HEMOSIL ACUSTAR ANTI-B2 GLYCOPROTEIN-I; IGG, IGM, IGG CONTROLS, IGM CONTROLS

K091556 · Instrumentation Laboratory CO · MSV · May 21, 2010 · Immunology

Device Facts

Record IDK091556
Device NameHEMOSIL ACUSTAR ANTI-B2 GLYCOPROTEIN-I; IGG, IGM, IGG CONTROLS, IGM CONTROLS
ApplicantInstrumentation Laboratory CO
Product CodeMSV · Immunology
Decision DateMay 21, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

HemosIL AcuStar Anti-β₂ Glycoprotein-I IgG: Fully automated chemiluminescent immunoassay for the semi-quantitative measurement of anti-β₂ Glycoprotein-I (anti-β₂GPI) IgG antibodies in human citrated plasma and serum on the ACL AcuStar™, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings. HemosILTM AcuStar Anti-β₂ Glycoprotein-I IgG Controls: For the quality control of the Anti-β₂ Glycoprotein-I IgG assay performed on the ACL AcuStar. HemosIL AcuStar Anti-β₂ Glycoprotein-I IgM: Fully automated chemiluminescent immunoassay for the semi-quantitative measurement of anti-β2 Glycoprotein-I (anti-β2GPI) IgM antibodies in human citrated plasma and serum on the ACL AcuStar™ as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings. HemosIL AcuStar Anti-β2 Glycoprotein-I IgM Controls: For the quality control of the Anti-β2 Glycoprotein-I IgM assay performed on the ACL AcuStar.

Device Story

HemosIL AcuStar Anti-B2GPI IgG/IgM are chemiluminescent immunoassays for use on the ACL AcuStar instrument. Input: human citrated plasma or serum. Process: magnetic particles coated with purified B2GPI capture specific antibodies; isoluminol-labeled anti-human IgG/IgM tracer binds captured antibodies; luminescent reaction triggered; ACL AcuStar optical system measures relative light units (RLUs). Output: semi-quantitative concentration of anti-B2GPI antibodies. Used in clinical laboratories by trained personnel. Results aid clinicians in diagnosing thrombotic disorders associated with APS. Benefits include automation of testing compared to manual ELISA methods.

Clinical Evidence

Clinical study evaluated 321 frozen citrated plasma samples from patients with primary APS (PAPS), secondary APS (SAPS), SLE, SLE-like, other cardiovascular disorders, and healthy normals. IgG assay sensitivity 64.1%, specificity 90.8%. IgM assay sensitivity 29.3%, specificity 95.2%. Cut-off (20 U/mL) established at 99th percentile of 262 healthy volunteers.

Technological Characteristics

Two-step chemiluminescent immunoassay. Reagents: magnetic particles coated with purified human β₂GPI, isoluminol-labeled anti-human IgG/IgM tracer, phosphate/borate buffers. Form factor: cartridge-based system for ACL AcuStar analyzer. Connectivity: barcode-based master curve and lot-specific calibration. Sterilization: not specified.

