VYSIS CLL FISH PROBE KIT (VYSIS LSI TP53 SPECTRUMORANGE/ATM SPECTRUMGREEN) AND LSI D135319 SPECTRUMORANGE/13Q34
K100015 · Abbott Molecular, Inc. · OVQ · Aug 9, 2011 · Immunology
Device Facts
Record ID
K100015
Device Name
VYSIS CLL FISH PROBE KIT (VYSIS LSI TP53 SPECTRUMORANGE/ATM SPECTRUMGREEN) AND LSI D135319 SPECTRUMORANGE/13Q34
Applicant
Abbott Molecular, Inc.
Product Code
OVQ · Immunology
Decision Date
Aug 9, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6040
Device Class
Class 2
Indications for Use
The Vysis CLL FISH Probe Kit is intended to detect deletion of the LSI TP53, LSI ATM, and LSI D13S319 probe targets and gain of the D12Z3 sequence in peripheral blood specimens from untreated patients with B-cell chronic lymphocytic leukemia (CLL). The assay may be used to dichotomize CLL (the 13q-, +12, or normal genotype group versus the 11q- or 17p- group) and may be used as an aid in determining disease prognosis in combination with additional biomarkers, morphology, and other clinical information. The Vysis CLL FISH Probe Kit is not intended for use in selection of therapy or in monitoring of residual disease.
Device Story
Device uses fluorescence in situ hybridization (FISH) to detect specific genomic aberrations in peripheral blood specimens from untreated CLL patients. Input: patient peripheral blood cells on microscope slides. Process: DNA denaturation; hybridization with fluorophore-labeled DNA probes (LSI TP53, LSI ATM, LSI D13S319, CEP 12, LSI 13q34); washing; DAPI counterstaining. Output: fluorescent signals visualized via fluorescence microscopy. Clinicians enumerate signals to determine copy number of probe targets; results used to classify patients into Döhner prognostic categories (good/intermediate vs. poor). Aids prognosis in conjunction with morphology and other clinical data.
Clinical Evidence
Clinical validity established via concordance with Reference FISH Test (RFT) used in Shanafelt et al. study. Concordance study of 64 specimens showed 97% agreement (62/64) for prognostic category assignment. Reproducibility demonstrated across three laboratories with generalized kappa statistic of 0.86 (almost perfect agreement). Analytical specificity 100% (200/200 loci); analytical sensitivity 97.98-98.94% for expected signal patterns.
Technological Characteristics
FISH DNA probe technology. Components: LSI TP53/ATM probe set, LSI D13S319/13q34/CEP 12 probe set, DAPI II counterstain, NP-40, 20X SSC salt. Detection via fluorescence microscopy using specific excitation/emission filters. Manual signal enumeration.
Indications for Use
Indicated for untreated patients with B-cell chronic lymphocytic leukemia (CLL). Used to detect specific chromosomal aberrations (del 17p13.1, del 11q22.3, del 13q14.3, and trisomy 12) to aid in disease prognosis. Not for therapy selection or monitoring residual disease.
Regulatory Classification
Identification
A gene expression profiling test system for breast cancer prognosis is a device that measures the ribonucleic acid (RNA) expression level of multiple genes and combines this information to yield a signature (pattern or classifier or index) to aid in prognosis of previously diagnosed breast cancer.
Special Controls
The device is classified as Class II under regulation 21 CFR 866.6040 with special controls. The special control guidance document "Class II Special Controls Guidance Document: Gene expression profiling test system for breast cancer prognosis" is available at http://www.fda.gov/cdrh/oivd/guidance/1627.html.
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Gene Expression Profiling Test System for Breast Cancer Prognosis.” See § 866.1(e) for the availability of this guidance document.
