THERATEST EL-TTG IGA/IGG AND THERATEST EL-GLIA IGA/IGG TEST SYSTEMS
K071692 · Theratest Laboratories, Inc. · MVM · Nov 5, 2007 · Immunology
Device Facts
Record ID
K071692
Device Name
THERATEST EL-TTG IGA/IGG AND THERATEST EL-GLIA IGA/IGG TEST SYSTEMS
Applicant
Theratest Laboratories, Inc.
Product Code
MVM · Immunology
Decision Date
Nov 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
The TheraTest EL-tTG™ IgA/IgG and TheraTest EL-GLIA™ IgA/IgG Kits are enzyme-linked immunosorbent assay (ELISA) test systems for the semi-quantitative measurement of IgA and IgG anti-tissue transglutaminase (tTG) and anti-gliadin antibodies in human serum. Detection and semi-quantitation of these antibodies is intended to aid the diagnosis of patients with gluten sensitive enteropathies: celiac disease and dermatitis herpetiformis, in conjunction with other clinical findings and laboratory tests.
Device Story
ELISA-based test system for semi-quantitative measurement of IgA and IgG antibodies against tissue transglutaminase (tTG) or gliadin in human serum; utilizes 96-well microtiter plates coated with recombinant human tTG or purified gliadin; patient serum samples incubated in wells; anti-tTG or anti-gliadin antibodies bind to solid-phase antigen; unbound material washed away; isotype-specific horseradish-peroxidase labeled anti-human immunoglobulin conjugate added; chromogenic substrate (TMB) added; color intensity proportional to bound antibody concentration measured via spectrophotometer at 450 nm; used in clinical laboratory settings by trained personnel; results aid diagnosis of celiac disease and dermatitis herpetiformis in conjunction with other clinical findings.
Clinical Evidence
No clinical data provided in the document; bench testing only.
Technological Characteristics
Solid-phase ELISA; 96-well microtiter plates coated with recombinant human tTG or purified gliadin; colorimetric detection using horseradish-peroxidase conjugate and TMB substrate; requires spectrophotometer (450 nm); semi-quantitative output in arbitrary units; manual or automated wash steps.
Indications for Use
Indicated for patients suspected of having gluten-sensitive enteropathies, specifically celiac disease and dermatitis herpetiformis. Used as an aid in diagnosis in conjunction with other clinical and laboratory findings.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Related Devices
K964344 — IGG ANTI-GLIADIN (IGG-AGA) ANTIBODY TEST KIT · Immco Diagnostics, Inc. · May 30, 1997
K052439 — AESKULISA GLIA A AND AESKULISA GLIA G · Aesku Diagnostics · Feb 7, 2006
K992878 — ANTI-TISSUE TRANSGLUTAMINASE (TTG)ANTIBODY TEST · Immco Diagnostics, Inc. · Oct 15, 1999
K123713 — IMMULISA ENHANCED CELIAC FUSION (TTG/DGP) IGA/IGG ANTIBODY ELISA · Immco Diagnostics, Inc. · Oct 25, 2013
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k071692
B. Purpose for Submission:
New device
C. Measurand:
IgA and IgG Anti-tissue Transglutaminase (tTG) antibodies
IgA and IgG Anti-Gliadin antibodies
D. Type of Test:
Semi-quantitative
E. Applicant:
TheraTest Laboratories, Inc.
F. Proprietary and Established Names:
TheraTest EL-tTG™ IgA/IgG
TheraTest EL-Glia™ IgA/IgG
G. Regulatory Information:
1. Regulation section:
21 § CFR 866.5660 Multiple autoantibodies immunological test system
21 § CFR 866.5750 Radioallergosorbent (RAST) immunological test system
2. Classification:
Class II
3. Product codes:
MVM, Autoantibodies, Endomysial (Tissue Transglutaminase)
MST, Antibodies, Gliadin
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
The TheraTest EL-tTG™ IgA/IgG and TheraTest EL-GLIA™ IgA/IgG Kits are enzyme-linked immunosorbent assay (ELISA) test systems for the semi-quantitative measurement of IgA and IgG anti-tissue transglutaminase (tTG) and anti-gliadin antibodies in human serum. Detection and semi-quantitation of these antibodies is intended to aid the diagnosis of patients with gluten sensitive enteropathies: celiac disease and dermatitis herpetiformis, in conjunction with other clinical findings and laboratory tests.
2. Indication(s) for use:
Same as Intended use.
3. Special conditions for use statement(s):
For prescription only.