Indications for Use

Indicated for semi-quantitative measurement of anti-β₂GPI IgG and IgM antibodies in human citrated plasma and serum to aid in diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS). For prescription use only.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k091556 B. Purpose for Submission: New device C. Measurand: Anti-β₂ Glycoprotein-I IgG and IgM autoantibodies D. Type of Test: Automated chemiluminescent immunoassay E. Applicant: Instrumentation Laboratory Co. F. Proprietary and Established Names: HemosIL AcuStar Anti-β₂ Glycoprotein-I IgG HemosIL AcuStar Anti-β₂ Glycoprotein-I IgM HemosIL AcuStar Anti-β₂ Glycoprotein-I IgG Controls HemosIL AcuStar Anti-β₂ Glycoprotein-I IgM Controls G. Regulatory Information: 1. Regulation section: 21 CFR § 866.5660, Multiple autoantibodies immunological test system 21 CFR § 862.1660, Single (specified) analyte controls (assayed and unassayed) 2. Classification: Class II (Assays) Class I (Controls) 3. Product code: MSV - Antibodies, β₂ Glycoprotein I JJX – Quality control material (assayed and unassayed) 4. Panel: Immunology (82) Chemistry (75) H. Intended Use: 1. Intended use(s): HemosIL AcuStar Anti-β₂ Glycoprotein-I IgG: Fully automated chemiluminescent immunoassay for the semi-quantitative measurement of anti-β₂ Glycoprotein-I (anti-β₂GPI) IgG antibodies in human citrated plasma and serum on the ACL AcuStar™, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings. HemosILTM AcuStar Anti-β₂ Glycoprotein-I IgG Controls: For the quality control of the Anti-β₂ Glycoprotein-I IgG assay performed on the ACL AcuStar. HemosIL AcuStar Anti-β₂ Glycoprotein-I IgM: Fully automated {1} chemiluminescent immunoassay for the semi-quantitative measurement of anti-β2 Glycoprotein-I (anti-β2GPI) IgM antibodies in human citrated plasma and serum on the ACL AcuStar™ as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings. HemosIL AcuStar Anti-β2 Glycoprotein-I IgM Controls: For the quality control of the Anti-β2 Glycoprotein-I IgM assay performed on the ACL AcuStar. 2. Indication(s) for use: See Intended Use above. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: For use on the ACL AcuStar automated chemiluminescent immunoassay analyzer (k083518). I. Device Description: The device consists of three components: 1) anti-β2GPI IgG or IgM Cartridge containing: a vial of magnetic particle suspension coated with purified human β2GPI, a vial of assay buffer, a vial of tracer consisting of a monoclonal mouse anti-human IgG or IgM antibody labeled with isoluminol, and a vial of sample diluent used for the regular pre-dilution of the sample and automatic dilution in rerun. The reagents are in a phosphate or borate buffer containing bovine serum albumin, stabilizers and preservative. 2) anti-β2GPI IgG or IgM Calibrator containing: one 1 mL barcoded tube of a solution with anti-β2GPI IgG or IgM in a phosphate buffer containing bovine serum albumin, stabilizers and preservative. 3) anti-β2GPI IgG or IgM Calibrator 2: one 1 mL barcoded tube of a solution with anti-β2GPI IgG or IgM in a phosphate buffer containing bovine serum albumin, stabilizers and preservative. The IgG and IgM control materials are sold separately. Each device consists of: three barcoded ready-to-use 1-mL Low anti-β2GPI IgG or IgM Low Control and three 1-mL Low anti-β2GPI IgG or IgM High Control. In addition to anti-β2GPI antibodies, each control contains phosphate buffer, bovine serum albumin, stabilizers and preservative. J. Substantial Equivalence Information: 1. Predicate device name(s): REAADS IgG anti-β2 GPI Test Kit REAADS IgM anti-β2 GPI Test Kit 2. Predicate K number(s): k031208 k031208 3. Comparison with predicate: {2} HemosIL™ AcuStar Anti-β2 GPI IgG | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use/Indications for Use | For the semi-quantitative measurement of anti-β_{2} Glycoprotein-I (anti-β_{2}GPI) IgG antibodies in human citrated plasma and serum on the ACL AcuStar, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other laboratory and clinical findings. | For the detection and semi-quantitation of IgG anti-β_{2} GPI antibodies in human serum or plasma as an aid for assessing the risk of thrombosis in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phosph | | Sample type | Serum or Citrated Plasma | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Technology | Two-step chemiluminescent immunoassay | ELISA | | Calibrator | Two calibrator levels (included in test kit) | Three calibrator levels (included