ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO

K051890 · Roche Diagnostics Corp. · JZO · Aug 4, 2005 · Immunology

Device Facts

Record IDK051890
Device NameROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO
ApplicantRoche Diagnostics Corp.
Product CodeJZO · Immunology
Decision DateAug 4, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.5870
Device ClassClass 2

Indications for Use

The Elecsys Anti-TPO immunoassay is for the in vitro quantitative determination of antibodies to thyroid peroxidase in human serum and plasma. The anti-TPO determination is used as an aid in the diagnosis of autoimmune thyroid diseases. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys 1010/2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers.

Device Story

In vitro diagnostic immunoassay; measures thyroid peroxidase (anti-TPO) antibodies in human serum/plasma. Uses electrochemiluminescence (ECLIA) on Roche Elecsys 1010/2010 or MODULAR ANALYTICS E170 analyzers. Competitive assay: patient sample competes with ruthenium-labeled anti-TPO antibodies for biotinylated TPO antigen; immunocomplexes bind to streptavidin-coated microparticles. Microparticles captured magnetically on electrode; voltage application induces chemiluminescent emission measured by photomultiplier. Results calculated via instrument-specific calibration curve and master curve from reagent barcode. Used in clinical laboratories by trained personnel to aid diagnosis of autoimmune thyroid diseases.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, including analytical performance characteristics, reagent stability, and interference studies (bilirubin, hemolysis, lipemia, rheumatoid factors, and biotin).

Technological Characteristics

Quantitative electrochemiluminescence immunoassay (ECLIA). Reagents: streptavidin-coated microparticles, ruthenium-labeled polyclonal anti-TPO antibody, biotinylated TPO. Measuring range: 5-600 IU/ml. Analyzers: Roche Elecsys 1010/2010 and MODULAR ANALYTICS E170. Calibration: 2-point calibration with master curve via reagent barcode.

Indications for Use

Indicated for in vitro quantitative determination of thyroid peroxidase antibodies in human serum and plasma to aid in the diagnosis of autoimmune thyroid diseases.

