K963498 · Dade Chemistry Systems, Inc. · JMG · Jan 24, 1997 · Immunology
Device Facts
Record ID
K963498
Device Name
FERR FLEX REAGENT CARTRIDGE
Applicant
Dade Chemistry Systems, Inc.
Product Code
JMG · Immunology
Decision Date
Jan 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5340
Device Class
Class 2
Indications for Use
The FERR Flex™ reagent cartridge is used on the Dimension™ RxL clinical chemistry system to quantitatively measure ferritin in human serum and plasma.
Device Story
Ferritin Flex™ Reagent Cartridge is a one-step enzyme immunoassay for use on the Dimension™ RxL clinical chemistry system. Input: human serum or plasma sample. Principle: sandwich immunoassay using monoclonal antibodies coated on chromium dioxide particles and β-galactosidase labeled monoclonal antibodies. Process: sample incubation forms particle-ferritin-conjugate sandwich; magnetic separation and washing remove unbound components; chromogenic substrate (CPRG) added; hydrolysis releases chromophore (CPR). Output: color change measured at 577 nm, proportional to ferritin concentration. Used in clinical laboratories by technicians. Results aid clinicians in assessing iron status.
Clinical Evidence
Split-sample comparison between FERR Flex™ on Dimension® RxL and Abbott IMx® assay. Sample size: 60 clinical patient samples. Results: correlation coefficient 0.983, slope 0.98, intercept -0.8 ng/mL.
Technological Characteristics
Sandwich enzyme immunoassay; chromium dioxide particles; β-galactosidase conjugate; chromogenic substrate (CPRG); colorimetric detection at 577 nm; automated system (Dimension® RxL).
Indications for Use
Indicated for the quantitative measurement of ferritin in human serum and plasma to aid in the diagnosis and monitoring of iron deficiency or iron overload states.
Regulatory Classification
Identification
A ferritin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ferritin (an iron-storing protein) in serum and other body fluids. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency amemia.
Predicate Devices
Abbott Laboratories IMx® Ferritin
Related Devices
K070552 — DIMENSION VISTA FERRITIN FLEX REAGENT CARTRIDGE (FERR) · Dade Behring, Inc. · Apr 30, 2007
K234052 — Access Ferritin · Beckman Coulter, Inc. · Mar 20, 2024
Submission Summary (Full Text)
{0}
DADE K9a3498
Date: 12/03/96
Interoffice Memorandum
JAN 24 1997
# Summary of Safety and Effectiveness Information
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
**Submitter's Name:** Cathy P. Craft
Dade International Inc.
P.O. Box 6101
Newark, DE 19714-6101
**Date of Preparation:** 12/03/96
**Name of Product:** Ferritin Flex™ Reagent Cartridge
**FDA Classification Name:** FerritinTest System
**Predicate Device:** Abbott Laboratories IMx® Ferritin
**Device Description:** The FERR Flex™ Reagent Cartridge is a one-step enzyme immunoassay based on the “sandwich” principle. Sample is incubated with chromium dioxide particles, coated with monoclonal antibodies specific for ferritin and conjugate reagent consisting of β-galactosidase labeled monoclonal antibodies specific for a second binding site on ferritin to form a particle-ferritin-conjugate sandwich. Unbound conjugate and analyte are removed by magnetic separation and washing. The sandwich bound β-galactosidase is combined with a chromogenic substrate chlorophenol red-b-d-galactopyranoside (CPRG). Hydrolysis of CPRG releases a chromophore (CPR). The color change at 577 nm is directly proportional to the concentration of ferritin in the original sample.
**Intended Use:** The FERR Flex™ reagent cartridge is used on the Dimension™ RxL clinical chemistry system to quantitatively measure ferritin in human serum and plasma.
6
{1}
# Comparison to Predicate Device:
| Item | Dimension® RxL Ferritin | Imx® Ferritin |
| --- | --- | --- |
| Technology | Sandwich format
monoclonal antibody
immunoassay | Sandwich format
monoclonal antibody
immunoassay |
| Detection | Colorimetric rate
measurement at 577nm
and 700nm | Fluorometric endpoint
measurement |
# Comments on Substantial
Equivalence: Split sample comparison between the FERR Flex™ reagent cartridge on the Dimension® RxL clinical chemistry system and the Abbott IMx® assay gave a correlation coefficient of 0.983, slope of 0.98, and an intercept of -0.8 ng/mL when tested with 60 clinical patient samples.
Conclusion: The FERR Flex™ Reagent Cartridge is substantially equivalent in principle and performance to the Imx® Ferritin Assay based on the split sample comparison discussed above.

Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.