in test kit) | | Quality control | Low and high controls (sold separately) | Normal and positive controls (included in test kit) | | Assay range | 6.4 – 6100 U/mL | Up to 100 GPL | | Cut-off | 20.0 U/mL | 20 GPL | HemosIL™ AcuStar Anti-β2 GPI IgM | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use/Indications for Use | For the semi-quantitative measurement of anti-β_{2} Glycoprotein-I (anti-β_{2}GPI) IgM antibodies in human citrated plasma and serum on the ACL AcuStar, as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome (APS) when used in conjunction with other | For the detection and semi-quantitation of IgM anti-β_{2} GPI antibodies in human serum or plasma as an aid for assessing the risk of thrombosis in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phosph | {3} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | laboratory and clinical findings. | | | Sample type | Serum or Citrated Plasma | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Technology | Two-step chemiluminescent immunoassay | ELISA | | Calibrator | Two calibrator levels (included in test kit) | Three calibrator levels (included in test kit) | | Quality control | Low and high controls (sold separately) | Normal and positive controls (included in test kit) | | Assay range | 1.1 – 841 U/mL | Up to 100 MPL | | Cut-off | 20.0 U/mL | 20 MPL | # K. Standard/Guidance Document Referenced (if applicable): CLSI EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline, Second Edition CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation Approved Guideline CLSI EP07-A2: Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition # L. Test Principle: The devices are chemiluminescent two-step immunoassays consisting of magnetic particles coated with human purified $\beta 2\mathrm{GPI}$ which capture, if present, the anti- $\beta 2\mathrm{GPI}$ antiphospholipid antibodies from the sample. After incubation, magnetic separation, and a wash step, a tracer consisting of an isoluminol-labeled anti-human IgG or IgM antibody is added and may bind with the captured anti- $\beta 2\mathrm{GPI}$ IgG or IgM on the particles. After a second incubation, magnetic separation, and wash step, reagents that trigger the luminescent reaction are added, and the emitted light is measured as relative light units (RLUs) by the ACL AcuStar optical system. The RLUs are directly proportional to the anti- $\beta 2\mathrm{GPI}$ IgG or IgM concentration in the sample. The ACL AcuStar anti- $\beta 2\mathrm{GPI}$ IgG and IgM assays utilize a 4 Parameter Logistic Curve (4PLC) fit data reduction method to generate a Master Curve. The Master Curve is predefined and lot dependent and it is stored in the instrument through the cartridge barcode. With the measurement of calibrators, the predefined Master Curve is transformed to a new, instrument specific 4PLC Working Curve. The concentration values of the calibrators are included in the calibrator tube barcodes. # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: HemosIL™ AcuStar Anti-β2-GPI IgG Precision testing was performed in accordance with CLSI Approved {4} Guideline EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition. | Sample | Mean (U/mL) | CV (%) (within run) | CV (%) (total) | | --- | --- | --- | --- | | Low anti-β2-GPI IgG Control | 19.1 | 7.8 | 11.2 | | High anti-β2-GPI IgG Control | 429 | 3.0 | 3.8 | | Anti-β2-GPI IgG plasma sample | 58.1 | 3.2 | 5.0 | An additional experiment was performed with two clinical plasma samples close to the cut-off and three clinical plasma samples in the upper end of the reportable range: | Sample | Mean (U/mL) | CV (%) (within run) | CV (%) (total) | | --- | --- | --- | --- | | A | 14.7 | 6.9 | 10.9 | | B | 20.9 | 4.7 | 7.3 | | C | 508 | 2.5 | 3.3 | | D | 1470 | 2.5 | 3.7 | | E | 2694 | 3.7 | 3.7 | # HemosIL™ AcuStar Anti-β2-GPI IgM Precision testing was performed in accordance with CLSI Approved Guideline EP05-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition. | Sample | Mean (U/mL) | CV (%) (within run) | CV (%) (total) | | --- | --- | --- | --- | | Low anti-β2-GPI IgM Control | 4.3 | 3.4 | 6.4 | | High anti-β2-GPI IgM Control | 63.0 | 2.4 | 4.3 | | Anti-β2-GPI IgM plasma sample | 11.0 | 3.6 | 5.8 | An additional experiment was performed with two clinical plasma samples close to the cut-off and three clinical plasma samples in the upper end of the reportable range: | Sample | Mean (U/mL) | CV (%) (within run) | CV (%) (total) | | --- | --- | --- | --- | | A | 13.6 | 4.5 | 8.3 | | B | 16.3 | 2.7 | 6.6 | | C | 91.9 | 2.4 | 5.7 | | D | 302 | 3.0 | 5.2 | | E | 510 | 4.1 | 6.0 | # b. Linearity/assay reportable range: Linearity of both the HemoSIL anti- $\beta 2$ -GPI IgG and IgM assays was assessed by preparing a series of dilutions of three different samples from patients suspected to have APS to cover the reportable range. The linearity tests were {5} performed with two different lots of reagents; results were comparable for both lots. Results from one lot are shown below: | Assay | Sample | Regression | r2 | Dilution Range (U/mL) | | --- | --- | --- | --- | --- | | IgG | 1 | y = 0.92x + 3.8 | 0.986 | 7.2 – 47.0 | | | 2 | y = 1.04x – 16.5 | 0.996 | 21.4 – 469.7 | | | 3 | y = 0.995x – 36.9 | 0.996 | 417 – 5632.0 | | | | | | | | IgM | 1 | y = 0.98x + 0.5 | 0.996 | 1.5 – 25.8 | | | 2 | y = 0.98x + 1.6 | 0.992 | 5.3 – 105.8 | | | 3 | y = 1.05x – 33.5 | 0.990 | 23.4 – 727.3 | The sponsor has established the linear range of the assays as $6.4 - 6100\mathrm{U / mL}$ for the anti- $\beta 2$ -GPI IgG assay and $1.1 - 841~\mathrm{U / mL}$ for the anti- $\beta 2$ -GPI IgM assay. c. Traceability, Stability, Expected values (controls, calibrators, or methods): An international reference material for anti- $\beta 2$ -GPI antibodies is not available. The assay is calibrated in relative arbitrary units $(\mathrm{U} / \mathrm{mL})$ The two calibrator values are assigned using in-house standards and a four-parameter Master Curve. The assignment values of the two calibrators are used to create a lot-specific four-parameter logistic curve, using two stored parameters from the Master Curve and two lot-specific parameters based on the calibrator values. Control values are assigned by analyzing the material with previous batches of reagents on two different ACL AcuStar instruments. The controls are analyzed in replicates of 10 on both instruments, then the batches of reagent are switched between instruments and the controls are re-analyzed. Stability studies were performed that support a shelf-life of 15 months at $2-8^{\circ}\mathrm{C}$ storage for the HemosIL AcuStar Anti- $\beta 2$ -GPI IgG and the HemosIL AcuStar Anti- $\beta 2$ -GPI IgM reagent cartridges, as well as calibrators and controls for both assays. The on-board stability of the reagent cartridge after opening on the AcuStar System was demonstrated to be at least 6 weeks. On-board stability of calibrators was shown to be $3.5\mathrm{h}$ . The sponsor recommends following the recommendations for storage conditions for samples in CLSI H18-A3 "Procedures for the Handling and Processing of Blood Specimens: Third Edition". {6} d. Detection limit: The Limit of Detection (LoD) was determined to be $6.4\mathrm{U / mL}$ for the anti- $\beta 2$ GPI IgG assay and $1.1\mathrm{U / mL}$ for the anti- $\beta 2$ -GPI IgM assay. The LoB and LoD were determined using the procedure described in CLSI EP17-A. Briefly, LoB was determined by performing 60 replicates of saline buffer. The results were ordered according to their value and the mean of position 57 and 58 (representing the $95^{\mathrm{th}}$ percentile) was determined as the LoB. The LoD was determined by reading 60 replicates of two low concentration samples and determining the standard deviation. LoD was calculated as $\mathrm{LoD} = \mathrm{LoB} + 1.65$ SD where SD is the pooled standard deviation of the two low positive sample readings. e. Analytical specificity: Both the IgG and the IgM assays were tested with the following endogenous interferents at two levels and showed less than $\pm 10\%$ interference at the concentration noted: Hemoglobin, $500~\mathrm{mg / dL}$ : Bilirubin, $18~\mathrm{mg / dL}$ : Triglycerides, $1250~\mathrm{mg / dL}$ : Heparin, $2~\mathrm{IU / mL}$ : Rheumatoid Factor (RF), $500~\mathrm{IU / mL}$ . Cross-reactivity: 10 samples positive for RF, 10 samples positive for antinuclear (ANA) antibodies, and 10 samples positive for rapid plasma reagin test (RPR, a form of syphilis antibody test) were tested with the HemosIL anti- $\beta 2$ GPI IgG and the IgM assays. None of the RF or RPR samples tested positive by the HemosIL anti- $\beta 2$ GPI assays. Two ANA samples (# 6 and #7) tested positive by the IgG assay and one sample (#7) tested positive by the IgM assay. Sample #7 tested positive by another anti- $\beta 2$ GPI commercial assay. f. Assay cut-off: The assay cut-offs were determined to be $20\mathrm{U / mL}$ in a reference range study described in the clinical cut-off section below. 