Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay
K220001 · Diazyme Laboratories, Inc. · DFH · Aug 4, 2022 · Immunology
Device Facts
Record ID
K220001
Device Name
Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay
Applicant
Diazyme Laboratories, Inc.
Product Code
DFH · Immunology
Decision Date
Aug 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5550
Device Class
Class 2
Indications for Use
The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only. The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
Device Story
Latex particle enhanced immunoturbidimetric assays for quantitative measurement of Kappa and Lambda free light chains in human serum; performed on validated clinical chemistry analyzers. Principle: antigen-antibody reaction between FLC in patient sample and anti-FLC antibodies coated on latex particles; resulting turbidity measured spectrophotometrically. Output: quantitative concentration values used by clinicians to assist in diagnosis and monitoring of multiple myeloma. Used in clinical laboratory settings by trained laboratory personnel. Results interpreted alongside other clinical/laboratory data to inform patient management decisions.
Clinical Evidence
No clinical data provided; substantial equivalence established through bench testing and performance validation of the immunoturbidimetric assay.
Technological Characteristics
Latex particle enhanced immunoturbidimetric assay; utilizes anti-FLC antibodies; spectrophotometric detection; intended for use on validated clinical chemistry analyzers; in-vitro diagnostic reagent kit.
Indications for Use
Indicated for the quantitative determination of Kappa and Lambda Free Light Chain (FLC) concentrations in human serum to aid in the diagnosis and monitoring of multiple myeloma in conjunction with other clinical and laboratory findings. For prescription use only.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Related Devices
K211648 — Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay · Diazyme Laboratories, Inc. · Sep 8, 2022
K253358 — Diazyme Human Kappa (k) Free Light Chain Assay; Diazyme Human Lambda (¿) Free Light Chain Assay · Diazyme Laboratories, Inc. · Dec 19, 2025
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K220001
B Applicant
Diazyme Laboratories Inc.
C Proprietary and Established Names
Diazyme Human Kappa (κ) Free Light Chain Assay
Diazyme Human Lambda (λ) Free Light Chain Assay
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DFH, DEH | Class II | 21 CFR 866.5550 - Immunoglobulin (Light Chain Specific) Immunological Test System | IM - Immunology |
## II Submission/Device Overview:
A Purpose for Submission:
Modification of previously cleared devices – addition of intended use as aid in monitoring of multiple myeloma
B Measurand:
Kappa Free Light Chain (FLC)
Lambda Free Light Chain (FLC)
C Type of Test:
Latex particle enhanced immunoturbidimetric, quantitative
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K220001 - Page 2 of 9
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Warning: The result of the Kappa FLC or Lambda FLC in a given specimen determined with assays and/or instrument platforms from different manufacturers can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the Kappa FLC or Lambda FLC assay used. Values obtained with different assay methods cannot be used interchangeably.
If, in the course of serially monitoring a patient, the assay method used for determining the Kappa FLC and Lambda FLC levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored.
