← Product Code [LRM](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LRM) · K062183

# FARRZYME HUMAN HIGH AVIDITY ANTI-DSDNA ENZYME IMMUNOASSAY KIT (K062183)

_The Binding Site, Ltd. · LRM · Nov 21, 2006 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LRM/K062183

## Device Facts

- **Applicant:** The Binding Site, Ltd.
- **Product Code:** [LRM](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LRM.md)
- **Decision Date:** Nov 21, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

This assay is intended for the in-vitro measurement of specific, high avidity IgG autoantibodies against double stranded deoxyribonucleic acid (dsDNA) present in human serum, as an aid to the diagnosis of systemic lupus erythematosus (SLE), in conjunction with other serological test results and clinical findings.

## Device Story

FARRZYME is an enzyme immunoassay (EIA) kit for detecting high avidity IgG autoantibodies against dsDNA in human serum. Used in clinical laboratories by trained personnel to aid in diagnosing systemic lupus erythematosus (SLE). The assay measures antibody-antigen binding; results are interpreted by clinicians alongside other serological tests and patient clinical presentation to support SLE diagnosis.

## Clinical Evidence

Performance evaluated using 252 samples (189 SLE, 28 healthy, 35 dsDNA positive). Comparison with BINDAZYME ELISA showed 85.3% overall agreement; comparison with Crithidia luciliae showed 87.3% overall agreement. Comparison with Farr RIA (n=166) showed 92.8% overall agreement (89.8% positive percent agreement). Analytical precision: intra-assay CV 2.2-5.4%, inter-assay CV 3.9-13.5%. Cut-off established at 30 IU/mL using 150 healthy donors. No interference observed from bilirubin, hemoglobin, chyle, or rheumatoid factor.

## Technological Characteristics

Enzyme immunoassay (EIA) kit; utilizes specific antigen-antibody binding principles for detection of high avidity IgG autoantibodies against dsDNA in human serum samples.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k062183

B. Purpose for Submission:
New device

C. Measurand:
Double-stranded DNA (dsDNA) autoantibodies (high avidity)

D. Type of Test:
Enzyme-linked immunosorbent assay (ELISA)

E. Applicant:
The Binding Site

F. Proprietary and Established Names:
FARRZYME Human High Avidity anti-dsDNA Enzyme Immunoassay Kit

G. Regulatory Information:
1. Regulation section:
CFR §866.5100 Antinuclear antibody immunological test system
2. Classification:
Class II
3. Product code:
LRM: Anti-DNA antibody (enzyme-labeled), antigen, control
4. Panel:
Immunology (82)

H. Intended Use:
1. Intended use:
The FARRZYME assay is intended for the in-vitro measurement of specific, high avidity IgG autoantibodies against double stranded deoxyribonucleic acid (dsDNA) present in human serum, as an aid to the diagnosis of systemic lupus erythematosus (SLE), in conjunction with other serological test results and clinical findings.
2. Indication for use:
Same as Intended Use
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
Microplate reader capable of measuring OD at 450nm referenced on air

I. Device Description:
The device consists of dsDNA (calf thymus) coated microwell strips; Sample Diluent; Wash Buffer; Calibrators containing anti-dsDNA autoantibodies at levels of 1000, 333, 111, 37, and 12.3 IU/mL; Positive Control containing diluted human serum and an assayed amount of anti-dsDNA autoantibodies; Negative Control containing diluted human serum; Single-Stranded DNA antibody Control; Conjugate with purified peroxidase labeled antibody to rabbit anti-human IgG; TMB Substrate; and 3M phosphoric acid Stop Solution.

{1}

2

J. Substantial Equivalence Information:

1. Predicate device name(s):
TBS BINDAZYME Human Anti-dsDNA Enzyme Immunoassay Kit
TBS CRITHIDIA LUCILIAE dsDNA Kit/Substrate Slides
Farr Radioimmunoassay (RIA) (pre-amendment, laboratory method)

2. Predicate 510(k) number(s):
K993727
K930987

3. Comparison with predicate devices:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | FARRZYME | BINDAZYME  |
|  Indication for Use | Aid in the diagnosis of SLE | Same  |
|  DNA antigen source | Calf thymus | Same  |
|  Method | ELISA | Same  |
|  Calibrators | 5 levels: 1000, 333, 111, 37, 12.3 IU/mL | Same  |
|  Controls | Positive and negative | Same  |
|  Second negative control | ssDNA antibody control | Same  |
|  Labeled conjugate | Peroxidase labeled rabbit anti-human IgG | Same  |
|  Substrate | TMB | Same  |
|  Stop solution | 3M phosphoric acid | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | FARRZYME | BINDAZYME  |
|  dsDNA antibodies detected | High avidity | High and low avidity  |
|  Wash buffer | More stringent | Conventional  |
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | FARRZYME | Crithidia luciliae  |
|  Indication for Use | Aid in the diagnosis of SLE | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | FARRZYME | Crithidia luciliae  |
|  DNA antigen source | Calf thymus | C. luciliae  |
|  Method | ELISA | Indirect immunofluorescence  |

