← Product Code [DBY](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DBY) · K022067

# CONVOY ADVANCED DELIVERY SHEATH (K022067)

_Boston Scientific · DBY · Sep 11, 2002 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DBY/K022067

## Device Facts

- **Applicant:** Boston Scientific
- **Product Code:** [DBY](/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DBY.md)
- **Decision Date:** Sep 11, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5520
- **Device Class:** Class 1
- **Review Panel:** Immunology

## Indications for Use

The Convoy Advanced Delivery Sheath Kit is intended to facilitate the intracardiac placement of interventional devices.

## Device Story

The Convoy Advanced Delivery Sheath Kit is a disposable system comprising an introducer sheath, vessel dilator, and guidewire with introducer. It is used by physicians in a clinical setting to facilitate the placement of various cardiovascular catheters into the heart. The device is available in multiple configurations, including varying diameters (8.5 F and 9.5 F), lengths (60 cm - 101.5 cm), and curve geometries (0° - 180° angles; 0.6" - 1.75" radii) to accommodate specific anatomical needs and physician preferences. By providing a stable conduit for interventional tools, the sheath assists in the precise delivery of devices during cardiac procedures, potentially improving procedural efficiency and patient outcomes.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

The kit consists of a disposable introducer sheath, vessel dilator, and guidewire. Sheaths are available in 8.5 F and 9.5 F diameters, lengths from 60 cm to 101.5 cm, and various curve configurations (0° - 180° angles; 0.6" - 1.75" radii). The device is sterilized via ethylene oxide (EO).

## Regulatory Identification

An immunoglobulin G (Fab fragment specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the Fab antigen-binding fragment resulting from breakdown of immunoglobulin G antibodies in urine, serum, and other body fluids. Measurement of Fab fragments of immunoglobulin G aids in the diagnosis of lymphoproliferative disorders, such as multiple myeloma (tumor of bone marrow cells), Waldenstrom's macroglobulinemia (increased immunoglobulin production by the spleen and bone marrow cells), and lymphoma (tumor of the lymphoid tissues).

## Predicate Devices

- Daig Corporation, Fast-Cath Introducing Sheath ([K964518](/device/K964518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SEP 1 1 2002

# 510(k) Summary

| Category:                                  | Comments                                                                                                  |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Sponsor:                                   | Boston Scientific Corporation/EP Technologies, Inc.<br>2710 Orchard Parkway<br>San Jose, CA 95134         |
| Correspondent:                             | Andrea L. Ruth, RAC<br>Senior Associate, Regulatory Affairs<br>2710 Orchard Parkway<br>San Jose, CA 95134 |
| Contact Numbers:                           | Phone: 408.895.3625<br>Fax: 408.895.2202<br>Email: rutha@bsci.com                                         |
| Device Common Name                         | Cardiac Introducer Sheath                                                                                 |
| Device Proprietary Name                    | Convoy™ Advanced Delivery Sheath Kit                                                                      |
| Device Classification Name                 | Introducer, Catheter                                                                                      |
| Device Classification                      | Class II, DYB                                                                                             |
| Predicate Device                           | Daig Corporation, Fast-Cath Introducing Sheath                                                            |
| Predicate Device<br>Manufacturer(s)        | Daig Corporation, A Division of St. Jude Medical                                                          |
| Predicate Device Classification<br>Name(s) | Introducer, Catheter                                                                                      |
| Predicate Device<br>Classification(s)      | Class II, DYB                                                                                             |

#### Date Summary Was Prepared: June 14, 2002.

#### Description of the Device:

The Convoy Advanced Delivery Sheath Kit consists of: (1) a disposable Introducer Sheath, (2) a Vessel Dilator and (3) Guidewire with Guidewire Introducer. These devices are designed for the introduction of various types of cardiovascular catheters to the heart.

The Introducer Sheaths are constructed in a range of curve reach configurations, diameters and lengths to respond to physician preferences. The Introducer Sheath configurations covered under the subject 510(k) Premarket Notification include 8.5 F

{1}------------------------------------------------

and 9.5 F diameter, angles ranging from 0° - 180°, radius of curvatures ranging from 0.6" - 1.75″, lengths of 60 cm - 101.5 cm, and may be configured with either one, two or three curves in a single or dual plane.

#### Intended Use:

The Convoy Advanced Delivery Sheath Kit is intended to facilitate the intracardiac placement of interventional devices.

## Comparison to Predicate Devices:

|                       | Predicate Device                                    | Modified Device    |
|-----------------------|-----------------------------------------------------|--------------------|
| 510(k) Reference      | K964518                                             | Current Submission |
| Intended Use          | Intracardiac Placement of<br>Interventional Devices | Same               |
| Device Description    | Intracardiac Introducing Sheath                     | Same               |
| Single Use?           | Yes                                                 | Same               |
| EO Sterilized?        | Yes                                                 | Same               |
| Manufacturer          | Daig Corp                                           | EPT/BSC            |
| Device Classification | Class II, DYB; 21 CFR §870.1340                     | Same               |

CONFIDENTIAL

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle head facing left, with three curved lines extending from the back of the head, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the eagle symbol.

Public Health Service

SEP 1 1 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Boston Scientific Corporation/ EP Technologies, Inc. Ms. Andrea L. Ruth Senior Associate, Regulatory Affairs 2710 Orchard Parkway San Jose, CA 95134

Re: K022067

> Convoy Advanced Delivery Sheath Kit Regulation Number: 870.1340 Regulation Name: Catheter introducer. Regulatory Class: Class II Product Code: DBY Dated: June 24, 2002 Received: June 25, 2002

Dear Ms. Ruth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 -- Ms. Andrea L. Ruth

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

H. Douglas Tillett

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Intended Use Statement

#### 510(k) Number (if known): K022067

Device Name:

.

Convoy™ Advanced Delivery Sheath Kit

Indication for Use:

Image /page/4/Picture/5 description: The image shows a close-up of a document with some text. On the right side, the text "Aug 13 1993" is vertically aligned. On the left side, the word "RECEIVED" is also vertically aligned. The rest of the image is blurry.

વિદ્ય નિર્મદા

The Convoy Advanced Delivery Sheath Kit is inter facilitate the intracardiac placement of intervent devices.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number K022067

Prescription Use √

OR

Over-The-Counter Use

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DBY/K022067](https://fda.innolitics.com/submissions/IM/subpart-f%E2%80%94immunological-test-systems/DBY/K022067)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