4. Special instrument requirements:
Microplate reader: Spectrophotometer with single (450 nm) or dual (450 nm, 620-690 nm reference) wavelength and ELISA plate washer (optional)
I. Device Description:
Each device contains the following: microwell plate with breakaway microwell
{1}
coated with either recombinant tTG or purified gliadin antigen; plate frame; assay controls (IgA or IgG positive and IgA or IgG negative); calibrators (IgA or IgG); tTG/ Gliadin goat anti-human IgA conjugate; tTG/ Gliadin goat anti-human IgG conjugate; TMB chromogen; wash buffer (10X) and stop solution.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
ImmuLisa™ Anti-human Tissue Transglutaminase IgA
ImmuLisa™ Anti-human Tissue Transglutaminase IgG
ImmuLisa™ Anti-Gliadin Antibody IgA
ImmuLisa™ Anti-Gliadin Antibody IgG
2. Predicate K number(s):
k992878 (Anti-hu tTG IgA)
k040095 (Anti-hu tTG IgG)
k964341 (Anti-Gliadin IgA)
k964344 (Anti-Gliadin IgG)
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | New Device | Predicate Device | |
| | TheraTest EL-tTG™ IgA/IgG and TheraTest EL-GLIA™ IgA/IgG | ImmuLisa™ Anti-hu tTG Antibody IgA/IgG | ImmuLisa™ Anti-Gliadin Antibody IgG or IgA |
| Antigen | Human recombinant tTG and purified gliadin | Recombinant human tTG | Purified gliadin |
| Method | ELISA | Same | Same |
| Assay Platform | 96 well microtiter plates coated with specific antigen | Same | Same |
| Measurement | Semi-quantitative | Same | Same |
| Sample | Serum | Same | Same |
| Sample volume required | 10 μL | Same | Same |
| Microplate Reader | Spectrophotometer | Same | Same |
| Detection Method | Colorimetric | Same | Same |
| Differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate | |
| | TheraTest EL-tTG™ IgA/IgG and TheraTest EL-GLIA™ IgA/IgG | ImmuLisa™ Anti-human Tissue Transglutaminase IgA/IgG | ImmuLisa™ Anti-Gliadin Antibody (IgG or IgA) |
| Intended use | For the semi-quantitative detection of | For the semi-quantitative detection of | For the semi-quantitative detection of IgA |
| | antigen | antigen | antigen |
| Method | ELISA | Same | Same |
| Assay Platform | 96 well microtiter plates coated with specific antigen | Same | Same |
| Measurement | Semi-quantitative | Same | Same |
| Sample | Serum | Same | Same |
| Sample volume required | 10 μL | Same | Same |
| Microplate Reader | Spectrophotometer | Same | Same |
| Detection Method | Colorimetric | Same | Same |
{2}
| Differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate | |
| | IgA/IgG antibodies to human recombinant tTG and purified gliadin in human serum | IgA/IgG antibodies to recombinant human tTG in serum | or IgG antibodies to gliadin in human serum |
| Indications for Use | Aid in the diagnosis of IgA/IgG celiac disease and dermatitis herpetiformis | Anti-hu tTG IgA: Aid in the diagnosis of celiac disease and dermatitis herpetiformis | Aid in the diagnosis of celiac disease and dermatitis herpetiformis |
| Positive and Negative Control | To be diluted 1:101 with 1X Wash buffer | Pre-diluted human serum. Ready to use. | Pre-diluted human serum. Ready to use. |
| Calibrator | One calibrator | Set of four calibrators | Set of four calibrators |
| Conjugate | Goat anti-human IgA (α chain specific) and IgG (Fcγ specific) horseradish peroxidase | Anti-human IgA/IgG alkaline phosphatase conjugate | Anti-human IgA or IgG alkaline phosphatase conjugate |
| Serum Dilution | 1:101 | 1:51 | 1:51 |
| Assay wash step | 3X wash twice | 4X wash twice | 4X wash twice |
| Incubation | 30-30-15 (IgA and IgG) | 60-30-30 (IgA) 30-30-30 (IgG) | 30-30-15 (IgA and IgG) |
| Reading after Stop solution | Within 30 min | Within 1 hour | Within 1 hour |
| Result Interpretation | ≥20 U/mL: Normal 21-25: Equivocal >25: Abnormal | <20 EU/mL: Neg 20-25: Borderline >25: Positive | Children: AGA-IgA: <23 EU/mL: Neg >23: Positive AGA-IgG: <28 EU/mL: Neg >28: Positive |
{3}
| Differences | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate | |
| | | | Adults:
AGA-IgA and
AGA-IgG:
<20 EU/mL: Neg
>20: Positive |
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