Predicate Devices
Agendia BV MammaPrint® (k062694)
Related Devices
P150041 — VYSIS CLL FISH PROBE KIT · Abbott Molecular, Inc. · Apr 11, 2016
K962873 — CEP 12 SPECTRUMORANGE DIRECT LABELED CHROMOSOME ENUMERATION DNA PROBE · Vysis · Jan 13, 1997
DEN170070 — MLL (KMT2A) Breakapart FISH Probe Kit; AML1 (RUNX1) Breakapart FISH Probe Kit, P53 (TP53) Deletion FISH Probe Kit; EVI1 (MECOM) Breakapart FISH Probe Kit, Del(20q) Deletion FISH Probe Kit; AML1/ETO (RUNX1/RUNXIT1)) Translocation, Dual Fusi, CBFB (CBFB) /MYH11 Translocation, Dual Fusion FISH Probe Kit,Del(5q)Deletion FISH Probe Kit; Del(7q)Deletion FISH Probe Kit · Cytocell, Ltd. · Dec 21, 2018
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k100015
B. Purpose for Submission:
New device
C. Measurand:
P53 deletion, ATM deletion, chromosome 11q deletion, chromosome 12 trisomy
D. Type of Test:
Fluorescence in situ hybridization (FISH)
E. Applicant:
Abbott Molecular Inc.
F. Proprietary and Established Names:
Vysis CLL FISH Probe Kit (Vysis LSI TP53 SpectrumOrange/ ATM SpectrumGreen and LSI D13S319 SpectrumOrange/ 13q34 SpectrumAqua/ CEP 12 SpectrumGreen Probes)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.6040 Gene expression profiling test system for breast cancer prognosis
2. Classification:
Class II
3. Product code:
OVQ, Chronic lymphocytic leukemia FISH probe kit
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
The Vysis CLL FISH Probe Kit is intended to detect deletion of the LSI TP53, LSI ATM, and LSI D13S319 probe targets and gain of the D12Z3 sequence in peripheral blood specimens from untreated patients with B-cell chronic lymphocytic leukemia (CLL). The assay may be used to dichotomize CLL (the 13q-, +12, or normal genotype group versus the 11q- or 17p- group) and may be used as an aid in determining disease prognosis in combination with additional biomarkers, morphology and other clinical information. The Vysis CLL FISH Probe Kit is not intended for use in selection of therapy or in monitoring of residual disease.
2. Indication(s) for use:
Same as Intended use.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Not indicated.
I. Device Description:
{1}
The Vysis CLL FISH Probe Kit consists of two DNA FISH probes sets and three general purpose reagents as follows:
1. LSI TP53 SpectrumOrange/ATM SpectrumGreen Probes:
- The SpectrumOrange-labeled LSI TP53 probe, approximately 172 kb in length (ch17: 7435119-7606823; March 2006 University of California at Santa Cruz (UCSC) Human Genome Browser), is located at 17p13.1 and contains the complete TP53 gene. It is used to detect p53 deletion on chromosome 17.
- The SpectrumGreen-labeled LSI ATM probe, approximately 732 kb in length (ch11: 107306249-108038407, March 2006 UCSC Human Genome Browser), is located at 11q22.3 and contains the ATM gene. It is used to detect ATM deletion on chromosome 11.
2. LSI D13S319 SpectrumOrange/13q34 SpectrumAqua/CEP 12 SpectrumGreen Probes:
- The SpectrumOrange-labeled LSI D13S319 probe, approximately 135 kb in length (ch13: 49500369-49635302, March 2006 USC Human Genome Browser), is located at 13q14.3. It is used to detect deletion of chromosome 13q.
- The SpectrumAqua-labeled LSI 13q34 probe, approximately 612 kb in length (ch13: 112739323-113351436, March 2006 USC Human Genome Browser), is located at 13q34. It serves as a quality control probe.
- The SpectrumGreen-labeled CEP 12 probe is located at the centromere of chromosome 12. It is used to detect chromosome 12 trisomy.
3. General Purpose Reagents:
- DAPI II Counterstain
- NP-40
- 20X SSC
J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s):
Agendia BV MammaPrint® (k062694)
2. Comparison with predicate:
Not applicable. Clearance is supported by published clinical study and method concordance bridging studies.
K. Standard/Guidance Document Referenced (if applicable):
American College of Medical Genetics, 2008. Standards and Guidelines for Clinical Genetics Laboratory - Clinical Cytogenetics.
