K062217 · Aesku, Inc. · MSV · Dec 13, 2006 · Immunology
Device Facts
Record ID
K062217
Device Name
AESKULSKA BETA2-GLYCO-A
Applicant
Aesku, Inc.
Product Code
MSV · Immunology
Decision Date
Dec 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
AESKULISA® β2-Glyco-A is a solid phase enzyme immunoassay employing native β2-Glycoprotein I highly purified from human plasma for the semiquantitative and qualitative detection of IgA antibodies against β2-Glycoprotein I in human serum. The presence of anti-β2-Glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and secondary antiphospholipid syndrome. AESKULISA® β2-Glyco-GM is a solid phase enzyme immunoassay employing native β2-Glycoprotein I highly purified from human plasma for the semiquantitative and qualitative detection of IgG and/or IgM antibodies against β2-Glycoprotein I in human serum. The presence of anti-β2-Glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and secondary antiphospholipid syndrome. AESKULISA® β2- Glyco-Check is a solid phase enzyme immunoassay employing native β2- Glycoprotein I highly purified from human plasma for the semiquantitative and qualitative detection of IgA, IgG and IgM antibodies against β2- Glycoprotein I in human serum. The presence of anti- β2- Glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and secondary antiphospholipid syndrome.
Device Story
AESKULISA® β2-Glyco series are ELISA-based diagnostic kits for detecting IgA, IgG, and IgM antibodies against β2-Glycoprotein I in human serum. The device uses microplate strips coated with purified human β2-Glycoprotein I antigen. Patient serum samples are incubated in the wells; specific antibodies bind to the antigen. After washing, horseradish peroxidase-conjugated anti-human immunoglobulin is added, followed by TMB substrate. The resulting enzymatic colorimetric reaction is stopped with acid and measured spectrophotometrically at 450 nm. The intensity of the yellow color is proportional to the concentration of antibodies in the sample. Results are interpreted qualitatively or semi-quantitatively against a cut-off control or standard calibrator curve. The device is intended for use in clinical laboratories by trained personnel to aid in the diagnosis of thrombotic disorders associated with antiphospholipid syndrome. The output provides clinicians with antibody levels, which, when combined with clinical evaluation, assist in confirming or ruling out APS-related thrombotic conditions.
Clinical Evidence
No clinical data provided in the document; substantial equivalence is based on technological characteristics and intended use.
Technological Characteristics
Solid-phase enzyme immunoassay (ELISA). Antigen: Native β2 glycoprotein I purified from human plasma. Form factor: Microtiter plate-based assay. Analyte: IgA, IgG, and IgM antibodies against β2 glycoprotein I. Intended for professional laboratory use.
Indications for Use
Indicated for patients suspected of having thrombotic disorders related to primary or secondary antiphospholipid syndrome. Used for the detection of IgA, IgG, and/or IgM antibodies against β2 glycoprotein I in human serum to aid in diagnosis.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Related Devices
K040463 — AESKULISA CARDIOLIPIN AGM · Aesku, Inc. · May 12, 2004
K980230 — QUANTA LITE B2 GPI SCREEN · Inova Diagnostics, Inc. · Mar 24, 1998
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k062217
B. Purpose for Submission:
New Devices
C. Measurand:
Anti-β2-glycoprotein IgA, IgG, IgM antibodies and combined IgAGM for screening
D. Type of Test:
Qualitative and Semi-quantitative ELISA
E. Applicant:
AESKU, Inc.