The TheraTest EL-tTG™ IgA/IgG or The TheraTest EL-Gliadin™ IgA/IgG Test System is a solid phase enzyme immunoassay in a 96-well plate format for the measurement of IgA and IgG antibodies against tissue transglutaminase or gliadin. Wells are coated with human recombinant tissue transglutaminase or purified gliadin, and incubated with Specimens, Calibrators, Positive and Negative Controls. During the incubation, anti-tTG or anti-gliadin antibodies present in the test sample are bound to the solid phase antigen. The wells are subsequently washed, and isotype-specific horseradish-peroxidase labeled anti-human immunoglobulin antibody (enzyme conjugate) is added. After incubating the wells with the enzyme conjugate, unbound labeled antibody is removed by washing. A chromogenic substrate solution is then added to the wells, and the presence of antibodies to tissue transglutaminase is detected by a color change produced by the conversion of the substrate by the enzyme. The reaction is stopped, and the intensity of the color, which is proportional to the amount of the bound antibody, is read by an ELISA reader. The absorbance value in the blank well (incubated with Specimen Diluent) is subtracted from the values obtained with Specimens, Calibrators and Controls.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Intra-assay studies on anti-tTG IgA/IgG and anti-Gliadin IgA/IgG:
Three specimens (high, moderate and near cut-off) were assayed for a total of 20 times and two runs per day. The mean units for anti-tTG IgA ranged from 26 to 187 units and the mean units for anti-tTG IgG ranged from 29 to 83 units. The percent CV for anti-tTG IgA ranged from 2.8% – 10.9% and for anti-tTG IgG, 2.9% – 5.3%. The mean units for anti-Gliadin IgA ranged from 29 to 154 units and the mean units for anti-Gliadin IgG ranged from 23 to 88 units. The percent CV for anti-Gliadin IgA ranged from 4.0% – 7.6% and for anti-Gliadin IgG 2.4% – 10.8%.
Inter-assay studies on anti-tTG IgA/IgG and anti-Gliadin IgA/IgG:
Three specimens (high, moderate and near cut-off) were assayed for a total of 20 different times in one or two runs. The mean units for anti-tTG IgA ranged from 30 to 193 units and for anti-tTG IgG ranged from 27 to 83 units. The percent CV for anti-tTG IgA ranged from 3.9% – 10.1% and 7.8% – 10.7%
{4}
for anti-tTG IgG. The mean units for anti-Gliadin IgA ranged from 34 to 134 units and the mean units for anti-Gliadin IgG ranged from 29 to 65 units. The percent CV for anti-Gliadin IgA ranged from 8.4% – 10.1% and 9.0% – 10.2% for anti-Gliadin IgG.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
There are no reference standards for tTG and gliadin. The results are reported in arbitrary units.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Interference by endogenous substances: No data provided. The package insert states that grossly hemolyzed, lipemic, icteric or microbially contaminated serum should not be used in this assay.
Crossreactivity with other autoantibodies: Testing was performed with 56 sera consisting of 40 inflammatory bowel disease (IBD) and 16 thyroid disease. The positive samples were: one IBD with anti-tTG IgA, one IBD with anti-tTG IgG, two IBD with anti-gliadin IgA, one IBD with anti-gliadin IgG and one thyroid disease with anti-gliadin IgG.
f. Assay cut-off:
The cut-off value of >25 U/mL was established based on testing 100 blood bank donors. There were 51 females and 49 males with a median age of 29 years and a range from 16 to 70 years. The race distribution was 14% Hispanic, 8% Black, 4% Asian, and 74% other. The 98th percentile was used to establish the upper limit of normal for IgA and IgG Anti-tTG and the 97.5th and 94th percentile were used to establish for IgA and IgG Anti-gliadin respectively.
2. Comparison studies:
a. Method comparison with predicate device:
Testing was performed on 106 samples (42 from IgA sufficient CD patients, 14 from IgA deficient patients and 50 healthy donors). The comparative study on anti-tTG IgA had a 100% Positive Percent Agreement (40/40) (95% CI: 91%-100%); 97% Negative Percent Agreement (64/66) (95% CI: 89%-100%); and 98% Overall Agreement (104/106) (95% CI: 93%-100%). Refer to table below:
| | ImmuLisa™ Anti-hu tTG IgA | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| EL-tTG™ IgA | Positive | 40 | 2 | 42 |
| | Negative | 0 | 64 | 64 |
| | Total | 40 | 66 | 106 |
The comparative study on anti-tTG IgG had a Positive Percent Agreement of 84% (27/32) (95% CI: 67%-95%); 95% Negative Percent Agreement (62/65) (95% CI: 87%-99%); and 92% Overall agreement (89/97) (95% CI: 84%-
{5}
96%). Refer to table below:
| | ImmuLisa™ Anti-hu tTG IgG | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Equivocal* | Negative | Total |
| EL-tTG™ IgG | Positive | 27 | (4) | 3 | 30 |
| | Equivocal* | 0 | (1) | (3) | (4) |
| | Negative | 5 | (1) | 62 | 67 |
| | Total | 32 | (6) | 65 | 97 |
*Equivocals were excluded from the calculation.
