K123880 · Inova Diagnostics, Inc. · LJM · Feb 7, 2014 · Immunology
Device Facts
Record ID
K123880
Device Name
QUANTA FLASH CENTROMERE
Applicant
Inova Diagnostics, Inc.
Product Code
LJM · Immunology
Decision Date
Feb 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
QUANTA Flash Centromere is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-centromere protein B autoantibodies in human serum. The presence of anti-centromere protein B autoantibodies is used as an aid in the diagnosis of Systemic Sclerosis, in conjunction with clinical finding and other laboratory tests. QUANTA Flash Centromere Calibrators are intended for use with the QUANTA Flash Centromere Reagents for the determination of IgG anti-centromere protein B autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate the unit values. QUANTA Flash Centromere Controls are intended for use with the QUANTA Flash Centromere Reagents for quality control in the determination of IgG anti-centromere protein B autoantibodies in human serum.
Device Story
QUANTA Flash Centromere is a chemiluminescent immunoassay (CLIA) for detecting IgG anti-centromere protein B autoantibodies in human serum. The device consists of reagents, calibrators, and controls used on an automated instrument platform. It functions by measuring light emission proportional to the concentration of specific autoantibodies in the patient sample. Used in clinical laboratory settings by trained personnel, the output provides semi-quantitative results to assist clinicians in diagnosing Systemic Sclerosis. By identifying specific autoantibodies, the device supports clinical decision-making when combined with other laboratory tests and clinical findings, potentially facilitating earlier or more accurate diagnosis for patients suspected of having systemic sclerosis.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on analytical performance testing, including precision, linearity, and method comparison studies against the predicate ELISA device.
Technological Characteristics
Chemiluminescent immunoassay using paramagnetic beads coated with recombinant CENP-B antigen. Detection via isoluminol-conjugated anti-human IgG monoclonal antibody. Energy source: BIO-FLASH instrument optical system. Connectivity: Barcode-based reagent/master curve loading. Software: Instrument-specific firmware for RLU measurement and unit calculation. Sterilization: Not applicable (reagents). Form factor: Reagent cartridge system.
Indications for Use
Indicated for the semi-quantitative determination of IgG anti-centromere protein B autoantibodies in human serum as an aid in the diagnosis of Systemic Sclerosis in conjunction with clinical findings and other laboratory tests. For prescription use only.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
QUANTA Lite Centromere ELISA (k003959)
Related Devices
K151559 — ImmuLisa Enhanced Centromere Antibody ELISA · Immco Diagnostics, Inc. · Mar 11, 2016
K250666 — Alegria Flash CTD Screen · Zeus Scientific · Oct 22, 2025
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k123880
B. Purpose for Submission:
New Device
C. Measurand:
Anti-centromere protein B anti-nuclear antibodies
D. Type of Test:
Semi-Quantitative, Chemiluminescent Immunoassay (CIA)
E. Applicant:
INOVA Diagnostics, Inc.
F. Proprietary and Established Names:
QUANTA Flash Centromere Reagents
QUANTA Flash Centromere Calibrators
QUANTA Flash Centromere Controls
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5100, Antinuclear Antibodies Immunological Test System
21 CFR §862.1150, Calibrator
21 CFR §862.1660, Quality Control Material (Assayed and Unassayed)
2. Classification:
Class II – Assay and Calibrator
{1}
Class I – Control
3. Product code:
LJM – Antinuclear antibody (enzyme-labeled), antigen, controls
JIX – Calibrator, Multi-Analyte Mixture
JJX – Single (Specified) Analyte Controls (Assayed and Unassayed)
4. Panel:
Immunology (82) (Assay)
Chemistry (75) (Calibrator and Controls)
H. Intended Use:
1. Intended use(s):
QUANTA Flash Centromere is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-centromere protein B autoantibodies in human serum. The presence of anti-centromere protein B autoantibodies is used as an aid in the diagnosis of Systemic Sclerosis, in conjunction with clinical finding and other laboratory tests.
QUANTA Flash Centromere Calibrators are intended for use with the QUANTA Flash Centromere Reagents for the determination of IgG anti-centromere protein B autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate the unit values.
