention and control of vancomycin-resistant Enterococcus (VRE) in healthcare settings. The test is performed on rectal swabs, or fecal specimens from patients to be screened for VRE colonization. VRESelect™ is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Results can be interpreted after 24 to 28 hours incubation. Subculture to non-selective media (e.g., trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.
2. Indications for use:
VRESelect™ is a selective and differential chromogenic medium, containing 8 µg/mL of vancomycin, for the qualitative detection of gastrointestinal colonization of vancomycin-resistant Enterococcus faecium (VREfm) and vancomycin-resistant Enterococcus faecalis (VREfs) and to aid in the prevention and control of vancomycin-resistant Enterococcus (VRE) in healthcare settings. The test is performed on rectal swab, or fecal specimens from patients to be screened for VRE colonization. VRESelect™ is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Results can be interpreted after 24 to 28 hours incubation. Subculture to non-selective media (e.g., trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.
3. Special conditions for use statement:
Prescription use.
The performance of VRESelect™ has not been evaluated with vancomycin resistant strains of Staphylococcus aureus.
After 24 to 28 hours incubation pink colonies can be reported as vancomycin-resistant Enterococcus faecium. Blue colonies should be confirmed by a direct catalase test and, if negative, by vancomycin susceptibility testing.
2
{2}
4. Special instrument requirements:
Not Applicable.
I. Device Description:
VRESelect™ is a selective medium for the detection of vancomycin-resistant Enterococcus (VRE). The selectivity of this medium is based on the presence of an antifungal/antibiotic mixture that inhibits the growth of most yeast, Gram negative and Gram positive bacteria, with the exception of VRE. Vancomycin-resistant E. faecium will appear as pink colonies on the agar medium. Blue colonies on the agar medium, suggestive of vancomycin-resistant E. faecalis should be confirmed by catalase and antimicrobial susceptibility testing.
J. Substantial Equivalence Information:
1. Predicate device name:
Remel Spectra VRE Chromogenic Media
Remel Bile Esculin Azide Agar w/ 6μg/mL Vancomycin
2. Predicate K number:
K092819
K972359
3
{3}
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate K092819 | Predicate K972359 |
| Intended Use | VRESelect™ is a selective and differential chromogenic medium, containing 8μg/mL of Vancomycin, for the qualitative detection of gastrointestinal colonization of vancomycin-resistant Enterococcus faecium (VREfm) and vancomycin-resistant Enterococcus faecalis (VREfs) and to aid in the prevention and control of vancomycin-resistant Enterococcus (VRE) in healthcare settings. The test is performed on rectal swabs or fecal specimens from patients to be screened for VRE colonization. VRESelect™ is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Results can be interpreted after 24 to 28 hours incubation. Subculture to non-selective media (e.g., trypticase soy agar with 5% sheep blood) is needed for susceptibility testing and epidemiological typing. | Remel Spectra VRE is a selective and differential chromogenic medium, containing 6μg/mL of Vancomycin, intended for use in the qualitative detection of gastrointestinal colonization with vancomycin-resistant Enterococcus faecium and Enterococcus faecalis (VRE) to aid in the prevention and control of VRE in healthcare settings. The test is performed with a rectal swab and fecal specimens from patients to screen for VRE colonization. Spectra VRE is not intended to diagnose VRE infection or to guide or monitor treatment for infections. Subculture to non-selective media (e.g. Tryptic Soy Agar with 5% sheep blood) is needed for further identification, susceptibility testing, and epidemiological typing. | Remel Bile Esculin Azide Agar w/ 6μg/mL vancomycin is a plated medium recommended for use in qualitative procedures as a selective and differential medium for the primary isolation of vancomycin-resistant enterococci from surveillance cultures. This product is not intended for use as [a] method of antimicrobial susceptibility testing. Confirmation of resistance by an approved method is recommended as some organisms on initial isolation may overcome the inhibitory effects of the medium. |
| Methodology | Enzymatic | Enzymatic | Enzymatic |
| Interpretation | Manual, Visual | Manual, Visual | Manual, Visual |
| Sample Type | Stool/fecal specimens | Rectal Swabs or Fecal Specimens | Rectal Swabs or Fecal Specimens |
| Differences | | | |
| Inoculation | Direct or Indirect | Direct Specimen | Direct Specimen |
K. Standard/Guidance Document referenced (if applicable):
Not Applicable.
{4}
L. Test Principle:
VRESelect™ is a selective medium for the detection of vancomycin-resistant Enterococcus (VRE). The selectivity of this medium is based on the presence of an antifungal/antibiotic mixture that inhibits the growth of most yeasts, Gram negative and Gram positive bacteria, with the exception of VRE.
Detection is based on the cleavage of chromogenic substrates by specific enzymes of Enterococcus faecium which produces pink colonies and Enterococcus faecalis which produces blue colonies.
