← Product Code [LLL](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL) · K190710

# EliA SymphonyS Immunoassay (K190710)

_Phadia AB · LLL · Nov 29, 2019 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL/K190710

## Device Facts

- **Applicant:** Phadia AB
- **Product Code:** [LLL](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL.md)
- **Decision Date:** Nov 29, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

EliA SymphonyS is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (Li-heparin, EDTA). EliA SymphonyS is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren’s syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA SymphonyS uses the EliA IgG method on the instrument Phadia 250. EliA SymphonyS is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (Li-heparin, EDTA). EliA SymphonyS is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren’s syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA SymphonyS uses the EliA IgG method on the instrument Phadia 2500/5000.

## Device Story

EliA SymphonyS is an automated in vitro diagnostic immunoassay for detecting antinuclear IgG antibodies. Input: human serum or plasma (Li-heparin, EDTA). Principle: wells coated with recombinant U1RNP, SS-A/Ro, SS-B/La, Centromere B, Scl-70, Jo-1 proteins, and synthetic SmD3 peptide. Patient antibodies bind to antigens; non-bound antibodies washed away; enzyme-labeled anti-IgG conjugate added; development solution added; fluorescence measured. Output: qualitative ratio compared to calibrators. Used in clinical laboratories on Phadia 250 or Phadia 2500/5000 instruments. Healthcare providers use results as an aid in diagnosing systemic autoimmune rheumatic diseases (SARDs) alongside clinical findings. Benefits: standardized, automated screening for multiple autoantibodies simultaneously.

## Clinical Evidence

Clinical performance evaluated using 444 serum samples (194 SARD, 250 differential diagnosis controls). Sensitivity ranged from 52.6% to 55.2% and specificity from 94.4% to 94.8% depending on treatment of equivocal results. Analytical performance included precision/reproducibility studies across multiple instruments and lots, and interference testing (endogenous/exogenous).

## Technological Characteristics

Solid-phase fluoroimmunoassay; polystyrene wells coated with recombinant proteins and synthetic peptide; beta-galactosidase labeled anti-human IgG conjugate; 4-Methylumbelliferyl-beta-D-galactoside (MUG) substrate; automated processing on Phadia 250/2500/5000 platforms; qualitative output via fluorometric measurement.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- Phadia EliA Symphony Immunoassay ([K072149](/device/K072149.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K190710

B. Purpose for Submission:
Revision of assay, migration to new instrument

C. Measurand:
Human IgG autoantibodies to:
- U1RNP/RNP70 (A/C)
- SS-A/Ro (52, 60 kDa)
- SS-B/La
- Centromere B (CENP-B)
- Scl-70 (topoisomerase I)
- Jo-1 (histidyl tRNA synthetase)
- SmD3 peptide

D. Type of Test:
Automated qualitative multianalyte solid-phase immunoassay

E. Applicant:
Phadia US, Inc.

F. Proprietary and Established Names:
Phadia EliA Symphony$ Immunoassay

G. Regulatory Information:
1. Regulation section:
21 CFR §866.5100– Antinuclear antibody immunological test system
2. Classification:
Class II

{1}

3. Product code:
LLL – Extractable Antinuclear Antibody, Antigen And Control

4. Panel:
Immunology (82)

H. Intended Use:

1. Intended use:
EliA Symphony$^{S}$ is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (Li-heparin, EDTA). EliA Symphony$^{S}$ is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren’s syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony$^{S}$ uses the EliA IgG method on the instrument Phadia 250.

EliA Symphony$^{S}$ is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (Li-heparin, EDTA). EliA Symphony$^{S}$ is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren’s syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony$^{S}$ uses the EliA IgG method on the instrument Phadia 2500/5000.

2. Indications for use:
Same as Intended Use

3. Special conditions for use statement(s):
For prescription use only

4. Special instrument requirements:
For use on the Phadia 250, Phadia 2500, or Phadia 5000 instruments

I. Device Description:
This IVD is a qualitative solid-phase fluoroimmunoassay, for the determination of autoantibodies against a panel of nine antigens: UIRNP/RNP70(A,C), SS-A/Ro (52,60), SS-

{2}

B/La, Centromere B (CENP), Scl-70, Jo-1, and  $\mathrm{SmD3}^P$  peptide. All individual analytes have all been previously cleared in standalone formats.

