← Product Code [LJM](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LJM) · K040200

# MESACUP-2 TEST CENP-B (K040200)

_Rhigene, Inc. · LJM · Mar 4, 2004 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LJM/K040200

## Device Facts

- **Applicant:** Rhigene, Inc.
- **Product Code:** [LJM](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LJM.md)
- **Decision Date:** Mar 4, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The MESACUP-2 Test CENP-B is a semi-quantitative enzyme-linked immunosorbent assay (ELISA) for the detection of anti-centromere (CENP-B) antibodies in human serum. The MESACUP-2 Test CENP-B is intended to be used by clinical (hospital and reference) laboratory personnel.

## Device Story

The device is an ELISA kit for detecting anti-CENP-B autoantibodies in human serum. It uses microtiter plates coated with recombinant CENP-B antigen to capture antibodies from patient samples. A polyclonal goat anti-human IgG, IgM, and IgA conjugate labeled with horseradish peroxidase (HRP) is added, followed by a TMB/H2O2 substrate. The resulting color change is measured photometrically. The kit includes calibrators and controls for semi-quantitative interpretation. It is intended for use by clinical laboratory personnel in hospital or reference settings. Results aid clinicians in diagnosing CREST syndrome. The device provides a standardized laboratory measurement to support clinical assessment of autoimmune connective tissue disease.

## Clinical Evidence

In-house studies compared MESACUP-2 Test CENP-B to predicate. Healthy donor population: 95% and 100% specificity. CREST syndrome population (previously positive by double immunodiffusion): 90% and 95% sensitivity. 95% relative agreement between methods.

## Technological Characteristics

Semi-quantitative indirect ELISA. Solid phase: microtiter plates coated with recombinant CENP-B antigen. Detection: HRP-conjugated polyclonal goat anti-human IgG/IgM/IgA. Substrate: TMB/H2O2. Photometric measurement. Reportable range: 5-300 U/mL. Cut-off: 16 U/mL. Manual or automated plate processing.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- Quanta Lite Centromere ELISA (k003959)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k040200

B. Analyte:
Anti-centromere antibody

C. Type of Test:
Semi-quantitative, ELISA

D. Applicant:
RhiGene, Inc

E. Proprietary and Established Names:
MESACUP-2 Test CENP-B

F. Regulatory Information:

1. Regulation section:
21 CFR §866.5100 Anti-nuclear Antibody Immunological Test System

2. Classification:
Class II

3. Product Code:
LJM (Anti-nuclear antibody, (enzyme-labeled), antigen, control)

4. Panel:
IM 82

G. Intended Use:

1. Intended use(s):
The MESACUP-2 Test CENP-B is a semi-quantitative enzyme-linked immunosorbent assay (ELISA) for the detection of anti-centromere (CENP-B) antibodies in human serum.

The MESACUP-2 Test CENP-B is intended to be used by clinical (hospital and reference) laboratory personnel.

2. Indication(s) for use:
The MESACUP-2 Test CENP-B is indicated as an aid in the diagnosis of CREST Syndrome (i.e., calcinosis, Raynaud's phenomenon, esophageal immotility, sclerodactyly, and talangiectasia).

{1}

Confidential

10/19/2011

3. Special condition for use statement(s): The device is for prescription use only.
4. Special instrument Requirements: None

# H. Device Description:

The device is an enzyme-linked immunosorbent assay (ELISA) using microtiter plates as the solid phase. The plate wells are coated with recombinant CENP-B antigen, which captures CENP-B autoantibodies present in the patient serum sample. The conjugate is polyclonal goat anti-human IgG, IgM and IgA (heavy chain specific) horseradish peroxidase (HRP) which uses 3,3'5,5' tetramethylbenzidine dihydrochloride/hydrogen peroxide  $(\mathrm{TMB} / \mathrm{H}_2\mathrm{O}_2)$  as substrate. The kit contains 2 levels of calibrators (0 units/mL and  $100~\mu \mathrm{mL}$ ) for interpretation of results. A positive and a negative control are included with the kit. The kit also contains sample diluent, wash buffer concentrate and stop solution.

