BIORAPID MONONUCLEOSIS

K030215 · Instrumentation Laboratory CO · KTN · Feb 24, 2003 · Immunology

Device Facts

Record IDK030215
Device NameBIORAPID MONONUCLEOSIS
ApplicantInstrumentation Laboratory CO
Product CodeKTN · Immunology
Decision DateFeb 24, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5640
Device ClassClass 2

Indications for Use

biorapid Mononucleosis is a one-step immunoassay for the qualitative detection of infectious mononucleosis heterophile antibodies in whole blood, serum, and plasma samples. The test aids in the diagnosis of infectious mononucleosis.

Device Story

biorapid Mononucleosis is a one-step immunochromatographic assay; detects heterophile antibodies associated with infectious mononucleosis in whole blood, serum, or plasma. Used in clinical settings, including physician office laboratories (POLs), by personnel with diverse educational backgrounds. Provides qualitative results to aid clinicians in diagnosing infectious mononucleosis. Benefits patients through rapid, point-of-care diagnostic confirmation.

Clinical Evidence

Bench testing and clinical evaluation of 622 specimens (296 serum, 261 plasma, 65 whole blood). Compared to predicate, showed 100% positive agreement and 99.6% negative agreement (overall 99.7%). Reproducibility study at three physician office laboratories using coded samples tested in triplicate over three days showed 100% agreement with expected results.

Technological Characteristics

Immunochromatography test; qualitative immunoassay; lateral flow format. No specific materials, energy sources, or software algorithms described.

Indications for Use

Indicated for the qualitative detection of infectious mononucleosis heterophile antibodies in whole blood, serum, and plasma samples to aid in the diagnosis of infectious mononucleosis.

Regulatory Classification

Identification

An infectious mononucleosis immunological test system is a device that consists of the reagents used to measure by immunochemical techniques heterophile antibodies frequently associated with infectious mononucleosis in serum, plasma, and other body fluids. Measurements of these antibodies aid in the diagnosis of infectious mononucleosis.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 4 2003 330215 # biorapid Mononucleosis - 510(k) Summary (Summary of Safety and Effectiveness) #### Submitted by: Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421 Phone: 781-861-4467 Fax: 781-861-4207 ## Contact Person: Carol Marble, Regulatory Affairs Director Phone: 781-861-4467 / Fax: 781-861-4207 ## Summary Prepared: January 17, 2003 ### Name of the device: biorapid Mononucleosis #### Classification name(s): | 866.5640 | Infectious Mononucleosis Immunological Test System | Class | |----------|----------------------------------------------------|-------| | 82KTN | System, Test, Infectious Mononucleosis | | #### Identification of predicate device(s): K972231 OSOM Mono Test #### Description of the Device/Intended Use(s): biorapid Mononucleosis is a one-step immunoassay for the qualitative detection of infectious mononucleosis heterophile antibodies in whole blood, serum, and plasma samples. The test aids in the diagnosis of infectious mononucleosis. #### Statement of Technological Characteristics of the Device Compared to Predicate Device: biorapid Mononucleosis uses the same methodology (immunochromatography test) as the predicate OSOM Mono Test and is substantially equivalent in performance, intended use and safety and effectiveness. #### Summary of Performance Data: A total of 622 specimens (296 serum, 261 plasma, and 65 whole blood) were evaluated internally and in an external study with biorapid Mononucleosis and the predicate device, OSOM Mono Test. When compared to the predicate device, biorapid Mononucleosis showed a positive agreement of 100% and a negative agreement of 99.6% (478/480). The overall agreement between the two tests was 99.7%. A reproducibility evaluation of the biorapid Mononucleosis test was conducted at three physician's offices laboratories (POL's) where testing was performed by personnel with diverse educational backgrounds. At each site, randomly coded samples (three positive and three negative) were tested in triplicate for three days. Results obtained showed a 100% agreement with the expected results. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads in profile, facing to the left. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. FEB 2 4 2003 Food and Drug Administration i 2098 Gaither Road Rockville MD 20850 Ms. Carol Marble Regulatory Affairs Director Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, Massachusetts 02421 Re: k030215 > Trade/Device Name: biorapid Mononucleosis Regulation Number: 21 CFR § 866.5640 Regulation Name: Infectious Mononucleosis Immunological Test System Regulatory Class: II Product Code: KTN Dated: January 17, 2003 Received: January 21, 2003 Dear Ms. Marble: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {2}------------------------------------------------ Page 2 - If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K030215 Device Name: biorapid Mononucleosis Indications for Use: biorapid Mononucleosis is a one-step immunoassay for the qualitative detection of infectious mononucleosis heterophile antibodies in whole blood, serum, and plasma samples. The test aids in the diagnosis of infectious mononucleosis. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) J. Plewes for J. Bautista Clinical Laboratory D Diveron J. 510(K) Number ________________________________________________________________________________________________________________________________________________________________ Prescription Use (Per 21 CFR 801.019) OR Over-The-Counter Use _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...