← Product Code [DFH](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFH) · K182098

# N Latex FLC kappa assay, N Latex FLC lambda assay (K182098)

_Siemens Healthcare Diagnostics Products GmbH · DFH · Nov 1, 2018 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFH/K182098

## Device Facts

- **Applicant:** Siemens Healthcare Diagnostics Products GmbH
- **Product Code:** [DFH](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFH.md)
- **Decision Date:** Nov 1, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5550
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma by means of particle-enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and as an aid in the diagnosis of amyloidosis (AL) in conjunction with other laboratory and clinical findings. The response category assignment of Complete Response for the monitoring of MM, is reliant upon the combination of clinical history and other tests including protein electrophoresis, immunofixation and bone marrow, imaging and urine assessments.

## Device Story

In-vitro diagnostic reagents for quantitative determination of free light chains (FLC) in human serum and EDTA-plasma. Principle: particle-enhanced immunonephelometry; polystyrene particles coated with monoclonal antibodies agglutinate with FLC in samples; light scattering intensity proportional to protein concentration. Used on BN II and BN ProSpec systems in clinical laboratory settings. Output: quantitative FLC concentration (mg/L). Healthcare providers use results alongside clinical history, protein electrophoresis, immunofixation, bone marrow, imaging, and urine assessments to monitor multiple myeloma therapy response and aid in amyloidosis diagnosis. Benefits include standardized monitoring of disease progression or response to therapy.

## Clinical Evidence

Retrospective clinical study of 98 subjects (104 MM, 21 LCMM) with 391 serum samples. Evaluated performance as an aid in monitoring MM disease status compared to IMWG clinical response criteria. Clinical sensitivity for detecting progressive disease was 26% (95% CI: 13.9%–40.3%) and clinical specificity was 97% (95% CI: 95.1%–98.8%). Comparison with predicate device showed 89% overall agreement in response category assignment.

## Technological Characteristics

Particle-enhanced immunonephelometry. Reagents: polystyrene particles coated with mouse monoclonal anti-human FLC kappa or lambda antibodies. Energy source: light beam for nephelometric scattering. Dimensions/form factor: liquid reagents for automated BN II/BN ProSpec systems. Connectivity: standalone/integrated into clinical laboratory instrument workflow.

## Regulatory Identification

An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.

## Predicate Devices

- The Binding Site Freelite® Human Kappa Free Kit for use on the Siemens BN™ II ([K031016](/device/K031016.md))
- The Binding Site Freelite® Human Lambda Free Kit for use on the Siemens BN™ II ([K031016](/device/K031016.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM

A. 510(k) Number:
K182098

B. Purpose for Submission:
Modification of previously cleared devices – addition of intended use as aid in monitoring of multiple myeloma

C. Measurand:
Kappa (κ) Free Light Chain (FLC)
Lambda (λ) Free Light Chain (FLC)

D. Type of Test:
Nephelometry, quantitative

E. Applicant:
Siemens Healthcare Diagnostics Products GmbH

F. Proprietary and Established Names:
N Latex FLC kappa assay
N Latex FLC lambda assay

G. Regulatory Information:
1. Regulation section:
21 CFR § 866.5550 – Immunoglobulin (light chain specific) immunological test system
2. Classification:
Class II
3. Product code:
DFH, Kappa, antigen, antiserum, control
DEH, Lambda, antigen, antiserum, control
4. Panel:
Immunology (82)

{1}

H. Intended Use:

1. Intended uses:

N Latex FLC kappa and N Latex FLC lambda assays

In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma by means of particle-enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and as an aid in the diagnosis of amyloidosis (AL) in conjunction with other laboratory and clinical findings.

The response category assignment of Complete Response for the monitoring of MM, is reliant upon the combination of clinical history and other tests including protein electrophoresis, immunofixation and bone marrow, imaging and urine assessments.