## General Description of the EliA Symphony $^S$  Test System

**Assay-specific reagents:**

- EliA Symphony $^S$  Wells are coated with nine human antigens (eight recombinant proteins, and one synthetic peptide) – 4 carriers (12 wells each), ready to use
- EliA ANA Positive Controls: Pooled human sera, containing IgG ANA against dsDNA, RNP, Sm, Ro, La, Scl-70, CENP, and Jo-1

**EliA Method-specific reagents:**

- EliA Sample Diluent: PBS containing BSA, detergent and  $0.095\%$  sodium azide – 6 bottles,  $48~\mathrm{mL}$  each, ready to use; or 6 bottles,  $400~\mathrm{mL}$  each, ready to use
- EliA IgG Conjugate 50 or 200:  $\beta$ -Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and  $0.06\%$  sodium azide – 6 wedge-shaped bottles,  $5\mathrm{mL}$  each, ready to use; or 6 wedge-shaped bottles,  $19\mathrm{mL}$  each, ready to use
- EliA IgG Calibrator Strips: Human IgG (0, 4, 10, 20, 100,  $600~\mu \mathrm{g / L}$ ) in PBS containing BSA, detergent and  $0.095\%$  sodium azide – 5 strips, 6 single-use vials per strip,  $0.3\mathrm{mL}$  each, ready to use
- EliA IgG Curve Control Strips: Human IgG  $(20\mu \mathrm{g} / \mathrm{L})$  in PBS containing BSA, detergent and  $0.095\%$  sodium azide - 5 strips, 6 single-use vials per strip,  $0.3\mathrm{mL}$  each, ready to use
- EliA IgG Calibrator Well: Coated with mouse monoclonal antibodies – 4 carriers (12 wells each), ready to use.
- EliA IgG/IgM/IgA Negative Control 2500/5000: Human sera from healthy donors in PBS containing BSA, detergent and  $0.095\%$  sodium azide – 6 single-use vials, 0.3 mL each, ready to use

**General Reagents:**

- Development Solution, Stop Solution, and Washing Solution

## Instrumentation

The EliA Symphony $^S$  test is to be performed on the Phadia 250 or Phadia 2500/5000 instruments. The instruments are automated platforms for EliA test procedures from sample and reagent handling up to processing of results.

## J. Substantial Equivalence Information:

1. Predicate device name(s):

Phadia EliA Symphony Immunoassay

2. Predicate 510(k) number(s):

K072149

{3}

# 3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended Use | EliA Symphonyδ is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (Li-heparin, EDTA). EliA Symphonyδ is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphonyδ uses the EliA IgG method on the instrument Phadia 250. EliA Symphonyδ is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (Li-heparin, EDTA). EliA Symphonyδ is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony uses the EliA IgG method on the instrument ImmunoCAP 250. | EliA Symphony is intended for the in vitro, qualitative measurement of antinuclear IgG antibodies in human serum and plasma (heparin, EDTA and citrate). EliA Symphony is based on human recombinant U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and native purified Sm proteins as antigen and is useful as an aid in the clinical diagnosis of patients with systemic lupus erythematosus (SLE), mixed connective tissue disease (MCTD), Sjögren's syndrome, scleroderma and polymyositis/dermatomyositis, in conjunction with other laboratory and clinical findings. EliA Symphony uses the EliA IgG method on the instrument Phadia 2500/5000  |
|  Assay type | Solid-phase fluoroimmunoassay | Same  |
|  Assay output | Qualitative | Same  |
|  Incubation temperature | 37°C | Same  |
|  Calibrators | 6 point, total IgG | Same  |
|  Calibration Method | EliA 6-point total human IgG calibration | Same  |
|  Positive | EliA ANA Positive Control 250, | EliA ANA Positive Control 100,  |

