VARELISA RECOMBI CTD SCREEN, MODEL 13096

K050967 · Sweden Diagnostics (Germany) GmbH · LJM · Jun 28, 2005 · Immunology

Device Facts

Record IDK050967
Device NameVARELISA RECOMBI CTD SCREEN, MODEL 13096
ApplicantSweden Diagnostics (Germany) GmbH
Product CodeLJM · Immunology
Decision DateJun 28, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2

Indications for Use

The Varelisa ReCombi CTD Screen EIA kit is for qualitative determination of 10 antinuclear antibodies in human serum or plasma to aid in the diagnosis of systemic rheumatic diseases such as SLE (systemic lupus erythematosus), drug induced lupus, scleroderma (progressive systemic sclerosis), MCTD (mixed connective tissue disease), SS (Sjogren’s syndrome) and Polymyositis/dermatomyositis. The Varelisa ReCombi CTD Screen detects antibodies against dsDNA, RNP, (RNP 70, A, C), Sm (B, B', D) SS-A/Ro (52 kDa, 60 kDa), SS-B/La, Scl-70, CENP-B, Histone, Ribosomal P protein and Jo-1 in a single microwell.

Device Story

ELISA-based diagnostic test; detects ten specific antinuclear antibodies in human serum or plasma. Microtiter plate wells coated with recombinant and native purified nuclear antigens and dsDNA. Patient sample added; specific antibodies bind to immobilized antigens. Rabbit anti-human IgG HRP conjugate added; forms enzyme-labeled conjugate-antibody-antigen complex. TMB substrate added; color formation proportional to antibody concentration. Performed in clinical laboratories by trained personnel. Requires microplate reader (450 nm/620 nm). Results aid clinicians in diagnosing systemic rheumatic diseases. Benefits include simultaneous screening for multiple autoantibodies in a single microwell.

Clinical Evidence

Clinical comparison study using 183 CTD samples and 100 disease controls. Positive agreement for CTD samples was 87.4% (95% CI: 81.7-91.9%) vs 85.2% (95% CI: 79.3-90.0%) for predicate. Negative agreement for control samples was 79.0% (95% CI: 69.7-86.5%) vs 84.0% (95% CI: 75.3-90.6%) for predicate. Data supports substantial equivalence.

Technological Characteristics

Indirect non-competitive EIA. Microtiter plate coated with recombinant/native nuclear antigens and dsDNA. Energy source: microtiter plate photometer (SLT Spectra II). Detection: HRP-conjugated anti-human IgG, TMB substrate. Qualitative output via OD ratio. Reagents: calibrator, positive/negative controls, wash buffer, sample diluent, stop solution.

Indications for Use

Indicated for qualitative determination of ten antinuclear antibodies in human serum or plasma to aid diagnosis of systemic rheumatic diseases including SLE, drug-induced lupus, scleroderma, MCTD, Sjögren's syndrome, and polymyositis/dermatomyositis. Prescription use only.