QUANTA Flash Centromere Controls are intended for use with the QUANTA Flash Centromere Reagents for quality control in the determination of IgG anti-centromere protein B autoantibodies in human serum.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
BIO-FLASH® Instrument System (k083518)
{2}
I. Device Description:
The QUANTA Flash Centromere Reagent cartridge contains one reagent cartridge with sufficient material for 50 tests. The QUANTA Flash Centromere reagent cartridge contains 4 reagent tubes:
- Centromere coated to paramagnetic beads in a buffered solution containing protein stabilizers and preservative;
- Assay buffer – phosphate-buffered saline solution containing Tween 20, protein stabilizers and preservatives;
- Tracer IgG – Isoluminol labeled anti-human IgG antibodies in a buffered solution containing protein stabilizers and preservative;
- Centromere Reagents include Resuspension Buffer and a transfer pipette for re-suspending the beads.
The QUANTA Flash Centromere Calibrator set includes two calibrators (Calibrator 1 and Calibrator 2). These are barcoded tubes containing 0.3 mL pre-diluted, ready-to-use reagent. Calibrators contain human antibodies to centromere in buffer. They are sold separately.
The QUANTA Flash Centromere Controls contain two vials (a Negative and a Positive) containing human antibodies to centromere in buffer, protein stabilizers and preservatives. They are sold separately.
The following additional reagents are required for the test and supplied separately; BIO-FLASH System Rinse, BIO-FLASH Triggers, and BIO-FLASH Cuvettes.
J. Substantial Equivalence Information:
1. Predicate device name(s):
QUANTA Lite Centromere ELISA
2. Predicate 510(k) number(s):
k003959
3. Comparison with predicate:
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | QUANTA Flash Centromere is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-centromere protein B autoantibodies in human serum. The presence of anti-centromere protein B autoantibodies is used as an aid in the diagnosis of connective tissue diseases including systemic sclerosis in conjunction with clinical finding and other laboratory tests. | Same |
| Assay Type | Semi-quantitative immunoassay | Same |
| Analyte Detected | Human anti-Centromere autoantibodies (IgG) | Same |
| Sample Matrix | Serum | Same |
| Cut-off between positive and negative | 20 units | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Methodology | Chemiluminescent Immunoassay utilizing magnetic particles | Enzyme-linked Immunosorbent Assay |
| Conjugate | Isoluminol conjugated monoclonal anti-human IgG antibodies | Horse radish peroxidase conjugated goat anti-human IgG antibodies |
| Signal Detected | Luminescence (visible light) | Absorbance at 450nm |
| Calibration and unit calculation | Instrument specific working curve based off a 6 point lot specific master curve used for unit calculations; stored on the instrument for 73 days for Centromere assay. | Single point calibrator run each time the assay is run |
| Centromere antigen | Recombinant CENP-B antigen | Recombinant CENP-A and CENP-B antigen |
{4}
5
K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods.
- CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation.
- CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach.
- CLSI H18-A4, Procedures for Handling and Processing Blood Specimens.
L. Test Principle:
The assay technology is similar to other solid phase indirect immunosorbent assays. The assay solid phase is paramagnetic beads centromere antigens coated onto paramagnetic beads. Patient samples are loaded into sample racks which are then placed into the sample carousel of the instrument. A patient’s serum is diluted by the instrument with sample dilution buffer in a disposable cuvette. A small amount of the diluted sample is combined with assay buffer and antigen coated beads in a second cuvette, and mixed. This reaction cuvette is incubated at 37°C. The cuvette is then exposed to a small magnet that holds the beads in place, the liquid is aspirated, and the beads are re-suspended as system rinse is added to the cuvette and the magnet is removed. This wash cycle is repeated one more time. During the third wash, no system rinse is added after the aspiration step.
The detecting reagent is an isoluminol-conjugated anti-human IgG monoclonal antibody. After the third wash, isoluminol conjugated monoclonal anti-human IgG is added to the beads in the cuvette, and mixed. Again, the cuvette is incubated at 37°C. Then, three wash steps, as described above, are performed on the beads. In the fourth wash step, no liquid is added to the beads after the aspiration. The cuvette is then placed in a light-tight luminometer and the beads are exposed to a base and an oxidizing agent. These two reagents, or “Triggers”, cause the isoluminol to produce a flash of visible light. The light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of autoantibodies bound to the antigen on the beads.