L. Test Principle:
The CLL FISH Probe Kit uses FISH DNA probe technology to determine deletion status of probe targets using CLL specimens for LSI TP53 SO (containing tumor protein p53 gene-location chromosome 17p), LSI ATM SGn (containing ataxia telangiectasia mutated gene-location at chromosome 11q), and LSI D13S319 SO (containing marker D13S319-location at chromosome 13q), as well as determining trisomy 12 with CEP12 SGn (D12Z3 alpha satellite at location chromosome 12p11.1-q11).
Peripheral blood cells from CLL patients are attached to microscope slides using
{2}
standard cytogenetic procedures. The resulting specimen DNA is denatured to single-stranded form and subsequently allowed to hybridize with the probes of the CLL FISH Probe Kit. Following hybridization, the unbound probe is removed by a series of washes, and the nuclei are counterstained with DAPI, a DNA-specific stain that fluoresces blue. Hybridization of the LSI TP53 SO, LSI ATM SGn, LSI D13S319 SO, LSI 13q34 SA, and CEP 12 SGn probes is viewed using a fluorescence microscope equipped with appropriate excitation and emission filters, allowing visualization of the orange, green, and aqua fluorescent signals.
In a cell with normal copy numbers of the LSI TP53 SO and LSI ATM SGn probe targets, two orange signals (LSI TP53 SO) and two green signals (LSI ATM SGn) will be expected. Since both the ATM and TP53 loci seem capable of losing only one copy while the other copy is retained, one orange signal (LSI TP53 SO) and/or one green signal (LSI ATM SGn) serves as an internal control of this probe set.
In a cell with normal copy numbers of LSI D13S319 SO and CEP 12 SGn probe targets, two orange signals (LSI D13S319 SO) and two green signals (CEP 12 SGn) will be expected. The presence (not the quantity) of aqua signals is used as an internal control of this probe set.
Enumeration of the LSI TP53 SO, LSI ATM SGn, LSI D13S319 SO, and CEP12 SGn signals provide a mechanism for determining absolute copy numbers of the probe targets and the presence of the aberrations of interest. Any aberrations detected are utilized to determine Döhner Classification and subsequently the dichotomization of prognosis.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
##### i. Precision:
The precision of the probes in the Vysis CLL FISH Probe Kit was evaluated in two (2) studies. The first study used interphase nuclei prepared from two (2) separate peripheral blood specimens lacking del(17p13.1), del(11q22.3), del(13q14.3), and trisomy 12, and eight additional specimens of which at least two specimens had one of the listed abnormalities at significant levels. A total of seven (7) tests were performed with three lots of Vysis CLL FISH Probe Kit on two days, and with one of the three lots on a third day. The second study consisted of nine (9) tests performed using three different lots of Vysis CLL FISH Probe Kit on three days on a blinded panel of eight different patient specimens with low levels of specific cytogenetic abnormalities. In both studies, the FISH signal patterns of 200 nuclei were evaluated by two technologists for each specimen. Percentage of cells with the signal pattern of interest was analyzed separately for each specimen per probe. The mean and standard deviations (SDs) of the observed percentage of abnormal cells for each specimen in both studies are summarized as follows:
{3}
| Probe and Signal Pattern | Study No | Sample ID | N | Mean | Between-Day Component SD[a] | Between-Lot Component SD | Total SD |
| --- | --- | --- | --- | --- | --- | --- | --- |
| TP53 [del(17p13.1) (1 signal)] | 2 | 7 | 9 | 13.6 | 3.11 | 0.00 | 3.11 |
| | 2 | 8 | 9 | 16.9 | 3.90 | 0.00 | 3.90 |