The comparative study on anti-Gliadin IgA had a $95\%$ Positive Percent Agreement (21/22) (95% CI: 77%-100%); 96% Negative Percent Agreement (78/81) (95% CI: 90%-99%); and 96% Overall Agreement (99/103) (95% CI: 90%-99%). Refer to table below:
| | ImmuLisa™ Anti-Gliadin IgA | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| EL-Glia™ IgA | Positive | 21 | 3 | 24 |
| | Equivocal* | (2) | (1) | (3) |
| | Negative | 1 | 78 | 79 |
| | Total | 22 | 81 | 103 |
*Equivocals were excluded from the calculation.
The comparative study on anti-Gliadin IgG had a $81\%$ Positive Percent Agreement (29/36) (95% CI: 64%-92%); 92% Negative Percent Agreement (58/63) (95% CI: 82%-97%); and 88% Overall Agreement (87/99) (95% CI: 80%-94%). Refer to table below:
| | ImmuLisa™ Anti-Gliadin IgA | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| EL-Glia™ IgG | Positive | 29 | 5 | 34 |
| | Equivocal* | (6) | (1) | (7) |
| | Negative | 7 | 58 | 65 |
| | Total | 36 | 63 | 99 |
*Equivocals were excluded from the calculation.
# b. Matrix comparison:
Both assays use serum as the sample matrix.
# 3. Clinical studies:
# a. Clinical Sensitivity and specificity
Samples used in the studies for determination of clinical sensitivity and specificity consisted of 44 samples from Celiac disease patients, 40 IBD, 16 thyroid disease patients and 100 blood bank donors. Samples from IBD, thyroid disease and the blood donors were combined as the control group. For anti-tTG IgA, clinical sensitivity was $95\%$ (42/44) and clinical specificity was $98\%$ (153/156) (refer to table below):
{6}
7
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Celiac disease* (n=44) | Control group (n=156) | Total (n=200) |
| EL-tTG™ IgA | Positive | 42 | 3 | 45 |
| | Equivocal | 0 | 0 | 0 |
| | Negative | 2** | 153 | 155 |
| | Total | 44 | 156 | 200 |
* The group includes patients on gluten-containing and gluten-restricted diet, as well.
** IgA deficient patients
For anti-tTG IgG, clinical sensitivity was 48% (21/44) and clinical specificity was 97% (152/156) (refer to table below):
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Celiac disease* (n=44) | Control group (n=156) | Total (n=200) |
| EL-tTG™ IgG | Positive | 21 | 3 | 24 |
| | Equivocal | 4 | 1 | 5 |
| | Negative | 19 | 152 | 171 |
| | Total | 44 | 156 | 200 |
* The group includes patients on gluten-containing and gluten-restricted diet, as well.
For anti-Gliadin IgA, clinical sensitivity was 55% (24/44) and clinical specificity was 95% (148/156) (see table below):
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Celiac disease* (n=44) | Control group (n=156) | Total (n=200) |
| EL-Glia™ IgA | Positive | 24 | 4 | 28 |
| | Equivocal | 3 | 4 | 7 |
| | Negative | 17** | 148 | 165 |
| | Total | 44 | 156 | 200 |
* The group includes patients on gluten-containing and gluten-restricted diet, as well.
** Two patients in this group are IgA deficient
For anti-Gliadin IgG, clinical sensitivity was 55% (24/44) and clinical specificity was 94% (146/156) (see table below):
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Celiac disease* (n=44) | Control group (n=156) | Total (n=200) |
| EL-Glia™ IgG | Positive | 24 | 6 | 30 |
| | Equivocal | 5 | 4 | 9 |
| | Negative | 15 | 146 | 161 |
| | Total | 44 | 156 | 200 |
{7}
* The group includes patients on gluten-containing and gluten-restricted diet, as well
b. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Same as assay cut-off.
5. Expected values/Reference range:
Expected values in the normal population should be negative.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.