2. Indications for use:

Same as Intended Use

3. Special conditions for use statements:

For prescription use only

Warning: The result of the FLC Kappa or FLC Lambda in a given specimen determined with assays and/or instrument platforms from different manufacturers can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay method used for determining the FLC Kappa and FLC Lambda levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored.

4. Special instrument requirements:

BN II (K943997) and BN ProSpec Systems (K001647)

I. Device Description:

The N Latex FLC kappa and N Latex FLC lambda assays are comprised of the following reagents in liquid form:

N latex FLC reagents: suspension of polysterene particles coated with mouse monoclonal antibodies to either human FLC Kappa or human FLC Lambda with preservative.

N FLC supplementary reagents: mouse immunoglobulin in buffered solution with preservatives.

{2}

N FLC standard SL and N FLC controls SL1 and SL2: human free light chains proteins, human serum albumin and protease inhibitors.

# J. Substantial Equivalence Information:

1. Predicate device names and 510(k) numbers:

The Binding Site Freelite® Human Kappa Free Kit and Freelite® Human Lambda Free Kit for use on the Siemens BN™ II - K031016

2. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device N Latex FLC kappa N Latex FLC lambda | Predicate Freelite® Human Kappa Free and Freelite® Human Lambda Free kits on the Siemens BN™II  |
|  Analyte | Kappa: Kappa FLC Lambda: Lambda FLC | Same  |
|  Measurement | Quantitative | Same  |
|  Detection Method | Nephelometric | Same  |
|  Calibrator | One level | Same  |
|  Reference material | Internal Reference preparation | Same  |
|  Units of measure | mg/L | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Indication for use | In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma by means of particle- enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and as an aid in the diagnosis of amyloidosis (AL) in conjunction with other laboratory and clinical findings.The response category assignment of ‘Complete Response’ for the monitoring of MM, is reliant upon | This kit is intended for the quantitation of kappa free light chains or lambda free light chains in serum and urine on the Siemens BN™ II. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom’s macroglobulinemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings.  |

{3}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | the combination of clinical history and other tests including protein electrophoresis, immunofixation and bone marrow, imaging and urine assessments. |   |
|  Sample type | Serum and EDTA plasma | Serum and urine  |
|  Detection antibody | Kappa: Monoclonal mouse anti-human FLC kappa antibody coated onto polysterene particles
Lambda: Monoclonal mouse anti-human FLC lambda antibody coated onto polysterene particles | Kappa: Polyclonal sheep anti-human kappa antibody coated onto latex particles
Lambda: Polyclonal sheep anti-human lambda antibody coated onto latex particles  |
|  Analytical measuring ranges | Kappa: 3.4 – 110 mg/L
Lambda: 1.9 – 60 mg/L | Kappa: 5.9 – 190 mg/L
Lambda: 5.0 – 160 mg/L  |
|  Instrument system | Siemens BN II and BN ProSpec Systems | Siemens BN II  |
|  Reference interval | Kappa: 8.24 – 28.90 mg/L
Lambda: 9.10 – 32.60 mg/L
Ratio: 0.53 – 1.51 | Kappa: 3.30 to 19.40 mg/L
Lambda: 5.71 to 26.30 mg/L
Ratio: 0.26 – 1.65  |

## K. Standard/Guidance Document Referenced (if applicable):

## L. Test Principle:

The FLC test systems are based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with monoclonal antibodies to human free light chains (FLCs), type kappa or lambda, are agglutinated when mixed with samples containing FLCs. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

## M. Performance Characteristics:

1. Analytical performance:

a. Precision/Reproducibility:

See K171742

b. Linearity/assay reportable range:

See K171742

{4}

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

See K171742

d. Detection limit:

See K171742

e. Analytical specificity:

See K171742

f. Assay cut-off:

See expected values/reference range below.