{2}

3

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Calibrators | 5 levels: 1000, 333, 111, 37, 12.3 IU/mL | None  |
|  Controls | Positive, negative and ssDNA control | Positive (apple-green kinetoplast staining) and negative  |
|  Labeled conjugate | Peroxidase labeled rabbit anti-human IgG | Fluorescein isothiocyanate (FITC) labeled sheep anti-human IgG  |
|  Counterstain | Not applicable | Evans Blue  |
|  Substrate | TMB | Not applicable  |
|  Stop solution | 3M phosphoric acid | Not applicable  |
|  Mounting medium | Not applicable | Necessary for reading results  |

K. Standard/Guidance Document Referenced (if applicable):
None referenced

L. Test Principle:
Calibrators, controls, and diluted patient samples are added to the wells and autoantibodies recognizing the dsDNA antigen bind during the first incubation. After washing the wells to remove all unbound proteins, purified peroxidase labeled rabbit anti-human IgG (γ chain specific) conjugate is added. The conjugate binds to the captured human autoantibody and the excess unbound conjugate is removed by a further wash step. The unbound conjugate is visualized with 3,3',5,5' tetramethylbenzidine (TMB) substrate which gives a blue reaction product, the intensity of which is proportional to the concentration of autoantibody in the sample. Phosphoric acid is added to each well to stop the reaction. This produces a yellow end point color, which is read at 450nm. The controls and patient results are compared to a 5 point calibration curve for interpretation.

M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The intra- and inter-assay precisions were measured using 6 samples covering the range of the calibration curve. Intra-assay precision was measured using 20 replicates in one assay. Inter-assay precision was measured by testing samples in duplicate in 6 assays performed over 3 days.

|  Intra-Assay Precision |   |   |   | Inter-Assay Precision  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  N=20 | IU/mL | SD | % CV | N=6 | IU/mL | SD | % CV  |
|  Sample 1 | 25.1 | 1.35 | 5.4 | Sample 1 | 24.6 | 3.32 | 13.5  |
|  Sample 2 | 39.0 | 1.86 | 4.8 | Sample 2 | 42.3 | 1.66 | 3.9  |
|  Sample 3 | 74.5 | 1.64 | 2.2 | Sample 3 | 61.3 | 7.13 | 11.6  |
|  Sample 4 | 205.5 | 6.74 | 3.3 | Sample 4 | 123.2 | 6.02 | 4.9  |

{3}

|  Intra-Assay Precision |   |   |   | Inter-Assay Precision  |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  N=20 | IU/mL | SD | % CV | N=6 | IU/mL | SD | % CV  |
|  Sample 5 | 361.2 | 15.60 | 4.3 | Sample 5 | 272.4 | 18.96 | 7.0  |
|  Sample 6 | 528.6 | 27.19 | 5.1 | Sample 6 | 474.1 | 32.46 | 6.9  |

# b. Linearity/assay reportable range:

The reportable range for the assay of 12.3-1000 IU/mL is based on the calibrators. In the linearity studies, high samples were first diluted at the original assay dilution of 1:100 then doubling dilutions were made.

![img-0.jpeg](img-0.jpeg)

![img-1.jpeg](img-1.jpeg)

{4}

![img-2.jpeg](img-2.jpeg)

c. Traceability, Stability, Expected values (controls, calibrators, or methods): The calibrators are standardized against the WHO reference material Wo-80.

d. Detection limit:

Confirmation that the FARRZYME assay can distinguish between two samples with values close to the bottom of the measuring range (120 and 165% of the lowest calibrator (12.3 IU/mL)) was obtained by statistical analysis (Student's t-test) of the results obtained from testing 20 replicates of each sample.

|   | Replicates | Mean (IU/mL) | Std. Dev. | % CV  |
| --- | --- | --- | --- | --- |
|  Sample 1 | 20 | 20.30 | 1.06 | 5.25  |
|  Sample 2 | 20 | 14.96 | 0.97 | 6.49  |

e. Analytical specificity:

Various elevated serum components (free and conjugated bilirubin, hemoglobin, chyle and rheumatoid factor) were added to low (normal) and high (positive) sera which were then assayed to test the possible effect of interfering substances provided in an Interference Check A plus kit (Kokusai, Japan). No interference was observed at the following concentrations.

|  Substance | Concentration  |
| --- | --- |
|  Bilirubin F (free) | 20.3 mg/dL  |
|  Bilirubin C (conjugated) | 20.2 mg/dL  |
|  Hemoglobin | 486 mg/dL  |
|  Chyle | 1460 Units  |
|  Rheumatoid factor | 45 IU/mL  |

Interference due to high IgG was evaluated in a separate study by testing 6 IgG myeloma sera on the FARRZYME assay and all gave negative results.

f. Assay cut-off:

The assay cut-off was established based on the results obtained from testing samples from healthy blood donors (n=150). At the established cut-off of ≤30

{5}

IU/mL as negative and &gt;30 IU/mL as positive, the 150 normal sera all gave results below 17.0 IU/mL, with 146 (97%) of the results below 12.3 IU/mL. Of the 224 SLE patients also tested, 36% of the patient sera gave a positive result which is similar to the range found in the furnished literature of 28-63% for an unselected (active versus inactive) population.

## 2. Comparison studies:

a. Method comparison with predicate device:

Two-hundred fifty-two samples including 189 from patients with SLE, 28 samples from healthy normals, and 35 samples positive for dsDNA antibodies by C. luciliae or ELISA were tested by FARRZYME, by the TBS conventional dsDNA ELISA and by the TBS CRITHIDIA LUCILIAE immunofluorescent assay. The low positive agreements with both predicate devices are due to the fact that both the conventional dsDNA ELISA and the C. luciliae detect low, moderate and high avidity antibodies whereas the FARRZYME only detects high avidity antibodies.

|   | BINDAZYME dsDNA ELISA  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  FARRZYME | + | 47 | 4* | 51  |
|   |  - | 33 | 168* | 201  |
|   |  Total | 80 | 172 | 252  |
|  Positive Percent Agreement |   | 58.8% (47/80)  |   |   |
|  Negative Percent Agreement |   | 97.7% (168/172)  |   |   |
|  Overall Agreement |   | 85.3% (215/252)  |   |   |

* Includes (3) Borderline BINDAZYME results as negative
** Includes (51) Borderline BINDAZYME results as negative

|   | Crithidia luciliae assay  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  FARRZYME | + | 37 | 14 | 51  |
|   |  - | 18 | 183 | 201  |
|   |  Total | 55 | 197 | 252  |
|  Positive Percent Agreement |   | 67.3% (37/55)  |   |   |
|  Negative Percent Agreement |   | 92.9% (183/197)  |   |   |
|  Overall Agreement |   | 87.3% (220/252)  |   |   |

b. Comparison to the Farr RIA

A comparison was performed between the FARRZYME ELISA and the Farr RIA assay which detects high avidity antibodies. One hundred sixty-six samples were included.

|   | Farr RIA  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  FARRZYME | + | 44 | 7 | 51  |
|   |  - | 5 | 110 | 115  |
|   |  Total | 49 | 117 | 166  |

{6}

7

Positive Percent Agreement | 89.8% (44/49)
Negative Percent Agreement | 94.0% (110/117)
Overall Agreement | 92.8% (154/166)

c. Matrix comparison:
The FARRZYME assay and all other assays in the comparison studies use serum as the matrix.

3. Clinical studies:
To check that the type of samples tested in a clinical laboratory did not give a higher incidence of positive samples than seen within the normal range, samples from patients with other diseases and samples known to contain other autoimmune antibodies were run with the FARRZYME assay. The study results were:

|  Disease controls | N | Result  |
| --- | --- | --- |
|  Ulcerative colitis | 13 | Negative  |
|  Crohn’s disease | 16 | Negative  |
|  Samples tested for Diphtheria toxoid | 9 | Negative  |
|  Known autoantibodies |  |   |
|  Scl-70 | 3 | Negative  |
|  Jo-1 | 4 | Negative  |
|  PR3 | 3 | Negative  |
|  MPO | 4 | Negative  |
|  GBM | 3 | Negative  |
|  Gliadin IgG | 3 | Negative  |
|  TPO | 3 | Negative  |
|  ASCA | 3 | Negative  |

4. Clinical cut-off:
See Assay cut-off

5. Expected values/Reference range:
See Assay cut-off

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LRM/K062183](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LRM/K062183)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
