VARELISA B2 GLYCOPROTEIN I ANTIBODIES SCREEN, MODELS 19048 AND 19096
K040452 · Pharmacia Deutschland GmbH · MSV · May 11, 2004 · Immunology
Device Facts
Record ID
K040452
Device Name
VARELISA B2 GLYCOPROTEIN I ANTIBODIES SCREEN, MODELS 19048 AND 19096
Applicant
Pharmacia Deutschland GmbH
Product Code
MSV · Immunology
Decision Date
May 11, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Indications for Use
The Varelisa B2-Glycoprotein I Antibodies Screen Kit is designed for the qualitative determination of β2-glycoprotein I antibodies in serum or plasma. The presence of B2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases.
Device Story
Varelisa B2-Glycoprotein I Antibodies Screen Kit is an ELISA-based diagnostic test; utilizes microplate wells coated with purified human β2-glycoprotein I antigen. Patient serum or plasma samples are applied; specific antibodies bind to antigen; detected via enzyme-labeled secondary antibody conjugate (horseradish peroxidase). Enzyme substrate added; color formation rate proportional to antibody concentration. Results reported as ratio of sample OD to cut-off OD. Used in clinical laboratories; performed by trained technicians. Output aids clinicians in diagnosing thrombotic disorders associated with Antiphospholipid Syndrome or SLE. Benefits include standardized qualitative assessment of autoimmune markers to support clinical decision-making.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated across 5 runs with 4 replicates per run; variances met specifications (within <12%, between <10%, total <15%). Analytical specificity tested against bilirubin, hemoglobin, chyle, and rheumatoid factor; no significant interference observed. Cut-off established using 432 healthy donors. Method comparison with predicate (n=40) showed 92.5% qualitative agreement. Matrix comparison (serum, citrate, heparin, EDTA plasma) showed no significant differences.
Technological Characteristics
Indirect noncompetitive enzyme immunoassay. Components: microplate strips coated with human purified B2-glycoprotein I, calibrator, negative control, enzyme-labeled conjugate, substrate, stop solution, buffered diluent, wash buffer. Detection principle: colorimetric enzyme-linked reaction. Sample type: serum or plasma.
Indications for Use
Indicated for qualitative detection of B2-Glycoprotein I antibodies (IgG/IgM/IgA) in human serum or plasma to aid in diagnosis of thrombotic disorders associated with primary Antiphospholipid Syndrome, systemic lupus erythematosus (SLE), or other autoimmune diseases.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
INOVA QUANTA Lite™ B2 GPI Screen
Related Devices
K040449 — VARELISA B2 GLYCOPROTEIN I IGG ANTIBODIES, MODELS 18748 AND 18796 · Pharmacia Deutschland GmbH · May 11, 2004
K991804 — BINDAZYME ANTI-B2 SCREEN EIA TEST KIT. · The Binding Site, Ltd. · May 5, 2000
K040450 — VARELISA B2 GLYCOPROTEIN I IGA ANTIBODIES, MODELS 18948 AND 18996 · Pharmacia Deutschland GmbH · May 11, 2004
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k040452
B. Purpose of Submission:
New device
C. Analyte:
B2-Glycoprotein I Antibodies Screen
D. Type of Test:
Qualitative ELISA
E. Applicant:
Pharmacia Deutschland GmbH/Sweden Diagnostics (Germany) GmbH
F. Proprietary and Established Names:
Varelisa B2-Glycoprotein I Antibodies Screen Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5660, Multiple Autoantibodies Immunological Test System
2. Classification:
Class II
3. Product Code:
MSV, System, Test, Antibodies β2-Glycoprotein I (β2-GPI)
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
The Varelisa B2-Glycoprotein I Antibodies Screen Kit is designed for the qualitative determination of β2-glycoprotein I antibodies in serum or plasma.
2. Indication(s) for use:
The presence of B2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases.
3. Special condition for use statement(s):
Prescription use only.
4. Special instrument Requirements:
None.