F. Proprietary and Established Names:
AESKULISA® β2-Glyco-A Protocol 30-30-30 REF 30-7205US
AESKULISA® β2-Glyco-A Protocol 30-15-15 REF 7205US
AESKULISA® β2-Glyco-GM Protocol 30-30-30 REF 30-7206US
AESKULISA® β2-Glyco-GM Protocol 30-15-15 REF 7206US
AESKULISA® β2-Glyco-Check Protocol 30-30-30 REF 30-7215US
AESKULISA® β2-Glyco-Check Protocol 30-15-15 REF 7215US
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5660, Multiple autoantibodies immunological test system
2. Classification:
II
3. Product code:
MSV, Antibodies, β2-Glycoprotein I (β2-GPI)
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
AESKULISA® β2-Glyco-A is a solid phase enzyme immunoassay employing native β2-Glycoprotein I highly purified from human plasma for the semiquantitative and qualitative detection of IgA antibodies against β2-Glycoprotein I in human serum. The presence of anti-β2-Glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and secondary antiphospholipid syndrome.
AESKULISA® β2-Glyco-GM is a solid phase enzyme immunoassay employing native β2-Glycoprotein I highly purified from human plasma for the semiquantitative and qualitative detection of IgG and/or IgM antibodies against β2-Glycoprotein I in human serum. The presence of anti-β2-Glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and
{1}
secondary antiphospholipid syndrome.
AESKULISA® β2- Glyco-Check is a solid phase enzyme immunoassay employing native β2- Glycoprotein I highly purified from human plasma for the semiquantitative and qualitative detection of IgA, IgG and IgM antibodies against β2- Glycoprotein I in human serum. The presence of anti- β2- Glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and secondary antiphospholipid syndrome.
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
Microplate reader measuring OD at 450 nm (reading filter), and optional 620 nm reference filter (600-690 nm)
Microplate washing device (300μL repeating or multichannel pipette or automated system).
I. Device Description:
Each device contains: 12x8 microplate strips with breakaway microwells coated with purified β2- Glycoprotein I antigen; six levels of calibrators (0, 3, 10, 30, 100, 300 U/mL); positive, negative, and cut-off controls (human serum, diluted); wash buffer concentrate; sample buffer concentrate; anti-human immunoglobulin (Ig A/G/M)) horseradish peroxidase conjugate; 3,3',5,5' tetramethylbenzidine (TMB)/H₂O₂ substrate; and 1M hydrochloric acid stop solution.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Varelisa β2- Glycoprotein I IgA Antibodies EIA KIT
Varelisa β2- Glycoprotein I IgG Antibodies EIA KIT
Varelisa β2- Glycoprotein I IgM Antibodies EIA KIT
Varelisa β2- Glycoprotein I Screen Antibodies EIA KIT
2. Predicate 510(k) number(s):
k040450 (IgA), k040449 (IgG), k040451 (IgM), k040452 (Screen Antibodies)
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | New Device | Predicate Device |
| Technology | ELISA | Same |
| Capure Antigens | Purified human beta-2 Glycoprotein I | Same |
| Platform | 96 well microtiter plate | Same |
| Stop Solution | Prediluted | Same |
| Diluted Sample Volume Required | 100 μl | Same |
{2}
| Differences | | |
| --- | --- | --- |
| Item | New Device | Predicate |
| Intended use: | The presence of anti-β2 glycoprotein I antibodies in conjunction with clinical findings and other laboratory results can be used as an aid in the diagnosis of thrombotic disorders related to primary and secondary Antiphospholipid Syndrome. | The presence of β-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases. |
| Controls | Positive, Negative and Cut-off Control | Positive and Negative controls |
| Stop Solution | Ready to use 1 M HCl | Ready to use 0.5M H_{2}SO_{4} |
| Calibrators | 6 levels for all AESKULISA (0, 3, 10, 30, 100 and 300 U/mL) | 6 levels for all (0, 4, 8, 20, 50 and 100 U/mL) |
| Assay Format | Qualitative and Semi-Quantitative | Same for single Conjugates, for mixed conjugate only Qualitative |
| Sample Diluent | 5 x concentrated | Ready to use |
K. Standard/Guidance Document Referenced (if applicable):
None referenced.
