EliA Celikey IgG Immunoassay; EliA GliadinDP IgA Immunoassay; EliA GliadinDP IgG Immunoassay

K181871 · Phadia AB · MVM · Mar 1, 2019 · Immunology

Device Facts

Record IDK181871
Device NameEliA Celikey IgG Immunoassay; EliA GliadinDP IgA Immunoassay; EliA GliadinDP IgG Immunoassay
ApplicantPhadia AB
Product CodeMVM · Immunology
Decision DateMar 1, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

EliA Celikey IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to tissue transglutaminase (tTG) in human serum and EDTA-plasma. EliA Celikey IgG is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical diagnosis of patients with celiac disease in conjunction with other laboratory and clinical findings. EliA Celikey IgG uses the EliA IgG method on the instrument Phadia 2500/5000. EliA GliadinDP IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA GliadinDP IgA uses the EliA IgA method on the instrument Phadia 2500/5000. EliA GliadinDP IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA GliadinDP IgG uses the EliA IgG method on the instrument Phadia 2500/5000.

Device Story

Automated semi-quantitative solid-phase fluoroimmunoassays for detection of celiac disease-associated autoantibodies (tTG IgG, gliadin IgA/IgG). Input: human serum or plasma samples. Process: samples incubated in polystyrene wells coated with recombinant tTG or synthetic deamidated gliadin peptides; bound antibodies detected via enzyme-labeled anti-IgG or anti-IgA conjugate; fluorescence measured after development/stop solution addition. Operation: performed on Phadia 2500/5000 instruments; random access; automated pipetting using disposable tips. Output: fluorescence response compared to calibrators to determine antibody concentration. Clinical use: aids diagnosis of celiac disease; results interpreted by clinicians alongside other laboratory/clinical findings.

Clinical Evidence

Bench testing only. Precision, linearity, LoD/LoQ, and instrument comparison studies performed. Instrument comparison (n=100-105 samples per assay) showed high agreement with predicate (R > 0.99). Reference range established using 400 healthy donors. Clinical performance data was previously reviewed in K062583 and K093459.

Technological Characteristics

Polystyrene wells coated with recombinant human tTG or synthetic deamidated gliadin peptides. Detection via beta-galactosidase labeled anti-IgG/IgA conjugate. Fluorescence-based measurement. Automated system (Phadia 2500/5000) with disposable pipette tips. Standards: CLSI EP05-A3, EP6-A, EP17-A2.

Indications for Use

Indicated for the in vitro semi-quantitative measurement of IgG antibodies to tissue transglutaminase (tTG) or IgA/IgG antibodies to gliadin in human serum or plasma to aid in the diagnosis of celiac disease in conjunction with other clinical findings.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY ONLY TEMPLATE A. 510(k) Number: K181871 B. Purpose for Submission: Adding previously cleared assays on a new instrument platform (Phadia 2500/5000) C. Measurand: IgG autoantibodies specific to tissue transglutaminase (tTG) IgA autoantibodies specific to gliadin IgG autoantibodies specific to gliadin D. Type of Test: Automated semi-quantitative solid-phase fluoroimmunoassays E. Applicant: Phadia AB F. Proprietary and Established Names: EliA Celikey IgG Immunoassay EliA GliadinDP IgA Immunoassay EliA GliadinDP IgG Immunoassay G. Regulatory Information: 1. Regulation section: 21 CFR §866.5660, Multiple autoantibodies immunological test system 21 CFR §866.5750, Radioallergosorbent (RAST) Immunological Test System 2. Classification: Class II 3. Product code: MVM, Autoantibodies, endomysial (tissue transglutaminase) MST, Autoantibodies, gliadin 4. Panel: Immunology (82) {1} H. Intended Use: 1. Intended uses: EliA Celikey IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to tissue transglutaminase (tTG) in human serum