{1}
Page 2 of 6
I. Device Description:
The Varelisa B2-Glycoprotein I Antibodies Screen kit consists of an ELISA strip wells (12 strips X 8 wells) coated with purified human β2-glycoprotein I antigen. It also contains ancillary reagents such as calibrators, negative controls, wash buffer, sample diluent, antibody-chromogen conjugate, enzyme substrate, stop solution for enzyme reaction, and a frame to hold ELISA strip wells. Also included are a package insert and a control certificate.
J. Substantial Equivalence Information:
1. Predicate device name(s):
INOVA QUANTA Lite B2 GPI Antibody Screen
2. Predicate K number(s):
k980230
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Qualitative determination of B2-glycoprotein I antibodies in serum or plasma for use in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases. | Qualitative detection of B2-GPI antibodies in serum for use in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of certain autoimmune disease thrombotic disorders, such as those secondary to systemic lupus erythematosis (SLE) or other lupus-like thrombotic diseases. |
| Antigen | Human affinity-purified β2-glycoprotein I | β2-glycoprotein I, unknown source |
| Conjugate | Anti-human IgG, IgA, IgM Antibody Conjugate, Horseradish peroxidase | Same |
| Assay Principle | Indirect noncompetitive enzyme immunoassay | Same |
| Sample dilution | 1:101 | Same |
| Differences | | |
| Item | Device | Predicate |
| Specimen | Serum and plasma | Serum |
| Assay evaluation | Ratio < 1.0 = negative
Ratio 1.0-1.4 = equivocal
Ratio > 1.4 = positive | Ratio < 1.0 = negative
Ratio ≥ 1.0 = positive |
{2}
# K. Standard/Guidance Document Referenced (if applicable):
None
# L. Test Principle:
The Varelisa B2-Glycoprotein I Antibodies EIA Kit is an indirect noncompetitive enzyme immunoassay. The wells of the microplate are coated with human purified $\beta 2$ -glycoprotein I antigen. Antibodies specific for $\beta 2$ -glycoprotein I present in the applied patient sample bind to the antigen and are detected by an enzyme labeled secondary antibody complex, which leads to the formation of an enzyme labeled conjugate-antibody-antigen complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen (substrate) is a function of the amount of conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample. The presence of antibodies is calculated by reference of the sample OD to the cut-off OD, and is reported as a ratio of OD sample/OD cut-off.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Five serum samples (low, medium, high, equivocal, negative) were selected from a serum bank. The samples were diluted 1:101 and measured in 5 runs, with 4 replicates per run. Standards and controls were analyzed in triplicate. Specified target values were: variance "within" $< 12\%$ ; "between" $< 10\%$ ; "total" $< 15\%$ . The specifications were met for each sample, with the following variances:
| Sample ID | | Run 1 | Run 2 | Run 3 | Run 4 | Run 5 | Mean (U/mL) | Variance | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | | Within | Between | total |
| ZSB 7170 | Mean (u/mL) | 1.8 | 2.0 | 1.9 | 1.9 | 1.9 | 1.9 | 2.60 | 3.49 | 4.35 |
| | CV% | 2.82 | 2.96 | 2.67 | 0.0 | 3.12 | | | | |
| ZSB 7262 | Mean (u/mL) | 2.7 | 3.0 | 2.7 | 2.9 | 2.8 | 2.8 | 2.12 | 4.50 | 4.97 |
| | CV% | 0.0 | 1.68 | 1.83 | 3.51 | 1.8 | | | | |
| ZSB 8300 | Mean (u/mL) | 3.0 | 3.2 | 3.1 | 3.2 | 3.1 | 3.1 | 2.04 | 2.48 | 3.21 |
| | CV% | 2.72 | 2.55 | 1.89 | 1.83 | 0.0 | | | | |
| negative | Mean (u/mL) | 0.7 | 0.6 | 0.6 | 0.7 | 0.7 | 0.7 | 11.9 | 4.9 | 12.9 |
| | CV% | 16.5 | 8.0 | 8.7 | 7.4 | 14.2 | | | | |
| equiv. | Mean (u/mL) | 1.1 | 1.1 | 1.0 | 0.7 | 1.1 | 1.1 | 4.9 | 4.1 | 6.4 |
| | CV% | 8.5 | 4.7 | 0.0 | 0.0 | 4.4 | | | | |
# b. Linearity/assay reportable range:
Not done.
