TFR

K981208 · Ramco Laboratories, Inc. · JNM · Oct 21, 1998 · Immunology

Device Facts

Record IDK981208
Device NameTFR
ApplicantRamco Laboratories, Inc.
Product CodeJNM · Immunology
Decision DateOct 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5880
Device ClassClass 2

Indications for Use

The TfR assay is an in vitro enzyme immunoassay for quantifying the concentration of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states.

Device Story

TfR assay is an in vitro enzyme immunoassay (double antibody sandwich method) for quantifying serum transferrin receptor (STR) in human serum or plasma. Process: patient samples diluted in buffer; pipetted into microwells pre-coated with polyclonal antibody to transferrin receptor; horseradish peroxidase (HRP) conjugated antibody added; incubation allows STR binding to capture and detection antibodies. Unbound material washed away; enzyme substrate added; stop solution added; intensity of yellow product measured via microplate reader at 450nm. Optical density is directly proportional to STR concentration; concentration determined by comparing sample optical density against a standard curve generated from provided STR standards. Used in clinical laboratory settings by trained personnel to aid in diagnosing iron deficiency anemia, especially when other disease states complicate diagnosis. Results assist clinicians in differentiating iron deficiency anemia from anemia of chronic disease, potentially guiding therapeutic decisions.

Clinical Evidence

Clinical evaluation compared TfR assay to predicate (QSI) using 283 samples (38 IDA, 175 ACD, 70 other). Linear regression of 155 paired samples yielded R2 of 88.2%. In IDA samples, TfR showed 78.9% agreement with clinical definition vs 78.3% for QSI. In ACD samples, TfR showed 82.9% agreement vs 68.1% for QSI. When results were refined by ferritin concurrence, TfR and QSI showed 100% agreement in IDA (n=17) and 98.8% in ACD (n=80). Overall, TfR and QSI results matched in 90.3% of all samples.

Technological Characteristics

In vitro enzyme immunoassay; double antibody sandwich method. Uses polyclonal capture antibody and monoclonal HRP-conjugated detection antibody. Calibrated using human placenta-derived transferrin receptor standards. Requires microplate reader with 450nm filter. Sample dilution 100:1. Single-stage assay format.

Indications for Use

Indicated for quantifying transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia, including in patients with other concurrent disease states.