and EDTA-plasma. EliA Celikey IgG is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical diagnosis of patients with celiac disease in conjunction with other laboratory and clinical findings. EliA Celikey IgG uses the EliA IgG method on the instrument Phadia 2500/5000. EliA GliadinDP IgA is intended for the in vitro semi-quantitative measurement of IgA antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA GliadinDP IgA uses the EliA IgA method on the instrument Phadia 2500/5000. EliA GliadinDP IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA GliadinDP IgG uses the EliA IgG method on the instrument Phadia 2500/5000. 2. Indications for use: Same as intended uses 3. Special conditions for use statement: For prescription use only 4. Special instrument requirements: For use on the Phadia 2500 and Phadia5000 instruments I. Device Description: The method-specific reagents are identical with K062583 (EliA Celikey IgG) and K093459 (EliA GliadinDP IgA and EliA GliadinDP IgG), but are filled in containers specific for the Phadia 2500/5000 instrument. Each device consists of the following: EliA Assay-specific Reagents: - EliA Celikey IgG Wells are coated with human recombinant tissue transglutaminase (tTG), ready to use - EliA GliadinDP IgA Wells are coated with synthetic deamidated gliadin peptides, ready to use - EliA GliadinDP IgG Wells are coated with synthetic deamidated gliadin peptides, ready to use EliA Method-specific Reagent: - EliA Sample Diluent: PBS containing BSA, detergent and 0.095% (w/v) sodium azide ready to use 2 {2} EliA IgG reagents (required for EliA Celikey IgG and EliA GliadinDP IgG): - EliA IgG Conjugate 50 or 200: β-Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide, ready to use - EliA IgG Calibrator Strips: Human IgG (0, 4, 10, 20, 100, 600 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use - EliA IgG Curve Control Strips: Human IgG (20 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use - EliA IgG Calibrator Well: Coated with mouse monoclonal antibodies, ready to use EliA IgA reagents (required for EliA GliadinDP IgA): - EliA IgA Conjugate 50 or 200: β-Galactosidase labeled anti-IgA (mousemonoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide, ready to use - EliA IgA Calibrator Strips: Human IgA (0, 0.3, 1.5, 5, 15, 80 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use - EliA IgA Curve Control Strips: Human IgA (20 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide, ready to use - EliA IgA Calibrator Well: Coated with mouse monoclonal antibodies, ready to use The EliA reagents are available as modular packages, each purchased separately. All packages are required to carry out EliA Celikey IgG, EliA GliadinDP IgA, and EliA GliadinDP IgG tests. J. Substantial Equivalence Information: 1. Predicate device names and Predicates: EliA Celikey IgG on Phadia 250 instrument, K062583 EliA GliadinDP IgA on Phadia 250 instrument, K093459 EliA GliadinDP IgG on Phadia 250 instrument, K093459 2. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device *Phadia 2500/5000 EliA Celikey IgG and EliA GliadinDP IgA/IgG (K181871) | Predicate Phadia 250 EliA Celikey IgG (K062583) EliA GliadinDP IgA/IgG (K093459) | | Intended Use: EliA Celikey IgG | EliA Celikey IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to tissue transglutaminase (tTG) in human serum and EDTA-plasma. EliA Celikey IgG is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the clinical | EliA Celikey IgG is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to tissue transglutaminase (tTG) in human serum and plasma. EliA Celikey IgG is based on recombinant human tissue transglutaminase as antigen and is useful as an aid in the | {3} | Similarities | | | | --- | --- | --- | | Item | Device *Phadia 2500/5000 EliA Celikey IgG and EliA Gliadin^{DP} IgA/IgG (K181871) | Predicate Phadia 250 EliA Celikey IgG (K062583) EliA