| | 1 | 9 | 7 | 29.8 | 5.15 | 4.54 | 6.87 |
| | 1 | 10 | 7 | 73.2 | 5.29 | 0.00 | 5.29 |
| ATM [del(11q22.3) (1 signal)] | 2 | 2 | 9 | 7.8 | 3.00 | 1.89 | 3.55 |
| | 1 | 1 | 9 | 8.4 | 3.12 | 0.91 | 3.25 |
| | 1 | 8 | 7 | 13.7 | 4.31 | 5.10 | 6.68 |
| | 2 | 8 | 9 | 18.7 | 3.46 | 0.00 | 3.46 |
| | 1 | 7 | 7 | 64.1 | 3.18 | 0.00 | 3.18 |
| D13S319 [del(13q14.3) (1 signal)] | 2 | 5 | 9 | 5.8 | 2.49 | 0.00 | 2.49 |
| | 2 | 6 | 9 | 14.9 | 2.46 | 1.15 | 2.72 |
| | 1 | 4 | 7 | 20.1 | 4.88 | 0.00 | 4.88 |
| | 1 | 10 | 7 | 53.6 | 8.57 | 0.00 | 8.57 |
| | 1 | 3 | 7 | 66.1 | 6.13 | 0.00 | 6.13 |
| | 1 | 7 | 7 | 84.9 | 4.14 | 0.00 | 4.14 |
| D13S319 [del(13q14.3) (0 signals)] | 1 | 10 | 7 | 18.4 | 6.32 | 0.87 | 6.38 |
| | 2 | 8 | 9 | 19.8 | 2.53 | 0.00 | 2.53 |
| CEP 12 [trisomy 12 (3 signals)] | 2 | 4 | 9 | 3.6 | 3.05 | 0.00 | 3.05 |
| | 2 | 3 | 9 | 8.3 | 2.77 | 0.00 | 2.77 |
| | 1 | 5 | 7 | 21.6 | 6.07 | 0.00 | 6.07 |
| | 1 | 8 | 7 | 67.3 | 4.84 | 0.00 | 4.84 |
| | 1 | 6 | 7 | 70.7 | 4.06 | 1.07 | 4.19 |
Note: [a]Between-Day (Within Lot) Component SD
# ii. Reproducibility:
Three individual laboratories tested a blinded 20-member slide panel consisting of specimens representing each of the five Döhner classifications:
13q- (monosomy or nullisomy) as sole abnormality
No cytogenetic abnormality
+12 without 17p- or 11q-
11q- without 17p-
17p-
The analysis was conducted with 10 slides per day for two days for a total of six runs across three testing sites. The same lots of the Vysis CLL FISH Probe Kit (Vysis LSI TP53 SpectrumOrange/ATM SpectrumGreen and LSI D13S319 SpectrumOrange/13q34 SpectrumAqua/CEP 12 SpectrumGreen Probes) were tested at each laboratory. Results show that there was no disagreement among the three testing sites in test results from LSI TP53 SpectrumOrange and CEP 12 SpectrumGreen probes and two disagreements among the three testing sites in test results from the other
{4}
three probes.
| Probe (Abnormality) | Disagree[a] | | | Overall | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | Site 2 | Site 3 | No of Agree | No of Tested | Percent Agreement |
| TP53 (17p-) | 0 | 0 | 0 | 20 | 20 | 100 |
| ATM (11q-) | 2 | 0 | 0 | 18 | 20 | 90 |
| CEP 12 (+12) | 0 | 0 | 0 | 20 | 20 | 100 |
| D13S319 1x (13q-) | 2 | 0 | 0 | 18 | 20 | 90 |
| D13S319 2x (13q-) | 0 | 1 | 1 | 18 | 20 | 90 |
Note: [a]Disagree = Number of specimens for which FISH result from one site did not agree with the results from the other two sites.
Additionally, Fisher's Exact Test showed that there was no significant difference within each probes or Prognostic Category among the three testing sites.
# b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
i. Real-time stability: One (1) lot of the Vysis CLL FISH probes and three (3) lots of other components of the Vysis CLL FISH Probe Kit were tested shortly after the respective expiration dating of the components under intended storage conditions as follows:
| Components | Storage Condition | Shelf life |
| --- | --- | --- |
| Vysis LSI TP53 SpectrumOrange (SO)/ ATM SpectrumGreen (SGn) | -20°C ±10°C | 24 months |
| Vysis LSI D13S319 SO/LSI 13q34 SpectrumAqua (SA)/CEP 12 SGn | -20°C ±10°C | 24 months |
| DAPI II Counterstain | -20°C ±10°C | 12 months |
| NP-40 | -20°C ±10°C | 36 months |
| 20X SSC | -20°C ±10°C | 60 months |
The functionality of the Vysis CLL FISH Probe Kit components was evaluated based on the following attributes: signal intensity, target background, cross-hybridization, specificity, and overall readability. The Vysis CLL FISH Probe Kit dating is determined by the component with the shortest expiration dating. The shelf-life of Vysis CLL FISH Probe Kit was determined to be 12 months.