L. Test Principle:
The AESKULISA® β2- Glyco-A, AESKULISA® β2- Glyco-GM and AESKULISA® β2- Glyco-Check devices are solid phase enzyme immunoassays for the semiquantitative and qualitative detection of IgA, IgG and/or IgM (separate), and IgA, IgG and IgM (mixed conjugate) antibodies respectively, against β2-Glycoprotein I in human serum. The wells of a microplate are coated with β2-Glycoprotein I antigen. Antibodies specific to β2- Glycoprotein I present in the patient sample bind to the antigen. Unbound fractions are washed off in the washing step. In the next step, anti-human immunoglobulins conjugated with horseradish peroxidase (conjugate) are incubated and react with the antigen-antibody complex of the samples in the microplates. Unbound conjugate is washed off in the washing step. Addition of TBM-substrate generates an enzymatic colorimetric blue reaction, which is stopped by diluted acid (color changes to yellow). The rate of color formation from the chromogen is a function of the amount of conjugate bound to the antigen-antibody
{3}
complex and is proportional to the initial concentration of the respective antibodies in the patient serum. The results are read spectrophotometrically and are interpreted by comparison to a cut-off control (qualitative) or a standard calibrator curve (semiquantitative).
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Three different samples (high, medium, near the cut-off) were assayed 18 times on three microplates of the specific antibody isotype, for two days for the inter-assay study. Three different samples (high, medium, near the cut-off) were assayed 24 times on one microplate of the specific antibody type for the intra-assay study. Both studies were performed on Protocol 30-15-15 incubation time. Target values for the studies were set at $\% \mathrm{CV} \leq 10$.
The inter-assay $\% \mathrm{CV}$ range for AESKULISA® $\beta 2$-Glyco-A was from $8.4\%$ to $9.6\%$, for AESKULISA® $\beta 2$-Glyco-GM was $4.4\%$ to $6.5\%$ and for AESKULISA® $\beta 2$-Glyco-Check was $8.7\%$ to $9.6\%$.
The intra-assay $\% \mathrm{CV}$ range for AESKULISA® $\beta 2$-Glyco-A was from $7.6\%$ to $8.7\%$, for AESKULISA® $\beta 2$-Glyco-GM was from $5.1\%$ to $8.8\%$ and for AESKULISA® $\beta 2$-Glyco-Check was $7.3\%$ to $8.4\%$. All the ranges were within the target values.
| AESKULISA® β2- Glyco-A | Inter-Assay Variation | | |
| --- | --- | --- | --- |
| | Sample 1 | Sample 2 | Sample 3 |
| CV (%) | 8.4 | 9.6 | 8.6 |
| Mean (U/mL) | 14.4 | 106.7 | 124.7 |
| AESKULISA® β2- Glyco-GM | | | |
| CV (%) | 4.4 | 6.9 | 6.5 |
| Mean (U/mL) | 19.6 | 70.3 | 172.1 |
| AESKULISA® β2- Glyco-Check | | | |
| CV (%) | 8.7 | 8.5 | 9.6 |
| Mean (U/mL) | 41.1 | 80.0 | 174.9 |
| AESKULISA® β2- Glyco-A | Intra-Assay Variation | | |
| --- | --- | --- | --- |
| | Sample 1 | Sample 2 | Sample 3 |
| CV (%) | 7.6 | 8.0 | 8.7 |
| Mean (U/mL) | 13.0 | 96.5 | 118.2 |
| AESKULISA® β2- Glyco-GM | | | |
| CV (%) | 5.1 | 8.8 | 6.6 |
| Mean (U/mL) | 17.2 | 54.3 | 228.5 |
| AESKULISA® β2- Glyco-Check | | | |
| CV (%) | 7.3 | 7.7 | 8.4 |
| Mean (U/mL) | 30.8 | 115.3 | 310.2 |
#### b. Linearity/assay reportable range:
Study design: Two samples known to contain different levels of $\beta 2$
{4}
Glycoprotein I IgA antibodies, two samples known to contain different levels of β2- Glycoprotein I IgG/M antibodies and two samples known to contain different levels of β2- Glycoprotein I Ig A/G/M antibodies were chosen and serially diluted to determine the linearity of the assay. From an initial dilution of 1/100, further dilutions of 1:200, 1:400 and 1:800 were made. The AESKULISA® β2- Glyco-A assay had a recovery range of 93.9% to 107.0%. AESKULISA® β2- Glyco-GM assay had a recovery range of 94.0% to 103.9%. AESKULISA® β2- Glyco-Check assay had a recovery range of 90.1% to 106.4% (see tables below).
