QUANTEX RF PLUS (LATEX, BUFFER, STANDARD, CONTROL)
K973464 · Instrumentation Laboratory CO · DHR · Oct 22, 1997 · Immunology
Device Facts
Record ID
K973464
Device Name
QUANTEX RF PLUS (LATEX, BUFFER, STANDARD, CONTROL)
Applicant
Instrumentation Laboratory CO
Product Code
DHR · Immunology
Decision Date
Oct 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5775
Device Class
Class 2
Indications for Use
quantex RF plus (latex, buffer, standard, control) is an in vitro diagnostic test for use with automated instrumentation in the quantitative determination of human rheumatoid factors in serum by turbidimetric immunoassay and is intended as an aid in the diagnosis of theumatoid arthritis.
Device Story
Quantex RF plus is an in vitro diagnostic reagent kit (latex, buffer, standard, control) used for the quantitative determination of human rheumatoid factors in serum. The device utilizes a turbidimetric immunoassay principle, designed for operation on automated clinical chemistry analyzers. By measuring turbidity changes in serum samples, the device quantifies rheumatoid factor levels. It is intended for use by laboratory professionals in clinical settings to aid in the diagnosis of rheumatoid arthritis. The output provides a quantitative concentration (IU/mL), which clinicians use to support diagnostic assessments. The device benefits patients by providing standardized, automated measurement of rheumatoid factors, facilitating timely clinical decision-making.
Clinical Evidence
Bench testing only. Method comparison studies were conducted on three automated platforms: Monarch (n=152, r=0.992), Hitachi (n=175, r=0.995), and Cobas Mira (n=261, r=0.990). Precision studies on the Monarch platform demonstrated within-run CVs < 3% and between-run CVs < 5%.
Technological Characteristics
In vitro diagnostic reagent kit consisting of latex, buffer, standard, and control. Operates via turbidimetric immunoassay on automated instrumentation. Analytical range: 0-200 IU/mL.
Indications for Use
Indicated for the quantitative determination of human rheumatoid factors in serum to aid in the diagnosis of rheumatoid arthritis using automated instrumentation.
Regulatory Classification
Identification
A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.
K160070 — Rheumatoid Factor (RF) Kit for use on SPAPLUS · The Binding Site Group , Ltd. · Dec 7, 2016
K060201 — OLYMPUS RF LATEX CALIBRATOR, AND OLYMPUS RF LATEX REAGENT WITH MODEL(S): ODC0028, AND OSR61105. · Olympus America, Inc. · Jun 6, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
K973464
Oct. 02, 1997
## SECTION 3
quantex RF plus - 510(k) SUMMARY (Summary of Safety and Effectiveness)
#### Submitted by:
Carol Marble Regulatory Affairs Engineer Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173 Phone: (617) 861-4467 (617) 861-4464 Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax: Fax:
#### Contact Persons:
Carol Marble Phone: (617) 861-4467
Betty Lane Alternate: Phone: (617) 861-4182
### Summary Prepared:
September 11, 1997
#### Name of the device:
quantex RF plus (latex, buffer, standard, control)
#### Classification name(s):
| | 866.5775 Rheumatoid factor immunological test systems | Class II |
|-------|-------------------------------------------------------|----------|
| 82DHR | System, test, rheumatoid factor | |
## Identification of predicate device(s):
K896271 quantex RF plus (latex, buffer, standard, control)
## Description of the device/intended use(s):
Biokit's quantex RF plus (latex, buffer, standard, control) is an in vitro diagnostic product for use with automated instrumentation in the quantitative determination of human theumatoid factors in serum by turbidimetric immunoassay and is intended as an aid in the diagnosis of Rheumatoid Arthritis.
## Statement of how the Technological Characteristics of the Device compare to the Predicate device:
The new quantex RF plus (latex, buffer, standard, control) is a modified version of the predicate quantex RF plus (latex, buffer, standard, control) with an analysis range extended from 0-100 IU/mL to 0-200 IU/mL. It is substantially equivalent in performance, intended use, safety and effectiveness to the predicate device as supported by the performance data and labeling.
## Summary of Performance Data:
In three different method comparison studies comparing the new quantex RF plus to the predicate, the correlation (r) on the Monarch (n=152) was 0.992, on the Hitachi (n=175) was 0.995 and on the Cobas Mira (n=261) was 0.990. Results from a precision study on a Monarch that accessed three serum samples with different levels of rheumatoid factors over multiple runs support package insert claims of within run CVs of < 3% and between run CVs of < 5%.
Page 1 of 1
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPT. OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping wave-like shapes, possibly representing human figures or abstract forms.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Carol Marble Requlatory Affairs Engineer Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173
OCT 2 2 1997
K973464 Re: Trade Name: quantex RF plus (layex, buffer, standard, control) Regulatory Class: II Product Code: DHR September 11, 1997 Dated: Received: September 12, 1997
Dear Ms. Marble: ""
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. this response to your premarket notification Please note: submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{2}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: quantex RF plus (latex, buffer, standard, control)
Indications for Use:
quantex RF plus (latex, buffer, standard, control) is an in vitro diagnostic test for use with automated instrumentation in the quantitative determination of human rheumatoid factors in serum by turbidimetric immunoassay and is intended as an aid in the diagnosis of theumatoid arthritis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Peter E. Mahari
Prescription Use
(Per 21 CFR 801.019)
✓
OR Over-The-Counter Use
ter Use
Page 1 of 1
Section 2
quantex RF plus 510(k)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.