EDIA ANTI-CCP

K062045 · Euro-Diagnostica AB · NHX · Dec 4, 2006 · Immunology

Device Facts

Record IDK062045
Device NameEDIA ANTI-CCP
ApplicantEuro-Diagnostica AB
Product CodeNHX · Immunology
Decision DateDec 4, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5775
Device ClassClass 2

Indications for Use

The EDIA™ anti-CCP test kit is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human sera and plasma. The assay is used to detect antibodies in a single specimen. The results of the assay are to be used as an aid to the diagnosis of Rheumatoid Arthritis (RA), in conjunction with other laboratory and clinical findings. The analysis should be performed by trained laboratory professionals. For in vitro diagnostic use.

Device Story

EDIA™ anti-CCP is an ELISA test kit for detecting IgG antibodies to Cyclic Citrullinated Peptides in human serum or plasma. Input: patient serum/plasma specimen. Principle: wells coated with CCP antigen capture specific antibodies; alkaline phosphatase-labeled anti-human IgG conjugate binds captured antibodies; substrate solution added to produce color intensity proportional to antibody concentration. Output: absorbance (optical density) measured and calculated against a calibrator curve to provide results in arbitrary units (U/mL). Used in clinical laboratories by trained professionals. Results aid physicians in diagnosing Rheumatoid Arthritis when combined with other clinical and laboratory findings.

Clinical Evidence

Clinical performance was evaluated using 416 RA patients and 793 non-RA controls (including healthy donors and various autoimmune/infectious disease states). Clinical sensitivity was 76% (95% CI: 72.1%-80.3%). Clinical specificity was 98.6% (782/793). Method comparison against the predicate device (n=678) showed 98.4% positive percent agreement and 99.4% negative percent agreement, with an overall agreement of 99.1%.

Technological Characteristics

ELISA-based immunoassay. Wells coated with Cyclic Citrullinated Peptides. Detection via alkaline phosphatase-labeled anti-human IgG conjugate and substrate solution. Measurement via optical density (absorbance). Standalone laboratory test kit. No specific software algorithm or connectivity described; manual or automated plate reader processing.

Indications for Use

Indicated for the detection and semi-quantitation of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human serum and plasma specimens to aid in the diagnosis of Rheumatoid Arthritis (RA). For use by trained laboratory professionals.