# I. Substantial Equivalence Information:

1. Predicate device name(s): Quanta Lite Centromere ELISA from INOVA Diagnostics
2. Predicate K number(s): k003959
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | MESACUP-2 Test CENP-B | Predicate  |
|  Indications for Use | For detection of anti-centromere antibodies as an aid in the diagnosis of CREST and related connective tissue diseases. | Same  |
|  Assay principle | Indirect ELISA | Same  |
|  Sample matrix | Serum | Same  |
|  Substrate | TMB | Same  |
|  Differences  |   |   |
|  Item | MESACUP-2 Test CENP-B | Predicate  |
|  Analyte | anti-CENP-B autoantibodies | Anti-CENP-A and anti-CENP-B autoantibodies  |
|  Cut-off | 16 U/mL | 20 Units  |
|  Detection range | 5-300 U/mL | 0-6 Units  |
|  Assay time | 150 minutes at Room Temp | 90 minutes at Room Temp  |
|  Conjugate | HRP-goat anti-human IgG/IgM/IgA | HRP-goat anti-human IgG  |

{2}

Confidential

10/19/2011

## J. Standard/Guidance Document Referenced (if applicable):

Not applicable

## K. Test Principle:

The assay involves enzyme-linked immunosorbent assay (ELISA) technology. Calibrators and patient sera are incubated with CENP-B antigens for a specified time and then washed. This is followed by incubation with horseradish peroxidase conjugated anti-human IgG, IgA, and IgM. The reaction is then stopped and the color is allowed to develop and measured photometrically.

## L. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

**a. Precision/Reproducibility:**

Three lots of the MESACUP-2 Test CENP-B were used to determine the intra-assay, inter-assay and inter-lot value precision for the test.

**Intra-assay**

Intra-assay precision (%CV) was determined by running 3 serum samples (i.e., negative, moderate and high positive) 8 times (i.e., 8 replicates per plate) on 3 separate assays. Three separate plates were randomly selected from each plate-coating run (kit-lot). The mean intra-assay precision for the 3 samples tested on 3 plates from each lot was 3.8% (Range: 2.4-5.4%).

**Inter-assay, intra-lot**

To determine the amount of variability between plates of the same lot, 3 samples in duplicate were tested on 6 separate assays. Six randomly selected plates randomly selected from the same plate lot were used for each one of 3 separate plate lots. The mean %CV for inter-assay, intra-lot precision was 4.5% with a range of 2.3-8.6%.

**Inter-assay, inter-lot**

The precision between lots was determined by comparing the values recovered for 3 different samples on 3 different pilot lots. Each of the 3 samples was tested in duplicate and by 2 operators in each assay. The mean inter-assay, inter-lot %CV was 2.3%.

**b. Linearity/assay reportable range:**

The reportable range of 5-300 U/mL was demonstrated by recovery studies.

**c. Traceability (controls, calibrators, or method):**

An international reference material for anti-CENP-B antibodies is not available. The assay is calibrated in relative arbitrary units based

{3}

Confidential

10/19/2011

on the upper-optical density detection limit using samples that were positive by both ELISA and Double Immunodiffusion (DID).

# d. Detection limit:

Not applicable.

# e. Analytical specificity:

Several substances were added to three patient specimens (i.e., negative, moderate, and high) each to test for interference. Based on the results summarized below, the addition of these substances at the levels tested did not affect the assay results.

|  Substance | Level Range (U/mL) | Low |   |   | Moderate |   |   | High  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Mean (U/mL) | SD | %CV | Mean (U/mL) | SD | %CV | Mean (U/mL) | SD | %CV  |
|  Hemoglobin | 0 – 480 | 53.9 | 2.06 | 3.8 | 84.8 | 4.17 | 4.9 | 191.8 | 4.79 | 2.5  |
|  Bilirubin C | 0 – 20.4 | 55.1 | 4.3 | 7.8 | 82.3 | 5.68 | 6.9 | 192.2 | 7.89 | 4.1  |
|  Bilirubin F | 0 – 18.7 | 48.9 | 3.54 | 7.2 | 74.7 | 1.97 | 2.6 | 191.8 | 2.99 | 1.6  |
|  Chyle | 0 – 2780 | 53.8 | 3.52 | 6.5 | 83.1 | 4.26 | 5.1 | 191.7 | 2.72 | 1.4  |
|  Rheumatoid Factor | 0 - 520 | 64.7 | 1.95 | 3.0 | 100.0 | 2.9 | 2.9 | 205.6 | 3.01 | 1.5  |

# f. Assay cut-off:

A healthy sample population consisting of 266 unselected serum samples (in duplicate) was tested for anti-CENP-B antibodies with both the MESACUP-2 Test CNEP-B and the DID method. In addition, a population of 691 collagen disease specimens were tested,  $5.1\%$  (35/691) of which were positive for ACA antibodies by DID. The cut-off was established by comparing the frequency distribution of values obtained for both populations with the MESACUP-2 Test CENP-B to the DID results for positive and negative agreement. Based on this comparison, the best overall accuracy compared to DID was established at  $16~\mathrm{U / mL}$ . Therefore, this was selected as the cut-off. Note: All 35 DID positive collagen specimens were also positive according to the MESACUP-2 Test CENP-2 results with this cut-off. There is no equivocal (gray) zone for this assay.