{4}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Control | 2500/5000 | 250 (Same material, different package for Phadia instrument)  |
|  Negative Control | EliA IgG/IgM/IgA Negative Control 250, 2500/5000 | EliA IgG/IgM/IgA Negative Control 100, 250 (Same material, different package for each Phadia instrument)  |
|  Conjugate (Detection antibody) | β-Galactosidase labeled monoclonal mouse anti-human IgG | Same  |
|  Detection substrate | 0.01% 4-Methylumbelliferyl-β-D-galactoside (MUG) | Same  |
|  Sample Dilution | 1:100 | Same  |
|  Cutoff, Interpretation | Negative (-): <0.7 Equivocal (Eq): 0.7–1.0 Positive (+): >1.0 | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Instrument platforms | Phadia 250 Phadia 2500 Phadia 5000 | Phadia 100 Phadia 250  |
|  Matrices | Serum Plasma (Li-heparin, EDTA) | Serum Plasma (Li-heparin, EDTA, citrate)  |
|  Measurands | Autoantibodies to: • U1RNP/RNP70 (A/C) • SS-A/Ro (52, 60 kDa) • SS-B/La • Centromere B (CENP-B) • Scl-70 • Jo-1 • SmD3 peptide | Autoantibodies to: • U1RNP/RNP70 (A/C) • SS-A/Ro (52, 60 kDa) • SS-B/La • Centromere B (CENP-B) • Scl-70 • Jo-1 • native purified Sm antigens  |
|  Reagent Stability: | Closed-kit shelf-life: 18 mo. Open-kit: 9 mo. Onboard: 28d | Shelf-life: 24 mo. Expiration date: 18 mo.  |

# K. Standard/Guidance Document Referenced (if applicable):

|  Org | Standard ID | Version | Title  |
| --- | --- | --- | --- |
|  CLSI | EP05 | A3 | Evaluation of Precision Performance of Quantitative Measurement Methods  |
|  CLSI | EP7 | ED3 | Interference in Clinical Chemistry  |

{5}

|  Org | Standard ID | Version | Title  |
| --- | --- | --- | --- |
|  CLSI | EP17 | A | Protocols for Determination of Limits of Detection and Limits of Quantification  |

# L. Test Principle:

EliA tests are fluorescence immunoassays for the detection and measurement of human antibodies based on EliA solid-phase components, which contain specific antigens for the antibodies to be measured. EliA uses a modular reagent system. The test specific, method specific, and general reagents are packaged and purchased as separate units.

EliA Symphony $^{S}$ : Polystyrene EliA wells are coated with a mixture of the target antigens: human U1RNP (RNP 70, A, C), SS-A/Ro (60 kDa, 52 kDa), SSB/La, Centromere B, Scl-70, Jo-1 proteins and a synthetic SmD3 peptide. Together, this grouping of autoantigenic targets are collectively referred to as extractable nuclear antigens (ENA). If present in the patient's specimen, autoantibodies to these proteins bind to the specific antigen. After washing away unbound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a fluorogenic development solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. Fluorometric measurements are converted to an arbitrary unitless ratio, by scaling to EliA total IgG calibration. The higher the ratio, the more autoantigen-specific IgG is present in the specimen. Qualitative interpretation of this ratio value is summarized in the table below.

|  Qualitative interpretation of Symphony\(^S\) results  |   |
| --- | --- |
|  Interpretation | [Ratio]  |
|  Negative (-) | < 0.7  |
|  Equivocal (Eq) | 0.7–1.0  |
|  Positive (+) | > 1.0  |

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

Acceptance criteria for all parameters was met.

# a. Precision/Reproducibility:

To evaluate reproducibility of the EliA Symphony $^{S}$  assay on the Phadia 250 instrument platform, six samples with a range of reactivities were tested. The samples were native samples or positive samples diluted with normal sera. Samples 2 through 6 were tested with three lots on three instruments  $\times$  seven days  $\times$  one run/day  $\times$  four

{6}

replicates for a total of 252 possible datapoints per sample while Sample 1 (negative sample) was tested with one lot on three instruments  $\times$  seven days  $\times$  one run/day  $\times$  four replicates for a total of 84 possible datapoints.

Reproducibility of EliA Symphony on Phadia 250 instrument:

|  Sample | Mean Ratio | Range of Ratios | Expected Result | Phadia 250 Results  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |   |   |  Negative/Equivocal/Positive | % Correct  |
|  1 | 0.48 | 0.38–0.59 | Negative | 0/0/84 | 100  |
|  2 | 0.76 | 0.63–0.96 | Equivocal | 47/205/0 | 81.35  |
|  3 | 1.02 | 0.91–1.13 | Positive | 0/100/152 | 60.32  |
|  4 | 4.35 | 3.82–5.09 | Positive | 0/0/252 | 100  |
|  5 | 23.21 | 20.79–26.20 | Positive | 0/0/252 | 100  |
|  6 | 42.92 | 34.41–53.97 | Positive | 0/0/250 | 100  |