Regulatory Classification

Identification

A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k062045 B. Purpose for Submission: New device C. Measurand: Anti-cyclic citrullinated peptide (CCP) D. Type of Test: Enzyme-linked immunosorbent assay (ELISA) E. Applicant: Euro-Diagnostica AB F. Proprietary and Established Names: EDIA™ Anti-CCP G. Regulatory Information: 1. Regulation section: CFR 866.5775 Rheumatoid Factor Immunological Test System 2. Classification: Class II 3. Product code: NHX Antibodies, anti-cyclic citrullinated peptide (CCP) 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): The EDIA™ anti-CCP test kit is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human sera and plasma. The assay is used to detect antibodies in a single specimen. The results of the assay are to be used as an aid to the diagnosis of Rheumatoid Arthritis (RA), in conjunction with other laboratory and clinical findings. The analysis should be performed by trained laboratory professionals. For in vitro diagnostic use. 2. Indication(s) for use: The EDIA™ anti-CCP test kit is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to Cyclic Citrullinated Peptides (CCP) in human sera and plasma. The assay is used to detect antibodies in a single specimen. The results of the assay are to be used as an aid to the diagnosis of Rheumatoid Arthritis (RA), in conjunction with other laboratory and clinical findings. The analysis should be performed by trained laboratory professionals. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: A microplate reader capable of reading 540-565nm. I. Device Description: {1} The EDIA™ anti-CCP test kit consists of CCP-coated microtiter wells (12strips x 8 wells), 6 levels of anti-CCP calibrators (human plasma in buffer with protein stabilizer and sodium azide), anti-CCP reference control (human plasma in buffer with sodium azide), positive and negative controls (human plasma in buffer and sodium azide, alkaline phosphatase-labeled goat polyclonal antibody to human IgG in Tris buffer with protein stabilizer and sodium azide, enzyme substrate PMP (phenolphthalein monophosphate) in buffer, stop solution (sodium hydroxide) in EDTA and carbonate buffer, 16x wash buffer concentrate (borate buffer with sodium azide) and 5x sample diluent concentrate (phosphate buffer with protein stabilizer and sodium azide). All reagents except wash buffer and sample diluent are ready-to-use. # J. Substantial Equivalence Information: 1. Predicate device name(s): Euro-Diagnostica Immunoscan RA anti-CCP 2. Predicate 510(k) number(s): k052133 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | EDIA anti-CCP | Immunoscan anti-CCP | | Indications for Use | Aid in the diagnosis of rheumatoid arthritis (RA) in conjunction with other laboratory and clinical findings. | Same | | Method | ELISA | Same | | Capture antigen | Citrullinated synthetic peptides | Same | | Qualitative cut-off: absorbance ratio | <0.95=negative, ≤0.95 to ≥1.0=borderline, >1.0=positive | Same | | Controls | Reference, positive and negative | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | EDIA anti-CCP | Immunoscan anti-CCP | | Microtiter plate configuration | 12 strips x 8 individual wells | 12 strips x 8 wells | | Conjugate/label | Goat anti-human IgG labeled with alkaline phosphatase | Rabbit anti-human IgG labeled with horseradish peroxidase | | Substrate | PMP | TMB | | Calibrators (arbitrary units) | 0, 2, 8, 30, 100 and 300 U/mL | 25, 50, 200, 800, and 1600 U/mL | {2} 3 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Semi-quantitative cut-off: arbitrary units/mL | ≤5 U/mL= negative, >5 U/mL= positive | ≤25 U/mL= negative, >25 U/mL= positive | | Sample dilution | 1:101 | 1:50 | K. Standard/Guidance Document Referenced (if applicable): None referenced. L. Test Principle: The assay principle is ELISA. The wells are coated with cyclic citrullinated peptides. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating. The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled goat antibodies to human IgG binds to the antibodies in the wells in this second incubation. After a further washing step, detection of specific antibodies is obtained by incubation with PMP substrate solution followed by the addition of a stop solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated against a calibrator curve and the results are given in arbitrary units. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Intra-assay precision was determined by testing six different samples eight times each. | | High (U/mL) | Medium (U/mL) | Low (U/mL) | Low (U/mL) | Low (U/mL) | Low (U/mL) | | --- | --- | --- | --- | --- | --- | --- | | Mean | 173.9 | 34.0 | 9.9 | 11.8 | 7.8 | 9.7 | | S.D. | 13.8 | 0.6 | 0.2 | 0.5 | 0.1 | 0.4 | | % C.V. | 7.9 | 1.9 | 2.1 | 4.0 | 1.9 | 4.4 | Inter-assay precision was determined by testing six different samples eight times each. Results were obtained for three different runs. | | High (U/mL) | Medium (U/mL) | Low (U/mL) | Low (U/mL) | Low (U/mL) | Low (U/mL) | | --- | --- | --- | --- | --- | --- | --- | | Mean. | 183.8 | 36.6 | 9.3 | 11.9 | 7.8 | 10.6 | | S.D. | 19.5 | 3.0 | 0.9 | 0.8 | 0.7 | 0.9 | | % C.V. | 10.6 | 8.2 | 9.8 | 6.3 | 9.5 | 8.9 | Batch to batch variation was determined by testing six different samples eight times each. Results were obtained for three different batches. {3} 4 | | High (U/mL) | Medium (U/mL) | Low (U/mL) | Low (U/mL) | Low (U/mL) | Low (U/mL) | | --- | --- | --- | --- | --- | --- | --- | | Mean | 232.5 | 41.6 | 11.5 | 14.1 | 9.8 | 13.0 | | S.D. | 30.9 | 4.5 | 1.3 | 1.2 | 1.0 | 1.2 | | % C.V. | 13.3 | 10.8 | 11.2 | 8.3 | 10.4 | 9.3 | b. Linearity/assay reportable range: The assay measuring range is 0-300 U/mL, based on the calibrator concentrations. Dilution recovery was determined by testing five serial dilutions of three different patient samples. | Sample | Dilution | Mean Measured Concentration (U/mL) | Calculated Concentration (U/mL) | Dilution Corrected % Recovery | | --- | --- | --- | --- | --- | | 1 | 1/100 | 205.0 | 205.0 | 100 | | | 1/200 | 110.5 | 102.5 | 108 | | | 1/400 | 47.3 | 51.3 | 92 | | | 1/800 | 24.8 | 25.6 | 97 | | | 1/1600 | 10.8 | 12.8 | 84 | | 2 | 1/100 | 138.9 | 138.9 | 100 | | | 1/200 | 70.3 | 69.5 | 101 | | | 1/400 | 40.4 | 34.7 | 116 | | | 1/800 | 18.3 | 17.4 | 105 | | | 1/1600 | 8.7 | 8.7 | 100 | | 3 | 1/100 | 47.3 | 47.3 | 100 | | | 1/200 | 26.7 | 23.6 | 113 | | | 1/400 | 13.0 | 11.8 | 110 | | | 1/800 | 6.3 | 5.9 | 107 | | | 1/1600 | 3.0 | 3.0 | 103 | c. Traceability, Stability, Expected values (controls, calibrators, or methods): There is no recognized standard for quantitatively expressing levels of anti-CCP antibodies. The new assay is calibrated in arbitrary units based on a positive patient serum pool. The recommended stability of 12 months is based on accelerated stability studies for plates, conjugate, calibrators, controls and substrate. Real time studies are ongoing. d. Detection limit: The detection limit of the assay was determined by running the zero calibrator 12 times on three different lots. The detection limit of 0.5 U/mL was calculated by finding the mean plus two standard deviations. e. Analytical specificity: {4} Three low positive samples were spiked to the following concentrations in diluted serum samples; bilirubin F at 0.188 mg/dL, bilirubin C at 0.2 mg/dL, hemoglobin at 453 mg/dL, chyle at 0.24 U/dL and rheumatoid factor at 200 IU/mL. The data indicate that the assayed concentrations do not interfere with the anti-CCP results. f. Assay cut-off: The assay cut-off level was determined by testing sera from 416 RA patients, 531 non-RA disease patients, and 262 healthy controls from adult subjects. A ROC analysis was performed. The analysis revealed a point with the lowest possible sum of false positives and false negative samples and the cut-off was set at 5 U/mL. Using this cut-off, 99.2% of the normal sera tested were negative, 98.3% of the non-RA disease sera were negative and 76.2% of the RA patient sera were positive. 2. Comparison studies: a. Method comparison with predicate device: A total of 678 frozen, retrospective sera were tested to determine the agreement between the new assay and the predicate device. The sera included 416 samples from patients with RA and 262 from apparently healthy blood donors. The study showed the following: | | Immunoscan RA anti-CCP | | | | | --- | --- | --- | --- | --- | | | | + | - | Total | | EDIA anti-CCP | + | 317 | 2 | 319 | | | - | 5 | 354 | 359 | | | Total | 322 | 356 | 678 | | Positive percent agreement | 317/322 = 98.4% | 95%CI: 96.4 - 99.5% | | --- | --- | --- | | Negative percent agreement | 354/356 = 99.4% | 95%CI: 98.0 - 99.9% | | Overall agreement | 672/678 = 99.1% | 95%CI: 97.9 - 99.6% | c. Matrix comparison: | Plasma | N | Range tested (U/mL) | Regression analysis | | --- | --- | --- | --- | | Heparin | 18 | 0 – 133 | y = 0.9859x + 3.6405, r² = 0.9032 | | EDTA | 15 | 1 – 84 | y = 1.0369x + 0.2723, r² = 0.9899 | 3. Clinical studies: a. Clinical sensitivity: | Patients with clinically defined RA | n | negative | positive | Clinical Sensitivity (95% CI) | | --- | --- | --- | --- | --- | | | 416 | 99 | 317 | 76% (72.1%-80.3%) | b. Clinical specificity: Clinical specificity for the new assay was determined by testing 793 samples {5} from 531non-RA disease patients and 262asymptomatic healthy blood donors. | | n | negative | positive | Clinical Specificity | | --- | --- | --- | --- | --- | | Healthy blood donors | 262 | 260 | 2 | 99% | | Crohn's disease | 10 | 10 | 0 | 100% | | Colitis ulcerosa | 10 | 10 | 0 | 100% | | Systemic lupus erythematosus (SLE) | 30 | 30 | 0 | 100% | | Sjögren's syndrome | 17 | 16 | 1 | 94% | | EBV | 5 | 5 | 0 | 100% | | Parvovirus | 5 | 5 | 0 | 100% | | Mycoplasma | 9 | 9 | 0 | 100% | | Toxoplasma | 6 | 6 | 0 | 100% | | Yersinia | 8 | 8 | 0 | 100% | | Chlamydia | 5 | 4 | 1 | 80% | | Malaria | 4 | 4 | 0 | 100% | | Borrelia | 9 | 9 | 0 | 100% | | Lues | 5 | 5 | 0 | 100% | | Rubella | 5 | 5 | 0 | 100% | | Anti-thyroid peroxidase (anti-TPO) | 20 | 20 | 0 | 100% | | Osteoarthritis | 21 | 21 | 0 | 100% | | Endocarditis | 3 | 3 | 0 | 100% | | Tuberculosis | 5 | 5 | 0 | 100% | | Legionella | 4 | 4 | 0 | 100% | | Salmonella | 3 | 3 | 0 | 100% | | AST/ASH (anti-streptococcal antibody) | 3 | 3 | 0 | 100% | | Schistosomiasis | 4 | 4 | 0 | 100% | | Chaga's disease | 3 | 3 | 0 | 100% | | Scleroderma | 17 | 16 | 1 | 94% | | Multiple sclerosis | 20 | 20 | 0 | 100% | | Insulin dependent diabetes mellitus | 20 | 20 | 0 | 100% | | Polymyositis / Dermatomyositis | 20 | 20 | 0 | 100% | | Mixed connective tissue disease | 20 | 19 | 1 | 95% | | MPO-ANCA positive | 20 | 20 | 0 | 100% | | Microscopic polyangiitis | | | | | | PR3-ANCA positive | 20 | 20 | 0 | 100% | | Wegener's granulomatosis | | | | | | Ds-DNA positive | 40 | 38 | 2 | 95% | | Inflammatory bowel disease | 80 | 79 | 1 | 99% | | Non-RA autoimmune patients | 80 | 78 | 2 | 98% | | TOTAL | 793 | 782 | 11 | 98.6% | 4. Clinical cut-off: See assay cut-off 5. Expected values/Reference range: The expected value in the normal population is negative. See assay cut-off. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...