K984191 · Dade Behring, Inc. · DDR · Dec 10, 1998 · Immunology
Device Facts
Record ID
K984191
Device Name
MYOGLOBIN FLEX REAGENT CARTRIDGE
Applicant
Dade Behring, Inc.
Product Code
DDR · Immunology
Decision Date
Dec 10, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5680
Device Class
Class 2
Indications for Use
The Myoglobin (MYO) method for the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module is used to quantitatively measure myoglobin (MYO) in human serum and plasma as an aid in the rapid diagnosis of acute myocardial infarction.
Device Story
The Myoglobin Flex™ reagent cartridge is an in vitro diagnostic assay for use on the Dimension® RxL clinical chemistry system. It utilizes a one-step sandwich enzyme immunoassay principle. Patient serum or plasma samples are incubated with chromium dioxide particles coated with myoglobin-specific monoclonal antibodies and a conjugate reagent consisting of β-galactosidase-labeled monoclonal antibodies. A particle/myoglobin/conjugate sandwich forms; unbound components are removed via magnetic separation and washing. The bound β-galactosidase catalyzes the hydrolysis of the chromogenic substrate chlorophenol red-β-d-galactopyranoside (CPRG) into chlorophenol red (CPR). The rate of color change, measured at 577nm, is directly proportional to the myoglobin concentration in the sample. The device is intended for use in clinical laboratory settings to assist healthcare providers in the rapid diagnosis of acute myocardial infarction.
Clinical Evidence
Clinical performance was evaluated via a split-sample comparison between the Dimension® RxL MYO method and the Stratus® Myoglobin assay. Testing involved 204 clinical patient samples with concentrations ranging from 10 to 1000 ng/mL. Results showed a correlation coefficient of 0.992, a slope of 1.04, and an intercept of -4.15 ng/mL.
Technological Characteristics
Sandwich format monoclonal antibody immunoassay. Solid support: chromium dioxide particles. Detection: colorimetric rate measurement at 577 nm and 700 nm. Energy source: Dimension® RxL system instrumentation. Connectivity: integrated into the Dimension® RxL clinical chemistry system.
Indications for Use
Indicated for the quantitative measurement of myoglobin in human serum and plasma to aid in the rapid diagnosis of acute myocardial infarction.
Regulatory Classification
Identification
A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart or renal disease.
K071474 — DIMENSION CLINICAL CHEMISTRY SYSTEM MPO FLEX REAGENT CARTRIDGE, CALIBRATOR,CONTROL, MODEL RF425, RC425,RC426 · Siemens Healthcare Diagnostics · Dec 10, 2008
K053576 — DIMENSION VISTA MYO REAGENT CARTRIDGE AND CALIBRATOR · Dade Behring, Inc. · Feb 7, 2006
K021229 — MYOGLOBIN AND MYOGLOBIN CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 973243, 973244 · Beckman Coulter, Inc. · Jun 28, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
#### DADE BEHRING INC. P O. Box 6101 Newark DE 19714
# Dade Behring
### Summary of Safety and Effectiveness Information
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Rebecca S. Ayash Submitter's Name: Dade Behring Inc. Building 500, Mailbox 514 P.O. Box 6101 Newark, DE 19714-6101
Date of Preparation: 11/20/98
Device Name: Myoglobin Flex™ reagent cartridge
Classification Name: Myoglobin Immunological Test System
Predicate Device: Stratus® Myoglogobin Fluorometric Enzyme Immunoassay
Device Description: The MYO method for the Dimension® RxL system with the heterogeneous immunoassay module is a one-step enzyme immunoassay based on the "sandwich" principle. Sample is incubated with chromium dioxide particles (CrO-) coated with monoclonal antibodies specific for myoglobin and conjugate reagent (ß-galactosidase labeled monoclonal antibodies specific for myoglobin). A particle/MYO/coniugate sandwich forms during the incubation period. Unbound conjugate and analyte are removed by magnetic separation and washing. The sandwich bound ß-galactosidase is combined with the chromogenic substrate chlorophenol redß-d-galactopyranoside (CPRG) and catalyzes the hydrolysis of CPRG to the chromophore, chlorophenol red (CPR). The concentration of MYO present in the patient sample is directly proportional to the rate of color change measured at 577nm due to formation of CPR.
Intended Use: The Myoglobin (MYO) method for the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module is used to quantitatively measure myoglobin (MYO) in human serum and plasma as an aid in the rapid diagnosis of acute myocardial infarction.
{1}------------------------------------------------
#### Comparison to Predicate Device:
| ltem | Dimension® RxL MYO | Stratus® Myoglobin |
|---------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Technology | Sandwich format monoclonal<br>antibody immunoassay | Sandwich format monoclonal<br>antibody immunoassay |
| Detection | Colorimetric rate measurement<br>at 577 nm and 700nm | Front surface fluorometry<br>measurement |
| Solid Support | Chrome | Glass fiber paper |
| Specimen Type | Serum or plasma | Serum or plasma |
| Intended Use | For the quantitative<br>determination of myoglobin in<br>serum and plasma | For the quantitative<br>determination of myoglobin in<br>serum and plasma |
| Indications for Use | To aid in the rapid diagnosis of<br>acute myocardial infarction | To aid in the rapid diagnosis of<br>acute myocardial infarction |
Comments on Substantial Equivalence: Split sample comparison between the MYO method on the Dimension® RxL system and the Stratus® Myoglobin assay gave a correlation coefficient of 0.992, slope of 1.04, and an intercept of -4.15 ng/mL when tested with 204 clinical patient samples ranging from 10 -- 1000 ng/mL.
Conclusion: The MYO Method for the Dimension® RxL system with the heterogeneous immunoassay module is substantially equivalent in principle and performance to the Stratus ® Myoglobin assay based on the split sample comparison summarized above.
Rebecca S. Ayash
Rebecca S. Avash Regulatory Affairs and Compliance Manager Date: 11/20/98
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, graphic representation of the department.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 0 1998
Ms. Rebecca Ayash Requlatory Affairs & Compliance Manager Dade Behring Inc. Building 500, Mailbox 514 P.O. Box 6101 Newark, Delaware 19714
K984191 Re : Myoglobin Flex Reagent Cartridge Trade Name: Requlatory Class: II Product Code: DDR Dated: November 20, 1998 Received: November 23, 1998
Dear Ms. Ayash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications Statement
Device Name: Myoglobin (MYO) Flex™ reagent cartridge
Indications for Use: The MYO method for the Dimension® RxL with the heterogeneous immunoassay module is a device used to quantitatively measure myoglobin in human serum and plasma as an aid in the rapid diagnosis of acute myocardial infarction.
Rebecca S. Cupsh
Rebecca S. Avash Regulatory Affairs and Compliance Manager Date: 11/20/98
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K984191
510(k) Number
At. An Awm
ision Sian-Off Office of Device Evaluation
Image /page/4/Picture/10 description: The image shows the text "prescription use" inside of an oval shape. There is an arrow pointing to a line that is next to the text. The text is in a sans-serif font. The oval shape is drawn with a thin line.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.