2. Comparison studies:

a. Method comparison with predicate device:

See K171742

b. Matrix comparison:

See K171742

3. Clinical studies:

The performance of the N Latex FLC kappa assay and the N Latex FLC lambda assay as an aid in monitoring of disease status in multiple myeloma (MM) and light chain multiple myeloma (LCMM) subjects was determined from a retrospective clinical study. Changes in FLC Kappa and FLC Lambda levels in serial serum samples from subjects diagnosed with MM and LCMM during treatment were compared to clinical response categories based on International Myeloma Working Group (IMWG) as determined by the clinicians.

The patient inclusion and exclusion criteria were as follows:

Inclusion criteria:

- Confirmed diagnosis of MM or LCMM
- At least four blood collections in intervals of ≥ three weeks
- Informed consent available
- Clinical information of response level available

Exclusion criteria:

{5}

- Less than four blood collections
- No informed consent available
- Patients subsequently diagnosed with another disease

A total of 125 (104 MM and 21 LCMM) subjects were enrolled in the study. Among them, 27 subjects were excluded: 17 subjects had less than four sample draws; four MM subjects were re-diagnosed to Waldenstrom Macroglobulinemia; two subjects had no informed consent available; and four subjects had no available clinical information about response level. Therefore, a total 98 subjects (81 IgG, IgA and IgM MM and 17 Kappa and Lambda LCMM subjects) with a total of 391 serum samples were used in the study analysis.

Of the 98 subjects, 47 (47%) were male and 51 (52%) were female. The mean age was 65 years and the range was 35 to 92 years. The majority of subjects (n=70, 71%), were Caucasian, the remaining subjects included 18 (18%) African American, one (1%) Asian and nine (9%) Unknown.

A total of 391 draws were collected from 98 subjects. The mean number of draws per subject was five, ranging from four to seven draws. The duration of the monitoring days ranged from 88 days to 1338 days.

The information collected for each sample included the following: (a) general information including patient's demographic information and date of blood sample drawn; (b) clinical information including patient's diagnosis report/medical notes, treatment initiated, and diagnostic response criteria; c) laboratory data including serum electrophoresis results, serum/ urine immunofixation electrophoresis, serum FLC Kappa and FLC Lambda, CRAB (Calcium, Renal Failure, Anemia, Bone lesions) status, and plasma cell count.

## Results:

The performance of the N Latex FLC kappa and N Latex FLC lambda assays as aid in monitoring MM was evaluated by testing a total of 391 samples from 98 subjects consisting of 98 baseline values and 293 monitoring observations and comparing to clinical disease response status.

At each follow-up visit, the percentage change of the Involved FLC Kappa (iFLC Kappa) or Involved FLC Lambda (iFLC Lambda) was calculated by comparing the test result to the test result from the previous visit. The monitoring response criteria based on the FLC results were based on the FLC ratio, % relative change of difference between both FLC (rd_FLC), M protein level and IFE monoclonal protein detection.

The clinical assessment for each monitoring observation was provided by the attending physician who was responsible for the corresponding therapy decisions based on therapy response criteria defined according to International Myeloma Working Group (IMWG) 2011/National Comprehensive Cancer Network (NCCN) Clinical Practice Guidelines in Oncology on Multiple Myeloma. Clinical status of each monitoring event was

6

{6}

categorized into one of six response categories: Stringent Complete Response (sCR), Complete Response (CR), Very Good Partial Response (VGPR), Partial Response (PR), Stable Disease (SD), and Progressive Disease (PD).

Data were analyzed based on the comparison of disease category determined from N Latex assays to clinical assessment.

Results of N Latex FLC kappa and N Latex FLC lambda assays in combination with other laboratory results were used to assign a response category and compared to the clinical assessment according to the table below:

|  Response Category | Response Criteria based on N Latex FLC, SPE, IFE Results | Clinical Assessment based on IMWG  |
| --- | --- | --- |
|  Good Response | FLC ratio normal, M protein not detectable and IFE negative* | sCR  |
|   |   |  CR  |
|  Moderate Response | ≥ 50 to ≥ 90% reduction of rd_dFLC** and ≥ 50 to ≥ 90% reduction of M protein | VGPR  |
|   |   |  PR  |
|  Stable Disease | 50% reduction to 25% increase of rd_dFLC and 50% reduction to 25% increase of M protein | SD  |
|  Progressive Disease | ≥ 25% increase of rd_dFLC and ≥ 25% increase of M protein AND increase of d_dFLC*** ≥ 100 mg/L and increase of M protein ≥ 0.5 g/L | PD  |