### D Special Instrument Requirements:
Roche cobas c501 analyzer
## IV Device/System Characteristics:
### A Device Description:
The Diazyme Human Kappa (κ) Free Light Chain Assay is comprised of the following reagents:
- Reagent R1: Tris buffer solution
{2}
- Reagent R2: Latex particles coated with polyclonal rabbit and goat anti-human kappa FLC antibodies
Material required but not provided
- Calibrators: A five-level set in liquid form (5 x 1 mL) with target concentrations: 10, 20, 40, 80, 160 mg/L, ready to use
- Controls Set: Two levels (15 mg/L and 30 mg/L), serum based, liquid form
The Diazyme Human Lambda (λ) Free Light Chain Assay is comprised of the following reagents:
- Reagent R1: Tris buffer solution
- Reagent R2: Latex particles coated with polyclonal rabbit anti-human lambda FLC antibodies
Material required but not provided
- Calibrators: A five-level set in liquid form (5 x 1 mL) with target concentrations: 10, 25, 55, 110, 220 mg/L, ready to use
- Control Set: Two levels (15 mg/L and 30 mg/L), serum based, liquid form
B Principle of Operation:
The Diazyme Human Kappa (κ) Free Light Chain Assay or the Diazyme Human Lambda (λ) Free Light Chain Assay is based on a latex enhanced immunoturbidimetric assay. Kappa FLC or Lambda FLC in the sample binds to specific anti-Kappa FLC or anti-Lambda FLC antibody, which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically and is proportional to the amount of Kappa FLC or Lambda FLC in the sample. The instrument calculates the Kappa FLC or Lambda FLC concentration by interpolation of obtained signal of a 6-point calibration curve prepared from calibrators of known concentrations.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Optilite Freelite Kappa Free Kit using Binding Site Optilite analyzer
Optilite Freelite Lambda Free Kit using Binding Site Optilite analyzer
B Predicate 510(k) Number(s):
K150658
C Comparison with Predicate(s):
K220001 - Page 3 of 9
{3}
K220001 - Page 4 of 9
| Device & Predicate Device(s): | K220001 | K150658 |
| --- | --- | --- |
| Device Trade Name | Diazyme Human Kappa Free Light Chain Assay
Diazyme Human Lambda Free Light Chain Assay | Freelite Human Kappa Free Kit
Freelite Human Lambda Free Kit |
| General Device Characteristic Similarities | | |
| Analyte | Kappa FLC
Lambda FLC | Same |
| Measurement | Quantitative | Same |
| Detection Method | Latex particle immunoturbidimetric | Same |
| Sample type | Serum | Same |
| General Device Characteristic Differences | | |
| Intended Use/Indications For Use | The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only. | The Optilite Freelite Kappa Free Kit is intended for the quantitative in vitro measurement of kappa free light chains in serum using the Binding Site Optilite turbidimetric analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings.
The Optilite Freelite Lambda Free Kit is intended for the quantitative in vitro measurement of kappa free light chains in serum using the Binding Site Optilite turbidimetric analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström’s macroglobulinemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. |
| Instrument | Roche cobas c501 | Binding Site Optilite analyzer |
{4}
K220001 - Page 5 of 9
| Detection Antibody | Polyclonal anti-Human Kappa/ anti-human Lambda Free light chain Ig fraction coated on polystyrene latex microparticles. | Polyclonal monospecific antibody coated onto polystyrene latex |
| --- | --- | --- |
| Analytical Measuring Range | Kappa FLC: 4.5–150 mg/L
90–3,000 mg/L (extended) | Kappa FLC: 2.9–127 mg/L
0.6–63500 mg/L (extended) |
| | Lambda FLC: 6.1–200 mg/L
122–4,000 mg/L (extended) | Lambda FLC: 52–1390 mg/L
1.3–139000 mg/L (extended) |
| Reference range | Kappa: 2.37 – 20.73 mg/L
Lambda: 4.23 – 27.69 mg/L
K/L Ratio: 0.22 – 1.74 | Kappa: 3.30 – 19.40 mg/L
Lambda: 5.71 – 26.30 mg/L
K/L Ratio: 0.26 – 1.65 |
VI Standards/Guidance Documents Referenced:
Not applicable
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Refer to K153394 and K181438
2. Linearity:
Refer to K153394 and K181438
3. Analytical Specificity/Interference:
Refer to K153394 and K181438
4. Assay Reportable Range:
Refer to K153394 and K181438
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Refer to K153394 and K181438
6. Detection Limit:
Refer to K153394 and K181438
{5}
7. Assay Cut-Off:
See expected values/reference range
## B Comparison Studies:
1. Method Comparison with Predicate Device:
Refer to K153394 and K181438
2. Matrix Comparison:
Not applicable
## C Clinical Studies:
1. Clinical Sensitivity and Specificity:
The clinical performance of the Diazyme Kappa FLC and Lambda FLC Assays as aid in monitoring of disease status in subjects diagnosed with multiple myeloma (MM) and light chain multiple myeloma (LCMM) was investigated in a clinical study using archived serial serum samples tested by the candidate devices in multicenter sites. A total of 169 MM subjects including 22 LCMM subjects from three sites were enrolled in the study. Changes in FLC Kappa and FLC Lambda levels in serial serum samples from subjects diagnosed with MM and LCMM during treatment were compared to clinical response categories based on International Myeloma Working Group (IMWG) guideline as determined by the clinicians.