Lot-to-lot reproducibility on the Phadia 250 was calculated using Samples 2 through 6:

|   | Mean Ratio | Expected Result | Lot 1 |   | Lot 2 |   | Lot 3 |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct  |
|  2 | 0.76 | Equivocal | 2/82/0 | 97.62 | 23/61/0 | 72.62 | 22/62/0 | 73.81 | 47/205/0 | 81.35  |
|  3 | 1.02 | Positive | 0/14/70 | 83.33 | 0/37/47 | 55.95 | 0/49/35 | 41.67 | 0/100/152 | 60.32  |
|  4 | 4.35 | Positive | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/252 | 100  |
|  5 | 23.21 | Positive | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/252 | 100  |
|  6 | 42.92 | Positive | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/250 | 100  |

Instrument-to-instrument reproducibility on the Phadia 250 was calculated using Samples 2 through 6:

|   | Mean Ratio | Expected Result | Instrument 1 |   | Instrument 2 |   | Instrument 3 |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct  |
|  2 | 0.76 | Equivocal | 1/83/0 | 98.81 | 7/77/0 | 91.66 | 39/45/0 | 53.57 | 47/205/0 | 81.35  |
|  3 | 1.02 | Positive | 0/24/60 | 71.43 | 0/21/63 | 75.00 | 0/55/29 | 34.53 | 0/100/152 | 60.32  |
|  4 | 4.35 | Positive | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/252 | 100  |
|  5 | 23.21 | Positive | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/252 | 100  |
|  6 | 42.92 | Positive | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/84 | 100 | 0/0/250 | 100  |

Reproducibility on the Phadia 2500/5000 instrument platform was tested using seven samples with a range of reactivities were tested with one lot on three instruments  $\times$  seven days  $\times$  one run/day  $\times$  four replicates for a total of 84 possible datapoints per sample:

{7}

Reproducibility of EliA Symphony$^{S}$ on Phadia 2500/5000 instrument:

|  Sample | Mean Ratio | Range of Ratios | Expected Result | Phadia 2500/5000Results  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |   |   |  Negative/Equivocal/Positive | % Correct  |
|  1 | 0.63 | 0.38–0.59 | Negative | 67/17/0 | 79.76  |
|  2 | 0.76 | 0.61–0.87 | Equivocal | 12/72/0 | 85.7  |
|  3 | 0.92 | 0.76–1.04 | Equivocal | 0/77/7 | 91.7  |
|  4 | 0.98 | 0.90–1.07 | Equivocal | 0/55/29 | 65.5  |
|  5 | 4.15 | 3.72–4.63 | Positive | 0/0/84 | 100  |
|  6 | 28.07 | 25.53–33.11 | Positive | 0/0/84 | 100  |
|  7 | 47.11 | 41.02–54.55 | Positive | 0/0/84 | 100  |

Instrument-to-instrument reproducibility on the Phadia 2500/5000 was calculated:

|   | Mean Ratio | Expected Result | Instrument 1 |   | Instrument 2 |   | Instrument 3 |   | Total  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct | Neg/ Equiv/Pos | % Correct  |
|  1 | 0.63 | Negative | 24/4/0 | 85.71 | 25/3/0 | 89.28 | 18/10/0 | 64.28 | 67/17/0 | 79.76  |
|  2 | 0.76 | Equivocal | 8/20/0 | 71.43 | 3/25/0 | 89.28 | 1/27/0 | 96.43 | 12/72/0 | 85.71  |
|  3 | 0.92 | Equivocal | 0/24/0 | 85.71 | 0/26/2 | 92.86 | 0/27/1 | 96.43 | 0/77/7 | 91.67  |
|  4 | 0.98 | Equivocal | 0/16/12 | 57.14 | 0/25/3 | 89.28 | 0/14/14 | 50.00 | 0/55/29 | 65.48  |
|  5 | 4.15 | Positive | 0/0/28 | 100 | 0/0/28 | 100 | 0/0/28 | 100 | 0/0/84 | 100  |
|  6 | 28.07 | Positive | 0/0/28 | 100 | 0/0/28 | 100 | 0/0/28 | 100 | 0/0/84 | 100  |
|  7 | 47.11 | Positive | 0/0/28 | 100 | 0/0/28 | 100 | 0/0/28 | 100 | 0/0/84 | 100  |

b. Linearity/assay reportable range:

Not applicable; the assay is qualitative.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability:

The EliA IgG calibration method and traceability were previously reviewed in K072149. There are no international standards for ANA antibodies. The positive and negative controls are prepared in house and arbitrary Ratio units are assigned during the development process.