Enterococcus gallinarum and Enterococcus casseliflavus are intrinsically resistant to vancomycin and may grow on the VRESelect™ medium as colorless or white colonies because they do not metabolize the chromogenic substrates. Vancomycin-susceptible enterococci are inhibited.
After 24 to 28 hours incubation pink colonies can be reported as vancomycin-resistant E. faecium and blue colonies should be confirmed by a catalase test and susceptibility testing.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility was demonstrated at three sites using a blinded panel of six ATCC reference strains including two vancomycin-resistant E. faecalis strains, three vancomycin-resistant E. faecium strains and one strain of vancomycin-susceptible E. faecalis. At each site, three technicians tested the panel on three lots of the VRESelect™ each day, for three days. The strains produced the expected results with the VRESelect™ 100% of the time both at 24 and 28 hours.
b. Linearity/assay reportable range:
Not Applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The following recommended quality control (QC) organisms were used: E. faecium (vanA) ATCC 700221 and E. faecalis (vanB) ATCC 51299 as positive controls, and E. faecalis ATCC 29212 as negative control. Quality control was performed on each day of testing by preparing a 0.5 McFarland suspension of each QC strain then diluting is 1:10 and transferring 10 μL of
{5}
this suspension to the VRESelect™ agar surface streaking for isolation. Quality control data was compiled across all three sites and overall QC results were acceptable.
## QC Data Summary
| QC Strain | Expected Results after 24- 28 hours at 35- 37°C |
| --- | --- |
| Enterococcus faecalis ATCC 51299 | Growth- small blue colonies |
| Enterococcus faecium ATCC 700221 | Growth- small pink colonies |
| Enterococcus faecalis ATCC 29212 | No growth |
## d. Detection limit:
The minimum concentration of VRE reliably detected by VRESelect™ is 10³ CFU/mL. Details of the recovery study are found in Section f below.
## e. Analytical specificity:
### Cross-reactivity Study
A Cross-reactivity study was performed using a total of 119 strains of Gram negative rods, Gram positive cocci and yeast. No cross-reactivity was observed with any of the organisms tested and no variation was seen between the 24 and 28 hour incubation time.
### Interference Study
The following substances were evaluated at physiologically or biologically relevant concentrations for potential interference with the performance of the VRESelect™ medium:
- Dulcolax, Adult Glycerin Suppositories, Vaseline, Preparation H, Original Boudreaux's Butt Paste, Tuck's Medicated Cooling Pads, Pepto-Bismol, Miconazole cream, Nonoxynol-9 (spermicide), KY Jelly, and Pepcid AC Max strength.
- Blood and Mucins
- Three commonly used transport media – Amies without charcoal, Cary Blair and LQ Stuart
The interfering substances tested caused no significant differences between the number of colonies observed on the control plates and the number of colonies observed on the VRESelect™ plates. The only exceptions were Tuck's Medicated Cooling Pads and Miconazole cream. Regarding Tuck's Medicated Pads, no pink coloration was observed after 24 hours on the VRESelect™ plates that had been inoculated with vancomycin-resistant E. faecium (ATCC 700221). Regarding Miconazole cream, an inhibitory effect on the growth of Enterococcus colonies on the VRESelect™ plates was
6
{6}
observed. Blood and mucin (3% to 5%) caused delayed colonial growth of one strain of vancomycin-resistant *E. faecalis* (ATCC 51299) tested. The growth of the same strain of vancomycin-resistant *E. faecalis* was inhibited at blood and mucin concentrations of 30% to 50%.
## Mixed Infection Study
A mixed infection study was conducted to demonstrate that high levels (1×10⁷ CFU/mL or higher) of *E. gallinarum* and/or *E. casseliflavus* will not suppress the growth of vancomycin-resistant *E. faecium* or *E. faecalis*. The results of the study revealed *E. faecium* or *E. faecalis* are still detected on the VRESelect™ medium in the presence of high levels of *E. casseliflavus* and *E. gallinarum*.
## f. Analytical Sensitivity
### Recovery Study
The minimum concentration of VRE reliably detected by VRESelect™ is 10³ CFU/mL.
To determine the percent recovery for the VRESelect™ media a panel of eighteen (18) vancomycin-resistant Enterococci (VRE)–(8) strains of vancomycin-resistant *E. faecium* and (10) strains of vancomycin-resistant *E. faecalis* – were tested at varying dilutions. A 0.5 McFarland suspension of each strain was prepared. A series of 10-fold serial dilutions in saline were carried out and inoculated onto three lots of VRESelect™ plates and one lot of Blood Agar plates. The plates were incubated at 35-37°C ambient air and read at 24 and 28 hours. The color and number of colonies were recorded. The Blood Agar plates were used to confirm the inoculum concentration at each dilution. Data confirm that the minimum concentration of VRE reliably detected by VRESelect™ is 10³ CFU/mL.
## g. Assay cut-off:
Not Applicable.