## AESKULISA® β2- Glyco-A
| Sample No. | Dilution Factor | measured concentration (U/ml) | expected concentration (U/ml) | Recovery (%) 90 - 110 % |
| --- | --- | --- | --- | --- |
| 1 | 1 / 100 | 123.7 | 122.0 | 101.4 |
| | 1 / 200 | 53.1 | 51.0 | 103.4 |
| | 1 / 400 | 29.3 | 36.5 | 98.1 |
| | 1 / 800 | 14.7 | 15.3 | 96.1 |
| 2 | 1 / 100 | 126.2 | 118.0 | 107.0 |
| | 1 / 200 | 61.1 | 59.0 | 103.6 |
| | 1 / 400 | 27.9 | 29.5 | 94.6 |
| | 1 / 800 | 13.9 | 14.6 | 93.9 |
## AESKULISA® β2- Glyco-GM
| Sample No. | Dilution Factor | measured concentration (U/ml) | expected concentration (U/ml) | Recovery (%) 90 - 110 % |
| --- | --- | --- | --- | --- |
| 1 | 1 / 100 | 151.4 | 206.0 | 95.7 |
| | 1 / 200 | 100.7 | 106.0 | 103.7 |
| | 1 / 400 | 48.7 | 56.0 | 97.4 |
| | 1 / 800 | 24.5 | 25.0 | 98.0 |
| 2 | 1 / 100 | 232.6 | 236.0 | 97.3 |
| | 1 / 200 | 124.2 | 119.5 | 103.9 |
| | 1 / 400 | 57.9 | 58.8 | 96.8 |
| | 1 / 800 | 28.1 | 29.9 | 94.0 |
## AESKULISA® β2- Glyco-Check
| Sample No. | Dilution Factor | measured concentration (U/ml) | expected concentration (U/ml) | Recovery (%) 90 - 110 % |
| --- | --- | --- | --- | --- |
| 1 | 1 / 100 | 189.2 | 198.0 | 95.1 |
| | 1 / 200 | 105.9 | 98.5 | 106.4 |
| | 1 / 400 | 47.0 | 46.8 | 94.4 |
| | 1 / 800 | 25.2 | 24.9 | 101.2 |
| 2 | 1 / 100 | 64.0 | 61.0 | 103.7 |
| | 1 / 200 | 39.8 | 40.5 | 98.3 |
| | 1 / 400 | 19.7 | 20.3 | 97.0 |
| | 1 / 800 | 9.1 | 10.1 | 93.1 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods): There is no reference standard for β2-glycoprotein I. The standards are
{5}
prepared in-house and assigned arbitrary units (U/ml) during the development process. The positive, cut-off, and negative controls are also prepared in-house.
d. Detection limit:
The sample buffer was tested 30 times for each assay and result in an analytical sensitivity of 1.0 U/mL. The value for the analytical sensitivity (detection limit) was calculated as the mean of the optical densities of the sample diluent.
e. Analytical specificity:
Interference by endogenous substances: No data provided. The package insert states that icteric, lipemic, hemolyzed or bacterially contaminated samples should not be used in these assays.
f. Assay cut-off:
The cut-off value for AESKULISA® β2- Glyco-A, AESKULISA® β2- Glyco-GM and AESKULISA® β2- Glyco-Check were determined in serum from 70 healthy donors. As a result, 15 U/mL has been defined as the cut-off value (Mean value +3SD).