2. Comparison studies: a. Method comparison with predicate device: HemoSIL AcuStar anti- $\beta 2$ GPI IgG: 150 clinically defined patient samples with results in the reportable range of both the test and the predicate (REAADS) were compared. There is no equivocal range for either test. Comparison of HemoSIL AcuStar anti- $\beta_{2}$ GPI IgG and Predicate | | Predicate Assay | | | | | --- | --- | --- | --- | --- | | | | + | - | Total | | HemoSIL AcuStar anti- β2 GPI IgG | + | 51 | 19* | 70 | | | - | 0 | 80 | 80 | | | Total | 51 | 99 | 150 | * 12 SLE, 7 SAPS {7} Positive agreement (51/51) = 100.0% (95% CI: 93.0% to 100.0%) Negative agreement (80/99) = 80.8% (95% CI: 71.7% to 88.0%) Overall agreement (131/150) = 87.3% (95% CI: 80.9% to 92.2%) HemoSIL AcuStar anti-β2GPI IgM: 205 clinically defined patient samples with results in the reportable range of both the test and the predicate (REAADS) were compared. There is no equivocal range for either test. Comparison of HemoSIL AcuStar anti-β₂GPI IgM and Predicate | | Predicate Assay | | | | | --- | --- | --- | --- | --- | | | | + | - | Total | | HemoSIL AcuStar anti-β2 GPI IgM | + | 30 | 5** | 35 | | | - | 17* | 153 | 170 | | | Total | 47 | 158 | 205 | * 3 SLE, 7 SAPS, 3 PAPS, 4 Control ** 1 SLE, 2 SAPS, 1 PAPS, 1 Control Positive agreement (30/47) = 63.8% (95% CI: 48.5% to 77.3%) Negative agreement (76/101) = 96.8% (95% CI: 92.8% to 99.0%) Overall agreement (183/205) = 76.5% (95% CI: 84.2% to 93.2%) b. Matrix comparison: The suitability of citrated plasma and serum was determined by assaying plasma/serum paired samples covering the assay range. The results of the Passing-Bablok regression analysis were following: | AcuStar anti-β2GPI | n = | regression | r value | | --- | --- | --- | --- | | IgG assay | 28 | y = 1.00x + 1.25 | 1.00 | | IgM assay | 48 | y = 0.99x + 0.02 | 0.996 | 3. Clinical studies: a. Clinical Sensitivity and Specificity: 321 frozen citrated plasmas were tested with the AcuStar anti-β2GPI IgG and IgM assays. These plasmas were from individuals diagnosed as primary APS (PAPS), secondary APS (SAPS), systemic lupus erythematosus (SLE) without APS and SLE-like by standard objective tests. Another group was patients with cardiovascular disorders but not classified in the previous four groups. A group of apparently healthy people was also included. The results summarized below are based on a cut-off of 20 U/mL: {8} 9 # AcuStar anti-β2GPI IgG: | AcuStar anti-β2GPI IgG | Clinical Diagnosis | | | | --- | --- | --- | --- | | | POS | NEG | Total | | POS | 59 | 21 | 80 | | NEG | 33 | 208 | 241 | | Total | 92 | 229 | 321 | Sensitivity (59/92) = 64.1% (95% CI: 53.5% to 73.9%) Specificity (208/229) = 90.8% (95% CI: 86.3% to 94.2%) The results for each clinical subgroup are shown in the table below: | Disease category | N | N Positive | % Positive | | --- | --- | --- | --- | | PAPS | 23 | 14 | 60.9% | | SAPS | 69 | 45 | 65.2% | | SLE | 115 | 20 | 17.4% | | SLE-like | 5 | 0 | 0% | | Others | 6 | 1 | 16.7% | | Normals | 103 | 0 | 0% | # AcuStar anti-β2GPI IgM: | AcuStar anti-β2GPI IgM | Clinical Diagnosis | | | | --- | --- | --- | --- | | | POS | NEG | Total | | POS | 27 | 11 | 38 | | NEG | 65 | 218 | 283 | | Total | 92 | 229 | 321 | Sensitivity (27/92) = 29.3% (95% CI: 20.3% to 39.8%) Specificity (218/229) = 95.2% (95% CI: 91.6% to 97.6%) | Disease category | N | N Positive | % Positive | | --- | --- | --- | --- | | PAPS | 23 | 7 | 30.4% | | SAPS | 69 | 20 | 29.0% | | SLE | 115 | 10 | 8.7% | | SLE-like | 5 | 0 | 0% | | Others | 6 | 1 | 16.7% | | Normals | 103 | 0 | 0% | b. Other clinical supportive data (when a. is not applicable): Not applicable. {9} 4. Clinical cut-off: The upper limits of normal (and cut-off) of the HemosIL anti-β2 GPI IgG and IgM assays were determined by assaying 262 citrated plasma samples from apparently healthy blood bank volunteers. Following the recommendations of the international consensus statement on an update of the classification criteria for definite antiphospholipid syndrome (APS)¹ the threshold for positive anti-β2 GPI antibodies was set at the 99th percentile of the tested population. This was determined to be 20 U/mL for both assays. 5. Expected values/Reference range: See Clinical cut-off discussion above. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. ¹ Miyakis S, et al. “International consensus statement on an update of the classification criteria for definite antiphospholipid syndrome (APS).” J Thromb Haemost. 4:295-306 2006.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...