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k050967 B. Purpose for Submission: New Device C. Measurand: dsDNA, RNP, (RNP 70, A, C), Sm (B, B', D) SS-A/Ro (52 kDa, 60 kDa), SS-B/La, Scl-70, CENP-B, Histone, Ribosomal P protein and Jo-1 D. Type of Test: Qualitative, ELISA E. Applicant: Sweden Diagnostics(Germany) GmbH F. Proprietary and Established Names: Varelisa®ReCombi CTD Screen EIA kit G. Regulatory Information: 1. Regulation section: 21CFR 866.5100, antinuclear antibodies, immunological test system 2. Classification: Class II 3. Product code: LJM, Antinuclear antibody (enzyme-labeled), antigen, control 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): The Varelisa ReCombi CTD Screen EIA kit is for qualitative determination of 10 antinuclear antibodies in human serum or plasma to aid in the diagnosis of systemic rheumatic diseases such as SLE (systemic lupus erythematosus), drug induced lupus, scleroderma (progressive systemic sclerosis), MCTD (mixed connective tissue disease), SS (Sjogren’s syndrome) and Polymyositis/dermatomyositis. The Varelisa ReCombi CTD Screen detects antibodies against dsDNA, RNP, (RNP 70, A, C), Sm (B, B', D) SS-A/Ro (52 kDa, 60 kDa), SS-B/La, Scl-70, CENP-B, Histone, Ribosomal P protein and Jo-1 in a single microwell. 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): Prescription use only 4. Special instrument requirements: Microtiter plate photometer SLT Spectra II (type F039002). I. Device Description: The kit contains: (a) Microplate strips (12 strips x 8 wells) coated with nuclear antigens (dsDNA, RNP, Sm, SS-A/Ro, SS-B/La, Scl-70, CENP-B, Histone, PM-Scl, {1} Rib-P and Jo-1); (b) Calibrator-BSA buffer, sodium azide and human serum - ready to use; (c) Positive and negative control; (d) Wash buffer; (e) Sample diluent; (f) IgG horseradish peroxidase (HRP) conjugate - ready to use; (g) Substrate Tetramethylbenzidine and (h) stop solution. # J. Substantial Equivalence Information: 1. Predicate device name(s): Varelisa ReCombi ANA screen 2. Predicate 510(k) number(s): k993108 1. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | Qualitative test for determination of ANA in serum and plasma to aid in the diagnosis of systemic rheumatic diseases. | Same | | Antigens used | SS-A/Ro, SS-B, RNP, Scl-70, CENP-B, Jo-1, dsDNA | Same | | Matrix | Serum and plasma | Same | | Assay principle | Indirect non competitive Enzyme immunoassay | Same | | Type | Qualitative | Same | | Sample dilution | 1:101 | Same | | Conjugate | Anti-human IgG (rabbit) horseradish peroxidase | Same | | Reagents | Calibrator, control, sample diluent, stop solution | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Sm antigen | Synthetic human SmD peptide plus recombinant SmBB | Sm antigen from calf thymus | | Additional antigen | Histone and Rib-P | | | Calibrator, control | 1.0 ml | 1.5 ml | | Wash buffer-vol used | 75 ml | 50ml | | Sample diluent | 100ml ready to use | 20 ml 5x concentrate | | Conjugate | 20ml | 15 ml | | Substrate | 20 ml | 15 ml | | Stop solution | 20 ml | 10 ml | | Incubation time | 30 minutes | 10 min | {2} K. Standard/Guidance Document Referenced (if applicable): None provided. L. Test Principle: Varelisa CTD screen test is an indirect non-competitive EIA. The microtiter plate wells are coated with human recombinant and native purified nuclear antigens and dsDNA. Antibodies specific for the nuclear antigens present in the patient sample binds to these nuclear antigens. In the second step the conjugate binds to the complex and converts the added substrate to form the colored solution. The amount of color produced is proportional to the concentration of the respective antibodies in the patient sample. Test result is expressed as ratio (OD patient sample/OD of cut off). M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: To determine reproducibility, variation was estimated within-run and between-runs using 4 samples (negative, low positive, moderate and high positive) with 8 replicates per run for 6 runs. Within one day one operator carried out the analyses. The intra-assay variability and inter-assay variability target values for positive samples (ratio 1.0-4.0) were &lt;12% and &lt;9% respectively. No target value was specified for the negative samples. For all samples tested, the specifications were met. Results are summarized below. | Sample | | Run 1 | Run2 | Run3 | Run4 | Run5 | Run6 | Overall Mean Ratio | %CV | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | | | Intra | Inter | | Negative | Mean ratio | 0.9 | 0.9 | 0.9 | 0.8 | 0.8 | 0.9 | 0.9 | 8.1 | 0.5 | | | %CV | 14.3 | 4.4 | 7.5 | 4.5 | 6.5 | 6.1 | | | | | Low positive | Mean ratio | 1.2 | 1.2 | 1.1 | 1.1 | 1.1 | 1.2 | 1.2 | 3.4 | 2.4 | | | %CV | 4.1 | 2.2 | 3.3 | 4.3 | 2.9 | 3.4 | | | | | Moderate positive | Mean ratio | 1.9 | 2.0 | 1.9 | 1.9 | 1.9 | 2.1 | 1.9 | 4.7 | 3.1 | | | %CV | 2.9 | 1.7 | 2.0 | 2.2 | 3.3 | 9.5 | | | | | High positive | Mean ratio | 4.3 | 4.3 | 4.2 | 4.2 | 4.0 | 4.1 | 4.2 | 2.2 | 2.4 | | | %CV | 2.8 | 2.5 | 2.4 | 1.8 | 1.5 | 2.0 | | | | b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): No information provided. d. Detection limit: The purpose of this study was to determine the detection limit. The ready-to-use sample