The ongoing real-time Kit Stability Study tests three (3) lots of Vysis LSI TP53 SO/LSI ATM SG probe and three (3) lots of the Vysis LSI D13S319 SO/LSI 13q34 SA/CEP12 SG probe. The functionality of the Vysis CLL FISH Probes was tested in triplicate at each time point using metaphase slides prepared from normal phytohemagglutinin (PHA)-stimulated peripheral blood lymphocytes. The interim test results have demonstrated 7-month stability at the ISC for the Vysis LSI TP53 SO/LSI ATM SG
{5}
probe and 4-month stability at the ISC for the Vysis LSI D13S319 SO/LSI 13q34 SA/CEP12 SG probe.
ii. Freeze-thaw Stability: A series of twenty freeze-thaw cycles (equal to the number of assays in the kit) was performed on each CLL FISH Probe Kit and the functionality of the CLL FISH Probe Kit components was evaluated based on the following attributes: signal intensity, target background, cross-hybridization, specificity, and overall readability. The stability test showed no decrease in performance.
iii. Shipping stability. The stability test showed no decrease in performance regardless of the method of shipment, (i.e. Ambient, Wet Ice, or Dry Ice).
iv. Hybridization signal stability: The slides prepared from normal male PHA-stimulated peripheral blood lymphocytes were hybridized with Vysis LSI CLL FISH probes (Vysis LSI TP53 SO/LSI ATM SG and Vysis LSI D13S319 SO/LSI 13q34 SA/CEP12 SG) and then stored at $-20^{\circ}\mathrm{C} \pm 10^{\circ}\mathrm{C}$ while protected from light. The hybridized slides were evaluated at Day 1, Day 21, and Day 42 post-hybridization with respect to signal intensity, target background, cross hybridization, specificity, and overall readability. The results showed that the slides can be stored with desiccant for up to three (3) weeks after the application of DAPI II Counterstain without significant loss in fluorescence signal intensity or the ability to count the signal patterns of 200 nuclei per slide.
d. Detection limit:
The analytical sensitivity of the probes in the Vysis CLL FISH Probe Kit for their respective chromosome target loci was established using interphase nuclei prepared from peripheral blood cultures of 25 karyotypically normal patients. For each specimen, the signal patterns of 200 nuclei were evaluated by counting the number of orange and green signals present for each probe target. Each of the two technologists evaluated 100 nuclei per specimen for a total of 200 nuclei per specimen and 5000 scoreable nuclei from normal specimens.
The sensitivity (with $95\%$ confidence intervals (CI) based on binomial distribution) was calculated as the percentage of scoreable interphase nuclei with the expected signal pattern of two signals per nucleus as follows:
| Probe | Total Number of Nuclei Scored | Number of Nuclei With Expected Signal Pattern | Sensitivity (95% CI) |
| --- | --- | --- | --- |
| Vysis LSI TP53 SpectrumOrange | 5000 | 4899 | 97.98% (97.55% – 98.35%) |
| Vysis LSI ATM SpectrumGreen | 5000 | 4934 | 98.68% (98.32% – 98.98%) |
| Vysis LSI D13S319 SpectrumOrange | 5000 | 4930 | 98.60% (98.23% – 98.91%) |
| Vysis CEP 12 SpectrumGreen | 5000 | 4947 | 98.95% (98.62% – 99.21%) |
e. Analytical specificity:
{6}
Analytical specificity is defined as the percentage of signals that hybridize to the correct locus and no other location. This test is a visual demonstration that each probe hybridizes specifically to the expected chromosome location. The analytical specificity of the Vysis CLL FISH probes for their respective chromosome target loci was established using metaphase chromosomes prepared from peripheral blood cultures of five karyotypically normal males on microscope slides. The hybridization location of each FISH signal on chromosomes of 20 consecutive metaphase nuclei on each of 5 slides was evaluated by one technologist for a total of 200 target loci.