The QUANTA Flash Centromere assay utilizes a predefined lot specific Master Curve that is loaded into the instrument through the reagent cartridge barcode. Based on the results of two calibrators, a working curve is created, which is used to calculate chemiluminescent units (CU) from RLU. The calibrators are not included as part of the QUANTA Flash Centromere Reagent Kit.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{5}
Precision of the QUANTA Flash Centromere assay was evaluated by running eight patient samples in duplicates, twice a day, for 20 days on one reagent lot (total of 80 replicates per sample). Controls were run as quality control during each run. The study results are summarized in the table below:
| | | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (CU) | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| A | 6.2 | 0.2 | 2.6% | 0.2 | 3.0% | 0.3 | 5.4% | 0.4 | 6.7% |
| B | 13.5 | 0.5 | 3.7% | 0.3 | 2.1% | 0.7 | 4.9% | 0.9 | 6.5% |
| C | 16.6 | 0.9 | 5.2% | 0.9 | 5.6% | 0.9 | 5.5% | 1.6 | 9.4% |
| D | 16.8 | 0.8 | 4.9% | 0.9 | 5.3% | 1.0 | 5.9% | 1.6 | 9.3% |
| E | 22.9 | 0.9 | 3.7% | 0.5 | 2.2% | 0.9 | 4.0% | 1.4 | 5.9% |
| F | 60.9 | 2.0 | 3.2% | 1.3 | 2.1% | 3.4 | 5.5% | 4.1 | 6.7% |
| G | 103.3 | 5.7 | 5.6% | 0.4 | 0.4% | 4.2 | 4.1% | 7.1 | 6.9% |
| H | 197.9 | 13.2 | 6.7% | 0.0 | 0.0% | 11.9 | 6.0% | 17.7 | 9.0% |
| I | 321.4 | 24.5 | 7.6% | 0.0 | 0.0% | 27.8 | 8.7% | 37.1 | 11.5% |
| J | 463.2 | 33.6 | 7.3% | 0.0 | 0.0% | 22.9 | 4.9% | 40.7 | 8.8% |
The lot-to-lot reproducibility of the assay was evaluated by testing samples with concentrations across the assay range. Five samples each in low, moderate, and high concentrations were tested in duplicate in three separate assay lots:
| Group | Sample Range (CU) | Average %CV of Group | Range of % CV between Lots |
| --- | --- | --- | --- |
| Low | 16.8 – 35.2 | 5.8% | 1.0 – 7.3% |
| Moderate | 60.2 – 121.3 | 5.0% | 2.3 – 11.5% |
| High | 214.9 – 568.9 | 7.1% | 1.2 – 11.6% |
## b. Linearity/assay reportable range:
The analytical measuring range (AMR) of the assays is defined by the lowest and highest points on the Master Curve (3.4 - 708.9 CU). The linearity across this range was determined by diluting a high positive samples with a negative samples by the dilution scheme recommended in CLSI EP6-A. To cover the entire range several dilution series (starting at different points on the curve) were generated and tested. The expected value was plotted against the observed value, and linear regression analysis was performed to get slope, intercept and $\mathbb{R}^2$ values. The results are summarized in the table below.
## Hook effect:
There is no hook effect observed for the Centromere assay.
{6}
| Sample | Test Range (CU) | Slope (95% CI) | Y-intercept (95% CI) | R² | Percent Recovery |
| --- | --- | --- | --- | --- | --- |
| 8 | 4.1 – 20.3 | 1.01 (0.98 –1.03) | 0.25 (-0.04 –0.56) | 1.00 | 99.8 –105.9% |
| 4 | 6.4 –63.7 | 0.98 (0.96 –1.00) | 1.89 (1.15 –2.64) | 1.00 | 80.3 –100.5% |
| 9 | 10.6 –106.1 | 1.01 (0.99 – 1.02) | -1.24 (-2.41 –-0.07) | 1.00 | 99.1 –109.5% |
| 2 | 12.7 –127.4 | 0.99 (0.97 –1.01) | 2.27 (0.89 –3.64) | 1.00 | 80.8 –101.0% |
| 5 | 24.4 –244.4 | 0.99 (0.97 –1.01) | -0.20 (-3.18 –2.78) | 1.00 | 89.7 –106.6% |
| 7 | 90.4 –632.7 | 1.04 (1.01 –1.08) | -25.39 (-38.69 –-12.09) | 1.00 | 86.1 –100.4% |
| All | 4.1 – 632.7 | 1.01 (1.00 –1.01) | -1.10 (-2.61 –0.41) | 1.00 | 80.3 –109.5% |
To assess the auto-rerun capability of the instrument, two samples greater than the highest point of the reportable range were run with the auto-rerun feature enabled. The manual dilution and the instrument re-run result variations were between -6% and 9% for the samples tested.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability and Value Assignment: There is no recognized reference standard for the measurement of anti-Centromere antibodies. Assay calibrators and controls are traceable to in-house standards that are used to create the lot-specific master curve for the QUANTA Flash Centromere assay. Calibrators and controls are manufactured by diluting human serum that contains high titer of anti-centromere antibodies with a buffer stabilizers and preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances. The target CU value is achieved through trial dilutions on a small scale. Once a dilution is selected, the calibrators and controls are bulked, tested, and adjusted. Upon completion of the manufacturing process, the calibrators and controls are tested on at least two instruments, on at least two lots, of reagent cartridge in replicates of 10 to determine final value assignment. Calibrators and control materials are specified in the labeling but are not supplied with the assay.