2. Comparison studies:
a. Method comparison with predicate device:
One hundred and fifteen (115) sera including 104 samples from patients diagnosed with APS, SLE and other autoimmune diseases (samples were from Germany); and 11 healthy donors were tested on the AESKULISA® β2- Glyco-A, AESKULISA® β2- Glyco-GM and AESKULISA® β2- Glyco-Check and the predicate devices. Results are summarized below.
AESKULISA® β2- Glyco-A
| | Predicate Device | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| AESKU β2GPI IgA | Positive | 17 | 7 | 24 |
| | Negative | 8 | 83 | 91 |
| | Total | 25 | 90 | 115 |
Positive Percent Agreement 68.8% (17/25)
Negative Percent Agreement 92.2% (83/90)
Overall Percent Agreement 87.0% (100/115)
AESKULISA® β2- Glyco-G
| | Predicate Device | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| AESKU β2GPI IgG | Positive | 30 | 18 | 48 |
| | Negative | 1 | 66 | 67 |
| | Total | 31 | 84 | 115 |
Positive Percent Agreement 96.8% (30/31)
Negative Percent Agreement 78.6% (66/84)
Overall Percent Agreement 83.5% (96/115)
{6}
AESKULISA® β2- Glyco-M
| | Predicate Device | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| AESKU β2GPI IgM | Positive | 14 | 11 | 25 |
| | Negative | 1 | 89 | 90 |
| | Total | 15 | 100 | 115 |
Positive Percent Agreement 93.3% (14/15)
Negative Percent Agreement 89.0% (89/100)
Overall Percent Agreement 89.6% (103/115)
AESKULISA® β2- Glyco-Check
| | Predicate Device | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| AESKU β2GPI Check | Positive | 52 | 6 | 58 |
| | Negative | 4 | 53 | 57 |
| | Total | 56 | 59 | 115 |
Positive Percent Agreement 92.9% (52/56)
Negative Percent Agreement 89.8% (53/59)
Overall Percent Agreement 91.3% (105/115)
Comparison of Protocol 30-15-15 and Protocol 30-30-30:
## AESKULISA β-2 Glyco A (REF7205US)
Comparability of the two protocols was assessed with 52 sera on both REF 7205US (30-15-15 minute protocol) and REF 30-7205US (30-30-30 minute protocol). The linear regression analysis is depicted in the large figure below with an $r^2 = 0.976$. Included in the sera are 26 sera close to the assay cut-off (<30 U/mL).
| Y = b[0] + b[1]X | value | range (CI95%) |
| --- | --- | --- |
| b[0] | -0.47 | -5.72 / 4.78 |
| b[1] | 0.984 | 0.940 / 1.028 |
| r² | 0.976 | |

{7}
# AESKULISA β-2 Glyco GM (REF7206US)
Comparability of the two protocols was assessed with 52 IgG sera and 60 IgM sera tested on both REF 7206US (30-15-15 minute protocol) and REF 30-7206US (30-30-30 minute protocol). The linear regression analysis is depicted in the large figure below (IgM) with an $r^2 = 0.973$ and the upper left small figure (IgG) with an $r^2 = 0.968$ . Included in these sera are 28/30 sera close to the cut-off (<30 U/mL).
| | IgG | | IgM | |
| --- | --- | --- | --- | --- |
| Y = b[0] + b[1]X | value | range (CI95%) | value | range (CI95%) |
| b[0] | -0.03 | -4.87 / 4.93 | -0.47 | -4.75 / 3.81 |
| b[1] | 0.99 | 0.94 / 1.04 | 1.04 | 0.99 / 1.09 |
| r2 | 0.968 | | 0.973 | |

# AESKULISA β-2 Glyco Chek (REF7206US)
Comparability of the two protocols was assessed with 60 sera tested on both REF 7215US (30-15-15 minute protocol) and REF 30-7215US (30-30-30 minute protocol). The linear regression analysis is depicted in the figure below with an $r^2 = 0.990$ . Included in these sera are 28 sera close to the cut-off (<30 U/mL).
