N LATEX IGE MONO REAGENT

K991787 · Dade Behring, Inc. · DGC · Jul 7, 1999 · Immunology

Device Facts

Record IDK991787
Device NameN LATEX IGE MONO REAGENT
ApplicantDade Behring, Inc.
Product CodeDGC · Immunology
Decision DateJul 7, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Indications for Use

In vitro diagnostic reagent for the quantitative determination of IgE in human serum or plasma by particle enhanced immunonephelometry using the Behring Nephelometer Systems, and aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Device Story

N Latex IgE mono Reagent is an in vitro diagnostic reagent used with Behring Nephelometer Systems. It measures IgE levels in human serum or plasma samples. The device utilizes particle-enhanced immunonephelometry to quantify IgE concentrations. The system is operated by laboratory professionals in clinical settings. The nephelometer measures light scattering caused by antigen-antibody complexes; the resulting quantitative output is used by clinicians to assess protein metabolism and immune function. This information assists in diagnosing conditions related to abnormal protein metabolism and the body's ability to resist infectious agents.

Clinical Evidence

Bench testing only. Correlation study compared N Latex IgE mono assay to N Latex IgE assay using 70 samples (range 14-855 IU/ml), yielding a correlation coefficient of 1.0, slope of 1.0, and y-intercept of 6.1. Precision studies conducted per NCCLS EP5-T2; inter-assay precision 1.5-4.1%, intra-assay precision 1.7-3.0%.

Technological Characteristics

Particle-enhanced immunonephelometry reagent for use with Behring Nephelometer Systems. Quantitative in vitro diagnostic assay. No specific materials or software architecture details provided.

Indications for Use

Indicated for the quantitative determination of IgE in human serum or plasma to aid in the diagnosis of abnormal protein metabolism and immune resistance deficiencies.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### JUL -7 1999 # 510(k) Summary For N Latex IgE mono Reagent This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 # The assigned 510(k) number is: 1 99 | 787 - Manufacture's Name, Address, Telephone, and Contact Person, Date of 1. Preparation: Manufacturer: Contact Information: Dade Behring Marburg GmbH Emil-von-Behring Str. 76 Marburg/Germany Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Rebecca S. Ayash Tel: 302-631-6276 Preparation date: May 24, 1999 #### 2. Device Name/ Classification: N Latex IgE mono Reagent: immunological test system Immunoglobulins A, G, M, D, and E Classification Number: Class II (866.5510) #### 3. ldentification of the Legally Marketed Device: N Latex IgE kit (K890530) #### Device Description: 4. N Latex IgE mono Reagent is intended to be used together with the Behring Nephelometer Systems for the quantitative determination of Immunoglobulin E (IgE) in human serum or plasma. #### 5. Device Intended Use: In vitro diagnostic reagent for quantitative determination of IgE in human serum or plasma by particle enhanced immunonephelometry using the Behring Nephelometer Systems. {1}------------------------------------------------ #### 6. Medical device to which equivalence is claimed and comparison information: There are a number of in vitro diagnostic products in commercial distribution, which employ immunoassay techniques for the quantitative determination of IgE in human serum or plasma. One such product is the Dade Behring N Latex IgE kit (K890530). N Latex IgE mono Reagent is substantially equivalent in intended use and results obtained to the N Latex IgE kit. The N Latex IgE mono Reagent, like the N Latex IgE kit is intended to be used for the quantitative determination of IgE in human serum or plasma using the Behring Nephelometer Systems. #### 7. Device Performance Characteristics: ## Correlation: The N Latex IgE mono assay was compared to the N Latex IgE assay by evaluating 70 samples ranging from 14 to 855 IU/ml. A correlation coefficient of 1.0 was obtained, with a y-intercept value of 6.1 and a slope of 1.0. ## Precision: Precision studies were performed by the evaluation of three levels of control material and two levels of human serum pools in a manner consistent with NCCLS Guideline EP5-T2. The inter-assay precision ranged from 1.5 to 4.1%, while the intra-assay precision ranged from 1.7 to 3.0%. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL -7 1999 Ms. Rebecca S. Ayash Manager, Regulatory Affairs, Biology DADE BEHRING INC. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Re: K991787 > Trade Name: N Latex IgE mono Reagent Regulatory Class: II Product Code: DGC Dated: May 24, 1999 Received: May 25, 1999 Dear Ms. Ayash: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ ### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally, 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Dade Behring Inc. N Latex IgE mono Reagent 510(k) Notification # Indications Statement K991787 Device Name: N Latex IgE mono Reagent Indications for Use: In vitro diagnostic reagent for the quantitative determination of IgE in human serum or plasma by particle enhanced immunonephelometry using the Behring Nephelometer Systems, and aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![Signature](signature.png) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K991787 | | Prescription Use (Per 21 CFR 801.109) | Over-The-Counter-Use (Optional Format 1-2-96) | |---------------------------------------|-----------------------------------------------| |---------------------------------------|-----------------------------------------------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...