For each probe and sample, the number of metaphase chromosome FISH signals hybridized to the correct locus and the number of metaphase chromosome FISH signals hybridized to the incorrect locus were enumerated. The specificity of each probe was calculated as the percentage of metaphase chromosome FISH signals hybridized to the correct locus as follows:
| Probe | Target | Total No of Metaphase Chromosome Hybridized | No of Metaphase Chromosome Hybridized to the Correct Target | Specificity (95% CI) |
| --- | --- | --- | --- | --- |
| Vysis LSI TP53 SpectrumOrange | 17p13.1 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis LSI ATM SpectrumGreen | 11q22.3 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis LSI D13S319 SpectrumOrange | 13q14.3 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis LSI 13q34 SpectrumAqua | 13q34 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis CEP 12 SpectrumGreen | 12p11.1-q11 | 200 | 200 | 100% (98.17% – 100%) |
f. Assay cut-off:
The normal cut-off values of the probes in the Vysis CLL FISH Probe Kit for their respective chromosome target loci were established using interphase nuclei prepared from peripheral blood cultures of 25 karyotypically normal patients. For each specimen, the signal patterns of 200 nuclei were evaluated by counting the number of orange and green signals present for each probe target.
The expected normal interphase signal pattern for all probes in the Vysis CLL FISH Probe Kit was two signals per nucleus. Since the specimen population does not fit a Gaussian distribution, the normal cut-off value was calculated using the beta inverse function as the maximum number or percentage of scoreable interphase nuclei with a specific abnormal signal pattern at which a specimen is considered normal for that signal pattern. The results are summarized as follows:
| Probe | Abnormal | Number of | Maximum | Normal Cut-off |
| --- | --- | --- | --- | --- |
| Vysis CLL FISH Probe Kit | 17p13.1 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis CLL ATM SpectrumGreen | 11q22.3 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis LSI D13S319 SpectrumOrange | 13q14.3 | 200 | 200 | 100% (98.17% – 100%) |
| Vysis LSI 13q34 SpectrumAqua | 13q34 | 200 | 200 | 100% (98.17% – 100%) |
{7}
| | Signal Pattern of Interest | Nuclei Evaluated | Number of False-Positive Patterns | Value Per 200 Nuclei (%) |
| --- | --- | --- | --- | --- |
| Vysis LSI TP53 SpectrumOrange | 1 signal | 200 | 8 | 14 (7.0%) |
| Vysis LSI ATM SpectrumGreen | 1 signal | 200 | 6 | 12 (6.0%) |
| Vysis LSI D13S319 SpectrumOrange | 1 signal | 200 | 5 | 11 (5.5%) |
| Vysis LSI D13S319 SpectrumOrange | 0 signal | 200 | 0 | 3 (1.5%) |
| Vysis CEP 12 SpectrumGreen | 3 signals | 200 | 1 | 5 (2.5%) |
# 2. Comparison studies:
a. Method comparison with predicate device Not applicable.
b. Matrix comparison: Not applicable.
c. Method concordance bridging studies:
A total of 64 de-identified specimens with Döhner classifications as determined previously based on results from the Reference FISH Test (RFT) were tested with the Vysis CLL FISH Probe Kit. They represent each of the five Döhner classifications, including no cytogenetic abnormality $(n = 12)$ , $13q-$ (monosomy or nullisomy) as sole abnormality $(n = 13)$ , $+12$ without $17p-$ or $11q-$ $(n = 14)$ , $11q-$ without $17p-$ $(n = 18)$ and $17p-$ $(n = 7)$ based on previous results using the RFT used in the Shanafelt study[a]. For each specimen tested, two technologists each scored 100 nuclei for each hybridization target by counting and recording the number of orange and green signals present in each nucleus. The results of the two technologists for each specimen were then combined to determine the Döhner class based on the specific abnormal signal patterns observed. The results showed that Vysis CLL FISH Probe Kit identified 38 of 39 specimens in Intermediate/Good prognostic category and 24 of 25 specimens in Poor prognostic category. The percent agreement between the Vysis CLL FISH Probe Kit and RFT was $97\%$ with a lower $95\%$ confidence boundary of $90\%$ .