{8}
| Y = b[0] - b[1]X | value | range
(CI95%) |
| --- | --- | --- |
| b[0] | -2.06 | -4.75 / 0.61 |
| b[1] | 1.03 | 1.00 / 1.06 |
| r² | 0.990 | |

b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity and specificity:
The clinical sensitivity and specificity study were evaluated on 79 samples from patients with the following diagnosis: 39 APS, 46 SLE, 17 SLE with secondary APS, 1 suspected APS, 1 with indeterminate connective tissue disease with APS and 11 healthy donors. Patients were tested on the AESKULISA® β2- Glyco-A, AESKULISA® β2- Glyco-GM and AESKULISA® β2- Glyco-Check and the results are summarized below.
AESKULISA® β2- Glyco-A
| Disease | # Tested | # positive AESKU (%) | # positive pred dev (%) |
| --- | --- | --- | --- |
| Antiphospholipid-Syndrome | 39 | 8 (20.5%) | 11 (26.2%) |
| Systemic-Lupus-Erythematosus | 46 | 7 (15.2%) | 5 (10.9%) |
| SLE with secondary APS | 17 | 9 (52.9%) | 9 (52.9%) |
| suspected APS | 1 | 0 (0%) | 0 (0%) |
| Indeterminate connective tissue disease with APS | 1 | 0 (0%) | 0 (0%) |
| healthy donors | 11 | 0 (0%) | 0 (0%) |
AESKULISA® β2- Glyco-M
{9}
| Disease | # Tested | # positive AESKU M (%) | # positive pred dev M (%) |
| --- | --- | --- | --- |
| Antiphospholipid-Syndrome | 39 | 18 (46.5%) | 10 (25.6%) |
| Systemic-Lupus-Erythematosus | 46 | 1 (2.2%) | 1 (2.2%) |
| SLE with secondary APS | 17 | 6 (35.3%) | 4 (23.5%) |
| suspected APS | 1 | 0 (0%) | 0 (0%) |
| indeterminate connective tissue disease with APS | 1 | 0 (0%) | 0 (0%) |
| healthy donors | 11 | 0 (0%) | 0 (0%) |
## AESKULISA® β2- Glyco-G
| Disease | # Tested | # positive AESKU G (%) | # positive pred dev G (%) |
| --- | --- | --- | --- |
| Antiphospholipid-Syndrome | 39 | 25 (64.1%) | 16 (41.0%) |
| Systemic-Lupus-Erythematosus | 46 | 8 (17.4%) | 3 (6.5%) |
| SLE with secondary APS | 17 | 13 (76.5%) | 9 (52.9%) |
| suspected APS | 1 | 1 (100%) | 1 (100%) |
| indeterminate connective tissue disease with APS | 1 | 1 (100%) | 0 (0%) |
| healthy donors | 11 | 0 (0%) | 0 (0%) |
## AESKULISA® β2- Glyco-Check
| Disease | # Tested | # positive AESKU (%) | # positive pred dev (%) |
| --- | --- | --- | --- |
| Antiphospholipid-Syndrome | 39 | 33 (84.6%) | 31 (79.5%) |
| Systemic-Lupus-Erythematosus | 46 | 8 (17.4%) | 9 (19.6%) |
| SLE with secondary APS | 17 | 15 (88.2%) | 14 (82.4%) |
| suspected APS | 1 | 1 (100%) | 1 (100%) |
| indeterminate connective tissue disease with APS | 1 | 1 (100%) | 1 (100%) |
| healthy donors | 11 | 0 (0%) | 0 (0%) |
Those numbers are in line with the ones found in literature studies.
b. Other clinical supportive data (when a. is not applicable):
Not applicable.
4. Clinical cut-off:
Same as assay cut-off.
5. Expected values/Reference range:
Expected values in the normal population should be negative.
{10}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.