Regulatory Classification

Identification

A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k051890 B. Purpose for Submission: Modifications to the manufacturer’s existing device. C. Measurand: Thyroid peroxidase anti-body (anti-TPO) D. Type of Test: Quantitative Electro-chemiluminescence immunoassay. E. Applicant: Roche Diagnostics Corp. F. Proprietary and Established Names: Roche Diagnostics COBAS Elecsys® Anti-TPO G. Regulatory Information: 1. Regulation section: 21 CFR 866.5870 Thyroid autoantibody immunological test system 2. Classification: Class II 3. Product code: JZO System, Test, Thyroid autoantibody 4. Panel: Immunology 82 H. Intended Use: 1. Intended use(s): The Elecsys Anti-TPO immunoassay is for the in vitro quantitative determination of antibodies to thyroid peroxidase in human serum and plasma. The anti-TPO determination is used as an aid in the diagnosis of autoimmune thyroid diseases. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the Roche Elecsys 1010/2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers. 2. Indication(s) for use: For in vitro diagnostic use. The immunoassay is for the in vitro quantitative determination of antibodies to thyroid peroxidase in human serum and plasma. The anti-TPO determination is used as an aid in the diagnosis of autoimmune thyroid diseases. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: For use on the Roche Elecsys® 1010/2010 and MODULAR ANALYTICS E170 (Elecsys® module) immunoassay analyzers previously cleared under k961481 and k961481/A003. I. Device Description: The COBAS Elecsys® Anti-Thyroid Peroxidase Antibody test system includes: {1} streptavidin-coated microparticles, polyclonal anti-TPO antibody coated with ruthenium complex (Anti-TPO-Ab-Ru(bpy)2+/3), biotinylated TPO, two polyclonal anti-TPO (A-TPO) antibody (sheep) calibrators (1 & 2) and two anti-TPO (PreciControl A-TPO 1 & 2) (human) antibody controls. # J. Substantial Equivalence Information: 1. Predicate device name(s): ELECSYS® Anti-Thyroid Peroxidase Antibody Test System 2. Predicate 510(k) number(s): k000155 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use/Indications for use | Immunoassay for the in vitro quantitative determination of antibodies to thyroid peroxidase in human serum and plasma. The anti-TPO determination is used as an aid in the diagnosis of autoimmune thyroid diseases. The electro-chemiluminescence immunoassay “ECLIA” is intended for use on the Roche Elecsys 1010/2010 and MODULAR ANALYTICS E170 (Elecsys® module) immunoassay analyzers. | Same | | Capture and detection reagents | Streptavidin-coated microparticles, polyclonal anti-TPO-Ab-Ru(bpy)2+/3 and biotinylated TPO | Same | | Sample Type | Human serum and plasma | Same | | Traceability/Standardization | WHO 66/387 | Same | | Test Principle | Competitive chemiluminescence | Same | | Measuring range | 5-600 IU/ml | Same | | Reagent Stability | Unopened: - at 2-8 °C up to the expiration date. Opened: - at 2-8 °C up to six weeks. - on E170/2010: two weeks. - on 1010: one weeks storage, alternately in refrigerator and on analyzer - at ambient temperature 20-25 °C; up to 20 hours total | Same | | Limitations | No interference from: Bilirubin up to 66 mg/dl Hemolysis up to 1.5 g/dl Lipemia up to 2100 mg/dl Rheumatoid factors up to 1500 U/ml | Same | {2} 3 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | R2 Composition | 20 ng/mL Biotin | 0 ng/mL Biotin | | PreciControl Anti-TPO | anti-TPO antibodies at approx 35 IU/mL | Level 1 = approx 20 IU/mL anti-TPO antibodies | K. Standard/Guidance Document Referenced (if applicable): Not referenced. L. Test Principle: This is an in vitro, prescription device to measure the amount of thyroid peroxidase antibodies in human serum. It is an electrochemiluminescence competitive assay. Patient samples and controls are incubated with anti-TPO antibodies labeled with a ruthenium complex. After addition of biotinylated TPO and streptavidin-coated microparticles, the anti-TPO antibodies in the sample compete with the ruthenium conjugated anti-TPO-Ab for the biotinylated TPO antigen. The immunocomplexes produced become bound to the solid phase via biotin-streptavidin interaction. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell buffer. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. The results are determined via a calibration curve which is instrument specifically generated by 2-point calibration and a master curve provided via the reagent bar code. M. Performance Characteristics (if/when applicable): 1. Analytical performance: Analytical performance of the modified device was supported by the following studies: | Performance characteristic | Acceptance Criteria | | --- | --- | | Precision | Intra-assay precision: ≤ 10% at 15-40 IU/ml ≤ 7% at > 40 IU/ml; Inter-assay precision: ≤ 18% at 15-40 IU/ml, ≤ 12% at > 40 IU/ml | | Analytical Sensitivity | 5.0 IU/ml | | Reportable Range | 5-600 IU/ml | | Linearity | Within 15% of expected value | | Specificity | Within 10% of control value | | Calibration Curve Stability | Within 15% of control values | a. Precision/Reproducibility: No change. b. Linearity/assay reportable range: No change. c. Traceability, Stability, Expected values (controls, calibrators, or methods): No change. d. Detection limit: No change. {3} e. Analytical specificity: No change. f. Assay cut-off: No change. 2. Comparison studies: a. Method comparison with predicate device: The modified device and the original device were compared by testing 50 clinical samples (naturally occurring and spiked samples) covering the reportable range. Results were within specified acceptance criteria for a slope of 0.97 to 1.03. Risk analysis studies to determine the impact for the changes to the original device and steps, if any, to minimize any potential hazard were provided. b. Matrix comparison: No change 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: No change. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...