Stability:

Closed-kit shelf-life stability at $2 - 8^{\circ}\mathrm{C}$ for EliA Symphony$^{S}$ wells was determined as 18 months, based on accelerated and real-time stability studies. Opened-kit shelf-life claim is 9 months, based on real-time stability studies.

Onboard storage stability claim is 28 days, based on real-time stability studies using the storage capabilities of the Phadia 250 instrument.

{8}

For sample stability, the assay directions refer to CLSI H18-A4: Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline – Fourth Edition.

d. Detection limit:

Not applicable; the assay is qualitative.

e. Analytical specificity:

Endogenous Interference:

Three serum samples (a negative, an equivocal, and a positive sample) were spiked with the five endogenous interfering test substances at the concentrations in undiluted samples listed, or blank vehicle solvent. The negative and equivocal samples were native samples and the positive sample was generated by mixing a positive sample with normal sera. Samples were tested in triplicate over two runs on a Phadia 250 instrument. The ratio of blank/spiked sample ranged from 0.88–1.16 for EliA Symphony⁵. None of the interferents changed the expected result. The substances tested and their tested concentrations are in the table below:

|  Interferent | Concentration  |
| --- | --- |
|  Bilirubin F | 19.2 mg/dL  |
|  Bilirubin C | 20.1 mg/dL  |
|  Hemoglobin | 496 mg/dL  |
|  Lipemic factor (ClinOleic) | 1%  |
|  Rheumatoid factor | 500 IU/mL  |

Exogenous Interference:

Three serum samples (a negative, an equivocal, and a positive sample) were spiked with the six exogenous interfering test substances in undiluted samples listed, or blank vehicle solvent at the concentrations listed in the table below. The negative and equivocal samples were native samples and the positive sample was generated by mixing a positive sample with normal sera. Samples were tested in triplicate over two runs on a Phadia 250 instrument. The ratio of blank/spiked sample ranged from 0.83–1.13 for EliA Symphony⁵. None of the interferents changed the expected result. The substances tested and their tested concentrations are in the table below:

|  Endogenous interference testing for EliA Symphony⁵ on Phadia 250  |   |
| --- | --- |
|  Interferent | Concentration  |
|  Ibuprofen | 0.219 mg/dL  |
|  Prednisone | 0.09 μg/dL  |
|  Hydroxychloroquine | 2.25 mg/dL  |
|  Azathioprine | 2.58 μg/dL  |
|  Losartan | 11.4 μg/dL  |
|  Infliximab¹ | 0.264 mg/dL  |

{9}

10

|  Endogenous interference testing for EliA Symphony$^{S}$ on Phadia 250  |   |
| --- | --- |
|  Interferent | Concentration  |
|  $^{1}$ tested in two runs, independent of other exogenous interferents  |   |

## Reference Sera:

Reference panels of qualified sera for ANA from the Centers for Disease Control and Prevention (CDC) and the Association of Medical Laboratory Immunologists (AMLI) were tested with the EliA Symphony$^{S}$ in duplicate 1:100 dilutions, using a single lot of EliA Symphony$^{S}$ reagents. Among the 12 CDC panel samples, 100% returned the expected result. Among the 10 AMLI panel samples, two reference samples expected to be positive (i.e. RNP/Sm and Scl-70) returned negative results.

## f. Assay cut-off:

To establish the assay cutoff, 70 samples from “apparently healthy” donors and 30 from SARD patients, known to be positive for the constituent single ANA autoantibody specificities represented in the EliA Symphony$^{S}$ assay were tested, using a Phadia 250 instrument. All known ANA positive samples were found to be positive on the EliA Symphony$^{S}$ assay, and the 95th percentile of the normal population lay below the equivocal range.

## 2. Comparison studies:

### a. Method comparison with predicate device:

A total of 633 serum samples were tested with the EliA Symphony$^{S}$ and the predicate assay. Samples were selected to represent intended use SARD diseases and differential diagnosis controls, summarized in the table below.

|  Disease/Indication  |   |   |   |
| --- | --- | --- | --- |
|  ANA-associated SARD samples |  | Differential diagnosis controls |   |
|  Systemic lupus erythematosus | 97 | Rheumatoid arthritis | 85  |
|  Sjögren’s syndrome | 96 | Hepatitis B virus | 36  |
|  Systemic sclerosis | 87 | Hepatitis C virus | 36  |
|  Polymyositis/dermatomyositis | 78 | HIV | 27  |
|  Mixed connective tissue disease | 46 | Cancer | 25  |
|   |  | Bacterial infection | 20  |
|  n = | 404 | n = | 229  |
|  Total n = 633  |   |   |   |