{3}
Page 4 of 6
c. Traceability (controls, calibrators, or method):
Not traceable to external sources. The consistency of the calibration material is given by a testing procedure comprising several internal target values that must be met. Each new production of calibration material is adapted to the corresponding master calibration material.
d. Detection limit:
Sample Buffer was diluted according to Directions for Use and measured 88 times on one plate. The Calibrator and the negative control were analyzed in four replicates. Analytical sensitivity was to verify that the background of the assay in less than 30% of the OD of the cut-off.
The sample buffer mean +3SD was 0.080 OD. The calibrator mean OD was 1.93, and negative control mean OD was 0.067. The factor was 0.35 and cut-off OD was 0.694. It was verified that the mean +3SD of the OD of the sample buffer was lower than 30% of the calculated OD of the cut-off.
e. Analytical specificity:
Interference was tested against potentially interfering substances found in blood: bilirubin, hemoglobin, chyle, and rheumatoid factor. Three samples from a serum bank were diluted 1:101 and spiked with different super-normal amounts of interferent. Spiked samples were analyzed in triplicate. Acceptance criteria were that spiked samples should show no more than 20% variation from unspiked sample.
| Additives (final sample concentration) | I | II | III | IV | V |
| --- | --- | --- | --- | --- | --- |
| Bilirubin F (mg/dL) | 0.0 | 4.8 | 9.5 | 14.3 | 19.0 |
| Bilirubin C (mg/dL) | 0.0 | 5.5 | 11.0 | 16.5 | 22.0 |
| Chyle (Units) | 0.0 | 357.5 | 715.0 | 1072.5 | 1430.0 |
| Hemoglobin (mg/dL) | 0.0 | 122.5 | 245.0 | 367.5 | 490.0 |
| RF (IU/mL) | 0.0 | 98.0 | 294.0 | 490.0 | - |
All samples met acceptance criteria and showed no significant influence on the test results. Additional interference tests on negative and equivocal sera with the same interferents demonstrated a negligible effect of the interferents on the equivocal and negative samples. The negative samples, however, remained negative, so the interference was judged to be not important.
f. Assay cut-off:
{4}
Page 5 of 6
Qualitative cut-points were determined by measuring 432 samples from apparently healthy Caucasian donors, equally distributed by sex and age. Diluted samples, standards, and controls were analyzed in duplicate. Specifications called for the 95th percentile to be smaller than or equal to the lower limit of the equivocal range. The following values were selected for negative, equivocal, and positive:
Ratio < 1.0 = negative
Ratio 1.0-1.4 = equivocal
Ratio > 1.4 = positive
The specifications were fulfilled and no difference was detected between sexes or ages.
2. Comparison studies:
a. Method comparison with predicate device:
The following samples were used in comparing the new device with 40 samples from a serum bank at Pharmacia, representing SLE, lupus-like, unknown clinical diagnosis, thrombosis, arthritis, malaria, thalassemia, and healthy individuals. The results were compared in a 6 field analysis. 37 of 40 sera agreed qualitatively (pos/neg) for overall agreement of 92.5%. Two sera were equivocal by the Varelisa assay and either positive (1) or negative (1) by the qualitative predicate. One serum was negative by Varelisa and positive by the predicate.
| Varelisa Abs Screen Lot SJ20 | n = 40 | INOVA QUANTA Lite Lot 170329 | |
| --- | --- | --- | --- |
| | | Positive | Negative |
| | positive | 20 | 0 |
| | equivocal | 1 | 1 |
| | negative | 1 | 17 |
b. Matrix comparison:
Twenty blood donor samples in the positive and negative range of the assay were tested in serum, citrate plasma, heparin plasma, and EDTA plasma, in the negative range of the assay. No significant differences were noted for any of the matrices.
3. Clinical studies:
a. Clinical sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a and b are not applicable): Not applicable
{5}
Page 6 of 6
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Expected value for normal healthy individuals is negative, or ratio <1.0. In a study of 432 healthy Caucasians, the following values were obtained:
N: 432
Mean: 0.54
Mean +2SD 1.24
Median: 0.5
95th percentile 1.0
N. Conclusion:
The submitted information in this premarket notification is complete, and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.