Kit Stability: The claimed stability of one year at $2 - 8^{\circ}\mathrm{C}$ for unopened kits is based on accelerated studies. Real-time stability is ongoing and the results are shown below. The BIO-FLASH software monitors the expiration dates of the onboard cartridges, as well as the reagent cartridge lots.
{7}
| Product | Real time testing | Specifications met |
| --- | --- | --- |
| QUANTA Flash Centromere Reagents | 10 months | yes |
| QUANTA Flash Centromere Calibrators | 12 months | yes |
| QUANTA Flash Centromere Controls | 12months | yes |
Opened kit components are stable as follows:
| Reagent Cartridge and beads | 73 days |
| --- | --- |
| Calibrators | 8 hours total use time onboard; up to 4 uses |
| Controls | 10 min per use, up to 2.5 hours total onboard or up to 15 uses (whichever is fulfilled first) |
The sponsor recommends that serum samples be stored according to the guidelines in CLSI H18-A4 and cautions against using samples that are microbially contaminated, heat-treated, or contain visible particulates.
# d. Detection limit:
Limit of Blank (LoB): 30 samples were run at least twice, in replicates to obtain at least 60 measurements using Immunoglobulin-stripped serum as the analyte-free material. The RLU value of these samples was lower than the bottom limit of the four parameter logistic curve that the instrument used to calculate CU, and therefore cannot be converted into CU. The Limit of Blank is claimed to be $< 3.4$ CU for Centromere antigen.
Limit of Detection (LoD): Five sera for the Centromere CIAs were serially diluted until the RLU values no longer showed a decreasing trend. Serum samples were diluted to a concentration that was between the LoB and approximately $4 \times$ LoB. These five samples were run in 4 replicates for 4 days; 2 days on 1 instrument and 2 more days on another instrument to obtain 16 measurements per sample for at least 80 measurements. The Limit of Detection for the assay is below the lower limit of the analytical measuring range (3.4 CU).
# e. Analytical specificity:
# Endogenous Interferents:
Sera with Centromere antibodies of different concentrations (negative, positive and around the cut-off) were spiked mixed with known quantities of bilirubin (10, 5 or $2.5\mathrm{mg / dL}$ ), hemoglobin (200, 100 or $50\mathrm{mg / dL}$ ), cholesterol (224.3, 112.2 or $56.0\mathrm{mg / dL}$ ), triglycerides (1000, 500 or 250 mg/dL) or rheumatoid factor (RF) (about 500, 300 or $100\mathrm{IU / mL}$ ). Each
{8}
sample was tested in triplicate and the recovery was calculated by comparing to control samples spiked with the same volume of diluents. No interference $(\leq 15\%)$ was detected in the samples up to the concentrations listed in the table below:
| Potential Interfering Compound | Maximum Concentration | Range of % Recovery |
| --- | --- | --- |
| Bilirubin | 10 mg/dL | 91 – 100% |
| Hemoglobin | 200 mg/dL | 89 – 100% |
| Triglycerides | 1000 mg/dL | 87 – 100% |
| Cholesterol | 224.3 mg/dL | 87 – 100% |
| Rheumatoid factor | 500 IU/mL | 100 – 115% |
Analytical Cross-reactivity:
A potential cross-reactivity of the QUANTA Flash Centromere CIA with other autoantibodies was evaluated by testing 51 SLE samples known to have high levels of various other antibodies (i.e., anti-dsDNA, anti-CCP, anti-chromatin, anti-DGP, etc.). All samples were negative by the QUANTA Flash Centromere CIA.
f. Assay cut-off:
The assay cut-off is 20 CU. Results $< 20$ CU should be reported as negative and results $\geq 20$ CU should be reported as positive.