| Prognostic Category by Vysis CLL FISH Probe Kit | Prognostic Category by RFT | | |
| --- | --- | --- | --- |
| | Intermediate/Good | Poor | Total |
| Intermediate/Good | 38 | 1 | 39 |
| Poor | 1 | 24 | 25 |
| Total | 39 | 25 | 64 |
Note: [a]Shanafelt TD, Witzig TE, Fink SR, et al., 2006. Prospective evaluation of clonal evolution during long-term follow-up of patients with untreated early-stage chronic lymphocytic leukemia. J Clin Oncol. 24(28):4634-41.
In addition, 20 blinded specimens representing each of the five Döhner
{8}
classifications were tested at three individual laboratories. Each site tested ten (10) slides per day for a total of four runs consisting of two runs using Vysis CLL FISH Probe Kit and two runs with their Reference FISH Tests. Percent agreement between test results for each probe was determined and summarized as follows:
| Probe and Signal Pattern | Number Tested | Number Agreed | Percent Agreement | Lower Boundary 95% CI |
| --- | --- | --- | --- | --- |
| TP53 [del(17p13.1) (1 signal)] | 60 | 60 | 100 | 95 |
| ATM [del(11q22.3) (1 signal)] | 60 | 60 | 100 | 95 |
| D13S319 [del(13q14.3) (1 signal)] | 60 | 58 | 97 | 90 |
| D13S319 [del(13q14.3) (0 signals)] | 60 | 56 | 93 | 85 |
| CEP 12 [trisomy 12 (3 signals)] | 60 | 60 | 100 | 95 |
3. Clinical studies:
a. Clinical Sensitivity and Specificity: Not applicable.
b. Other clinical supportive data (when a. is not applicable): Prognostic claim of the Vysis CLL FISH Probe Kit is supported by literature. In a prospective study, Shanafelt et al established the relationship between Vysis CLL FISH results and overall 5-year survival (OS) in a cohort of untreated, predominantly early-stage patients with CLL/small lymphocytic lymphoma (SLL). A total of 159 patients with previously untreated CLL (83% Rai stage 0/1) were enrolled in a prospective trial evaluating clonal evolution by interphase FISH. Patients provided baseline and follow-up specimens for interphase FISH testing during 2 to 12 years. Chromosome abnormalities were detected by FISH testing at study entry in 120 of the 159 patients. Sixty-one patients had a single abnormality, 52 patients had two abnormalities, and seven patients had three or more abnormalities. All deletions were monoallelic except for the 13q14 region. In 41 (44%) of the 93 patients who had 13q- deletions there were biallelic or concomitant monoallelic and biallelic deletions of 13q-. Chromosomal abnormalities detected by Vysis CLL FISH Probe at study entry predicted overall 5-year survival. Patients with 11q- or 17p- on baseline FISH had shorter expected OS from diagnosis or baseline FISH analysis as compared to other FISH categories. The results showed that patients with an initial FISH diagnosis of 17p- or 11q- had a poor prognosis and those with a FISH classification of normal, 13q-, or +12 had a good/intermediate FISH prognosis. Multivariable Cox regression analyses showed that FISH categories associated with survival remained associated significantly with OS (poor v good FISH, P <0.00022)
{9}
when accounting for other relevant factors such as Rai stage at study entry (P<0.00012) and age at diagnosis (P<0.000024).
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
| Abnormality | Cut-off Value (%) | Normal Range (%) | Number of Specimens in Normal Database | Abnormal Range (%) | Number of Specimens in Abnormal Database |
| --- | --- | --- | --- | --- | --- |
| 11q deletion | >6.0 | 0-3.0 | 58 | 8.0-95.0 | 16 |
| Trisomy 12 | >2.5 | 0-0.5 | 56 | 16.5-89.0 | 18 |
| 13qx1 | >5.5 | 0-4.0 | 43 | 6.0-97.5 | 31 |
| 13qx2 | >1.5 | 0-1.0 | 62 | 2.5-84.5 | 12 |
| 17p deletion | >7.0 | 0-6.0 | 66 | 10.0-97.5 | 8 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.