All samples were run in singlicate on a Phadia 250 instrument. Qualitative agreement of positive and negative results were compared between the EliA Symphony$^{S}$ test and

{10}

the predicate. Agreement measures were calculated, treating equivocal results as either positive or negative:

|   | Predicate |   |   | Total  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive | Equivocal |   | Negative  |
|  EliA SymphonyS | Positive | 279 | 4 | 2 | 285  |
|   |  Equivocal | 2 | 2 | 4 | 8  |
|   |  Negative | 5 | 19 | 316 | 340  |
|   | Total | 286 | 25 | 322 | 633  |
|  Equivocal results considered positive (95% CI)  |   |   |   |
| --- | --- | --- | --- |
|  PPA: | 287/311 | 92.3% | (88.7 - 95.0%)  |
|  NPA: | 316/322 | 98.1% | (96.0 - 99.3%)  |
|  Equivocal results considered negative (95% CI)  |   |   |   |
|  PPA: | 279/286 | 97.6% | (95.0 - 99.0%)  |
|  NPA: | 341/347 | 98.3% | (96.3 - 99.4%)  |

To evaluate comparison of qualitative results for the EliA Symphony $^S$  test between the two instrument platforms, 81 positive, 10 equivocal, and 19 negative samples were tested with the EliA Symphony $^S$  in singlicate on both Phadia 250 and Phadia 2500/5000 instruments. Agreement measures were calculated, treating equivocal results as either positive or negative:

|   | Phadia 250 |   |   | Totals  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive | Equivocal |   | Negative  |
|  Phadia 2500/5000 | Positive | 80 | 3 | 0 | 83  |
|   |  Equivocal | 1 | 7 | 2 | 10  |
|   |  Negative | 0 | 0 | 17 | 17  |
|   | Totals | 81 | 10 | 19 | 110  |
|  Equivocal results considered positive (95% CI)  |   |   |   |
| --- | --- | --- | --- |
|  PPA: | 93/91 | 98.8% | (93.3–100%)  |
|  NPA | 17/19 | 89.7% | (72.6–97.8%)  |
|  Equivocal results considered negative (95% CI)  |   |   |   |
|  PPA: | 83/81 | 100% | (96.0–100%)  |
|  NPA: | 27/29 | 89.5% | (66.9–98.7%)  |

{11}

# b. Matrix comparison:

To evaluate the commutatability of matrices intended for use with the EliA Symphony $^S$  assay; 62 matched sets of serum, EDTA plasma, and lithium heparin plasma were tested on a Phadia 250 instrument. The cohort composed 29 negative, 10 equivocal, and 23 positive samples (as defined by serum results). The samples covered a range of ratios of the serum values (0.09–69.70).

Qualitative agreement for comparisons of serum samples to the plasma matrices are summarized below. No sample changed its evaluation from negative to positive or vice versa.

Matrix Comparison: Agreement for EliA Symphony between serum and EDTA plasma:

|   | Serum |   |   | Totals  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive | Equivocal |   | Negative  |
|  EDTA plasma | Positive | 22 | 0 | 0 | 22  |
|   |  Equivocal | 1 | 9 | 0 | 10  |
|   |  Negative | 0 | 1 | 29 | 30  |
|   | Totals | 22 | 10 | 29 | 62  |
|  Equivocal results considered positive (95% CI)  |   |   |   |
| --- | --- | --- | --- |
|  PPA: | 32/33 | 97.0% | (84.7–99.5%)  |
|  NPA | 29/29 | 100% | (88.3–100%)  |
|  Equivocal results considered negative (95% CI)  |   |   |   |
|  PPA: | 22/23 | 95.6% | (79.0–99.2%)  |
|  NPA: | 39/39 | 100% | (91.0–100%)  |