diluent was measured 80 times per run (2 runs per lot). Analytical sensitivity was calculated as the mean of the optical density (OD) of the sample diluent plus 3SD (expressed as ratios). The specification for detection limit is &lt;0.3. For the first run, the (mean OD + 3SD) = [0.02 + 3(0.009)] = 0.047 and ratio 0.1 and for the second run, the (mean OD + 3SD) = [0.017 + 3(0.010)] = 0.046 and ratio 0.1. The results met the specifications. e. Analytical specificity: Diluted samples were spiked and tested to determine interference from 4 different concentrations each of Bilirubin F (4.6-18.5mg/dL), Bilirubin C (5.4-21.6 mg/dL), Chyle (530-2120 units), Hemoglobin (116.8-467 mg/dL) and {3} rheumatoid factor (100-500 IU/mL). The target value for the deviation of the value of the spiked sample with the interfering substance should be ± 20% of the value of the sample spiked with a buffer blank. The results obtained were within the specification. The above noted substances did not adversely affect the results. f. Assay cut-off: The assay cut-off is set to 1.0 based on the results of normal population study (400 samples from Caucasian subjects), the reference sera and correlation study. The specification was that the 95th percentile should lie below the ratio 1.0. Results are considered negative if the ratio &lt; 1.0, low positive if the ratio is between 1.0 and 1.4, moderate to high positive if the ratio &gt; 1.4. 2. Comparison studies: a. Method comparison with predicate device: Substantial equivalence to the predicate was determined by testing 183 CTD samples and 100 disease control samples. Equivocal results of predicate ANA assay is considered as negative and CTD low and moderate positive as positive. Analysis of agreement between the current and predicate assay is shown below | | Varelisa ANA screen | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | | Positive | Equivocal | Negative | Total | | Varelisa CTD | CTD Samples | >1.4 | 144 | 1 | 4 | 149 | | | | 1.0-1.4 | 9 | 2 | 0 | 11 | | | | <1.0 | 3 | 5 | 15 | 23 | | | | Total | 156 | 8 | 19 | 183 | | | Control Samples | >1.4 | 11 | 2 | 0 | 13 | | | | 1.0-1.4 | 3 | 1 | 4 | 8 | | | | <1.0 | 2 | 5 | 72 | 79 | | | | Total | 16 | 8 | 76 | 100 | | | Varelisa CTD | | Varelisa ANA screen | | | | --- | --- | --- | --- | --- | --- | | | % agreement | 95% CI | % agreement | 95% CI | Difference | | Positive (CTD) | 87.4 | 81.7-91.9% | 85.2 | 79.3-90.0% | 2.2 % | | Negative (Control) | 79.0 | 69.7-86. 5% | 84.0 | 75.3-90.6% | -5.0 % | The new device was also compared to the predicate device using the ANA reference panel from CDC, reference panel 2002 from AMLI and WHO International Standard. Samples were analyzed in duplicate. The specificities of the CDC, AMLI and WHO reference samples and the predicate/dew device comparative results are listed below. | CDC Reference Panel | | | AMLI Reference Panel | | | | --- | --- | --- | --- | --- | --- | | Target | New Device (Ratio) | Predicate (Ratio) | Target | New Device (Ratio) | Predicate (Ratio) | | dsDNA & weak Sm | 2.7 | 4.7 | Negative | 0.8 | 1.1 | | SSB & week SSA | 5.5 | 8.0 | Sm | 4.4 | 5.3 | {4} | CDC Reference Panel | | | AMLI Reference Panel | | | | --- | --- | --- | --- | --- | --- | | Target | New Device (Ratio) | Predicate (Ratio) | Target | New Device (Ratio) | Predicate (Ratio) | | Weak Sm, U1RNP, SSA, SSB | 6.1 | 7.7 | RNP | 5.7 | 6.9 | | U1-RNP | 5.3 | 5.3 | SSA | 3.3 | 5.1 | | Sm, histone | 5.2 | 5.0 | SSB | 5.6 | 7.9 | | Nucleolar pattern | 0.3 | 0.3 | Scl-70 | 3.0 | 2.8 | | SSA | 3.3 | 5.3 | Jo-1 | 3.3 | 4.6 | | Cenp | 4.6 | 7.1 | Centromere | 2.4 | 3.6 | | Scl-70 | 4.5 | 4.9 | dsDNA | 4.8 | 6.9 | | Jo-1 | 3.7 | 5.4 | Negative | 0.2 | 0.2 | | PM-Scl | 1.5 | 0.9 | WHO | | | | Ribosomal P | 1.1 | 0.5 | dsDNA | 2.8 | 3.7 | b. Matrix comparison: To demonstrate that serum and plasma (EDTA, heparin and citrate) samples gave same results, four samples for the different anticoagulants were collected from each of 10 blood donors and assayed in duplicates. The samples were then spiked with 10 different ANA antibody positive samples. All spiked samples were run in four replicates. The specification was set that the difference between the serum and plasma results should be lower than $\pm 20\%$ for positive samples. Negative samples should be negative. For EDTA and citrate plasmas, the specifications were met. For heparin plasma, a significant difference between serum and plasma was found. Heparin plasma samples are not recommended for this assay and the limitation is stated in the package insert. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Same as assay cut-off. 5. Expected values/Reference range: Expected value in the normal population was determined by measuring samples from 400 healthy blood donors with equal numbers of both genders and equal distribution in the defined age categories (&lt;30 years to &gt;60 years). The mean ratio for the female cohort is higher than that of the male. The overall mean and median ratios for both cohorts were 0.44 and 0.37 respectively. The mean ratio and 2 SD was 1.08 and the $95^{\text{th}}$ Percentile was 0.83. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: 1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...