Gliadin^{DP} IgA/IgG (K093459) | | | diagnosis of patients with celiac disease in conjunction with other laboratory and clinical findings. EliA Celikey IgG uses the EliA IgG method on the instrument Phadia 2500/5000. | clinical diagnosis of patients with Celiac disease. EliA Celikey IgG uses the EliA IgG method on the instrument ImmunoCAP 250*. (* Former name of Phadia 250) | | Intended Use: EliA Gliadin^{DP} IgA/IgG | EliA Gliadin^{DP} IgA/IgG is intended for the in vitro semi-quantitative measurement of IgA/IgG antibodies directed to gliadin in human serum or plasma (Li-heparin, EDTA) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA Gliadin^{DP} IgA/IgG uses the EliA IgA/IgG method on the instrument Phadia 2500/5000. | EliA Gliadin^{DP} IgA/IgG is intended for the in vitro semi-quantitative measurement of IgA/IgG antibodies directed to gliadin in human serum and plasma (heparin, EDTA, citrate) to aid in the diagnosis of celiac disease in conjunction with other laboratory and clinical findings. EliA Gliadin^{DP} IgA/IgG uses the EliA IgA/IgG method on the instrument Phadia 250. | | Signal | Immuno-fluorescence measurement | Same | | Assay type | ELISA | Same | | Type of test | Semi-quantitative | Same | | Antigen: Celikey IgG | Human recombinant transglutaminase (tTG) | Same | | Gliadin^{DP} IgA/IgG | Synthetic deaminated gliadin peptides | Same | | Common, dedicated Phadia reagents | Same. Introduction of new article numbers for Development Solution, Stop Solution and Washing Solution is only due to larger filling volumes which are required for the bigger instruments Phadia 2500/5000 | Same | | Time to 1st result | About two hours | Same | | Detection antibody | | | {4} | Similarities | | | | --- | --- | --- | | Item | Device *Phadia 2500/5000 EliA Celikey IgG and EliA Gliadin^{DP} IgA/IgG (K181871) | Predicate Phadia 250 EliA Celikey IgG (K062583) EliA Gliadin^{DP} IgA/IgG (K093459) | | (conjugate): Celikey IgG and Gliadin^{DP} IgG | β-Galactosidase anti-IgG, mouse monoclonal antibodies | Same | | Gliadin^{DP} IgA | β-Galactosidase anti-IgA, mouse monoclonal antibodies | Same | | Incubation temperature | 37°C | Same | | Sample matrix: Celikey IgG | Serum and plasma (EDTA) | Same | | Calibrator strips: Celikey IgG and Gliadin^{DP} IgG | Human IgG 0, 4, 10, 20, 100, 600 μg/L | Same | | Gliadin^{DP} IgA | Human IgA 1, 0.3, 1.5, 5, 15, 80 μg/L | Same | | Control: Celikey IgG and Gliadin^{DP} IgG | EliA IgG curve control strips, 20 μg/L | Same | | Gliadin^{DP} IgA | EliA IgA curve control strips, 5μg/L | Same | | Cutt-off | Negative <7 Equivocal 7-10 Positive >10 | Same | | Assay set-up | Random access | Same | | Result calculation software | Phadia Information Data Manager (IDM) | Same | | Sample volume | 90 μL (20 μL of non-diluted sample) | Same | | Conjugate volume | 90 μL | Same | | Development solution | 90 μL | Same | | Stop solution | 200 μL | Same | {5} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Instrument | Phadia 2500/5000 | Phadia 250 | | Sample matrix: GliadinDP IgA/ IgG | Serum and plasma (Li-heparin, EDTA) | Serum and plasma (Li-heparin, EDTA, citrate ) | | Measuring range: Celikey IgG GliadinDP IgA GliadinDP IgG | LoQ to upper limit 1.7 to 227 EliA U/mL0.4 to 142 EliA U/mL 1.4 to 302 EliA U/mL | LoD to upper limit 0.6 to ≥ 600 EliA U/mL 0.1 to ≥ 142 EliA U/mL 0.4 to ≥ 132 EliA U/mL | | Sample dilution, pipets | Disposable Pipette Tips in Racks for pipetting samples in Dilution Well | Steel pipette to dilute the samples in Dilution Plates | | Risk for carry-over | When running EliA tests on the Phadia 2500/5000 instruments, there is no need for this warning statement because these instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, and carry-over from samples to conjugate is impossible. | The warning “DO NOT REUSE” in the Phadia 250 DFU for EliA Conjugates is due to the fact that a low risk of conjugate