2. Comparison studies:
a. Method comparison with predicate device:
Three hundred and fifty eight (358) samples were tested by QUANTA Lite Centromere ELISA and QUANTA Flash Centromere CIA. Summarized in the table below, the positive and negative agreements are $84.5\%$ and $99.3\%$ , respectively and the overall agreement is $95.4\%$ . Of the 358 samples, only 107 were within the analytical measuring range:
| | Centromere ELISA | | | Percent Agreement (95% CI) | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | | Total |
| QUANTA Flash Centromere | Positive | 41 | 1 | 42 | Pos. Agree = 85.4% (72.2-93.9%) |
| | Negative | 7 | 58 | 65 | Neg. Agree = 98.3% (90.9-100.0%) |
| | Total | 48 | 59 | 107 | Total Agree = 92.5% (85.8-96.7%) |
To increase the number of samples in the analytical measuring range of both
{9}
assays, an additional 27 contrived samples were tested. Contrived samples were made by diluting clinically diagnosed systemic sclerosis samples that are positive on the QUANTA Flash Centromere assay with normal serum to desired reactivity. Total number of samples for this second method comparison was 134 and results are summarized below:
| | Centromere ELISA | | | Percent Agreement (95% CI) | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | | Total |
| QUANTA Flash Centromere | Positive | 53 | 7 | 60 | Pos. Agree = 84.1% (72.7-92.1%) |
| | Negative | 10 | 64 | 74 | Neg. Agree = 90.1% (80.7-95.9%) |
| | Total | 63 | 71 | 134 | Total Agree = 87.3% (80.5-92.4%) |
b. Matrix comparison:
Not applicable since human serum is the only claimed specimen matrix.
3. Clinical studies:
a. Clinical sensitivity and specificity:
To determine the sensitivity and specificity of the QUANTA Flash Centromere assay, a total of 831 samples from patients diagnosed with systemic sclerosis and various disease controls were tested. No demographic information was available for the samples. See below for the number of subjects for each disease tested in this analysis. Samples from normal blood donors were excluded from the calculations. The sensitivity and specificity results are summarized below:
| All Patients (n=831) | Diagnosis – Systemic Sclerosis | | | Percent Agreement (95% confidence interval) | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | | Total |
| QUANTA Flash® Centromere CIA | Positive | 32 | 6 | 38 | Sensitivity = 23.4% (16.6-31.3%) |
| | Negative | 105 | 688 | 793 | Specificity = 99.1% (98.1-99.7%) |
| | Total | 137 | 694 | 831 | |
{10}
The table below shows the results for each clinical subgroup and each analyte.
| Cohort | n= | No (%) pos |
| --- | --- | --- |
| Systemic sclerosis (SSc) | 137 | 32 (23.4 %) |
| Other Inflammatory Diseases | 622 | 6 (1.0%) |
| Systemic lupus erythematosus (SLE) | 170 | 1 (0.6%) |
| Rheumatoid arthritis | 92 | 1 (1.1%) |
| Inflammatory bowel disease | 49 | 0 (0.0%) |
| Crohn’s disease | 27 | 0 (0.0%) |
| Ulcerative colitis | 30 | 0 (0.0%) |
| Multiple sclerosis | 19 | 0 (0.0%) |
| Hashimoto’s thyroiditis | 33 | 0 (0.0%) |
| Grave’s disease | 58 | 0 (0.0%) |
| Interstitial cystitis | 40 | 0 (0.0%) |
| Asthma | 27 | 1 (3.7%) |
| Other Disease Controls* | 77 | 3 (4.6%) |
| Infectious Diseases: | 72 | 0 (0.0%) |
| Hepatitis C virus infection | 24 | 0 (0.0%) |
| Hepatitis B virus infection | 31 | 0 (0.0%) |
| Human immunodeficiency virus infection | 8 | 0 (0.0%) |
| Syphilis | 9 | 0 (0.0%) |
non-systemic autoimmune disease, diagnosis pending (30), Osteoarthritis (14), Late Onset SLE (6), Polymyalgia Rheumatica (6), Connective Tissue Disease (5), Raynaud's Disease (3), SLE/Raynaud's Overlap (2), Gout (2), Psoriasis Arthritis (2), Primary Biliary Cirrhosis (1), Rosacea (1), Sakroiliitis (1), Fibromyalgia (1), Carcinoma of the Colon (1), QT syndrome (1), Incomplete CREST (1)
b. Other clinical supportive data (when a. is not applicable):
Not applicable.
4. Clinical cut-off:
There is no clinically accepted cutoff defined for this analyte.
5. Expected values/Reference range:
400 samples from random blood donors were tested of which four samples (1.0%) were positive. The sample average value was 4.0 CU; 95% of the samples had results between 3.5 – 4.5 CU.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
{11}
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.