The patient inclusion and exclusion criteria were as follows:
**Inclusion criteria:**
- Confirmed diagnosis of MM or LCMM
- At least three blood collections per patient
- Clinical information of response level available
**Exclusion criteria:**
- Less than three blood collections
- MGUS patients
Of the 169 MM subjects, 94 subjects were diagnosed with IgG Kappa (IgGK), 32 IgG Lambda (IgGL), 20 IgA Kappa (IgAK), 11 IgA Lambda (IgAL), 1 IgM Kappa (IgMK), 1 IgM Lambda (IgML), and 22 Light Chain Multiple Myeloma (13 LCMM-Kappa and 9 LCMM-Lambda). Twelve of 169 subjects had two subtypes (clones) as follows: six LCMM Kappa/IgGK; one LCMM Lambda/IgGL; one IgGK/LCMM Kappa; one IgGL/LCMM Lambda; one IgAK/LCMM Kappa; one IgAK/IgGK; and one IgML/LCMM Lambda.
Ninety (90) subjects (53%) were male and 79 (47%) were female. The median age was 66 years old with the age ranging from 43 to 93 years old. The majority of the subjects were
K220001 - Page 6 of 9
{6}
African American (n=77; 45.6%). The remaining subjects were: Caucasian (n=42; 24.9%), Asian (n=20; 11.8%); Caucasian/Hispanic (n=4; 2.4%); African/Asian (n=1; 0.6%); South Asian Indian (n=1; 0.6%); Other (n=9; 5.3%) and unknown (n=15; 8.9%). A total of 541 blood draws including 169 baselines and a total of 372 monitoring observations were tested by the Diazyme Human Kappa FLC and Lambda FLC Assays. The number of draws per subject ranged from three to nine draws. The duration of the monitoring days ranged from 10 days to 1002 days (with a median of 203 days).
The information collected for each sample included the following: (a) general information including patient's demographic information and date blood sample was drawn; (b) clinical information including patient's diagnosis and treatment initiated, and diagnostic response criteria; c) bone imaging when available and laboratory data including hypercalcemia, renal failure, anemia, and bone disease (CRAB), bone marrow plasma cell count greater than 10 percent, serum electrophoresis results (M-protein), serum immunofixation electrophoresis.
The clinical assessment for each monitoring observation as determined by the clinicians according to International Myeloma Working Group (IMWG) Version 1.2020/National Comprehensive Cancer Network (NCCN) Clinical Practice Guidelines in Oncology on Multiple Myeloma and categorized into one of seven response categories: Stringent Complete Response (sCR), Complete Response (CR), Very Good Partial Response (VGPR), Partial Response (PR), Minimal Response (MR), Stable Disease (SD), and Progressive Disease (PD).
For each of 372 monitoring observations, the percentage change of the Involved FLC Kappa (iFLC Kappa) or Involved FLC Lambda (iFLC Lambda) to not involved FLC (niFLC) was calculated by comparing the test result to the test result from the previous visit. The results of Diazyme Kappa and Lambda FLC assay in combination with other laboratory results were used to assign a response category and compared to the clinical assessment according to the following table:
| | Response Criteria based on Diazyme FLC, SPE, IFE Results |
| --- | --- |
| Good Response | FLC ratio normal, M protein not detectable and IFE negative# |
| Moderate Response | ≥ 50 to ≥ 90% reduction of rd_dFLC* and ≥ 50 to ≥ 90% reduction of M protein |
| Stable Disease | 49% reduction to 25% increase of rd_dFLC and
49% reduction to 25% increase of M protein |
| Progressive Disease | ≥ 25% increase of rd_dFLC and ≥ 25% increase of M protein
AND
increase of d_dFLC^ ≥ 100 mg/L and increase of M protein ≥ 0.5 g/L |
# Applies only to MM population
* rd_dFLC = (dFLC t2 - dFLC t1) / dFLC t1 (while rd = relative difference; dFLC = iFLC - ni FLC; t1 = time point 1; t2 = time point 2;
^ d_dFLC = dFLC t2 - dFLC t1; dFLC = iFLC - niFLC
(while iFLC = involved FLC; niFLC = not involved FLC)
K220001 - Page 7 of 9
{7}
# Results:
The clinical performance of the Diazyme Human Kappa and Lambda FLC assays in combination with other laboratory tests as an aid in monitoring of MM patient was evaluated by comparing the clinical status assessed by physician (taking into account all available clinical and laboratory information) with the response criteria derived from the Diazyme Human Kappa and Lambda FLC based on changes between the consecutive blood draws. The results are summarized in the following tables.