contamination by carry-over from samples was identified. In order to reduce the risk, the single use statement for the conjugate was included in the Phadia 250 DFU. | | Loading of EliA Carriers | The Phadia 2500/5000 instruments do not have Loading Tray. The EliA carriers are loaded into racks which are directly transferred to the cooled storage compartment | EliA carriers are loaded manually on the Loading Tray from where they can be processed directly or transferred to the cooled storage compartment. | | Barcode reader | On the Phadia 2500/5000 the reagents are on a moving belt which conveys them past the barcode reader. The lot-specific information will be read automatically by the instrument during loading. | The Phadia 250 instrument has a built-in barcode reader at the front of the instrument, but the operator needs to scan the barcodes manually by showing the reagents to the barcode reader. Alternatively, the operator can also enter the characters below the barcode manually. | | Process time / Time to patient result | Phadia 2500/5000 instruments process two Wells in parallel in 48 seconds. Phadia 2500/5000 provides the results at a 24 seconds interval. | Phadia 250 needs one minute to process one Well. Phadia 250 provides the results at a one minute interval. | | Daily throughput | About 2500/5000 tests | About 250 tests | {6} *Phadia 2500/5000 – Phadia 2500 and Phadia 5000 are identical instruments except for sample throughput. The difference is that Phadia 2500 consists of one process module (two process lines) and Phadia 5000 consists of 2 process modules (2x2 process lines). For that reason Phadia uses the Phadia 2500 and Phadia 5000 instrument names interchangeably. ## K. Standard/Guidance Document Referenced: CLSI EP05-A3; Evaluation of Precision Performance of Quantitative Measurement Methods CLSI EP6-A Evaluation of the Linearity of Quantitative Measurement Procedures CLSI EP17-A2, Protocols for Determination of Limits of Detection and Limits of Quantification ## L. Test Principle: The EliA wells are molded cups comparable to excised wells from a microtiter plate. They are made of polystyrene and are coated with the respective antigen. The EliA wells are coated with human recombinant tissue transglutaminase for Celikey IgG, or with synthetic deamidated gliadin peptides for Gliadin$^{\mathrm{DP}}$ IgA and Gliadin$^{\mathrm{DP}}$ IgG. If present in the patient's specimen, antibodies to these proteins bind to the specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgA or IgG antibodies (EliA IgA Conjugate or EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the more specific IgA or IgG is present in the specimen. To evaluate test results, the response for patient samples is compared directly to the response for calibrators. ## M. Performance Characteristics: 1. Analytical performance: All results presented below met the manufacturer’s predetermined acceptance criteria. a. Precision: The imprecision of the assay was assessed in a study with a total of 21 runs (3 instruments × 7 runs, 28 replicates/instrument/per sample) using serum native patient samples. The studies were performed with 1 run/day over a period of 7 days. Each sample was tested in four replicates/run giving in total 84 replicates per sample. The data was calculated against the calibration curve from Day 1. The results are summarized in the table below: {7} EliA Celikey IgG on Phadia 2500/5000 | Mean (EliA U/mL) | Within-run | | Between-run | | Between-Instruments | | Total Total Imprecision | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 1.6 | 0.2 | 10.0 | 0.1 | 7.2 | 0.4 | 25.0 | 0.4 | 27.9 | | 7.6 | 0.2 | 2.8 | 0.2 | 3.1 | 0.3 | 4.2 | 0.4 | 5.9 | | 9.6 | 0.3 | 2.7 | 0.3 | 2.6 | 0.5 | 5.4 | 0.6 | 6.6 | | 100.4 | 2.7 | 2.6 | 3.1 | 3.0 | 3.4 | 3.3 | 5.3 | 5.1 | | 274.6 | 8.4 | 3.1 | 6.9 | 2.5 | 9.5 | 3.5 | 14.5 | 5.3 | EliA Gliadin$^{\text{DP}}$ IgA on Phadia 2500/5000 | Mean (EliA U/mL) | Within-run | | Between run | | Between Instruments | | Total Imprecision | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 0.8 | 0.1 | 8.9 | 0.0 | 5.7 | 0.1 | 14.5 | 0.1 | 18.2 | | 7.4 | 0.2 | 3.0 | 0.1 | 1.5 | 0.1 | 1.3 | 0.3 | 3.6 | | 8.7 | 0.2 | 2.8 | 0.2 | 2.6 | 0.2 | 2.3 | 0.4 | 4.5 | | 42.8 | 1.5 | 3.6 | 0.9 | 2.1 | 0.2 | 2.8 | 2.1 | 5.0 | | 135.3 | 5.8 | 4.3 | 3.8 | 2.8 | 10.4 | 7.7 | 12.5 | 9.3 | EliA Gliadin$^{\text{DP}}$ IgG on Phadia 2500/5000 | Mean (EliA U/mL) | Within-run | | Between run | | Between Instruments | | Total Imprecision | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 3.6 | 0.3 | 8.2 | 0.2 | 5.2 | 0.3 | 8.6 | 0.5 | 13.0 | | 7.2 | 0.2 | 3.3 | 0.4 | 5.4 | 0.2 | 3.1 | 0.5 | 7.0 | | 9.3 | 0.3 | 2.8 | 0.5 | 5.2 | 0.0 | 0.0 | 0.5 | 5.9 | | 73.7 | 2.6 | 3.5 | 4.9 | 6.7 | 2.2 | 3.0 | 6.0 | 8.1 | | 219.6 | 8.2 | 3.7 | 13.2 | 6.0 | 6.8 | 3.1 | 17.0 | 7.7 | b. Linearity/assay reportable range: Four patient serum samples were tested with one batch of EliA Celikey IgG, EliA Gliadin$^{\text{DP}}$ IgA and EliA Gliadin$^{\text{DP}}$ IgG Immunoassays and one set of system reagents on Phadia 2500/5000. The samples were serially diluted using the EliA Sample Diluent and tested in triplicates. The ratio of observed/expected (O/E) values were calculated. Observed values were graphed against the calculated values and a linear regression was performed. The results of the regression analysis are summarized below: {8} EliA Celikey IgG on Phadia 2500/5000 | Dilution range (EliA U/mL) | Slope | 95% CI | Intercept | 95% CI | R² | | --- | --- | --- | --- | --- | --- | | 1.3–35.5 | 1.03 | 0.95–1.10 | 0.49 | −0.71–1.69 | 0.99 | | 1.8–2.9 | 1.02 | 0.93–1.11 | 0.69 | −0.69–2.07 | 0.99 | | 2.2–96.6 | 1.02 | 0.98–1.06 | 1.68 | −1.18–4.54 | 1.00 | | 7.1–174.0 | 1.01 | 0.97–1.05 | 2.26 | −1.11–5.63 | 1.00 | | 2.8–227.8 | 1.04 | 0.98–1.10 | 2.48 | −2.75–7.72 | 0.99 | The reportable range (Limit of Detection, upper limit) for EliA Celikey IgG is from 0.6–227 EliA U/mL. The measuring range (Limit of Quantitation, upper limit) is from 1.7–227 EliA U/mL. EliA Gliadin$^{\text{DP}}$ IgA on Phadia 2500/5000 | Dilution range (EliA U/mL) | Slope | 95% CI | Intercept | 95% CI | R² | | --- | --- | --- | --- | --- | --- | | 0.4–28.5 | 1.00 | 0.99–1.01 | −0.04 | −0.16–0.09 | 1.00 | | 1.3–137.8 | 0.99 | 0.95–1.02 | −1.69 | −3.39–0.00 | 1.00 | | 0.6–26.9 | 1.00 | 0.98–1.02 | −0.30 | −0.54–−0.05 | 1.00 | | 2.1–162.0 | 0.99 | 0.98–1.00 | 0.79 | 0.02–1.57 | 1.00 | The reportable range (Limit of Detection, upper limit) and the measuring range (Limit of Quantitation, upper limit) for EliA Gliadin$^{\text{DP}}$ IgA is from 0.4–142 EliA U/mL. The measuring range (Limit of Quantitation, upper limit) is from 0.4 to 142 EliA U/mL. EliA Gliadin$^{\text{DP}}$ IgG on Phadia 2500/5000 | Dilution range (EliA U/mL) | Slope | 95% CI | Intercept | 95% CI | R² | | --- | --- | --- | --- | --- | --- | | 0.6–26.3 | 0.99 | 0.96–1.02 | −0.32 | −0.65–0.02 | 1.00 | | 0.8–54.4 | 1.00 | 0.98–1.01 | −0.50 | −0.86–−0.13 | 1.00 | | 7.0–334.9 | 0.98 | 0.92–1.03 | −5.65 | −13.85–2.56 | 0.99 | | 7.3–288.3 | 1.00 | 0.99–1.01 | 1.02 | −0.31–2.35 | 1.00 | The reportable range (Limit of Detection, upper limit) for EliA Gliadin$^{\text{DP}}$ IgG is from 0.6 to 302 EliA U/mL. The measuring range (Limit of Quantitation, upper limit) is from 1.4 to 302 EliA U/mL. The following statements are included in the package inserts: "Please note that concentration values between LoD and LoQ may show a higher uncertainty." "Please note that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range." High dose hook effect: The hook effect was previously reviewed in the predicate devices (K062583 and K093459). c. Traceability, Stability, Expected values (controls, calibrators, or methods): The EliA IgG stability and calibration method was previously reviewed in