# 2. Comparison studies:

# a. Method comparison with predicate device:

The tables below show the results of comparison of serum samples  $(N = 80)$  that were tested with the MESACUP-2 Test CENP-B and the predicate device.

|   | Quanta Lite + | Quanta-Lite --  |
| --- | --- | --- |
|  MESACUP-2 + | 18 | 1  |
|  MESACUP-2 - | 3 | 58  |

{4}

Confidential

10/19/2011

|  STATISTIC | Value | 95% CI  |
| --- | --- | --- |
|  Prevalence | 0.2625 | 0.1733 - 0.3648  |
|  Positive agreement | 0.9474 | 0.7189 - 0.9972  |
|  Negative agreement | 0.9508 | 0.8540 - 09872  |
|  Total agreement | 0.9490 | 0.7864 - 0.9922  |

# b. Matrix comparison:

Serum is the only recommended matrix.

# 3. Clinical studies:

# a. Clinical sensitivity:

Clinical sensitivity for the MESACUP-2 Test CENP- was determined by testing a population of CREST Syndrome patient serum specimens  $(n = 20)$ . Using the cut-off of  $16\mathrm{U / mL}$ ,  $90\%$  (18/20) of the samples were positive for anti-CENP-B antibodies. The mean value for the CREST Syndrome samples was  $79.1\mathrm{U / mL}$ . Single Factor ANOVA analysis that compared this value to the mean for the healthy controls gives a p-value of  $2.95\times 10^{-74}$ . Therefore, at a level of  $p &lt; 0.05$  for statistical significance, the results of this population were determined to be statistically different compared to the healthy controls.

# b. Clinical specificity:

The applicant evaluated 266 samples, in duplicate, from 168 consecutive healthy blood donors and used these samples as the normal population. The mean value was  $3.0\mathrm{U / mL}$ $(\mathrm{SD} = \pm 1.9)$ . Only 2 samples tested positive  $(0.75\%)$  in this sample population. Therefore, the specificity was  $99\%$ .

Similar studies were performed with samples from patients from various autoimmune diseases. The table below provides a summary of the results obtained in these subgroups and in samples from patients with CREST Syndrome.

|  Disease or Disease Status | N | Mean (U/mL) | SD | Specificity (%)  |
| --- | --- | --- | --- | --- |
|  CREST Syndrome | 31 | 104.8 | 35.3 | 97  |
|  Rheumatoid Arthritis | 47 | 4.0 | 8.8 | 4  |
|  Sjögren's Syndrome | 47 | 12.8 | 36.3 | 9  |
|  Systemic Lupus Erythematosus | 103 | 7.3 | 18.3 | 7  |
|  Systemic Sclerosis | 23 | 3.0 | 1.6 | 0  |
|  Mixed Connective Tissue Disease | 59 | 7.6 | 20.2 | 7  |
|  Polymyositis | 9 | 5.3 | 3.9 | 10.3  |
|  Dermatomyositis | 17 | 25.3 | 51.8 | 24  |

The minimal cross-reactivity does not seem to be clinically significant, and it is expected to be due to "cross-over of the various

{5}

Confidential

10/19/2011

autoimmune diseases" (Applicant's conclusion, p. 51). Results similar to those seen here for Dermatomyositis have been reported in the literature where a positive test result for ACA was seen in patients with Raynaud's disease. The data support the specificity of the test for the intended indication

c. Other clinical supportive data (when  $a$  and  $b$  are not applicable): Not applicable.

4. Clinical cut-off: See assay cut-off.

5. Expected values/Reference range: The expected value in the normal population is negative.

# M. Conclusion:

The purpose of this  $510(\mathrm{k})$  was to seek clearance to market the MESACUP-2 Test CENP-B ELISA for the detection of anti-CENP-B antibodies in human serum. The data submitted in this  $510(\mathrm{k})$  substantiate a good level of agreement between the results obtained with the MESACUP-2 Test CENP-B and the predicate device, for those conditions specified in the Indications for Use statement. The applicant provided evidence that the devices are substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LJM/K040200](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/LJM/K040200)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