| Response based on Diazyme FLC Assay | Clinical Assessment | | | | |
| --- | --- | --- | --- | --- | --- |
| | Good Response** | Moderate Response*** | Stable Disease | Progressive Disease | Total |
| Good Response | 44 | 6 | 8 | 0 | 58 |
| Moderate Response | 2 | 27 | 27 | 6 | 62 |
| Stable Disease | 25 | 12 | 173 | 20 | 230 |
| Progressive Disease | 0 | 0 | 5 | 17 | 22 |
| Total | 71 | 45 | 213 | 43 | 372 |
| Concordance (n/N) (95% CI)* | 62% (44/71) (50.7%; 73.2%) | 60% (27/45) (46.7%; 73.3%) | 81% (173/213) (76.1%; 86.4%) | 40% (17/43) (25.6%; 55/8%) | 70% (261/372) (65.5%; 74.5%) |
* 95% CI was calculated by bootstrap method
** Including subjects with sCR and CR
*** Including subjects with VGPR, PR and MR.
The clinical sensitivity and specificity based on "Progression" and "No-progression" are summarized in the following table. Subjects with 'Progression' consist of those monitoring events defined as 'Progressive Disease'. Subjects with 'No-Progression' consist those monitoring events defined as 'Good Response, Moderate Response, and Stable Disease Response' categories.
| | Clinical Assessment | | | |
| --- | --- | --- | --- | --- |
| | | Progression | No Progression | Total |
| Change in Diazyme FLC | PD* | 17 | 5 | 22 |
| | No-PD | 26 | 324 | 350 |
| | Total | 43 | 329 | 372 |
| Clinical Sensitivity: 40% (17/43) (95% CI: 25.6% to 55.8%) | | | | |
| Clinical Specificity: 98% (324/329) (95% CI: 97.3% to 99.7%) | | | | |
* PD: ≥25% increase of rd_dFLC and ≥25% increase of M protein AND increase of d_dFLC ≥ 100 mg/L and increase of M protein ≥ 0.5 g/L
2. Other Supportive Data:
In this clinical study, all samples were also tested with the predicate device (Binding Site Optilite Freelite Kappa and Freelite Lambda Free Kit). The clinical performance of the Diazyme Human Kappa/Lambda FLC Assays as an aid in monitoring of MM patients was
K220001 - Page 8 of 9
{8}
compared to the predicate device based on "Progression" and "No-progression". The results are summarized in the following table.
| Diazyme Human Kappa/Lambda FLC | | Optilite Freelite Kappa/Lambda FLC | |
| --- | --- | --- | --- |
| Sensitivity (n/N) (95% CI*) | Specificity (n/N) (95% CI*) | Sensitivity (n/N) (95% CI*) | Specificity (n/N) (95% CI*) |
| 40% (17/43) (25.6%; 55.8%) | 98% (324/329) (97.3%; 99.7%) | 44% (19/43) (30.2%; 60.5%) | 97% (320/329) (95.4%; 98.8%) |
* 95% CI was calculated by bootstrap method
## D Clinical Cut-Off:
Not applicable
## E Expected Values/Reference Range:
Refer to K153394 and K181438
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220001 - Page 9 of 9
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.