K061165. {9} The EliA IgA stability and calibration method was previously reviewed in K062787. # Controls: Controls are sold separately from the assay. Calibrators are traceable to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A,G and M from World Health Organization (WHO). # d. Detection limit: The limit of blank (LoB), limit of detection (LoD), and limit of quantitation studies were performed on the Phadia 2500/5000 instrument. One blank sample was run in 33 replicates in two runs for a total of 66 observations. The LoD was measured by assaying three low level samples (11 replicates for each sample in each run) using two assay runs, for a total of 66 observations. Three additional low level samples were used for the LoD determination of the EliA Celikiey IgG for a total of 132 determinations. The LoB and LoD were calculated according to CLSI guideline EP17-A2. To estimate the LoQ, a target level for the CV% of $20\%$ has been defined. The mean value of the sample with the CV% closest to and below the precision target of $20\%$ CV is selected as LoQ. The results are summarized in the table below: | Phadia 2500/5000 | Units | LoB | LoD | LoQ | | --- | --- | --- | --- | --- | | EliA Celikey IgG | EliA U/mL | 0.0 | 0.6 | 1.7 | | EliA GliadinDP IgA | EliA U/mL | 0.0 | 0.2 | 0.4 | | EliA GliadinDP IgG | EliA U/mL | 0.3 | 0.6 | 1.4 | # e. Analytical specificity: Interference: Previously reviewed in K062583 for Celikey $^{\text{R}}$ IgG and in K093459 for Gliadin $^{\text{DP}}$ IgA and IgG. Carry-over: Phadia 2500/5000 instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, therefore carry-over from samples to conjugate was not evaluated. # f. Assay cut-off: The three assays' cut-offs are the same as in the predicate devices, K062583 Celikey $^{\text{R}}$ IgG and in K093459 for Gliadin $^{\text{DP}}$ IgA and IgG and are summarized in the table below: | EliA Celikey IgG, EliA GliadinDP IgA, EliA GliadinDP IgG | | | --- | --- | | Negative | < 7 EliA U/mL | | Equivocal | 7 – 10 EliA U/mL | | Positive | > 10 EliA U/mL | {10} 11 2. Comparison studies: a. Method comparison with predicate device: Instrument comparison was performed with the predicate device, see 2c below. b. Matrix comparison: Matrix comparison between serum and plasma was previously reviewed in K062583 for Celikey IgG and in K093459 for GliadinDP IgA and IgG. c. Instrument comparison The purpose of this study is to evaluate conformance and show comparability of the EliA assays on the instruments Phadia 250 (PH250) versus Phadia 2500/5000 (PH2500/5000). The samples were run in single replicates on one PH250 instrument and three PH2500/5000 (A, B, C) instruments for a total of four measurments per sample. Only samples inside the measuring range were included in the calculations. Results were analyzed by Passing-Bablock regression. EliA Celikey IgG: 100 samples were used in the study, 77 positive, 23 negative and 10 equivocal samples. The results are summarised below: | Instrument | Intercept | 95% CI | Slope | 95% CI | R | | --- | --- | --- | --- | --- | --- | | PH2500/5000 A | 0.49 | 0.26–0.73 | 0.93 | 0.91–0.94 | 0.998 | | PH2500/5000 B | 0.14 | –0.19–0.46 | 0.96 | 0.93–1.00 | 0.992 | | PH2500/5000 C | –0.28 | –0.55––0.00 | 0.98 | 0.97–1.00 | 0.998 | Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) were evaluated with equivocal results considered positive in the table below: | Criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C | | --- | --- | --- | --- | | PPA | 97.4% | 97.4% | 96.1% | | 95% CI | 90.9% – 99.7% | 90.9% – 99.7% | 89.0% – 99.2% | | NPA | 82.6% | 87.0% | 91.3% | | 95% CI | 61.2% – 95.0% | 66.4% – 97.2% | 72.0% – 98.9% | | TPA | 94.0% | 95.0% | 95.0% | | 95% CI | 87.4% – 97.8% | 88.7% – 98.4% | 88.7% – 98.4% | {11} Results when equivocal results are considered negative are presented in the table below: | Criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C | | --- | --- | --- | --- | | PPA | 95.5% | 97.0% | 94.0% | | 95% CI | 87.5% – 99.1% | 89.6% – 99.6% | 85.4% – 98.3% | | NPA | 100% | 100% | 100% | | 95% CI | 89.4% – 100% | 89.4% – 100% | 89.4% – 100% | | TPA | 97.0% | 98.0% | 96.0% | | 95% CI | 91.5% – 99.4% | 93.0% – 99.8% | 90.1% – 98.9% | # EliA Gliadin $^{\text{DP}}$ IgA: 105 samples were used in the study, 79 positive, 21 negative and 5 equivocal samples. The results are summarised below: | Instrument | Intercept | 95% CI | Slope | 95% CI | R | | --- | --- | --- | --- | --- | --- | | PH2500/5000 A | 0.60 | 0.40–0.93 | 0.95 | 0.92–0.98 | 0.993 | | PH2500/5000 B | 0.55 | 0.28–0.83 | 0.97 | 0.95–1.00 | 0.994 | | PH2500/5000 C | -0.24 | -0.71–-0.06 | 1.09 | 1.06–1.11 | 0.994 | Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) were evaluated with equivocal results considered positive in the table below: | Criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C | | --- | --- | --- | --- | | PPA | 100% | 100% | 100% | | 95% CI | 95.7% – 100% | 95.7% – 100% | 95.7% – 100% | | NPA | 90.5% | 85.7% | 90.5% | | 95% CI | 69.6% – 98.8% | 63.7% – 97.0% | 69.6% – 98.8% | | TPA | 98.1% | 97.1% | 98.1% | | 95% CI | 93.3% – 99.8% | 91.9% – 99.4% | 93.3% – 99.8% | Results when equivocal results are considered negative are presented in the table below: | Criteria | PH2500/5000 A | PH2500/5000 A | PH2500/5000 A | | --- | --- | --- | --- | | PPA | 98.7% | 100.0% | 98.7% | | 95% CI | 93.2% – 100% | 95.4% – 100% | 93.2% – 100% | | NPA | 96.2% | 96.2% | 96.2% | | 95% CI | 80.4% – 99.9% | 80.4% – 99.9% | 80.4% – 99.9% | | TPA | 98.1% | 99.0% | 98.1% | | 95% CI | 93.3% – 99.8% | 94.8% – 100% | 93.3% – 99.8% | {12} EliA GliadinDP IgG: 104 samples were used in the study, 80 positive, 19 negative and 5 equivocal samples. The results are summarised below: | Instrument | Intercept | 95% CI | Slope | 95% CI | R | | --- | --- | --- | --- | --- | --- | | PH2500/5000 A | −0.14 | −0.41–0.13 | 1.06 | 1.04–1.09 | 0.997 | | PH2500/5000 B | 0.43 | 0.13–1.16 | 0.99 | 0.94–1.02 | 0.994 | | PH2500/5000 C | −0.06 | −0.20–0.20 | 1.08 | 1.07–1.10 | 0.998 | Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) were evaluated with equivocal results considered positive in the table below: | Criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C | | --- | --- | --- | --- | | PPA | 100.0% | 100.0% | 100% | | 95% CI | 95.8% – 100% | 95.8% – 100% | 95.8% – 100% | | NPA | 89.5% | 100.0% | 94.7% | | 95% CI | 66.9% – 98.7% | 82.4% – 100.0% | 74.0% – 99.9% | | TPA | 98.1% | 100.0% | 99.0% | | 95% CI | 93.2% – 99.8% | 96.5% – 100% | 94.8% – 100% | Results when equivocal results are considered negative are presented in the table below: | Criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C | | --- | --- | --- | --- | | PPA | 98.8% | 100% | 100% | | 95% CI | 93.2% – 100% | 95.5% – 100% | 95.5% – 100% | | NPA | 100% | 95.8% | 91.7% | | 95% CI | 85.8% – 100% | 78.9% – 99.9% | 73.0% – 99.0% | | TPA | 99.0% | 99.0% | 98.1% | | 95% CI | 94.8% – 100% | 94.8% – 100% | 93.2% – 99.8% | 3. Clinical studies: Clinical performance was previously reviewed in Previously reviewed in K062583 for Celikey $^{\text{R}}$ IgG and in K093459 for Gliadin $^{\text{DP}}$ IgA and IgG. 4. Clinical cut-off: See assay cut-off {13} 14 5. Expected values/Reference range: A total of 400 apparently healthy Caucasian blood donors matched by age and sex were tested on the Phadia 2500/5000 instrument to evaluate expected values in the normal population. The results are summarized below: | Test (n=400) | Mean (EliA U/mL) | Median (EliA U/mL) | Range | 95th percentile | 99th percentile | | --- | --- | --- | --- | --- | --- | | EliA Celikey IgG | 2.3 | 2.3 | 0.1–6.1 | 3.3 | 3.9 | | EliA Gliadin^{DP} IgA | 1.8 | 1.4 | 0.0–16.1 | 3.8 | 7.1 | | EliA Gliadin^{DP} IgG | 1.6 | 0.9 | 0.1–51.5 | 2.7 | 13.3 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...