## 2. Comparison studies:
### a. Method comparison with predicate device:
The performance of the VRESelect™ agar medium was evaluated at three laboratory sites. A total of 946 fecal samples were evaluated. Fecal samples were inoculated using the indirect inoculation method where approximately 0.5 g of stool is placed in 1 mL of saline, then mixed by vortexing. Using a swab or a disposable loop, approximately 10-50 μl is transferred onto the VRESelect™ agar medium and the Bile Esculin Azide Agar with 6 μg/ml of
7
{7}
Vancomycin (BEAV). The BEAV plates were observed for growth at 24 and 48 hours. The VRESelect™ plates were observed at 24 and 28 hours. Colonies with blue or pink pigment (VRESelect™) or brown to black pigment diffusing into the medium (BEAV), were identified by a combination of conventional reference methods to include Gram stain, catalase, PYR, VITEK 2 identification, and vancomycin MIC E-Test.
Percent agreement of the VRESelect™ compared to BEAV and conventional methods is presented in Table 1. Specimens that were positive on the VRESelect (i.e. grew pink or blue colonies between 24 and 28 hours incubation) were compared to specimens that were confirmed positive based on growth and pigment on BEAV combined with Gram stain, catalase test, PYR test, biochemical identification, and E-test MIC)
| VRESelect™ | Table 1. BEAV +Confirmation | | |
| --- | --- | --- | --- |
| | | % Positive Agreement | % Negative Agreement |
| VRESelect™ | 24 hrs | 96% (182/189, [CI. 0.92, 0.98]) | 96% (727/757, [CI. 0.94, 0.97]) |
| | 28 hrs | 98% (186/189, [CI. 0.95, 0.99]) | 95% (721/757, [CI. 0.93, 0.96])* |
* Thirty-three (33) of the 36 specimens that were BEAV plus Confirmation negative and that grew pink and/or blue colonies on VRESelect™ media, after subculture to blood agar plates (BAPs), were confirmed to be vancomycin resistant *E. faecium* and/or *E. faecalis* by biochemical identification and vancomycin E-Test. Three (3) specimens that were BEAV plus Confirmation negative and that grew pink and/or blue colonies on VRESelect™ media, after subculture to blood agar plates (BAPs), were not confirmed biochemical identification and vancomycin E-Test to be *E. faecium* and/or *E. faecalis* and represent false positive results.
Performance data for the VRESelect™ agar medium compared to biochemical identification, for both isolates of *E. faecalis* (VREfs) and *E. faecium* (VREfm) is presented in Table 2.
| | Table 2. Vitek 2 Biochemical Identification | | |
| --- | --- | --- | --- |
| | % Positive Agreement | | % Negative Agreement |
| VREfm | | | |
| VRESelect™ @ 24 hours | 94% (171/181, [CI. 0.90, 0.97]) | | 97% (740/765, [CI. 0.95, 0.98]) |
| VRESelect™ @ 28 hours | 97% (175/181, [CI. 0.93, 0.99]) | | 96% (734/765, [CI. 0.94, 0.97]) |
| | | | |
| VREfs | | | |
| VRESelect™ @ 24 hours | 94% (15/16, [CI. 0.70, 0.99]) | | 98% (910/930, [CI. 0.97, 0.99]) |
| VRESelect™ @ 28 hours | 94% (15/16, [CI. 0.70, 0.99]) | | 98% (909/930, [CI. 0.97, 0.99]) |
{8}
Performance data for the VRESelect™ agar medium compared to the vancomycin MIC E-test, for both isolates of E. faecalis and E. faecium are presented in Table 3. Specimens that were identified in the reference arm of the study as vancomycin-resistant and identified as E. faecium or E. faecalis by biochemical identification and grew pink or blue colonies on VRESelect™ were considered in positive agreement.
| | Table 3. Vancomycin Resistance (E-Test) | |
| --- | --- | --- |
| | % Positive Agreement | % Negative Agreement |
| VREfm | | |
| VRESelect™ @ 24 hours | 96% (171/178, [CI. 0.92, 0.98]) | 97% (743/768, [CI. 0.95, 0.98]) |
| VRESelect™ @ 28 hours | 98% (175/178, [CI. 0.95, 0.99]) | 96% (737/768, [CI. 0.94, 0.97]) |
| | | |
| VREfs | | |
| VRESelect™ @ 24 hours | 100% (12/12, [CI. 0.82, 1.00]) | 98% (911/934, [CI. 0.96, 0.99])* |
| VRESelect™ @ 28 hours | 100% (12/12, [CI. 0.82, 1.00]) | 97% (910/934,[ CI. 0.96, 0.98]) |
b. Matrix comparison:
Not Applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable.
4. Clinical cut-off:
Not Applicable.
{9}
5. Expected values/Reference range:
Vancomycin resistant *E. faecalis* presents as blue color colonies which should undergo catalase and susceptibility testing.
Vancomycin resistant *E. faecium* presents as pink color colonies.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence evaluation.
10
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.