Matrix Comparison: Agreement for EliA Symphony between serum and Li-Heparin plasma:

|   | Serum |   |   | Totals  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Positive | Equivocal |   | Negative  |
|  Li-Heparin plasma | Positive | 22 | 0 | 0 | 22  |
|   |  Equivocal | 1 | 8 | 0 | 10  |
|   |  Negative | 0 | 2 | 29 | 30  |
|   | Totals | 23 | 10 | 29 | 62  |
|  Equivocal results considered positive (95% CI)  |   |   |   |
| --- | --- | --- | --- |
|  PPA: | 31/33 | 93.9% | (80.4–98.3%)  |
|  NPA | 29/29 | 100% | (88.3–100%)  |
|  Equivocal results considered negative (95% CI)  |   |   |   |
|  PPA: | 22/24 | 91.7% | (74.2–97.7%)  |
|  NPA: | 37/37 | 100% | (90.6–100%)  |

{12}

# 3. Clinical studies:

# a-b. Clinical Sensitivity and Specificity:

A total of 444 clinical serum samples were tested with the EliA Symphony in singlicate on a Phadia 250 instrument. These samples included 194 samples from SARD diseases, and 250 from control disease categories that could be expected to be found in the differential diagnosis of the diseases in the Indications for Use.

|  Disease/Indication | Total n | N (%) Positive*  |
| --- | --- | --- |
|  ANA-associated SARD samples | 194 |   |
|  Systemic lupus erythematosus | 83 | 37 (44.6%)  |
|  SLE lupus nephritis | 19 | 13 (68.4%)  |
|  Sjögren's syndrome | 28 | 17 (60.7%)  |
|  Systemic sclerosis | 32 | 16 (50.0%)  |
|  Mixed connective tissue disease | 22 | 16 (72.7%)  |
|  Polymyositis/dermatomyositis | 10 | 3 (30.0%)  |
|  Autoimmune disease controls | 190 |   |
|  Rheumatoid arthritis | 40 | 0  |
|  Primary anti-phospholipid syndrome | 28 | 4 (14.3%)  |
|  Graves' disease | 28 | 1 (3.6%)  |
|  Hashimoto's thyroiditis | 10 | 1 (10.0%)  |
|  Celiac disease | 28 | 0  |
|  Crohn's disease | 22 | 0  |
|  Ulcerative colitis | 22 | 1 (4.5%)  |
|  Autoimmune hepatitis | 6 | 3 (50.0%)  |
|  Primary biliary cholangitis | 5 | 0  |
|  Granulomatosis with polyangiitis | 1 | 0  |
|  Infectious disease controls | 60 |   |
|  bacterial | 30 | 2 (6.7%)  |
|  viral | 30 | 1 (3.3%)  |
|  Total: | 444 |   |

* Equivocal results evaluated as negative

Diagnostic performance, calculated as sensitivity and specificity values, is summarized in the table below. Two sets of clinical sensitivity and specificity measures were calculated, treating equivocal results as either positive or negative.

|   | Diagnoses |   | Totals  |   |
| --- | --- | --- | --- | --- |
|   |   |  SARD |   | Differential  |
|  EliA SymphonyS | Positive | 102 | 13 | 115  |
|   |  Equivocal | 5 | 1 | 6  |
|   |  Negative | 87 | 236 | 323  |
|   | Totals | 194 | 250 | 444  |

{13}

|  Equivocal results considered positive (95% CI)  |   |   |   |
| --- | --- | --- | --- |
|  Sensitivity: | 107/194 | 55.2% | (47.9–62.3%)  |
|  Specificity | 236/250 | 94.4% | (90.8–96.9%)  |
|  Equivocal results considered negative (95% CI)  |   |   |   |
|  Sensitivity: | 102/194 | 52.6% | (45.3–59.8%)  |
|  Specificity | 237/250 | 94.8% | (91.3–97.2%)  |

# 4. Clinical cut-off:

See Assay cutoff (M.1.f)

# 5. Expected values/Reference range:

The frequency for ENA autoantibodies was investigated in a cohort of 558 apparently healthy subjects equally distributed across age and sex, using sera banked by the sponsor, as well as commercially obtained from the U.S. to enrich for racial and ethnic minority populations. Samples were tested with the EliA Symphony assay on a Phadia 250 instrument. Three samples  $(0.5\%)$  were equivocal and seven samples  $(1.3\%)$  were positive on Phadia 250. The results for the total cohort  $(n = 558)$  are depicted in the table below:

|  Reference Ranges for EliA SymphonyS  |   |
| --- | --- |
|  Median | 0.1  |
|  Mean | 0.3  |
|  95thPercentile | 0.3  |
|  99thPercentile | 1.1  |
|  Min | 0.0  |
|  Max | 25.9  |

# N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.

# O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL/K190710](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LLL/K190710)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