Regulatory Classification

Identification

A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 10/21/98 Image /page/0/Picture/1 description: The image shows the logo and contact information for Ramco Laboratories Inc. The address is 4507 Mt. Vernon, Houston, Texas 77006. The phone number is (713) 526-9677, and the toll-free number is 1-800-231-6238. The fax number is 713-526-1528, and they are available 24 hours. ## 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 The assigned 510(k) number is: __ 981208 | Submitter: | Ramco Laboratories, Inc.<br>4507 Mt. Vernon<br>Houston, TX 77006 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------| | Phone #:<br>Fax #: | (713) 526-9677<br>(713) 526-1528 | | Contact Person: | Jeffrey B. Grubb, President | | Date Prepared: | March 27, 1998 | | Device Name: | Transferrin Receptor Assay | | Trade Name: | <i>TfR</i> | | Common Name: | TfR Assay | | Classification Name: | sTfR Immunological Test System | | Predicate Device Name: | Quantikine IVD sTfR Immunoassay<br>R&D Systems, Inc.<br>614 McKinley Place N.E.<br>Minneapolis, MN 55413<br>510(k) #K970718 | The TFR assay is an in vitro enzyme immunoassay based upon the double antibody sandwich method. Plasma or serum samples are diluted in buffer and pipetted into microwells pre-coated with polyclonal antibody to transferrin receptor. Horseradish peroxidase conjugated antibody specific for serum transferrin receptor (STR) is added to the wells and incubated. During this incubation, the STR binds to the polyclonal antibodies adsorbed to the wells and the HRP-conjugated second antibodies bind to the captured STR. {1}------------------------------------------------ Any unbound STR and excess HRP-conjugate are washed from the wells. Enzyme substrate is added to the wells and allowed to incubate, a stop solution is then added to stop the reaction and the intensity of the yellow product is measured in a microplate reader. The optical density of the resulting solution is directly proportional to the concentration of the STR in the standard samples. A standard curve is generated from the STR standards provided in the assay and the concentration of STR in the unknown sample is determined by comparing the unknown's optical density reading with the standard curve graph. The TFR assay is an in vitro enzyme immunoassay for quantifying the concentration of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states. Ramco Laboratories, Inc. is claiming substantial equivalence of its TFR assay the R&D Systems (Minneapolis, Minnesota) Quantikine IVD sTfR Immunoassay ("QSI") which obtained 510(k) clearance on May 27, 1997 (K970718, Class II). Both TFR and OSI's intended use is to quantitatively determine the level of transferrin receptor in human serum or plasma to aid in the diagnosis of iron deficiency anemia. Both are based on the microplate sandwich enzyme immunoassay technique, use calibrators to generate a log-log curve from which the quantity of an unknown sample is read, use a monoclonal transferrin antibody conjugated to horseradish peroxidase, use a microplate reader with a 450nm filter to read the results, and supply controls which, like the unknown samples, must be diluted prior to use. Finally, both assays offer a prescribed normal range, the upper limit of which is the indicator for the presence of iron deficiency anemia. Both TFR and OSI use a monoclonal antibody in its conjugate. However, TFR uses a polyclonal transferrin receptor antibody to coat the microplate, making it a polyclonalmonoclonal assay, while QSI uses a monoclonal antibody to coat the plate making it a double monoclonal. TFR provides 7 standards consisting of TFR, isolated from human placenta, contained in a phosphate buffer with BSA and normal rabbit serum and calibrated in ne/ml while QSI provides 6 standards, consisting of human serum transferrin receptor, contained in buffered animal serum and calibrated in nmol/L. The assay results are reported in ug/ml in the TFR assay compared to nmol/L for OSI. Samples and controls are diluted 100:1 prior to use in the 7fR assay compared to 50:1 for QSI. Finally, TFR is a single-stage compared to QSI which is a two-stage assay. Over an 18 month period of time, patient samples were collected. These samples were segregated into one of three categories: 1) Anemia of Chronic Disease ("ACD"), clinically defined as Serum Iron < 60, Iron Binding Capacity < 350 and Hemoglobin < 12; 2) Iron Deficiency Anemia ("IDA"), clinically defined as Serum Iron < 60, Iron Binding Capacity > 350 and Hemoglobin < 12; and 3) neither IDA nor ACD, samples that had either Serum Iron > 60 or Hemoglobin > 12. Of the above samples, TFR results were obtained for 38 of the IDA, 175 of the ACD, and 70 of the "Neither IDA nor ACD" samples (283 total), while QSI results were obtained for 23 of the IDA, 119 of the IDA and 13 of the "Neither IDA nor ACD" samples (155 total). {2}------------------------------------------------ Performing a linear regression of all values for the 155 samples for which a TFR and OSI value has been obtained yields an R2 of 88.2% (See Table 17). Of the 38 IDA samples, QSI results were obtained for 23. IffR is in agreement (>8.3ug/ml) with the clinical definition of IDA 30 out of 38 times or 78.9% while QSI is in agreement with the clinical definition 18 out of 23 times or 78.3%. Of these 23 samples, TfR delivered the same individual result as OSI for 22 out of 23 samples. The IDA population was further refined to those samples where QSI and ferritin results concurred (STR > 28.1 and ferritin < 30). This reduced the population to 17 samples. Of these 17 samples, TFR concurred with the individual QSI results 17 out of 17 times or 100.0% . Of the 175 ACD samples, QSI results were obtained for 119. TfR is in agreement ( < 8.3 ug/ml) with the clinical definition of ACD 145 out of 175 times or 82.9% while OSI is in agreement with the clinical definition 81 out of 119 times or 68.1%. Of these 119 samples, TfR delivered the same individual result as OSI. The ACD population was further refined to those samples where OSI and ferritin results concurred (STR < 28.1 and ferritin > 30). This reduced the population to 80 samples. Of these 80 samples, TFR concurred with the individual OSI results 79 out of 80 times or 98.8%. Of the 70 "Neither IDA nor ACD" samples. OSI results were obtained for 13. Of these 13 samples, TFR delivered the same individual result as OSI 12 out of 13 times or 92.3% When all samples results are taken as a whole, TFR delivered the same individual result as QSI 140 out of 155 times or 90.3%. When only those cases are used where OSI and ferritin are in agreement. TFR and OSI give the same result 96 out of 97 times or 98.97%. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES OCT 27 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Jeffrey B. Grubb President Ramco Laboratories, Inc. 4507 Mt. Vernon 77006 Houston, Texas - Re: K981208/S1 Trade Name: ની ને જ II Requlatory Class: Product Code: JNM Dated: July 29, 1998 Received: July 31, 1998 Dear Mr. Grubb: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Image /page/3/Picture/9 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The eagle is composed of three curved lines that form the shape of the bird's body and wings. The text is in all capital letters and is evenly spaced around the circle. {4}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial in your 510 (x) promodevice to a legally marketed predicate device equiration or your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro regaractic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and 394 1500. That would be please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, (301) 391 1009. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . be obtained from the Division of Small Manufacturers Assistance at its be extained more (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number: K981208 Device Name: IfR The IJR assay is an in vitro enzyme immunoassay for quantifying Indications for Use: The 1918 assay is an in weeptor in human serum or plasina to aid in the diagnosis of iron deficiency anemia, particularly in the presence of other disease states. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|---------| | (Division Sign-Off) | | | Division of Clinical Laboratory Devices | | | 510(k) Number | K981208 | | Prescription Use | ✓ | OR | Over-The-Counter Use | _____ | |----------------------|-------| |----------------------|-------|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...