*Applies only to MM population
**rd_dFLC = (dFLC t2 - dFLC t1) / dFLC t1
\*\*\*d_dFLC = dFLC t2 - dFLC t1 (with 'dFLC = iFLC - ni FLC')

# Definitions:

- t1 = time point 1; t2 = time point 2
- rd = relative difference
- iFLC = involved FLC; niFLC = not involved FLC
- Relative change of involved FLC: rd_iFLC = (iFLC t2 - iFLC t1)/iFLC t1
- Relative change of difference between both FLC: dFLC = iFLC - niFLC

The performance of N Latex FLC Kappa and FLC Lambda in combination with other laboratory tests compared to the clinical assessment is summarized in the following table:

|  Response Category | Clinical Assessment  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |  Good Response | Moderate Response | Stable Disease | Progressive Disease | Total  |
|  Good Response | 21 | 3 | 6 | 1 | 31  |

{7}

8

|  Moderate Response | 10 | 104 | 38 | 16 | 168  |
| --- | --- | --- | --- | --- | --- |
|  Stable Disease | 3 | 24 | 126 | 18 | 171  |
|  Progressive Disease | 0 | 0 | 9 | 12 | 21  |
|  Total | 34 | 131 | 179 | 47 | 391  |
|  Agreement | 62% | 79% | 70% | 26% | 67%  |

## Discordance results:

The response criteria established by clinicians are often based on multiple factors besides FLC results, therefore FLC results alone might lead to a different classification.

The clinical disease response categories were condensed into two clinical status categories: “Progression” and “No-progression”. Subjects with ‘Progression’ consist of those monitoring events defined as ‘Progressive Disease’. Subjects with ‘No-Progression’ consist those monitoring events defined as ‘Good Response, Moderate Response, and Stable Disease Response’ categories. The performance N-Latex FLC results compared to the clinical status are summarized as follows:

|   | Clinical Status  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Progression | No Progression | Total  |
|  Change in N Latex FLC | Positive* | 12 | 9 | 21  |
|   |  Negative** | 35 | 335 | 370  |
|   |  Total | 47 | 344 | 391  |
|  Clinical Sensitivity: 26% (12/47) (95% CI: 13.9% – 40.3%)
Clinical Specificity: 97% (335/344) (95% CI: 95.1%– 98.8%)  |   |   |   |   |

*Positive: ≥25% rd_dFLC and d_dFLC ≥100 mg/L
**Negative: &lt;25% rd_dFLC and d_dFLC FLC &lt;100 mg/L

b. Other clinical supportive data (when a. and b. are not applicable):

The performance of N Latex FLC kappa and FLC lambda assays were also compared to the predicate device based on the disease response categories described above.

In this retrospective clinical study, all samples were tested with the N Latex FLC kappa and FLC lambda assays and with the predicate device. The performance of N Latex FLC kappa and N Latex FLC lambda assays compared to the predicate device based on the same disease response category is summarized as follows:

|  Response based on N Latex FLC results | Response Based on Freelite Results  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |  Good Response | Moderate Response | Stable Disease | Progressive Disease | Total  |
|  Good Response | 27 | 2 | 0 | 1 | 31  |
|  Moderate Response | 0 | 66 | 15 | 0 | 81  |
|  Stable Disease | 0 | 13 | 238 | 7 | 258  |
|  Progressive Disease | 0 | 0 | 5 | 16 | 21  |
|  Total | 27 | 81 | 259 | 24 | 391  |
|  Agreement | 100% | 81% | 92% | 67% | 89%  |

{8}

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
See K171742

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

9

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFH/K182098](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DFH/K182098)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
